BayPort
TM
6-Bay Charger Operating Manual
Table of Content
Symbol Description...2 Warnings and Cautions...3 Warnings and Cautions...4
Intended Use... 4 Unintended Uses... 4
Overview...5
System Description... 5
Operation...6
Charging Batteries... 6
Cleaning and Maintenance...7
Charging Unit... 7
Specifications...8 Electromagnetic Compatibility...9 Troubleshooting and Service...13
Symbol Description These important symbols may appear on your Long Island Technology GroupTM BayPortTM 6-Bay Charger. Please note their meaning.
CE marking of conformity indicates this device complies with European Directive 93/42 EC on medical devices and Directive 2002/95/EC on the restriction of the use of certain hazardous substances in electrical and electronic equipment.
The UL mark indicates this product has been tested to, and conforms to applicable standards.
More information for this product can be found in this operating manual.
AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IEC 60601-1 Edition 3.1, IEC 60601-1-2 Edition 4, ANSI/AAMI E-S60601-1 (2005), CAN/CSA-C22.2 No. 60601-1 (2008)
2
Warnings and Cautions Users of this product should be thoroughly trained in the appropriate medical procedures. Also they should read and understand this owner’s manual for this product and all equipment used with it. No other training is required to use this device.
!! !! !! !! !! !! !! !!
Read and understand the Operating Manual before using the unit. Use BayPortTM 6-Bay Charger only with BFWTM/Long Island Technology GroupTM batteries. DO NOT subject the BayPortTM 6-Bay Charger to strong shocks, which include but are not limited to dropping the unit on the floor. DO NOT use this unit for anything other than its intended use. DO NOT modify this equipment. DO NOT open unit. No serviceable components are inside device. DO NOT submerge this device or any of its components in liquid. WARNING Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
!!
USE ONLY components and accessories listed on page 5 and 7 of this manual. Failure to do so may decrease system performance, may lead to unsafe operation, may negatively affect EMC performance and could result in non-compliance and could void the warranty
!! !!
DO NOT use if packaging is damaged.
WARNING Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the BayPortTM 6-Bay Charger, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
!! !!
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put in service according to the EMC information in the Electromagnetic Capability of this manual. WARNING Falling Hazard: avoid sitting or standing under the unit when it is mounted.
3
Warnings and Cautions
!! !! !!
DO NOT use non-VESA mounts. Follow local disposal rules for electronic equipment. WARNING: To avoid the risk of electric shock, this equipment must only be connected to a supply main with protective earth
Intended Use
The BFWTM BayPortTM 6-Bay Charger is intended to only charge BFWTM/LITGTM batteries used in BFWTM/ LITGTM products. The charger is intended to be VESA compatible. The product is intended to be used by medical professionals.
4
Overview System Description
The BFWTM BayPortTM 6-Bay Charger is constructed of a durable metal housing. Six charging slots allow up to six BFWTM/LITGTM 3 or 4 cell batteries to be charged simultaneously and stored easily. LEDs indicate the charging status of each battery. The principle of operation is charging smart batteries. The intended environment for use is for Professional Health Care except for near active HF Surgical Equipment and the RF shielded room of an ME System for magnetic resonance imaging.
D C
A
B
A. B. C. D.
BFWTM BayPortTM 6-Bay Charger AC Cable 4-Cell Battery 3-Cell Battery
5
Operation Charging Batteries
1. Plug the female end of the AC Cable (B) into the AC Power Receptacle (C) on the bottom of the unit. 2. Plug the male end of the AC Cable (B) in an AC Outlet. 3. Flip the main power switch into the On position on the AC Power Receptacle (C). 4. Fully charging the Battery before use, will ensure the longest Battery life.
A
B
C
A. B. C.
Charging Indicators (See Table Below)* AC Cable AC Power Receptacle Charging Indicator* Green Blinking Green Solid Red Blinking
Battery is currently being charged Battery is fully charged An error has occurred during charing
5. Flip the main power switch into the OFF position on the AC Power Receptacle (C) to turn the unit off.
!!
The appliance inlet shall not be blocked during operation. This will make it difficult to disconnect the device
6
Cleaning and Maintenance Charging Unit 1. Clean the BayPortTM 6-Bay Charger components with an alcohol wipe. 2. Use a can of compressed air to clean out any dust accumulations.
!! !! !!
DO NOT pour cleaning solution directly on the surface of the BayPortTM 6-Bay Charger DO NOT use any sterilization process or cleaning process using excessive heat or humidity as it will damage the device NEVER immerse the unit in any type of liquid
Note: Damaging any part of the system with the use of an improper cleaning agent or cleaning process will void all warranties.
7
Specifications Classification
Classification & Type
Environment
Operation Storage Relative Humidity Pressure/Altitude
Performance
Charge Time from Empty
Mechanical LxWxH Weight
Electrical
Input Voltage Input Current Line Cord (IEC 60320)
None 60 Degrees F (15 C) to 80 Degrees F (26.7 C) 23 Degrees F (0 C) to 104 Degrees F (40 C) 45% - 75% 860 - 1,060 hPa/2,000m ~3 hours 19.25” x 8.125” x 2.5” 11.7 lbs 100-240 VAC 50/60 Hz 1.0 A Max Maximum Length 10ft
8
Electromagnetic Compatibility Portable and mobile RF communications equipment can affect medical electrical equipment Guidance and manufacturer’s declaration-electromagnetic emissions The BayPort
6-Bay Charger is intended for use in the electromagnetic environment specified below. The customer or the user of the BayPortTM 6-Bay Charger should assure that it is used in such an environment
Immunity Test
IEC 60601 Test Level
TM
Compliance Level
Electromagnetic Environment Guidance Portable and mobile RF communications equipment should be used no closer to any part of the BayPortTM 6-Bay Charger, including cables, than the recommended separation distance calculated from the equation application to the frequency of the transmitter. Recommended Separation Distance: Battery Operated Device
Conducted RF IEC 61000-4-6
3 Vrms 150kHz - 80 MHz
3 Vrms
Radiated RF IEC 61000-4-3
3V/meter 80 MHz 2.5 GHz
3V/m from 30MHz to 1GHz, 3V/m for 1 GHz to 25GHz; (1000 Hz 80% Modulated Test Signal)
d = 1.17√P d = 1.17√P 80 MHz to 800 MHz d = 2.23√P 800 MHz to 2.5 GHz Where P is the maximum output power rating of the transmitter in watts(W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey (1), should be less than the compliance level in each frequency range (2). Interference may occur in the vicinity of equipment marked with the following symbol:
9
Electromagnetic Compatibility NOTE 1: At 80 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from surfaces, objects, and people. (1) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy to asses the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the BayPortTM 6-Bay Charger is used exceeds the applicable RF compliance level above, the BayPortTM 6-Bay Charger should be observed to verify normal operation. IF abnormal performance is observed, additional measure may be necessary, such as re-orienting or relocating the BayPortTM 6-Bay Charger. (2) Over the frequency range 150kHz to 80MHz, field strengths should be less than 3 V/m. Recommended separation distances between portable and mobile RF communications equipment and the BayPortTM 6-Bay Charger The BayPortTM 6-Bay Charger is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the BayPortTM 6-Bay Charger can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the BayPortTM 6-Bay Charger as recommended below, according to the maximum output power of the communications equipment. Rate maximum output power of transmitter
Separation distance according to frequency of transmitter in meters
W
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.17 P
d = 1.17 P
d = 2.23 P
.01
0.12
0.12
0.23
.1
0.37
0.37
0.737
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separate distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. NOTE 3: An additional factor of 10/3 has been incorporated into the formulae used in calculated the recommended separation distance for transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from surfaces, objects, and people.
10
Electromagnetic Compatibility Guidance and manufacturer’s declaration - electromagnetic immunity The BayPort 6-Bay Charger is intended for use in the electromagnetic environment specified below. The customer or the end user of the BayPortTM 6-Bay Charger should assure that it is used in such an environment. TM
Immunity test Electrostatic discharge (ESD)
IEC 60601 test level IEC 61000-4-2
IEC 61000-4-2
± 8 kV contact ± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
± 8 kV contact ± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
IEC 61000-4-3 Radiated RF electromagnetic field 3 V/m 80 % AM at 1 kHz Proximity fields from RF wireless communications equipment Electrical Fast Transient/burst
Surge
Conducted Disturbances induced by RF fields
Compliance level
IEC 61000-4-3 3 V/m 80 % AM at 1 kHz
Electromagnetic environment - guidance Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %. The user of the (BayPortTM 6-Bay Charger) should avoid situations that could result in excess electrostatic discharge. Radiated RF levels should be that of a typical commercial or hospital environment
IEC 61000-4-3
IEC 61000-4-3
See Table 9 on page 22.
See Table 9 on page 22.
IEC 61000-4-4
IEC 61000-4-4
± 2 kV for power supply lines 100kHz repetition frequency
Mains power quality should be that of a typi± 2 kV for power supply lines cal commercial or hospital environment. 100kHz repetition frequency
IEC 61000-4-5
IEC 61000-4-5
± 0,5 kV, ± 1 kV line(s) to line(s) ± 0,5 kV, ± 1 kV, ± 2 kV line(s) to earth
± 0,5 kV, ± 1 kV line(s) to line(s) ± 0,5 kV, ± 1 kV, ± 2 kV line(s) to earth
IEC 61000-4-6
IEC 61000-4-6
3V 0,15 MHz - 80MHz 6 V in ISM bands between 0,15 MHz and 80 MHz 80% AM at 1 kHz
3V 0,15 MHz - 80MHz 6 V in ISM bands between 0,15 MHz and 80 MHz 80% AM at 1 kHz
IEC 61000-4-11
IEC 61000-4-11
0 % UT; 0,5 cycle At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0 % UT; 0,5 cycle At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
Voltage dips, short interruptions and voltage variations on 0 % UT; 1 cycle power supply input and lines 70 % UT; 25/30 cycles Single phase: at 0°
0 % UT; 250/300 cycle
0 % UT; 1 cycle and 70 % UT; 25/30 cycles Single phase: at 0°
RF wireless communications equipment should be used no closer than 30 cm to the BayPortTM 6-Bay Charger.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the BayPortTM 6-Bay Charger requires continued operation during power mains interruptions, it is recommended that the BayPortTM 6-Bay Charger be powered from an uninterruptible power supply or a battery.
0 % UT; 250/300 cycle
NOTE UT is the a.c. mains voltage prior to application of the test level.
11
Electromagnetic Compatibility Guidance and manufacturer’s declaration - Electromagnetic Emissions The BayPortTM 6-Bay Charger is intended for use in the electromagnetic environment specified below. The customer or the end user of the BayPortTM 6-Bay Charger should assure that it is used in such an environment.
Emissions Test RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emissions IEC 61000-3-2 Voltage fluctuations/ flicker emissions IEC 61000-3-3
IEC 60601 test level
Compliance level
Group 1
Group 1
Class A
Class A
Class A
Class A
Complies
Complies
Electromagnetic environment - guidance The BayPortTM 6-Bay Charger uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The BayPortTM 6-Bay Charger is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
NOTE The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment. Table 9 - Test specifications for Enclosure Port Immunity to RF wireless communications equipment Test frequency (MHz)
Band (MHz)
Service
Modulation
Maximum power (W)
Distance (m)
Immunity Test Level (V/m)
385
380 - 390
TETRA 400
Pulse modulation 18Hz
1,8
0,3
27
450
430 - 470
GMRS 460, FRS 460
FM ± 5Hz deviation 1 kHz sine
2
0,3
28
704 - 787
LTE Band 13, 17
Pulse modulation 217 Hz
0,2
0,3
9
800 - 960
GSM 800/900, TETRA 800, iDEN 820, CDMA 850, LTE Band 5
Pulse modulation 18Hz
2
0,3
28
1700 - 1990
GSM 1800; CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4, 25; UMTS
Pulse modulation 217 Hz
2
0,3
28
2400 - 2570
Bluetooth, WLAN, 802.11 b/g/n, RFID 2450, LTE Band 7
Pulse modulation 217 Hz
2
0,3
28
5100 - 5800
WLAN 802.11 a/n
Pulse modulation 217 Hz
0,2
0,3
9
710 745 780 810 870 930 1720 1845 1970
2450
5240 5500 5785
12
Troubleshooting and Service Symptom
Possible Issue
Battery(ies) not charging
Main power cord not plugged in Main power switch is not fully engaged Bad battery(ies) Internal component failure
Solution Plug in cord Toggle ON/OFF switch Replace battery(ies) Return unit to manufacturer for service
Warranty and Service Warranty against manufacturer’s defects applies to the following components of the BayPort™ 6-Bay Charger under normal use for one (1) year from the date of sale from Long Island Technology Group (includes parts and labor). The warranty does not cover products damaged by the following: • Accident, misuse, abuse, or alteration • Servicing by unauthorized persons • Use with unauthorized accessories In all cases Long Island Technology Group, reserves the right to determine the cause of all malfunctions and at its sole discretion will determine the damage and/or repairs that are covered under this warranty. Send All Inquires To: Long Island Technology Group, LLC 60 Carolyn Blvd Farmingdale, New York 11735 USA www.litgp.com [email protected] Tel: (631)-270-4463 Fax: (631)-414-7078
Obelis s.a. Blvd Géné ral Wahis 53 1030 Brussels, Belgium Tel: +(32) 2.732.59.54 Fax: +(32)2.732.60.03 Email: [email protected]
13
Scan Here to See All Long Island Technology GroupTM Manuals Or Visit: http://www.litgp.com/manuals
Contents are fragile. Handle with care.
Conforms to: IEC 60601-1 Edition 3.1, IEC 60601-1-2 Edition 4 AAMI ES60601-1 Edition 1, CSA C22.2 NO 60601-1:08 Edition 2 Manufactured by Long Island Technology Group, LLC 60 Carolyn Blvd, Farmingdale, NY 11735
60 Carolyn Blvd, Farmingdale NY 11735 (631) 270-4463 litgp.com Revision 1.1