Checklist
1 Page
Preview
Page 1
LUCAS® 3 Service Checklist Work Order #:
Type of inspection:
Customer:
Annual
LUCAS Serial #:
Post-Repair
1. 2. 3.
Exterior Physical Inspection Device exterior condition (cleaned – no exterior damage) Check Bellows and Suction Cup Check Fabrics (Patient straps, Stabilization strap and carrying bag)
1.
Maintenance Procedure Perform if necessary Software version (Fill in version/checksum below OR add print-out) CPU: Main SW: • • • •
2. 3. 4. 5. 6.
Control Protective Charger COM Board App
Clean Electric Fan and Mesh Grill Lubricate Carry Ball Screw Lubricate Claw Lock Mechanism Test run Attach Service Label (optional)
Pass Fail NA
Complete Complete Complete Complete Complete
Pass Fail NA
“Transmit” mode •
Confirm that Bluetooth connection to the PC is working
•
Last Service Date is set to PC Date for PM service
Remarks:
LUCAS® 3 Chest Compression System – Service PIP Checklist 3326789 Rev C, © 2017 Physio-Control, Inc.
Print-out added Checksum:
Remarks
Remarks
Electronics Tests: 1. Internal Function Test and Adjust LED is green 2. “30:2” mode • Device runs without unusual sound. • Confirm active 30:2 LED is green. • Confirm audible alert prior to ventilation pause and intermittent LED 3. Check that PAUSE LED is green. 4. “Continuous” mode • Confirm continuous LED is green. • Confirm LED blinks every 10 strokes. 5. “Adjust” mode • Confirm that Suction Cup runs smoothly. 6. “Charge” mode • Connect a PSU and confirm that the charging sequence starts.
Performed by:
Remarks
J J J J
Fu Function Check Mechanics Tests: 1. Check that the Decompression Spring moves smoothly and without unusual noise. 2. Check that the Carry Ball Nut runs smoothly and without unusual noise. 3. Check for Release Ring play when Claw Mechanism is unlocked. 4. Check that Claw Mechanism remains locked when pressure is applied.
7.
Routine
Date:
Remarks