Service Manual
94 Pages
Preview
Page 1
HEALTHCARE
Sonicaid Team Service Manual This Service Manual is applicable to the Sonicaid Team Cardio-tocograph (Fetal Monitor) System Base Unit, Printer and S8000/8002 Printer. All servicing and repairs to these units must only be performed by authorised and qualified engineering staff; and strictly in accordance with the information provided in this manual. Huntleigh Healthcare Ltd cannot accept responsibility for safety, reliability and performance if modifications or repairs are carried out by unauthorised personnel. Sonicaid replacement parts must be used. NOTE: As with all Sonicaid instruments the Team Fetal Monitors are subject to continuous development and improvement and consequently may incorporate minor changes from the information contained in this manual. Changes from the previous issue are indicated by a vertical line on the right or left-hand side of the paragraph or drawing area affected. For information on the current issue of the manual, and any relevant amendment sheets, contact Huntliegh Healthcare Ltd. Service Department.
The Sonicaid™ Team is in conformity with the Medical Device Directive (93/42/EEC) and has been subject to the conformity assurance procedures laid down in the Council Directive.
Huntleigh Healthcare is certified by LRQA as an approved medical device manufacturer.
No part of this publication may be reproduced, stored in a retrieval system, or transmitted (by electronic, mechanical, photocopying, recording or any other means), without prior written permission from Huntleigh Healthcare Ltd. Always follow the instructions in the Operator’s and Service Manuals. Repairs and modifications must be carried out by authorised personnel.
1
1
General Introduction Two generations of the Sonicaid Team fetal monitor have been produced. This Service Manual has been written for the second generation. See section 1.1.1 “Identifying your system”, to identify the generation of your Team unit.
Note:
A Service Manual is available for the first generation of Sonicaid Teams - please contact Huntleigh Healthcare for details.
1.1 Introduction The Sonicaid Team fetal monitor consists of a base unit which collects the monitored information and a printer unit. This Service manual is applicable to the following base unit models:
Team Single fetal heart rate monitoring with an ultrasound transducer, and uterine activity with an external toco transducer.
Team Duo As Team, above, but with a second ultrasound transducer for monitoring twin fetal heart rates.
Team FECG Twin fetal heart rate monitoring by ultrasound transducers and invasively by a fetal ECG scalp electrode. Uterine activity can be measured either with an external toco transducer or an intra-uterine catheter pressure transducer.
Team IP Twin fetal heart rate monitoring either by two ultrasound transducers, or invasively by a fetal ECG scalp electrode and an ultrasound transducer. Uterine activity can be measured either with an external toco transducer or an intra-uterine catheter pressure transducer. Team IP can also measure the maternal heart rate (this feature is not currently available in the USA).
Team DM For use in a remote clinic or the patient’s home, Team DM provides the same facilities as Team, but includes a modem for transmitting stored CTG data.
This Service manual is applicable to the following printer unit models:
Standard Thermal printer for a continuous paper record of monitored CTG data.
Team Care Incorporates CTG analysis for use during the antepartum period.
IP Trend This printer incorporates a CTG analysis system for use during the intrapartum period. 2
1.1.1 Identifying your system The second generation of Sonicaid Team has the following distinguishing features compared to the first generation: 1 2 3 4
The base unit display is backlit The base unit keypad has tactile buttons The toco zero button is on the base unit keypad The symbols and other markings on the base unit and printer are on labels, rather than being silkscreened onto the case
1.1.2 Product description For a full description of the function and operation of the Sonicaid Team fetal monitor please refer to the Sonicaid Team User Manual.
1.1.3 Standards compliance Team is designed to comply with: • • •
IEC601-1 (1988) BS5724 Part 1 (1989) EN60601-1
1.1.4 Electromagnetic interference Make sure the environment in which Team is installed is not subject to any sources of strong electromagnetic interference, such as radio transmitters, mobile telephones, etc.
1.1.5 CE Mark The CE Mark on this product denotes conformity with the Euro-pean Council Directive 93/42/EEC concerning medical devices.
This fetal monitoring device is a prescription device in the USA.
3
2
Servicing MOD STATE The Mod State label on the underside indicates the modification state of the instrument i.e. the changes which have been made. A figure is struck through when the changes for that Mod. State have been done. Information in this manual which is applicable to certain Mod. States is indicated as such.
IMPORTANT: Servicing policy Due to the complexity of the electronics in the Team fetal monitor, the servicing policy is to locate, and return, the printed circuit board which is causing the problem. Huntleigh Healthcare does not supply surface mount components for the Team PCBs. For this reason, circuit diagrams are not included in this manual. Block diagrams and fault finding sections are included to make fault finding to leaded component level possible. If the problem cannot be located with absolute certainty, return the Team unit with full details of the symptoms. Boards which have been mutilated due to attempts at repair, or incorrectly packaged will not qualify for an allowance against the cost of a good board. The following procedure should be used: 1 2 3 4
Check and make a note of all the visible and/or audible symptoms of the problem. Check the power supply circuit, and the power supplies to each board. Check the clock and gating signals, and other waveforms, where indicated. Check the input and output signals or logic conditions where indicated.
Units within the warranty period must not be dismantled and should be returned to Huntleigh Healthcare Ltd for repair. Any units returned showing signs of tampering or accidental damage will not be covered under the warranty (refer to user manual for further details Antistatic Handling, Electro Static Discharge (ESD) The Sonicaid Team Series uses Electrostatic Discharge Sensistive Devices ESD’s) in its manufacture. The damage they suffer when handled incorrectly can be catastrophic. More often and potentially even worse, the damage may be partial or latent, seriously impairing the reliability of the unit. Due to the nature of the components used within the unit, special precautions must be taken to avoid damage to the circuitry. Static damage may not be immediately evident but could cause premature failure. The Sonicaid Team series must only be dismantled and serviced within an ESD protected area (EPA) as defined by CECC00015 (published by CENELEC) to avoid damage to the assembles.
4
2.1 Identifing your system software version Every Team base unit has a Standard Menu. Your Team may have been purchased with the optional Advanced Menu. To see the menu, press the [MENU] button on the Team base unit keypad (top right). <<<TIME/DATE <<<USER NAME <<<VERSION
mmHg/kPa>>> EXIT>>>
Figure 2.1 Team Standard Menu
2.1.1 Using the Team Menu <<<TIME/DATE <<<ALARM <<<ANNOTATE
NEXT>>> ELAPSED TIME >>> EXIT>>>
Figure 2.2 Team Advanced Menu (first level) Each menu item has arrows (>>>) pointing at the buttons on the Team keypad. To select a menu item, press the keypad button indicated by the arrows. To obtain information about the facilities installed in your Team unit, press select [VERSION] from the Team menu. The following screen appears, which tells you the firmware revision (version), and facilities fitted to your Team. Your Team unit may have different facilities from those shown in the example. Press the [EXIT] key to return to normal the display. VERSION 7.0 MODEM FITTED DUO FITTED 360 mins STORE FREE
EXIT>>>
Fig 2.3 Information about the facilities installed in the Team unit
5
2.2 Access for service 2.2.1 Team unit 1. First switch off and detach all electrical connections. 2. Disconnect all cables from the front and rear of the Team unit and, if fitted, remove the Printer by undoing the captive screw located in the paper well, beneath the paper. 3. With the Team unit upside down undo the seven captive screws on the base with a flat-bladed screwdriver. With the unit upright, and the display facing you, lift the case top up and pivot to the left, taking care not to strain the internal wiring, and rest on its side. 4. The top section of the case may be lifted off and placed to one side, with its components still connected to the bottom section. 5. On the Team Duo version, the Duo board is mounted above the main board. 6. On the Team IP version, the ECG board is mounted above the Duo board.
2.2.2 Team unit PCB access 1. Remove all plug and socket connections. 2. Disconnect and remove the loudspeaker. 3. Remove the two screws and four hexagonal pillars securing the PCB. 4. Remove or loosen the central transformer mounting screw. 5. Spring the board out of its base clips and lift up from the left side.
2.2.3 Printer 1. Take the Printer and remove the platen and any paper. 2. Remove the countersunk screw in the well with a Posidrive screwdriver, keeping it safe. 3. Turn the Printer upside down, and undo the four captive screws with a flat bladed screwdriver and the other two screws with a Posidrive screwdriver, keeping them safe. 4. Turn the Printer upright and, with the button facing you, carefully lift the case top up and pivot to the right, taking care not to strain the internal wiring, and rest on its side.
2.3 Connector numbering convention The connectors in the Team unit are identified with the letters SK (for Socket) and a number, for example, SK3 for socket number 3. Plugs are identified with the letters PL (for Plug) and a number, for example, PL3 for plug number 3. Each connection on a connector is indicated by the addition of a / and a number, for example, SK3/3 for socket number 3, connection number 3. 6
2.4 Installation of software upgrades From time to time new versions of software may become available, to add facilities or improvements. The software is contained in EPROMs and the procedures below detail how these are changed.
2.4.1 Team unit 1. Unplug all transducers, connecting cables and power cord from the Team unit. 2. Turn the unit upside-down. Using a flat-bladed screwdriver undo the seven screws in the base. 3. Turn the unit back upright. Carefully lift up the case top, hinging it from the left. 4. Remove the IC in SKT U6 on the PCB. 5. Taking anti-static precautions, re-fit the new EPROM in SKT U6 on the PCB. Check EPROM orientation. 6. Carefully re-fit the case lid and tighten up the seven base screws. 7. Re-configure the unit’s language, date format, modem/alarm settings as required.
2.4.2 Printer 1. Unplug all connecting cables from rear of the Team Printer unit. 2. To separate the Printer from the Team unit, undo the captive screw located in the paper-well beneath the paper. 3. Turn the Printer unit upside-down. Using a flat-bladed screwdriver undo the four captive case screws in the base. Remove and keep the remaining 2 screws from the bottom and 1 screw inside the paperwell with a cross-head screw driver. 4. With the Printer unit upright remove the case top by pivoting it to the right. Do not strain the internal cables. 5. Remove the IC in SKT U1, located at the rear of the PCB. 6. Taking anti-static precautions, re-fit the new EPROM into the socker. Check the EPROM for security of fitting and orientation. 7. Carefully re-fit the case top and tighten all the screws prior to re-fitting to the Team base unit.
2.5 Printer compatibility The latest version Printer shipped with the IP model Team unit must remain with it. Earlier versions of Printer which you may have in your establishment will not function fully with the IP model of the Team unit. For Printer version see the Mod. State label on the underside - if mod state 3 or above, this printer will function with all models of the Team unit.
7
3. Transducers Monitor must be serviced by Huntleigh Healthcare Ltd. Information is supplied here on the cable and connections for the transducers.
3.1
Ultrasound transducers These multi-crystal, wide-angle ultrasound transducers are used for monitoring fetal heart rate. The ultrasound transducers operate at a frequency of either 2 MHz (colour-coded blue), or 1.5 MHz (colourcoded yellow), and have a maximum ultrasonic power output density of 12mW/cm2 (2 MHz) or 4mW/cm2 (1.5 MHz). The following simple procedures will enable you to test the operation of the ultrasound transducer. The tests are performed with the ultrasound transducer connected to the fetal monitor, and the monitor switched on, as for normal use. Holding the ultrasound transducer in one hand, with the transducer face against the palm, stroke the back of the hand repeatedly with one finger. See diagrams below. Note the audio output, pulse indicator, FHR display and recorder trace on the fetal monitor is in synchronism with the finger movement. Water or gel may be necessary to obtain good contact between the palm and transducer.
The crystal elements in a transducer can be damaged if the transducer is dropped. If one or more crystals have been damaged, this can leave non-receptive areas on the transducer face, reducing the beam coverage. The positions of the crystals behind the transducer face are shown below:
To test the integrity of a suspect transducer: 1.
Squeeze a small amount of Aquasonic gel on to the transducer face over each crystal.
2.
Move the gel tube rapidly up and down over each crystal, keeping the tip of the gel tube in contact with the transducer. Check that you get an audio signal synchronised with the tube movement.
If a transducer problem is confirmed, contact Huntleigh Healthcare Ltd, Service department for advice. 8
3.2
External contractions transducers This transducer is a tocodynamometer whose central section is depressed by the patient’s abdominal muscles during a contraction. To check the operation, connect the external toco transducer to the Team unit. Press the ‘Toco Zero’ pushbutton on the Team unit keypad. The display should show 10% +/- 1%.
3.2.1 Field repair and calibration 1. Remove the stud to release the case back. 2. Refer to the following diagram. Adjust the 20k potentiometer to minimum resistance. + EXCITATION 200R
STRAIN GAUGE OUTPUT STRAIN GAUGE
20K
100R - EXCITATION
3. Set the 200R potentiometer to mid-position. 4. Check the resistance between output and negative excitation and make a note of the reading. 5. Measure the resistance between the output and positive excitation and make a note of the reading. 6. Adjust the 200R potentiometer so that the reading equals that recorded in 3 above.
3.2.2 Test 1. Plug the transducer into the Team unit and press the ‘Toco Zero’ key (10% deflection). 2. With the transducer face-up and horizontal, place a 100g weight on the centre of the plunger. Note he reading on the display. Take the weight off. Adjust the 20k potentiometer slightly. Re- zero the transducer. Place the 100g weight on the button and note the reading. 3. Repeat as above until a reading of 80% ±5% is obtained with the 100g weight. 4. Fit the case back. Fit the membrane. 5. Re-check the transducer with the 100g weight on the button. Make sure the reading is now 70–90%.
9
3.3
FECG leg electrode assembly This assembly is strapped to the thigh of the patient and provides a convenient connection point for the fetal ECG scalp electrode.
1 FECG socket: pin 1 M REF pin 2 FECG REF pin 3 FECG electrode Key: 1 FECG socket on Team 2 M REF 3 FECG REF 4 FECG electrode
017D065 900X260 2
900X259 4
3
Connection diagram for FECG using Safelinc electrodes
3.4
IUP transducer The TeamIP model is designed to make internal IUP measurements with the Intran disposable cathetertransducer. This transducer is connected to the Team unit by the Intran Extension Lead. Transducer sensitivity is 5mV/V/mmHg.
3.5
Patient-held event marker The event marker is operated by a patient operated hand-held push switch. The event mark (a solid sharp triangle) is printed at the top end of the fetal heart rate scale on the chart. When the event marker is operated, the Team unit bleeps and the words “FETAL EVENT NOTED” are displayed for a short time. The audio bleep is selectable in the Setup/Alarm menu.
10
1
2
8
9
3
10
7
11
12
4
6
3
ALL DIMENSIONS IN MM. Tolerances (Unless Stated Otherwise) Linear ±0.5mm X ±0.3mm X.X X.XX ±0.15mm Angular ±15' Internal Diameters +0.05/-0.03 To Ø6 +0.10/-0.04 Ø6-Ø25 Over Ø25 +0.50/-0.10
4
5
6
7
1
1
8
2
8
2
8
9
9
3
10
7
3
10
7
11
12
11
12
9
4
6
4
6
5
5
FINISH
CLEAN
SEE BILL OF MATERIALS
MATERIAL
7
BLACK
RED
BLUE
YELLOW
10
14
13
11
12
AF No
430
DATE
20AUG97
16
PINK
GREEN
1:1
SCALE (CAD. REF.)
15
RED BLACK
AMENDMENT
20AUG97
RC
CHKD
RESTR'R'D
REDRAWN+
DRAWN
2
3
4
5
6
7
8
9
10
17 19
SCALE REFERS TO C.A.D. ORIGINAL DO NOT SCALE FROM THIS PRINT
1
Huntleigh Healthcare Ltd Diagnostic Products Division 35 Portmanmoor Road, Cardiff CF24 5HN Tel: 02920 485885 © Huntleigh Healthcare Ltd.
5
2
REMOVE BURRS & SHARP EDGES
1
ISS
2
3
4
18 19
8909-6904
DRAWING No:
STANDARD NICOLAY LOOM
TITLE
RED
BLACK
YELLOW
BLUE
IF IN DOUBT ASK
1
HEALTHCARE
1
DRAWN TO BS308
A3
Figure __ Standard Nicolay Loom Assembly
11
12
DRAWN TO BS308
1
Huntleigh Healthcare Ltd Diagnostic Products Division 35 Portmanmoor Road, Cardiff CF24 5HN Tel: 02920 485885 © Huntleigh Healthcare Ltd.
HEALTHCARE
1
2
ALL DIMENSIONS IN MM. Tolerances (Unless Stated Otherwise) Linear ±0.5mm X ±0.3mm X.X X.XX ±0.15mm Angular ±15' Internal Diameters +0.05/-0.03 To Ø6 +0.10/-0.04 Ø6-Ø25 Over Ø25 +0.50/-0.10
175±5
FINISH
CLEAN
3
493
1:1
SCALE (CAD. REF.)
AF No
DATE
4SEP97
5AUG97
RC
DRAWN
CHKD
RESTRCT'R'D
REDRAWN +
AMENDMENT
1
ISS
IF IN DOUBT ASK
8909-6958
DRAWING No:
SPEAKER LOOM BARE
TITLE
STRIPPED AND TINNED 4
SCALE REFERS TO C.A.D. ORIGINAL DO NOT SCALE FROM THIS PRINT
SEE BILL OF MATERIALS
MATERIAL
REMOVE BURRS & SHARP EDGES
A3
Figure __ Speaker Loom - Bare
C
APP'D
CHKD
DRAWN
INTLS
RC
DATE
22 FEB 1991
FINISH & COLOUR
PARTS LIST:-
8900-6952
PIN 1 INDICATED BY ON SOCKET
1900-0557, QTY 1 HIROSE IDC SOCKET
MATERIAL
THIRD ANGLE PROJECTION
HOLE <Ø7mm = X X.X X.XX
TOLERANCE
±0.15
-0.00 ±0.5 ±0.3
DIMENSIONS ARE ALL IN MILLIMETERS
= = =
UNLESS OTHERWISE STATED +0.15
TO BS308 & BS3939
NTS
ORIGINAL SCALE
MODIFICATION
1000-0915, QTY 200mm 20WAY RIBBON CABLE
185±5
22 FEB 1991
2 1
ISSUE DATE
4 APR 1991
HEALTHCARE
A3
8900-2504
DRAWING No.
DISPLAY TO PCB RIBBON CABLE
TITLE CABLE DRAWING
REMOVE ALL SHARP EDGES & BURRS
PIN 1
1900-0558, QTY 1 HIROSE HIF-2B 20 WAY CONNECTOR
TO BE SOLDERED TO DISPLAY PCB
Figure __ Cable Drawing - Display to PCB ribbon cable
13
Figure __
CCFL Inverter PCB Assy
DRAWN TO BS308
SCALE REFERS TO C.A.D. ORIGINAL DO NOT SCALE FROM THIS PRINT
REMOVE BURRS & SHARP EDGES
R1
C2 C5
C4
SW1
C6
Q1
R2
P1 P2 P4 P3
IF IN DOUBT ASK
L1
C3
Q2 1
10
2
9
T1
3
P6
8
4
7
5
6
C1
ALL DIMENSIONS IN MM. Tolerances (Unless Stated Otherwise) Linear
Internal Diameters
X
±0.5mm
To Ø6
+0.05/-0.03
X.X
±0.3mm
Ø6-Ø25
+0.10/-0.04
X.XX
±0.15mm
Over Ø25 +0.50/-0.10
Angular
±15'
06-04-98
582
NEW LAYOUT
3
DATE
AF No.
AMENDMENT
ISS
MATERIAL SEE NOTE ABOVE
TITLE
CCFL INV. PCB ASSY.
FINISH CLEAN
HEALTHCARE
14
Huntleigh Healthcare Ltd Diagnostic Products Division 35 Portmanmoor Road, Cardiff CF24 5HN Tel: 02920 485885 © Huntleigh Healthcare Ltd.
DRAWING No. SCALE (CAD.REF.)
N.T.S
DRAWN
CHKD
8909-0002
P5
A4
SH 1 OF 1
11
A2
1
TO BS308
DSJ
DATE
040203
2.8M WAS 1.8M. ECR. 0126 REFERS.
SCREENING IMPROVED
DRG
11
CONSIDER THE AFFECT OF ANY CHANGES ON:
DO NOT SCALE
ISSUE
84002517
ISSUE
DRAWING NO
3
4
BLACK
BLACK
RED
GREEN
7
5
11
10
2
4
3
WEIGHT
COMMODITY CODE
POLARISING TAB POSITION
DESIGN RIGHT Huntleigh HEALTHCARE.
THIS DRAWING CONTAINS PROPRIETARY INFORMATION ANY REPRODUCTION DISCLOSURE OR USE OF THIS DRAWING IS EXPRESSLY PROHIBITED EXCEPT AS AGREED IN WRITING BY THE COMPANY
LIN ±0.2 ANG ±1°
1:1
SCALE
DATE
ORIGINAL
2
CLEAN TOLERANCES UNLESS OTHERWISE STATED
(PART OF UPPER PLUG BODY)
10
2
18 17
FINISH
CHK
6
7
TITLE
11
HEALTHCARE
35 Portmanmoor Road, Cardiff, CF24 5HN All Rights Reserved
c Huntleigh HEALTHCARE.
2 7
9 7
DRAWING NUMBER
84002517
11 A2
OF 1
ISSUE SH 1
PCB ADHERED TO WEBS ON TRANSDUCER CASE WITH EPOXY ADHESIVE NOTE ORIENTATION
BLACK
RED
CABLE SCREENING, SLEEVED
FOLD BACK BRAID EQUALLY (360°) AND UNDER CABLE CLAMP, TO CONNECT WITH RFI COATING
8
ULTRASOUND TRANS. 1.5MHz Huntleigh
mm
UNITS
CLAMP RADII TO FACE TOWARDS CABLE
MATERIAL
EXTERNAL VIEW OF ASSEMBLED PLUG
5 SCALE REFERS TO C.A.D. ORIGINAL DO NOT SCALE FROM THIS PRINT
8400-6919M (2.8M)
1K3 RESISTOR MRS25
22.5°
M2.5 CRINKLE WASHERS UNDER SCREW HEADS
FOLD BACK BRAID EQUALLY (360°) AND UNDER CABLE CLAMP,
POLARISING KEY POSITION (PART OF PLUG INSERT)
SOLDER SIDE PIN POSITIONS
6
9
12
8
1
SCREEN EARTH (GREEN)
RED
NICOLAY PLUG (YELLOW)
(3 METRE AND 4 METRE LENGTHS NO LONGER REQUIRED.)
CABLE LENGTH 8400-6919M 2.8M ± 10mm
2
F
E
D
C
B
A
Figure __ 1.5MHz ULT Transducer Wiring
15
A2
ISSUE
SH 1 OF 1
11
1
TO BS308
DSJ
DATE
040203
2.8M WAS 1.8M. ECR. 0126 REFERS.
SCREENING IMPROVED
DRG
11
CONSIDER THE AFFECT OF ANY CHANGES ON:
DO NOT SCALE
ISSUE
84002515
16
DRAWING NO
3
4
BLACK
RED
GREEN
7
5
11
10
2
4
3
LINK
WEIGHT
COMMODITY CODE
SOLDER SIDE PIN POSITIONS
6
9
12
8
1
BLACK SCREEN EARTH (GREEN)
LINK RED
NICOLAY PLUG (BLUE)
POLARISING TAB POSITION
DESIGN RIGHT Huntleigh HEALTHCARE.
THIS DRAWING CONTAINS PROPRIETARY INFORMATION ANY REPRODUCTION DISCLOSURE OR USE OF THIS DRAWING IS EXPRESSLY PROHIBITED EXCEPT AS AGREED IN WRITING BY THE COMPANY
LIN ±0.2 ANG ±1°
1:1
SCALE
DATE
ORIGINAL
CLEAN
9
2
FINISH
TOLERANCES UNLESS OTHERWISE STATED
POLARISING KEY POSITION (PART OF PLUG INSERT)
EXTERNAL VIEW OF ASSEMBLED PLUG
(PART OF UPPER PLUG BODY)
2
17 16
MATERIAL
CHK
6
TITLE
8
10
HEALTHCARE
35 Portmanmoor Road, Cardiff, CF24 5HN All Rights Reserved
c Huntleigh HEALTHCARE.
1 7
8 7
DRAWING NUMBER
84002515
11 A2
OF 1
ISSUE SH 1
PCB ADHERED TO WEBS ON TRANSDUCER CASE WITH EPOXY ADHESIVE NOTE ORIENTATION
BLACK
RED
CABLE SCREENING, SLEEVED
FOLD BACK BRAID EQUALLY (360°) AND UNDER CABLE CLAMP TO CONNECT WITH RFI COATING
7
ULTRASOUND TRANS. 2.0MHz. Huntleigh
mm
UNITS
CLAMP RADII TO FACE TOWARDS CABLE
8400-6920M
22.5°
5 SCALE REFERS TO C.A.D. ORIGINAL DO NOT SCALE FROM THIS PRINT
M2.5 CRINKLE WASHERS UNDER SCREW HEADS
FOLD BACK BRAID EQUALLY (360°) AND UNDER CABLE CLAMP
(3 METRE AND 4 METRE LENGTHS NO LONGER REQUIRED.)
CABLE LENGTH 8400-6920M 2.8M ± 10mm
2
F
E
D
C
B
A
Figure __ 2 MHz ULT Transducer Wiring
2065-0005 1901-0223 2215-0007
PLASTIC WASHER WAVY WASHER TERMINAL
REMOVE EXISTING WHITE CAPS
ABOVE PARTS 2 POSNS.
[CHASSIS]
2080-0001 1901-0211 1901-0273 2080-0001 1901-0289
NUT S.P. WASHER SOLDER TAG NUT SHOULDER WASHER
NUTS 2080-0001 (4 POSNS) TO BE TIGHTENED TO A TORQUE OF 0.17Nm ±0.015Nm
DRAWN TO BS308
RED
HEALTHCARE
2065-0003 1901-0275 1901-0214 1901-0255
8400-6134 2065-0002
TIE WRAP
ALL DIMENSIONS IN MM. Tolerances (Unless Stated Otherwise) Linear ±0.5mm X ±0.3mm X.X X.XX ±0.15mm Angular ±15' Internal Diameters +0.05/-0.03 To Ø6 +0.10/-0.04 Ø6-Ø25 Over Ø25 +0.50/-0.10
1
6
CLEAN
FINISH
2
11
10
4
3
39K RESISTOR MRS25
2:1
IF IN DOUBT ASK
AF No
A###
DRAWN
CHKD
CABLE CHG
REDRAWN/
AMENDMENT
22.5°
7
ISS
8400-2510
DRAWING No:
FECG TDX ASSY (PRE-ENCAP.)
TITLE
POLARISING KEY POSITION (PART OF PLUG INSERT)
POLARISING TAB POSITION (PART OF UPPER PLUG BODY)
NICOLAY PLUG (BLUE)
EXTERNAL VIEW OF ASSEMBLED PLUG
SCALE (CAD. REF.)
20JUL95
DATE
NOTE: RESISTOR MUST NOT BE ALLOWED TO COME WITHIN 3mm OF OTHER WIRES
12
9
PARTS LIST: 8400-6962
MATERIAL
7
8
SOLDER SIDE PIN POSITIONS
PVC SLEEVE 3mm ID QTY 25mm
BLACK
SCREEN
RED
CABLE 2.3 METRES
SCALE REFERS TO C.A.D. ORIGINAL DO NOT SCALE FROM THIS PRINT
REMOVE THIS SECTION OF GROMMET
Huntleigh Healthcare Ltd Diagnostic Products Division 35 Portmanmoor Road, Cardiff CF24 5HN Tel: 02920 485885 © Huntleigh Healthcare Ltd.
PLASTIC WASHER SOLDER TAG S.P. WASHER NUT
[CHASSIS]
PATIENT EARTH DISC SHOULDER WASHER
SCREEN
BLACK
USE ASSEMBLY JIG 8400-6007/8/9 TO ENSURE CORRECT ORIENTATION OF TERMINALS 2215-0007 & EARTH DISC 8400-6134 (AS SHOWN ON THIS DRAWING)
REMOVE BURRS & SHARP EDGES
A2
Figure __ FEGC TDX ASSY (PRE-ENCAP)
17
A2
ISSUE
1
TO BS308
WEIGHT
COMMODITY CODE
FOR INFORMATION SEE DRG 8400-3602 FOR TOCO TRANSDUCER STRUCTURE
PARTS LIST: 8330-6990 CONTRACTIONS TRANSDUCER
Used with: FM6
FISCHER PLUG
PARTS LIST: 8310-6902 CONTRACTIONS TRANSDUCER
Used with: TEL2
4 PIN LEMO 0B-SERIES PLUG (RED)
8300-6973 CONTRACTIONS TRANSDUCER - FM5L
PARTS LISTS: 8300-6920 CONTRACTIONS TRANSDUCER - FM5
Used with: FM5, FM5L
SIGNAL (GREEN) 4
1 7
5
6
3
2
3
2
39K
TOLERANCES UNLESS OTHERWISE STATED
IMPORTANT:
CHK
DATE
DESIGN RIGHT Huntleigh HEALTHCARE.
THIS DRAWING CONTAINS PROPRIETARY INFORMATION ANY REPRODUCTION DISCLOSURE OR USE OF THIS DRAWING IS EXPRESSLY PROHIBITED EXCEPT AS AGREED IN WRITING BY THE COMPANY
LIN NTS ANG ±1°
7 6
12
9
5
11
10
1
4
2 3
1:1
HEALTHCARE
Huntleigh
mm
UNITS
60 ±2
+ EXCITATION (RED)
NEOPRENE SLEEVING IN 3 POSITIONS HELD IN PLACE ON PCB WITH ADHESIVE
- EXCITATION (BROWN)
SIGNAL
+ EXCITATION (RED)
All Rights Reserved
c Huntleigh HEALTHCARE.
35 Portmanmoor Road, Cardiff, CF24 5HN
84012513
DRAWING NUMBER
6 A2
OF 1
ISSUE SH 1
WIRING DIAG TOCO STRUCTURE
TITLE
8
- EXCITATION (BROWN)
SIGNAL (GREEN)
DOUBLE LAYER OF INSULATING TAPE
POLARISING KEY POSITION (PART OF PLUG INSERT)
EXTERNAL VIEW OF ASSEMBLED PLUG
POLARISING TAB POSITION (PART OF UPPER PLUG BODY)
2500 ±25
RESISTOR (MRS25)
39K
8
CUT BACK ANY SPARE CORES AT BOTH ENDS, CLOSE TO CABLE SHEATH
SCALE
RESISTOR (MRS25)
NOTE WIRE ROUTING
TWIST WIRES TOGETHER BEFORE ASSEMBLING CASE
PARTS LIST: 8401-6902M
ORIGINAL
4
7
PIN VIEW (SOLDER SIDE)
MERIDIAN-STYLE TOCO TRANSDUCER HEAD ASSEMBLY
PARTS LIST: 8400-6921 CONTRACTIONS TRANSDUCER
CLEAN
5
- EXCITATION (BROWN)
+ EXCITATION (RED)
+ EXCITATION (RED)
- EXCITATION (BROWN)
Used on: FM6, FM7, MERIDIAN, TEL3, TEAM, FM800
FINISH
12
9
DISPLAY OFF (FM3R)
6 SCALE REFERS TO C.A.D. ORIGINAL DO NOT SCALE FROM THIS PRINT
NICOLAY PLUG (PINK)
5
MATERIAL
10
PIN VIEW (SOLDER SIDE)
4
1
PIN VIEW (SOLDER SIDE)
4
1
[]
2 RESISTORS (FC55), SLEEVED LINKING PIN 1 TO PIN 4 & PIN 2 TO PIN 7
LINK WIRE
+ EXCITATION (RED)
SIGNAL (GREEN)
- EXCITATION (BROWN)
4
SEE NOTES ABOVE
PIN 11 NOT CONNECTED
SIGNAL (GREEN)
- EXCITATION (BROWN)
+ EXCITATION (RED)
SIGNAL (GREEN)
4
PIN VIEW (SOLDER SIDE)
3
1K 5
+ EXCITATION (RED)
SIGNAL (GREEN)
2
1
PIN VIEW (SOLDER SIDE)
3 5
PIN VIEW (SOLDER SIDE)
2
- EXCITATION (BROWN)
3
1K5
4 PIN LEMO 0B-SERIES PLUG (RED)
PARTS LIST: 3820-6902 CONTRACTIONS TRANSDUCER (MK II)
7 PIN LEMO 3B-SERIES PLUG (RED)
PARTS LIST: 3820-6903 CONTRACTIONS TRANSDUCER (MK I)
Used with: FM3R, FM4
200504
AF No. A289 REFERS.
RC
6
DATE
Used with: FM2, FM3, FM4
5 PIN (180°) DIN PLUG
2
NICOLAY STRAIGHT LEAD
DRG
CONSIDER THE AFFECT OF ANY CHANGES ON:
DO NOT SCALE
ISSUE
DRAWING NO
SH 1 OF 1 OF
6
84012513
18 22.5°
w w
F
E
D
C
B
A
Figure __ Wiring Diagram - Toco Transducer
8400-2513-2.dwg
W:DRAWINGSOBTEAMDRAWINGS
HEALTHCARE
5K6 RESISTOR MRS 25
7
8
6
12
9
5
11
10
1
4
2
3
CONNECTIONS VIEW FROM SOLDER SIDE
+0.05/-0.03 To Ø6 +0.10/-0.04 Ø6-Ø25 Over Ø25 +0.50/-0.10
Internal Diameters
ALL DIMENSIONS IN MM. Tolerances (Unless Stated Otherwise) Linear ±0.5mm X ±0.3mm X.X ±0.15mm X.XX
BLACK (- EXCITATION)
RED (+ EXCITATION)
WHITE (+ SIGNAL)
GREEN (- SIGNAL)
REMOVE BURRS & SHARP EDGES
NICOLAY PLUG BLUE
DRAWN TO BS308
Huntleigh Healthcare Ltd Diagnostic Products Division 35 Portmanmoor Road, Cardiff CF24 5HN Tel: 02920 485885 © Huntleigh Healthcare Ltd.
FM588-1
FINISH
MATERIAL
CLEAN
BOM 55-8400-6937
POLARISING KEY POSITION (PART OF PLUG INSERT)
22.5°
POLARISING TAB POSITION (PART OF UPPER PLUG BODY)
SCALE (CAD. REF.)
23.09.99
DATE
A695
AF No
DSJ
2
ISS
8400-2513
DRAWING No:
INTRAN EXTENSION LEAD
TITLE
RED
GREEN
WHITE
BLACK
IF IN DOUBT ASK
FOR REFERENCE ONLY PIN VIEW OF INTRAN CONNECTOR (NON SOLDER SIDE)
CHKD
REDRAWN STRAIGHT GRIPPER
AMENDMENT
DRAWN
EXTERNAL VIEW OF ASSEMBLED PLUG
REF: SUPPLIED CABLE LENGTH 3.85M TRIM CABLE LENGTH TO 2000mm ± 10mm.
PURCHASE SPEC DRG No. A4 8400-8010
INTRAN EXTENSION LEAD ASSY
SCALE REFERS TO C.A.D. ORIGINAL DO NOT SCALE FROM THIS PRINT
A3
Figure __ Cable diagram - INTRAN Extension Lead - Wiring
19
4. Team Unit Main PCB This section refers to circuit diagrams and component layout fault location diagrams. Checks for correct functioning are included in the following pages.
4.1 Guide to Main PCB circuits TEAM Main PCB Circuits: Top sheet: interconnection sheet 1 Analogue circuits 2 Clock Generator circuits 3 Digital circuits 4 Isolated 232/485 5 Power Supplies 6 Signal Processing 7 Team Telemetry
20
Fig. 4.3 page 4.7 Fig. 4.7 page 4.19 Fig. 4.15, page 4.41 Fig. 4.16, page 4.43 Fig. 4.13, page 4.35 Fig. 4.5, page 4.13 Fig. 4.10, page 4.27 Fig. 7.1, page 7.7