BARD
Sherlock 3cg Tip Confirmation System
Sherlock 3cg Tip Confirmation System for use with Site-Rite Instructions for Use Rev Date April 2012
Instructions for Use
28 Pages
Preview
Page 1
For use with Site~Rite Vision* Ultrasound System Instructions for Use
INSTRUCTIONS FOR USE
Sherlock 3CG* Tip Confirmation System
Keep Dry
Fragile
Manufacturer
CF Type Equipment
Attention, see instructions for use
Storage humidity limitation
Operating humidity limitation
40o C -18 C o
38o C 10 C o
Rx Only
1
Operating temperature limitation
Federal (U.S.A.) law restricts this device to sale by or on the order of a physician
Storage temperature limitation
Do not dispose with ordinary municipal waste
Instructions For Use
- TABLE OF CONTENTS Section 1
Section
Overview 3
5
Magnetic Navigation Only Mode
1.1 Indications for Use
Step 1: Prepare Device
1.2 Post Market Clinical Trial
Step 2: Position Patient and Perform Ultrasound Pre-scan
1.3 Sherlock 3CG* TCS Description
Step 3: Measure Catheter Length
1.4 System Components
Step 4: Prepare Sensor
1.5 Warnings and Precautions
Step 5: Position Sensor
Step 6: Perform Initial Magnet Tracking Calibration
Step 7: Prepare Catheter Sterile Field
Step 8: Access the Vein
Step 9: Perform Final Magnet Tracking Calibration
Step 10: Insert Catheter
Step 11: Catheter Tip Guidance and Positioning
2
Assembling the Sherlock 3CG* TCS
2.1 Attaching the Sherlock* 3CG TCS Sensor Holster to the Roll Stand
2.2 Connecting the Sherlock 3CG* TCS Sensor to the Site~Rite Vision* Ultrasound System
3
Sherlock 3CG* TCS Information
3.1 Switching Between Ultrasound and TCS on 3CG
3.2 Sherlock 3CG* TCS Graphical Interface
3.3 Sherlock 3CG* TCS Controls and Indicators
5
6
6
3.4 Sherlock 3CG* TCS Audio Information
4
3.5 Parallax
Sherlock 3CG* TCS Catheter Guidance
4.1 Verifying Catheter Compatibility
4.2 Catheter Placement
Step 1: Prepare Device
Step 2: Position Patient and Perform Ultrasound Pre-scan
Step 3: Measure Catheter Length
Step 4: Prepare Sensor
Step 5: Position Sensor and ECG electrodes
Step 6: Evaluate External ECG waveform
Step 7: Perform Initial Magnet Tracking Calibration
Step 8: Prepare Catheter Sterile Field
Step 9: Access the Vein
Step 10: Attach Catheter Stylet to Fin Assembly
Step 11: Perform Final Magnet Tracking Calibration
Step 12: Insert Catheter
Step 13: Catheter Tip Guidance and Positioning
Step 14: Complete Catheter Placement
Step 15: Procedural Record
15
Troubleshooting and Error Messages
6.1 Error Screens
6.2 ECG Troubleshooting
6.3 Magnetic Navigation Troubleshooting
17
19
7
Magnetic Tracking Printing
23
8
File Management
24
9
Cleaning and Disinfection
24
9.1 Cleaning Procedure
9.2 Disinfection Procedure
10
Warranty 24
11
Service and Repair
25
12
Technical Specifications
25
13
11.1 Sherlock 3CG* TCS Sensor and Display Operational and Storage Conditions
Disposal Information
25
2
Sherlock 3CG* Tip Confirmation System
1 Overview 1.1 Indications for Use The Sherlock 3CG* Tip Confirmation System (TCS) is indicated for guidance and positioning of Peripherally Inserted Central Catheters (PICCs). The Sherlock 3CG* TCS provides real-time PICC tip location information by using passive magnet tracking and the patient's cardiac electrical activity (ECG). When relying on the patient’s ECG signal, the Sherlock 3CG* TCS is indicated for use as an alternative method to chest X-ray and fluoroscopy for PICC tip placement confirmation in adult patients. Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P wave as in atrial fibrillation, atrial flutter, severe tachycardia, and pacemaker driven rhythm. In such patients, who are easily identifiable prior to PICC insertion, the use of an additional method is required to confirm PICC tip location.
1.2 Post Market Clinical Trial The Sherlock 3CG* Tip Confirmation System (TCS) is Bard’s next generation, fully integrated magnetic tracking and ECG-based peripherally inserted central catheter (PICC) tip confirmation technology, which represents the next evolution of the Sherlock* II Tip Location System and the previouslymarketed Sapiens Tip Confirmation System. The Sapiens TCS Post-Market Clinical trial was a prospective, single arm, single center study designed to assess the efficacy of the ECG method for correctly positioning the tip of catheters in adult patients. The primary endpoints were to assess the performance of the Sapiens technology with respect to: (1) the accuracy of the Sapiens technology with regard to correct positioning of the catheter tip when compared to postoperative chest x-ray, (2) safety of using an invasive intracardiac electrode, and (3) compatibility with peripherally inserted central catheters. One hundred fourteen (114) subjects received a PICC. Final PICC tip location was confirmed at the cavoatrial junction or within +/- 1 cm in 99.1% (113/114) of the subjects. No adverse events were reported.
1.3 Sherlock 3CG* TCS Description The Sherlock 3CG* TCS is designed to aid in PICC tip positioning through magnet tracking and ECG technology. It is designed to operate with Bard Access Systems’ catheter kits labeled [
] and “Sherlock 3CG* TPS Stylet”.
Note: When used in conjunction with catheter kits labeled [ through ECG technology.
] the device provides magnetic navigation information but does not allow positioning
ECG Positioning The Sherlock 3CG* TCS displays an ECG signal detected by the intravascular and body electrodes, which can be used for catheter tip positioning. In patients with a distinct P-wave, the P-wave will increase in amplitude as the catheter approaches the top of the cavo-atrial junction. As the catheter advances into the right atrium, the P-wave will decrease in amplitude and may be biphasic or invert.
Magnetic Navigation Permanent magnets are encapsulated within the tip of the Sherlock 3CG* TPS Stylet. No magnetic energy is generated by the Site~Rite Vision* Ultrasound System or the sensor. The Sherlock 3CG* TCS displays the relative position of the magnet-tipped stylet to the sensor. It does this in two steps: 1. Sherlock 3CG* TCS takes a background measurement of the ambient magnetic field during the calibration cycle. 2. Sherlock 3CG* TCS senses changes in the magnetic field. When the Sherlock 3CG* TCS detects the stylet, it displays the stylet tip location and orientation.
3
Instructions For Use
1.4 System Components Authorized accessories include: - Site~Rite Vision* Ultrasound System - Sherlock 3CG* TCS Sensor - Sherlock 3CG* TPS Stylet (included with specially marked central venous catheter kits [ - Sherlock* TLS Stylet (included with specially marked central venous catheter kits [
]
]
- Fin Assembly (included with specially marked central venous catheter kits) - Sherlock* Sensor Holder (ordered separately or with specially marked central venous catheter kits)
or
.
- Sherlock* 3CG TCS Sensor Holster - Bard Access Systems' supplied printer (purchased seperately)
1.5 Warnings and Precautions This section specifies warnings and precautions specific to the functionality of the Sherlock 3CG* TCS. - See the Bard Access Systems' catheter Instructions for Use (IFU) for possible complications associated with peripherally inserted central catheter (PICC) placements and ECG positioning.
Warnings Warning: This product should only be operated by qualified medical personnel. Warning: Do not power the Sherlock 3CG* TCS in the presence of flammable anesthetic gases. Explosion may result. Warning: Do not attempt to sterilize the sensor. Damage to the equipment may occur. Warning: The following actions void the warranty of the Sherlock 3CG* TCS and may result in injury or equipment damage. - Opening or servicing the Site~Rite Vision* Ultrasound System with Sherlock 3CG* TCS by anyone other than Bard Access Systems’ authorized service personnel. - Removing system labels by anyone other than Bard Access Systems’ authorized service personnel. - Connecting the sensor or applied patient components to any unauthorized system or accessory. Refer to Section 1.4 for complete components. - Installation of unauthorized software. - Modification of system software settings without authorization by Bard Access Systems. Warning: If the Site~Rite Vision* Ultrasound System with Sherlock 3CG* TCS is visibly damaged, discontinue use immediately. Use of the damaged system may result in injury or equipment damage. Warning: Do not submerge the sensor in liquid or allow fluid to enter the connectors. Damage to the equipment may occur. Warning: Sherlock 3CG* TCS is not intended to diagnose or treat disease. Warning: Only Bard Access Systems’ authorized service personnel should attempt to service this equipment. Site~Rite Vision* Ultrasound System contains static sensitive components and circuits. Failure to observe proper static control procedures may result in damage to the system. Warning: Do not rely on ECG signal detection for catheter tip positioning when interpretation of the external or intravascular ECG P-wave is difficult. For example, when: - P-wave is not present - P-wave is not identifiable - P-wave is intermittent These conditions may be a result of heart rhythm abnormalities, atrial fibrillation, atrial flutter, severe tachycardia or presence of cardiac rhythm devices. In these cases, rely on magnetic navigation and external measurement for tip positioning and use chest x-ray or fluoroscopy to confirm catheter tip location, as indicated by institutional guidelines and clinical judgment. Warning: Do not rely on ECG signal detection for catheter tip positioning when there are no observable changes in the intravascular P-wave. In this case, rely on magnetic navigation and external measurement for tip positioning and use chest X-ray or fluoroscopy to confirm catheter tip location, as indicated by institutional guidelines and clinical judgment. Warning: Do not place and/or use the Sherlock 3CG* TCS in the presence of strong magnetic fields such as Magnetic Resonance Imaging (MRI) devices. The high magnetic fields created by an MRI device will attract the equipment with a force sufficient to cause death or serious injury to persons between the equipment and the MRI device. This magnetic attraction may also damage the equipment. The magnetic and the RF fields associated with the MRI environment may interfere with the display of ECG waveforms. Consult the MRI manufacturer for more information. Warning: Do not remove Sherlock 3CG* TCS enclosures. To avoid electrical shock, use only the power cord supplied with the system and connect only to properly grounded wall outlets. Only Bard Access Systems qualified personnel should service the system
4
Sherlock 3CG* Tip Confirmation System
Warning: Ensure all connecting cables and connections are electrically insulated and do not come into contact with other electrical cables or metal surfaces. Warning: Ensure that the patient does not directly or indirectly contact non-insulated metal surfaces. Warning: Place skin electrodes carefully at locations indicated in these Instructions for Use and ensure good skin-electrode contact. Failure to do so may cause unstable ECG waveforms and/or ECG waveforms that are not described in these Instructions for Use. In such a case, rely on magnetic navigation and external measurement for tip positioning and use chest X-ray or fluoroscopy to confirm catheter tip location, as indicated by the institutional guidelines and clinical judgment.
Precautions Caution: Federal (U.S.A) law restricts this device to sale by or on the order of a physician. Caution: Do not pull the cables to disconnect from the system. Pulling the cable may damage the cable or cable connection. Caution: Excessive twisting or bending of the sensor cable may cause system failure. Caution: Use only Bard Access Systems’ cleaning and disinfection procedures. Failure to do so may damage the device. Caution: Do not use excessive force when connecting or disconnecting the Fin Assembly to or from the sensor or equipment damage may occur. Caution: When the sensor is not in use, store in the holster, roll stand basket or other secure location to avoid damage. Caution: Do not allow any ferromagnetic objects, e.g., wired undergarments, metal instruments, watches, jewelry, electronics, metal bedrails, etc. to be
within 12 in (30 cm) of the sensor once the calibration process is complete. These items may interfere with the sensor's ability to accurately locate the Sherlock* Stylet tip.
Caution: Equipment operating in close proximity may emit strong electromagnetic or radio frequency interference which could affect the performance of this device. Avoid operating the device near pumps, cauterizers, diathermy equipment, cellular phones, or other portable and mobile radio frequency communications equipment. Maintain equipment separation of at least 5 ft (1.5 m).
Caution: Electrodes should be applied only to intact, clean skin (e.g., not over open wounds, lesions, infected or inflamed areas). Caution: Placement of red electrode outside of the correct region may result in reduced ECG performance. See Section 4. Caution: Discontinue electrode use immediately, if skin irritation occurs. Caution: To avoid damage to the device operating system, shutdown the device through the Power Control button from the Application Toolbar.
2 Assembling the Sherlock 3CG* TCS 2.1 Attaching the Sherlock 3CG* TCS Sensor Holster to the Roll Stand
The Sherlock 3CG* TCS Sensor can be placed in the holster when not in use. To attach the sensor holster to the roll stand see illustrations below.
Caution: When the sensor is not in use, store in the holster, roll stand basket or other secure location to avoid damage. 2.2 Connecting the Sherlock 3CG* TCS Sensor to the Site~Rite Vision* Ultrasound System
The Sherlock 3CG* TCS Sensor connects to the USB port on the display.
5
Instructions For Use
3 Sherlock 3CG* TCS Information 3.1 Switching Between Ultrasound and TCS To switch from ultrasound to TCS mode, select [
] by:
1. Pressing the button on the touch screen. 2. Clicking on the button with the mouse pointer. 3. Navigating to the button using the arrow buttons on the probe and pressing the enter button.
To switch from TCS mode to ultrasound, select [
] via the touch screen, mouse pointer, or buttons on the probe.
Note: For an explanation of the other controls on the above toolbar, refer to the Site~Rite Vision* Ultrasound System Hardware and Operating System Instructions for Use.
3.2 Sherlock 3CG* TCS Graphical Interface Intravascular ECG
External (or baseline) ECG
Maximized magnet tracking window.
Magnet tracking requires calibration.
6
Sherlock 3CG* Tip Confirmation System
SOFTWARE VERSION
ECG MENU
ECG Display Adjustment
PATIENT INFORMATION Patient Information will be shown on the procedural record.
SITE~RITE VISION* ULTRASOUND SYSTEM APPLICATION TOOLBAR
7
FREEZE
MENU: Select to open or close the ECG options menu.
Instructions For Use
SEARCHING FOR SIGNAL ICON Stylet tip had not been detected. DEPTH ICON Relative stylet depth under sensor.
OFF SCREEN ICON Location of the approaching stylet tip
STYLET ICON Location and direction of stylet tip under the sensor. The circle represents the tip of the stylet and the tail represents the body of the stylet.
MAIN SCREEN
REFERENCE SCREEN
Real-time external and intravascular ECG waveforms.
Used for freezing the external and intravascular ECG waveforms to assist in maximum P-Wave identification.
Minimized magnet tracking window.
8
Sherlock 3CG* Tip Confirmation System
3.3 Sherlock 3CG* TCS Controls and Indicators
Front Panel Controls/Indicators Freeze: Select to copy the current ECG waveforms from the Main Screen to the Reference Screen.
Print to file: Select to save and print the current ECG waveforms in the reference screen. (Saves to both the hard-drive and USB external storage device if connected).
Menu: Select to open/close the ECG menu.
Magnet Tracking Controls
Calibrate: Select to calibrate magnet tracking.
Audio: : Select to toggle sound on or off.
9
Instructions For Use
ECG Menu Controls
Save: Select to stop and save the ECG waveform recording.
Demo: Select to initiate demonstration mode. For training, contact your Bard Access Systems’ Sales Representative.
Channel Channel Selection: Select either the external (white) or intravascular (yellow) ECG waves for manipulation.
Scale Channel Scale: Select to adjust the ECG signal amplitude of the external (white) or intravascular (yellow) ECG waveforms on the Main Screen.
Level Channel Level: Select to adjust the vertical position of the external (white) or intravascular (yellow) ECG waveforms on the Main Screen.
Speed Display Speed: Select to adjust the speed of the ECG playback.
Close: Closes the ECG menu
-
10
Sherlock 3CG* Tip Confirmation System
3.4 Sherlock 3CG* TCS Audio Information When the audio is on, there are three possible tones: - Tone 1: stylet tip detected and the tip is not under the sensor. - Tone 2: stylet tip is under the sensor and at or above the Bard logo. - Tone 3: stylet tip is under the sensor and below the Bard logo.
1 2
3
Note: Audio signals are only applicable to the magnetic navigation capability of the device.
3.5 Parallax When a patient's chest is not flat, the sensor will rest at an angle, causing an effect known as parallax. Parallax is the difference in the apparent location of an object from two different points of view. The difference between the point of view of the sensor and the user can be several centimeters.
Note: Parallax should only be considered in relation to magnetic navigation.
11
Instructions For Use
4 Sherlock 3CG* TCS Catheter Guidance
4.1 Verifying Catheter Compatibility - To verify or change Sherlock 3CG* TCS catheter settings, select the gear icon highlighted below:
- Under the general tab in the configuration window, go to PICC Compatibility Options. - Select the appropriate box for catheter tracking purposes. - Prior to closing the menu, select Accept Changes or Reject Changes.
12
Sherlock 3CG* Tip Confirmation System
4.2 Catheter Placement Step 1: Prepare Device - Connect the sensor to the Site~Rite Vision* Ultrasound System via USB cord. - Verify that the ECG flatline signal is scrolling. - Verify the battery charge is sufficient for the procedure. - Enter Patient Information as needed.
Step 2: Position Patient and Perform Ultrasound Pre-scan - Refer to the Bard Access Systems’ Site~Rite* Ultrasound System Instructions for Use.
Step 3: Measure Catheter Length - Refer to Bard Access Systems’ Catheter Instructions for Use.
Step 4: Prepare Sensor - Slide the Fin Assembly onto the sensor until fully seated.
- Place sensor in the sensor holder with the fin and ECG electrodes remaining outside the holder and tighten the cinch ring.
13
[1]
[2]
[3]
[4]
Instructions For Use
Step 5: Position Sensor and ECG electrodes - Remove the adhesive backing from the sensor holder and place the sensor directly on the patient's skin with the adhesive side down. Place the sensor as flat as possible for best results. Note: The sensor should be positioned the same for left or right side placements. - Prepare and attach external ECG electrodes per the following steps: •
Ensure electrode locations are oil-free and completely dry.
Caution: Electrodes should be applied only to intact, clean skin (e.g., not over open wounds, lesions, infected or inflamed areas). -
Attach electrodes to all lead wires.
-
Remove backing and press electrodes firmly onto skin at the specified locations. •
Place black electrode on patient’s lower right shoulder.
•
Place red lead on patient’s lower left side, inferior to the umbilicus and laterally along the mid-axillary line.
Caution: Placement of red electrode outside of this region may result in reduced ECG performance.
Place high on the patient’s chest -touching the neck, if possible. Bard logo face up.
Cord routed toward the patient’s feet.
Warning: Place skin electrodes carefully at locations indicated in these Instructions for Use and ensure good skin-electrode contact. Failure to do so may cause unstable ECG waveforms and/or ECG waveforms that are not described in these Instructions for Use. In such a case, rely on magnetic navigation and external measurement for tip positioning and use chest X-ray or fluoroscopy to confirm catheter tip location, as indicated by the institutional guidelines and clinical judgment. Caution: Discontinue electrode use immediately, if skin irritation occurs. Tips: - Prior to securing the sensor holder to the patient, it may be necessary to clean the skin and remove excess hair. - Do not move the sensor after it is secure. Best results will be achieved if the patient remains still and the sensor is not placed on open wounds, over bandages, drapes, gowns or other coverings.
14
Sherlock 3CG* Tip Confirmation System
Step 6: Evaluate External ECG waveform - Refer to the Bard Access Systems’ catheter Instructions for Use.
Step 7: Perform Initial Magnet Tracking Calibration - Calibrate the Sherlock 3CG* TCS by selecting CALIBRATE [
] prior to setting up the sterile field to ensure there is no environmental interference.
Tip: If calibration fails, remove any items that may be causing magnetic interference (e.g., active motor driven equipment, monitor leads, cell phones, name tags, jewelry, etc.).
Step 8: Prepare Catheter Sterile Field - Refer to the Bard Access Systems’ catheter Instructions for Use.
Step 9: Access the Vein - Refer to the Bard Access Systems’ Site~Rite* Ultrasound System Instructions and Bard Access Systems’ catheter Instructions for Use.
Step 10: Attach Catheter Stylet to Fin Assembly - Refer to Bard Access Systems’ catheter Instructions for Use.
Step 11: Perform Final Magnet Tracking Calibration - Ensure the catheter tip is at least 12 inches (30 cm) away from the sensor before calibrating. - Select CALIBRATE [
] immediately prior to catheter insertion.
- Once calibration is complete, ask the patient to remain still and do not reposition the patient. - Refer to Bard Access Systems' catheter instructions for use for catheter insertion.
Step 12: Insert Catheter - Refer to Bard Access Systems’ catheter Instructions for Use for catheter insertion.
Step 13: Catheter Tip Guidance and Positioning - Refer to Bard Access Systems’ catheter Instructions for Use for catheter insertion. - Initially a searching magnifying glass will indicate that the stylet tip is outside the sensor range. - Use a slow steady motion while advancing the catheter. Magnetic Navigation and Catheter Tip Tracking - As the stylet tip approaches the sensor, an icon appears at the edge of the screen indicating the approach of the stylet tip. - When the stylet is under the sensor, the stylet and depth icons will display the location, orientation, and depth of the stylet in relation to the sensor. - Advance the catheter slowly to achieve optimal performance (1 cm per second). There may be a slight delay from the time the catheter is moved until the stylet icon moves on the display. Advancing the catheter too quickly may result in erratic movements of the stylet icon on the display. - Insert the catheter until the magnetic navigation shows the stylet moving consistently downward. - Continue to slowly advance catheter until the catheter is inserted to the external measurement as determined in the Bard Access Systems’ catheter instructions for use. - Select [
] to minimize the magnetic navigation window and freeze / save the current ECG waveforms in the reference screen.
Note: select the minimized magnetic navigation window to return to a maximized state ECG Positioning - In patients with a distinct P-wave, the P-wave will increase in amplitude as the catheter approaches the cavo-atrial junction. As the catheter advances into the right atrium, the P-wave will decrease in amplitude and may become biphasic or inverted. Note: If the intravascular ECG waveform is not displayed, flush the catheter with saline. If the problem continues, check the stylet-to-fin connection. To freeze and compare ECG waveforms select [
] to copy ECG waveforms in the reference window.
Note: Pause to let the rhythm settle before freezing the reference ECG waveforms. - For final catheter positioning, refer to Bard Access Systems' catheter instructions for use.
Step 14: Complete Catheter Placement - Complete catheter insertion, securement and remaining procedure according to the catheter Instructions for Use and facility protocol.
15
Instructions For Use
Step 15: Procedural Record - Select [ Note: Selecting [
] to print / save a procedural record. ] will send the procedural record to a storage device and approved printer, if connected. A typical printout is shown below.
16
Sherlock 3CG* Tip Confirmation System
5 Magnetic Navigation Only Mode Step 1: Prepare Device - Connect the sensor to the Site~Rite Vision* Ultrasound System via USB cord. - Verify the battery charge is sufficient for the procedure. - Enter Patient Information as needed.
Step 2: Position Patient and Perform Ultrasound Pre-scan - Refer to the Bard Access Systems’ Site~Rite* Ultrasound System Instructions for Use.
Step 3: Measure Catheter Length - Refer to Bard Access Systems’ Catheter Instructions for Use.
Step 4: Prepare Sensor -Place the Sherlock 3CG* Sensor in the sensor holder and tighten the cinch ring
1]
2]
3]
4]
Step 5: Position Sensor - Remove the adhesive backing from the sensor holder and place the sensor directly on the patient's skin with the adhesive side down. Place the sensor as flat as possible for best results. Note: The sensor should be positioned the same for left or right side placements.
Place high on the patient's chest touching the neck if possible. Bard logo face up.
Cord routed towards the patient's feet.
17
Instructions For Use
Step 6: Perform Initial Magnet Tracking Calibration - Calibrate the Sherlock 3CG* TCS by selecting CALIBRATE [
] prior to setting up the sterile field to ensure there is no environmental interference.
Tip: If calibration fails, remove any items that may be causing magnetic interference (e.g. active motor driven equipment, monitor leads, cell phones, name tags, jewelry, etc.).
Step 7: Prepare Catheter Sterile Field Refer to the Bard Access Systems’ catheter Instructions for Use.
Step 8: Access the Vein - Refer to the Bard Access Systems’ Site~Rite* Ultrasound System Instructions and Bard Access Systems’ catheter Instructions for Use.
Step 9: Perform Final Magnet Tracking Calibration - Ensure the catheter tip is at least 12 inches (30 cm) away from the sensor before calibrating. - Select CALIBRATE [
] immediately prior to catheter insertion.
- Once calibration is complete, ask the patient to remain still and do not reposition the patient. - Refer to Bard Access Systems' catheter instructions for use for catheter insertion.
Step 10: Insert Catheter - Refer to Bard Access Systems’ catheter Instructions for Use for catheter insertion.
Step 11: Catheter Tip Guidance and Positioning - Refer to Bard Access Systems’ catheter Instructions for Use for catheter insertion. - Initially a searching magnifying glass will indicate that the stylet tip is outside the sensor range. - Use a slow steady motion while advancing the catheter. Magnetic Navigation - As the stylet tip approaches the sensor, an icon appears at the edge of the screen indicating the approach of the stylet tip. - When the stylet is under the sensor, the stylet and depth icons will display the location, orientation, and depth of the stylet in relation to the sensor. - Advance the catheter slowly to achieve optimal performance (1 cm per second). There may be a slight delay from the time the catheter is moved until the stylet icon moves on the display. Advancing the catheter too quickly may result in erratic movements of the stylet icon on the display. - Insert the catheter until the magnetic navigation shows the stylet moving consistently downward. - Continue to slowly advance catheter until the catheter is inserted to the external measurement as determined in the Bard Access Systems’ catheter instructions for use. - Below is an image of Magnetic Navigation only mode.
18
Sherlock 3CG* Tip Confirmation System
6 Trouble Shooting and Error Messages 6.1 Error Screens Sensor Disconnected
Sensor Error
Cause:
Sensor not connected to the display.
Solution:
Ensure the sensor is properly connected. Disconnect and reconnect the sensor.
Cause:
The display power supply is connected to a noisy electrical outlet.
Solution:
Disconnect the display power supply from the electrical outlet to run the system on battery power.
Cause:
Sensor cannot be detected by the display.
Solution:
Call the technical support hotline (800) 443-3385.
Cause:
Interference caused by magnetic field changes or by sensor movement.
Solution:
1. Move all metal objects at least 12 inches (30 cm) away from the sensor. 2. Lower bed rails. 3. Move all active electronic devices at least 5 feet (1.5 m) away. 4. Do not move the sensor and ask the patient to remain still. 5. Calibrate. 6. If the error continues, discontinue use and call the technical support hotline (800) 443-3385.
Cause:
Sensor malfunction.
Solution:
1. Turn off and restart the system. 2. If the error continues, discontinue use and call the technical support hotline (800) 443-3385.
Magnet Error
Cause:
Interference caused by magnetic field changes or by sensor movement.
Solution:
1. Move all metal objects at least 12 inches (30 cm) away from the sensor. 2. Lower bed rails. 3. Move all active electronic devices at least 5 feet (1.5m) away. 4. Do not move the sensor and ask the patient to remain still. 5. Ensure catheter tip is at least 12 inches (30 cm) away from the sensor. 6. Calibrate. 7. If the error continues, discontinue use and call the technical support hotline (800) 443-3385.
19