transcapsule_v-707_operation_manual.pdf
Page 1
0123
Atom Transcapsule
V-707
Operation Manual
TO THE OPERATOR AND THE PERSON IN CHARGE
OF MAINTENANCE AND CARE OF THE UNIT:
앫 Read this Manual carefully before operating the unit.
앫 Keep this Manual where it is readily accessible for reference
in case of need.
Page 2
The unit for 100V region doesn’t suit the CE mark.
For further information or questions, please contact the following representative if you are in an EU
member state. In other countries, please contact your distributor or Atom Medical International, Inc.
Atom Medical International Inc. EU Office
Via Francesco Croce, 65 20081 Abbiategrasso (MI) ITALY
Tel: +39-02-9976-3101, Fax: +39-02-9976-3110
Page 3
INTRODUCTION
This Operation Manual deals with the specifications, operation and maintenance of the Atom Transcapsule V-707.
Atom is by no means responsible for any malfunction arising from a user ignoring the instructions for operation
and maintenance described in this Manual as well as for any accident attributable to repair by someone other than
technical personnel belonging to Atom.
Read this Manual carefully and familiarize yourself thoroughly with its contents before operating the unit. Keep
this Manual where it is readily accessible for reference in case of need. If any technical problem should arise,
please contact your local Atom representative.
CAUTION
This product is shipped without being disinfected. Be sure to clean and disinfect the unit before
using it for the first time after purchase.
Used parts and products past their lives should be disinfected and/or sterilized and then disposed
of as medical waste.
INTENDED USE
The Atom Transcapsule V-707 is a closed-type transport incubator for neonatal and premature infants intended to transport low birth weight infants and seriously ill neonates.
SAFETY INFORMATION
Instructions to ensure the safe operation of the unit are found throughout this Manual. Please read the Manual
carefully before operating the unit. Please follow the instructions in operating it.
[1] Basic Instructions
1. Follow the instructions for the safe use of the unit.
Follow the operating instructions described in this Manual for the safe use of the unit.
2. Inspect the unit on a periodical basis.
Periodical inspection is needed to use the unit under optimum conditions. (See “[9] Maintenance Inspection” of
this Manual.)
3. Never use the unit when faulty.
If any damage or malfunction of the unit should be noticed, stop using it immediately and contact your local
Atom representative.
4. EMC
Medical electric equipment needs special precautions regarding EMC. It needs to be installed and put into
service according to the EMC information provided in this Manual.
❉ EMC stands for electromagnetic compatibility. EMC requirements specify measures against emitting and
receiving unwanted electric waves.
−1−
Page 4
[2] Definition of Warning Indication
Three levels of warning indication are used throughout this Manual and on the unit. They are defined as
follows.
A DANGER notice indicates an immediately hazardous situation which, if not
avoided,
will result in death or serious injury, serious damage to property such as
DANGER:
total loss of use of equipment, and a fire.
A WARNING notice indicates an indirectly (potentially) hazardous situation
WARNING: which, if not avoided, will result in death or serious injury, serious damage to
property such as total loss of use of equipment, and a fire.
A CAUTION notice indicates a hazardous situation which, if not avoided, can reCAUTION: sult in minor or moderate injury, partial damage to property, and loss of
data stored in computers.
[3] Definition of Marks
1. Marks to indicate caution or warning
Mark
Title
and Indication
General attention
Indicates unspecified general caution or warning.
Caution: Hot surface
Indicates that the surface can be dangerously hot.
2. Marks to prohibit action
Mark
Title
and Indication
General prohibition
Indicates unspecified general prohibition.
Prohibition of disassembly
Indicates prohibition of disassembly of the unit where it may cause an electric shock or other hazards.
Prohibition of use of fire
Indicates prohibition of use of fire where an external use of fire may cause the unit to ignite
under certain conditions.
−2−
Page 5
3. Symbols to give instructions for action
Mark
Title
and Indication
General instruction
Indicates unspecified general action on the part of the user.
Remove the power plug from the power outlet
Instructs the user to remove the power plug from the power outlet in the case of malfunction or
where lighting may occur.
Connect a ground wire
Instructs the user to connect the ground wire without fail where the unit is provided with a
ground terminal.
4. Symbols of international standards (IEC)
Mark
Title
and Indication
Equipment partially on
Indicates that a part of the device is “ON”.
Equipment partially off
Indicates that a part of the device is “OFF”.
Type BF applied part
Indicates that the device is classified as Type BF in terms of the degree of protection against an
electric shock.
Alarm setting
Indicates a switch for setting the alarm function.
Audible alarm silenced
Indicates that an audible alarm is silenced.
Caution
Indicates that the user needs to see a relevant accompanying document before operating the
unit.
Fan
Indicates a control switch that controls the fan.
Lighting
Indicates a switch that controls lighting.
Recycle mark (battery)
Indicates that recycling is recommended.
−3−
Page 6
5. Other symbols
Mark
Title
and Indication
Setting
Indicates that a setting is increased.
Setting
Indicates that a setting is decreased.
Alarm
Indicates a power failure alarm.
AC power
Indicates that the device is connected to AC power.
Battery
Indicates that the device is being powered by a battery or batteries.
External DC power source
Indicates that the unit is operating on an external DC power source.
Alarm
Indicates that an audible alarm is silenced.
Alarm
Indicates that a system of the unit is in an abnormal condition.
Alarm
Indicates that the incubator air temperature has exceeded or fallen below the set temperature by
3˚C or more.
Alarm
Indicates that the rotation of the fan is abnormal or the fan has not been attached.
Alarm
Indicates a high temperature condition with the incubator air temperature exceeding 39˚C, and
the high temperature adjusting volume.
Setting
Indicates a setting value of the incubator air temperature.
Incubator air temperature
Indicates a present incubator air temperature and the connecting port of the incubator air temperature sensor cable.
Heater
Indicates a heater output.
Skin temperature
Indicates the infant’s skin temperature, the connecting port of the skin temperature probe and
the connecting port of the skin temperature monitor cable.
−4−
Page 7
Mark
Title
and Indication
POWER
Power switch
Indicates the power switch of the main body.
BREAKER
Breaker
Indicates the breaker of the main body.
12V
DC power connecting port
Indicates the DC power connecting port of the main body.
Lighting lamp
Indicates the lighting lamp cable connecting port.
Test switch
Indicates the test switch.
−5−
Page 8
WARRANTIES
This equipment is guaranteed by Atom Medical Corporation to be free of any defects in both materials and workmanship for a period of one year from the date of delivery when used normally for its intended purpose. Any parts
of this equipment proving to be so defective will be repaired or replaced at no charge during the warranty period.
In the following cases, however, actual expenses need to be paid even during the warranty period.
(1) Wear and tear of expendables
(2) Trouble and damage due to improper handling, such as dropping the unit during transport or transfer
(3) Trouble and damage due to a fire, salt, gas, extraordinar y voltage, earthquake, thunder, storm and
flood, or other natural calamities
(4) Travel expenses in the case of a trip to an isolated island, a remote place, etc. for the purpose of repair
Damage in shipment should be reported promptly to Atom accompanied by the certificate of the carrier concerned.
All correspondence concerning the equipment should specify the model name and the serial number.
WARNING
Atom is by no means held responsible for compensation for death, injury or damage to property if such loss
should occur due to any of the following causes.
1. Trouble or damage due to installation, relocation, maintenance or repair by someone other than personnel belonging to Atom or a sales representative or an installation company authorized by Atom.
2. Trouble or damage to Atom products caused by a product of another company other than that delivered
by Atom.
3. Trouble or damage due to modification, maintenance or repair using a part other than that specified by
Atom.
4. Trouble or damage due to neglecting the operating precautions or operating instructions described in
the Operation Manual of the unit.
5. Trouble or damage due to operation under ambient conditions, including electrical requirements and
installation requirements, other than those described in the Operation Manual.
6. Trouble due to careless or improper modification.
7. Trouble or damage due to using equipment sold or repaired by a secondhand goods dealer or a repair
shop without notifying Atom. Trouble or damage due to using equipment sold or repaired by the above
dealer or shop without following the instructions specified by Atom.
8. Trouble or damage due to using the unit connected to a peripheral device which is not specified by
Atom, which does not meet the applicable requirements of the international standards IEC-60601-1,
IEC-60601-1-1 and IEC-60601-1-2, or which is not equipped with equivalent safety features.
−6−
Page 9
[1] Precautions on Jamming
WARNING
Electric surgical knives, portable and mobile communication equipment, and other devices which
generate high-frequency noise can cause jamming to various electric equipment for medical use
and thus result in malfunction.
Since portable and mobile communication equipment, etc. are often used in medical facilities,
some measures should be taken to prevent jamming due to such devices.
Portable and mobile communication equipment and other devices which generate high frequency
should not be used near the unit during its operation to prevent malfunction of the unit due to
jamming.
[2] Responsibility for Care of Equipment
CAUTION
It is the user (a hospital, a doctor’s office, a clinic) that is responsible for the operation, maintenance and care of the electric equipment for medical use. The equipment should be used only by
medical personnel.
[3] Prohibition of Modification
WARNING
Do not disassemble or modify the unit. Otherwise a fire, an electric shock or injury may result.
[4] Periodical Inspection
CAUTION
Proper periodical inspection is needed to use the unit under optimum conditions.
[5] In Case of Trouble
CAUTION
If any abnormal condition or trouble should occur to the unit, indicate on the unit that it is out of
order and contact immediately your local Atom representative or service engineer. See this Operation Manual for whom to make contact.
If any abnormal condition or trouble should occur, do not use the unit until it has been repaired
completely by a service engineer so as to prevent possible danger.
−7−
Page 11
C
O
N
T
E
N
T
S
INTRODUCTION ............................................................................................. 1
[1] Operating Precautions
1-1.
DANGER .......................................................................................... 10
1-2.
WARNING ........................................................................................ 12
CAUTION ........................................................................................ 14
1-3.
[2] Parts Identification ...................................................................................... 15
[3] Setting up the Incubator
3-1. Assembling ............................................................................................. 21
3-2. Locking the Casters ............................................................................... 23
3-3. Adjusting the Height of the HL Stand ................................................. 23
3-4. Connecting the Power Pack to the Main Body .................................. 24
3-5. Where to Place the Incubator. .............................................................. 24
3-6. Fixing the Incubator to transport vehicle ........................................... 24
3-7. Power Outlet and Grounding ............................................................... 25
3-8. Power Cord and Power Switch ............................................................. 26
3-9. Inspection before Use ........................................................................... 27
[4] Power Source
4-1. Using the AC Power (AC220-240V) ..................................................... 28
4-2. Using the Battery Power ...................................................................... 29
4-3. Using the External DC Power (DC12V) ............................................. 30
[5] Incubator Air Temperature Control and Placing the Infant in the Incubator
5-1. Setting the Incubator Air Temperature ............................................... 31
5-2. Placing the Infant in the Incubator ...................................................... 33
5-3. Skin Temperature Monitoring ............................................................. 35
[6] Oxygen Supply ............................................................................................. 37
[7] Other Operating Procedures
7-1. Inserting the Cords and Tubes ............................................................ 39
7-2. Lighting Lamp ........................................................................................ 39
7-3. I.V. Pole (Option) ................................................................................... 39
7-4. Test Switch ............................................................................................. 40
7-5. Check Switch ......................................................................................... 40
7-6. Unlocking the Hood Lock (for opening the rear hood) .................... 40
[8] Cleaning and Disinfection
8-1. Skin Temperature Probe ....................................................................... 41
8-2. Hood ........................................................................................................ 42
8-3. Mattress Platform .................................................................................. 43
8-4. Replacing the Filter ............................................................................... 44
[9] Maintenance Inspection
9-1. Daily Inspection ..................................................................................... 45
9-2. Three-month Inspection ....................................................................... 46
9-3. Periodical Replacement Parts .............................................................. 46
9-4. Inspection Checklist .............................................................................. 47
9-5. Life ........................................................................................................... 48
9-6. Disposal .................................................................................................. 48
[10] Alarms and Alerts ........................................................................................ 49
[11] Troubleshooting ........................................................................................... 51
[12] Technical Data .............................................................................................. 52
Page 12
[1] Operating Precautions
Please follow the operating instructions described in this Manual for the safe use of the unit. The unit should
be operated only by those who have been trained and instructed properly in its operation. The unit should be
operated only for its intended use.
1-1.
DANGER
Death or serious injury, damage to equipment or a fire will result if the instructions given below are not followed.
Be sure to keep the front admittance panel and the access ports closed
when using the unit.
If the front admittance panel or an access port is left open when the unit is in use, the infant may fall
out of the baby compartment. Be sure to keep the front admittance panel and the access ports closed
during phototherapy also when it is performed on an infant placed in the incubator.
Do not leave the unit unattended when the front admittance panel or an
access port is open.
If the front admittance panel or an access port is left open, the infant may fall out of the baby compartment and even lose its life. Never leave the unit unattended when the front admittance panel or an
access port is open.
If the front admittance panel, a snap-open access port or a hook slider should
be found loose or faulty, stop using the unit immediately and ask for repair.
The infant may fall.
Always check the front admittance panel operating retainer and the hook
sliders of the snap-open access ports for proper operation. If the front admittance panel or a snap-open port should not close securely, don’t place
an infant in the unit, and ask for repair.
Do not open the front admittance panel or an access port in order to lower
the incubator air temperature.
Because the incubator air temperature is controlled automatically at a set temperature, the heater
output will increase dangerously to make up for any temperature drop.
Open and close the snap-open access ports gently.
If you open or close them roughly, they may not open or close properly, which is dangerous.
After the infant is placed in the incubator, check again that the front admittance panel and the access ports are all closed securely.
Do not fasten the infant fixing bands too tightly.
If they are fastened too tightly, they will press the infant dangerously.
Never place a body warmer or other possible ignition sources in or near the
unit.
Use of oxygen will increase a risk of explosion or a fire. A body warmer or other devices in which fire
is used or which will generate a spark may cause an explosion or a fire if used near the unit.
− 10 −
Page 13
Do not use the unit in the presence of a flammable anesthetic gas.
The unit may cause an explosion or a fire if used in the presence of such a gas.
Do not use ether, alcohol or other ignitable substances.
Even a small amount of ether, alcohol or other ignitable substances may cause a fire when mixed
with the oxygen in the incubator.
If water or any liquid should get into the unit, stop using the unit and request a repair service.
If the power supply should be short-circuited, it may cause the incubator to ignite or burn out. Chemical
fluids, having entered the unit, may stick to the incubator parts and cause abnormal operation of the
incubator.
Ground the unit securely.
Otherwise, a leakage current may cause an electric shock. In order to complete the ground connection, connect the power cord only to a properly grounded 3P power outlet including a ground terminal. Do not operate the unit if you have any doubt about its ground connection.
Do not use a device generating high frequency near the unit.
To prevent malfunction of the unit due to jamming, do not use electric surgical knives, portable and
mobile communication equipment and other devices which generate high frequency near the unit
during its operation.
Analyze arterial gas levels repeatedly when a high oxygen environment is
required.
It is reported to be extremely important and essential, when the infant’s conditions call for a high
oxygen environment, to repeatedly analyze arterial gas levels in order to maintain the oxygen concentration in the incubator at a desired level. Follow the doctor’s instructions in measuring the oxygen concentration because ignoring essential requirements may increase a risk of retinopathy of
prematurity and other adverse effects.
Do not give a shock to the unit or allow it to be struck by other objects.
The screws or fixed parts may become loose.
− 11 −
Page 14
1-2.
WARNING
Death or serious injury due to a fire or an electric shock will result if the instructions given below are not
followed.
Always maintain the incubator air temperature at a fixed level during the
standby.
Place the infant in the incubator only when the incubator air temperature
has stabilized.
Be sure to follow the doctor’s instructions in setting the incubator air temperature.
Be sure to follow the doctor’s instructions in setting the oxygen concentration in the incubator.
Be sure to use oxygen intended for medical use.
Be sure to bear in mind the following precautions during oxygen supply.
쎲 Do not place a body warmer, a flashlight, oils and fats, or flammable vaporizable matters in the
incubator.
쎲 Use pure cotton for the infant’s clothing, bed sheets, etc. Do not use any material that is easily
charged with static electricity.
쎲 Use pure cotton or fire-proof materials for the clothing of doctors, nurses and those ambulance
men who handle this unit.
Bear in mind the following precautions while using oxygen supply equipment.
쎲 If oil, grease or a grease-like substance should get in contact with pressurized oxygen, a violent
spontaneous ignition may occur. Do not let such substances stick to the oxygen pressure regulator, the oxygen cylinder valve, piping, connections and other parts of oxygen supply equipment.
쎲 On a high-pressure oxygen cylinder, use only a tested pressure reducing valve or pressure regulating valve indicated specifically for oxygen supply. Do not use such a valve for any gas other than
air or oxygen. It is dangerous to use a valve to supply a gas other than air or oxygen first and then
to supply oxygen again.
Smoking is prohibited in the room where the unit is installed. Do not place
any possible ignition sources in the room.
Avoid damaging the power cord.
A damaged power cord may cause a fire or an electric shock.
쎲 Do not pinch the power cord between the unit and the wall, a shelf or the floor.
쎲 Do not place the power cord near a heating apparatus or heat it.
쎲 Do not put anything heavy on the power cord.
쎲 Always grasp the power plug with your hand to remove the power cord from the power outlet.
A damaged power cord should be replaced immediately with a new one.
− 12 −
Page 15
Use only the power cord supplied with the unit.
Otherwise a fire or an electric shock may result.
Before cleaning or disinfecting the unit, be sure to turn off the unit, remove
the power plug and make sure that the incubator air temperature and the
heater temperature have sufficiently fallen.
Do not touch the power plug with a wet hand.
Touching the power plug with a wet hand may cause an electric shock.
Do not disassemble or modify the unit.
Disassembling or modifying the unit may cause a fire, an electric shock or injury.
Do not install the unit where it will be exposed to excessive humidity, dust
or steam.
Installing the unit in such a place may cause a fire or an electric shock.
The power outlet should be located near the unit to prevent accidental contact with a trailing power cord. Use a separate power outlet for each unit.
Do not put many loads on one power outlet.
In order to complete the ground connection, connect the power cord only to
a properly grounded 3P power outlet including a ground terminal.
Do not operate the unit if you have any doubt about its ground connection.
Ground peripheral electric equipment securely.
Electrical ratings of this unit are as follows:
q AC220-240V; power consumption 220VA; frequency 50/60Hz; operating voltage range AC198264V
w DC12V; 10A; operating voltage range 11.0-14.0V
The unit should not be connected to any other power source.
The unit should be serviced only by qualified personnel.
Be sure to inspect the unit before starting a day’s work.
Operating the unit without inspecting it before starting a day’s work may let a defect pass unnoticed
and cause a serious accident.
− 13 −
Page 16
1-3.
CAUTION
Injury or damage to ambient objects may result if the instructions given below are not followed.
Be sure to clean and disinfect the unit before using it for the first time after
purchase.
The unit is shipped without being disinfected.
When inserting a cord or a tube into the unit, be careful not to let it wind
around or squeeze the infant inside.
Do not twist or pull the cords by force.
If any defect should be found, ask an expert for repair without attempting to repair it yourself.
Remove the power plug from the power outlet before moving the unit to
another place or when the unit is not going to be used for a long time.
Moving the unit to another place with the power plug connected to the power outlet will damage the
power cord and may cause a fire or an electric shock.
Remove the power plug from the power outlet before cleaning or disinfecting the unit.
Cleaning or disinfecting the unit with the power plug connected to the power outlet may cause an
electric shock.
Place the unit on a stable surface.
Placing the unit on an unstable platform or a tilted surface will cause it to fall or drop and may hurt
someone. Check the strength of the place where the unit is to be placed or installed.
Install the unit out of reach of small children.
Do not install the unit in direct sunshine or near a heating apparatus.
Do not expose the unit to an extraordinarily high temperature or an excessive humidity.
Do not place anything heavy on the unit.
The unit uses rechargeable batteries (a lead-acid battery and a nickel metal
hydride battery). To dispose of the unit, remove the lead-acid battery and
the nickel metal hydride battery from the unit and send them to authorized
recycling facilities for recycling.
Check the operation of the peripheral devices.
If a device transmitting or receiving weak signals is installed near the unit, it may be affected by the
electromagnetic waves generated by the latter. Check the operation of peripheral devices for any
effect before using the unit in clinical settings. Stop using the unit immediately if any trouble is
detected.
When you transfer the incubator by holding the handles, make sure you do
so with the help of another person.
The power pack weighs about 13kg. Please use caution when handling it.
− 14 −
Page 17
[2] Parts Identification
2-1. Main body
e
r
t
q
y
w
u
i
o
!0
!1
!2
!3
!4
❉ The I.V pole (option) can be attached to either side of the incubator.
No.
q
w
e
r
t
y
u
No.
Name
Front admittance panel
Mattress platform (placed in the unit)
Front admittance panel packing
Lighting lamp
Front admittance panel retainer
I.V. pole (option)
Skin temperature probe connecting port
− 15 −
Name
i
o
!0
Oxygen supply port
Handle
Mattress (placed in the unit)
!1
!2
!3
!4
Snap-open access port
Control panel
Power panel
Hood fixing device
Page 18
2-2. Right side of the main body (the power supply)
POWER
q
BREAKER
15A
w
No.
q
w
e
12V
e
Name
Power switch
Breaker
DC power connecting port
2-3. Rear side of the main body
w
q
No.
q
w
e
e
Name
Filter
High temperature adjusting volume
Test switch
− 16 −
Page 19
2-4. Inside of the main body
q
w
e
r
t
y
u
No.
q
w
e
r
t
y
u
Name
Mattress
Hood packing
Mattress platform
Fan cover
Fan
Fan cover catch
Heater
− 17 −
Page 20
2-5. Stand
쎲 HL stand
q
w
e
No.
q
w
e
Name
Main body
HL stand
Power pack for V-707
쎲 Cabinet stand for internal hospital use
q
w
e
No.
q
w
e
❉ The HL stand, Cabinet stand for internal hospital use and
Oxygen gas cylinder in the figure are not accessories.
− 18 −
Name
Main body
Cabinet stand for internal hospital use
Power pack for V-707
Page 21
2-6. Operation panel
u
q
i
o
!0
w
!1
!2
e
!3
r
!4
!5
t
!6
y
!7
No.
q
w
e
r
t
y
u
i
o
No.
Name
!0
!1
!2
!3
!4
!5
!6
!7
Skin temperature display
Incubator air temperature display
Set temperature display
Heater output indicator
Incubator air temperature selector switch
Setting switch 왓•왕
Alarm silence/reset switch
Alarm silence indicator
System failure alarm indicator
− 19 −
Name
Power failure alarm indicator
Set point alarm indicator
Fan alarm indicator
High temperature alarm indicator
AC power indicator
Battery power indicator
External DC power indicator
Lighting switch
Page 22
2-7. Power pack for V-707
q
w
FUSE
e
r
t
No.
q
w
e
r
t
Name
Display panel
Fuse (for the battery)
Power pack cable
Power cord
AC power connecting port
2-8. Display panel (the power pack)
q
w
No.
q
w
e
e
Name
Battery charge level indicator
Power (charging) indicator
Check switch
− 20 −
Page 23
[3] Setting up the Incubator
3-1. Assembling
Assemble the separate components of the incubator correctly. If you use the HL stand or the Cabinet stand for
internal hospital use with the incubator, attach it correctly
to the incubator as instructed below.
CAUTION
The incubator is heavy. At least two able
people are required to assemble or move the
incubator.
3-1-1. HL stand
(1)
Place the incubator on the HL stand in such a way
that the front side of the incubator and the front side
of the HL stand will face in the same direction. At
this time, be sure to set the rubber feet attached to
the four corners of the incubator bottom securely
on the top surface of the stand so that the incubator
will not tilt or wobble.
(2)
Put the main body fixing bands around the handle
of the incubator and the upper frame of the stand on
both sides of the incubator as illustrated and fasten
each fixing band securely.
(3)
Main body
fixing band
Place the power pack on the power pack holder at
the bottom of the stand. The stand is also equipped
with oxygen cylinder holders on the front side and
the rear side at the bottom. Two 500L cylinders can
be laid in the cylinder holders when needed.
Power pack
WARNING
Cylinder holder
Be sure to put together the incubator and
the stand securely with the main body fixing
bands. Otherwise the incubator may tip over
when force is applied to open the hood, or it
may fall down when a shock is given during
transport. Never operate the incubator before it is fixed securely to the stand.
− 21 −
Page 24
3-1-2. Cabinet stand for internal hospital use
(1)
Place the incubator on the Cabinet stand for internal hospital use in such a way that the front side of
the incubator and the front side of the stand will face
in the same direction. In doing so, be sure to set the
rubber feet attached to the four corners of the incubator bottom securely on the top surface of the stand
so that the incubator will not tilt or wobble.
(2)
Put the main body fixing band around the handle of
the incubator as illustrated and fasten the band securely by means of the fixing device on either side
of the stand.
Main body
fixing band
Fixing device
(3)
Place the power pack on the lower right surface of
the stand base and fix it securely in place with the
fixing band provided for the purpose. When needed,
two 500L cylinders can be laid on the lower left surface of the stand base. Fix the cylinders securely in
place with the fixing band provided for the purpose.
WARNING
Be sure to put together the incubator and
the stand securely with the main body fixing
bands. Otherwise the incubator may tip over
when force is applied to open the hood, or it
may fall down when a shock is given during
transport. Never operate the incubator before it is fixed securely to the stand.
− 22 −
Power pack
Page 25
3-2. Locking the Casters
The casters of the HL stand and the cabinet stand for internal hospital use are provided with stoppers. To use
the incubator on the HL stand and the cabinet stand for internal hospital use, take the following steps.
(1)
Install the incubator in a desired place and step on
the stoppers on the casters of the stand to lock them.
(2)
To move the incubator to another place, unlock the
stoppers of the casters.
Lock
CAUTION
Install the HL stand and the cabinet stand for
internal hospital use on a horizontal and stable
place. Lock the stoppers on the casters to fix
the stand securely. To move the incubator to
another place, be sure to unlock the casters.
Stopper
Lock
Cabinet stand for
internal hospital use
Stopper
3-3. Adjusting the Height of the HL Stand
The height of the HL stand is adjustable freely within the range of 48~80cm. Take the following steps to adjust
the height.
(1)
Check that the incubator is fixed securely to the
stand with the main body fixing bands.
(2)
Hold the upper frame with both hands and pull up
the operating levers toward the upper frame. The
stand will be unlocked and can be moved up or down.
(3)
Release the lever when a desired height is achieved,
and the stand will be locked at that height.
Unlock the
stand
Operating lever
CAUTION
The stand is provided with a spring for elevation. Be sure to adjust the height of the stand
with the incubator fixed to the stand. For
safety’s sake, two people should cooperate in
adjusting the height.
− 23 −
Page 26
3-4. Connecting the Power Pack to the Main Body
쐽 Insert the connector of the power pack cable into the
DC power connecting port on the side of the main body
and turn the ring to attach the connecting cord securely.
DC power connecting port
Power pack cable
Ring
3-5. Where to Place the Incubator.
Except during transport, place the incubator on a horizontal surface convenient for work. Avoid placing the
incubator near a heating apparatus, by the window or where fire is used.
CAUTION
Avoid placing the incubator in direct sunshine, near a stove or a radiator, in the direct airflow of an
air conditioner, or by a cold window. The incubator will be affected directly by these external thermal conditions.
3-6. Fixing the Incubator to transport vehicle
The method to fix the incubator to an ambulance varies with the type of ambulance. Consult the institutions
concerned (hospitals, the fire station) about the fixing method.
WARNING
Do not use the HL stand and the Cabinet stand for internal hospital use in an ambulance.
− 24 −