Arthrex

Arthroscopy Systems

Model AR-6480 DualWave Arthroscopy Pump Troubleshooting Guide Rev B

Troubleshooting Guide

2 Pages

Troubleshooting PROBLEM  SOLUTION  ** Tubing Out **  1. Tubing sensor indicator not seated 2. Ensure that the tubing pressure plug is seated completely 3. Change tubing * Return to Arthrex for repair  Console fails Self Diagnostic Test  1. Ensure no tubing is connected to the pump during power on sequence * Return to Arthrex for repair  Console will not power up  1. Check AC main power cord 2. Try alternate power outlet 3. Check AC main fuses * Return to Arthrex for repair  Lack of distention liquid bloody/cloudy  Does not pump when activated  ** Door Open **  1. Roller housing not secured – ensure locking lever is properly secured * Return to Arthrex for repair  No/Low Flow  To Adjust the Shaver Detection Settings:  1. Activate LAVAGE mode 2. Increase pressure 3. Decrease outflow 1. Check for error messages 2. Open all tubing clamps and shut-off valves 3. Ensure actual pressure is below target pressure 4. Check if the tubing is pinched, kinked, or blocked * Return to Arthrex for repair  Inadequate pressure  Arthrex DualWave Set-up and Operation  1. Press the MENU Button  2. Press the DEFAULTS Button  3. Press the SHAVER Button  4. Choose alternate DETECTION FILTER setting 5. Press DONE to save changes  1. Increase pressure. 2. Activate LAVAGE mode 3. Reduce outflow 4. Use high-flow cannulas 1. Check for error messages 2. Check that all tubing clamps are open 3. Check the settings for flow and pressure 4. Check if the tubing is pinched, kinked or blocked 5. Check that tubing seats correctly over the rollers 6. Verify use of high-flow cannulas 7. If failure remains, return to Arthrex for repair  ** Overpressure **  1. Increase or open outflow 2. Manipulate joint to stress-free position  ** Pressure Fault **  1. Ensure adequate fluid supply 2. Decrease outflow 3. Check tubing for damage and if it is pinched, kinked or blocked 4. Check tubing for proper connections 5. Replace tubing 6. If failure remains, return to Arthrex for repair  No Shaver Detection  1. Ensure universal shaver cable is connected 2. Turn on and off DualWave while the shaver system is powered on 3. Change SHAVER DETECTION PRESET to alternate setting 4. Isolate the power source 5. If failure remains, return to Arthrex for repair  ©2013, Arthrex Inc. All rights reserved. LM1-0625-EN_B  http://www.arthrex.com/imaging-resection/fluid-management
File Type: PDF
File Size: 191 KB
File Name: model_ar-6480_dualwave_arthroscopy_pump_troubleshoo.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions