ATOM Medical Corporation

Incubators

Power Pack i Battery Replacement Instructions

Instructions

3 Pages

Atom Power Pack i Battery Replacement  Note: These instructions apply to the following Atom Power Pack i models: • 61581 Atom Power Pack i with Lithium-ion battery - uses 95755 Lithium-ion Battery Pack (1 x battery) • 61624 Atom Power Pack i with Lead-acid battery - uses 98755 Lead-acid Battery Pack (2 x batteries)  Caution: • Battery replacement may cause electric shock or burns if a short circuit occurs. Please read the following instructions carefully to ensure minimum risk when replacing batteries. • Before replacement, remove any jewellery including watches, rings, necklaces or anything else which may be made of, or contain, metal. • Only use tools with insulated handles. • Do not place tools or metal objects on the batteries. •	As soon as the battery has been disconnected, the Power Pack i will begin to discharge voltage. To avoid a spark while reconnecting the battery, it is recommend to complete the following process within 30 minutes.  What you’ll need • Replacement battery • Suitable cutting knife, eg box cutter  PD348  • Screwdriver
File Type: PDF
File Size: 1.68 MB
File Name: PD348 - Atom Power Pack battery removal instructions.pdf