Technical Training Guide
47 Pages
Preview
Page 1
Technical training guide
D190231-0
Training program for Volumat Agilia
Presentation
Usage exercise
Safety features
Disassembling procedures
Tests and Calibration
Kits presentation D190231-0
Training program for Partner Agilia service software
Connecting a device
Performing a complete configuration
Calibration
Detailed history reading
Other functions
D190231-0
Introduction Volumat Agilia / Volumat MC Agilia:
Up to 1500 events are recorded
A large display for "at a glance“ monitoring
Compatible with Vigilant central infusion monitoring
Internal battery 2.2Ah / 7.4 Volt ( LI-ion ) 8 hours autonomy at 125 ml/h External power 9 Volt CC, power 15 Watt
Swing lock pole / rail clamp D190231-0
General view
D190231-0
Front housing 3 Safety features Medium position calibration of the door, to detect
in minimum distance if the door is open or closed to avoid a free flow. A OCS “ Occlusivity Control System” when the
pump is used without a drop detector, the system starts to detect if the pump is occlusive or not before running the infusion procedure and avoid any free flow. A dedicated set including an automatic clamp
driven by a specific motor to open or close clamp when it removed from the pump. To avoid a free flow. D190231-0
Volumat Agilia start up
Auto-test
OCS system starts
The programming mode displays
Confirmation or modification of infusion mode
Without Drop detector => Volume limit / Time limit are mandatory With Drop detector => limit volume or time are not mandatory
Press bolus key to prime the line if this option is activated
Validate the selection to start the infusion D190231-0
OCS System Tube clamp Mobile clamp zone A
Mobile clamp zone D OCS finger pressure zone
Mobile clamp zone C
Mobile clamp zone B
α° D190231-0
OCS System Motor clamp design
D190231-0
OCS System Obturator / sensors Two Optical switches identifies 4 positions for the OCS finger
D190231-0
OCS System Position 1: Clamp absent
The linear actuator is back to its maximum The fixed mechanical part is back to its maximum (position 0) The mobile mechanical part is on position 1 The OCS finger is out the housing for 2 mm
No clamp, position 1 : state opto1 = 0, state opto2 = 1 The presence and the absence of the clamp should be detected
D190231-0
OCS System Position 0: Clamp present and closed
The linear actuator is back to its maximum The fixed mechanical part is back to its maximum (position 0) The mobile mechanical part is also back to its maximum ( position 0) The OCS finger in the limit surface of the housing
Clamp detected and closed, position 0 : state opto1 = 1, state opto2 = 1 The presence and the absence of the clamp should be detected Closing angle α° should permit to detect that the clamp is in the close position (with the OCS finger).
D190231-0
α°
OCS System Position 3: Clamp open, OCS finger out of the housing at its maximum
The linear actuator is out at its maximum The fixed mechanical part is out to its maximum (position 3) The mobile mechanical part is also out to its maximum ( position 3)
Clamp open, finger left to the maximum, position 3: state opto1 = 1, state opto2 = 0
D190231-0
OCS System Position 2: Finger in intermediary position
The linear actuator is in intermediary position The fixed mechanical part is in intermediary position (position 2) The mobile mechanical part is also in intermediary (position 2)
Clamp open, finger in intermediary position; position 2: state opto1 = 0, state opto2 = 0
D190231-0
OCS System Clamp damage
The linear actuator is in unknown position The fixed mechanical part is in unknown position The mobile mechanical part is in the back to its maximum (position 0)
Blown mechanical fuse: state opto1 = 1,state opto2 = 1 The OCS electro-mechanical mechanism system fails when closing and opening the door, That should not damage the tube or the device
D190231-0
OCS System Summary Optical switch status according the actuator position OCS System
Position
Optical switch 1 status
Optical switch 2 status
Clamp present and closed
0
1
1
Clamp absent
1
0
1
Finger in intermediary position
2
0
0
Clamp open, OCS finger out of the housing at its maximum
3
1
0
Damage clamp
0
1
1
D190231-0
OCS System Actuator position according to the OCS system function
Position
W/O clamp
0 1
With clamp Before OCS starts
With clamp After OCS
OCS Sequences
X
2
±
3
°
X
X
D190231-0
Air bubble detector
D190231-0
Pump rework Modification
Soft CPU / power supply
Version
1.1c
New part number
Starting since s/n: All except: 19830250 => 19830253 19830255 => 19830259 19830261 => 19830268 19830270 19830272 19830274 19860340 => 19860343
OCS failure + the membrane material could be damaged in contact with the silicone tube. A drift of the pressure on upstream and downstream
20103343
????
Air bubble board Membrane
Old part number
1.1a / 1.2a
Soft CPU
Door lever with over moulded magnet
Reasons
(1)
20042276 => 20042287 20042714 => 20042714 20134572 Gain improvement of the sensor
179004
179004
OCS failure: New membrane
179055
179069
(1) More details concerning this modification is available on the product Maintenance protocol PM0194: Soft rework to 1.1e with a replacement of the membrane
D190231-0
20012133 20227436
Disassembling procedure Remark : Disconnect the battery before starting any disassembling procedure. • Unscrew the 2 fixing screws ( marker 1) holding the battery cover • Remove the battery cover ( marker 2 ).
D190231-0