BOWA-electronic
ERGO 300 Operating Manual Aug 2014
Operating Manual
48 Pages
Preview
Page 1
OPERATING MANUAL MORCELLATOR
Table of contents 1
General information
4
1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 1.10
Indications Principle Contraindications Medical application On receipt Scope of delivery, accessories and spare parts Technical Data «ERGO 300 - Morcellator» Ambient conditions Warranty coverage Right to make technical changes reserved
4 4 4 4 5 6 7 7 7 7
2 3
Explanation of symbols Safety information
8 8
3.1 3.2 3.3 3.4
EMC Manufacturer’s Declaration of Conformity Modification and misuse Essential requirements During use
8 8 9 9
4
Device overview
4.1
ERGO 300 Morcellator system chart
12
5
Before use
13
5.1 5.2 5.2.1 5.2.2 5.3
Connection to the power supply Device preparation Assembly of the Morcellator set with protective tube (option 1) Assembly of the Morcellator set with Trocar Cannula (option 2) Mains connection
13 13 14 17 19
6
Use
20
6.1 6.2 6.2.1 6.2.2 6.2.3 6.3 6.4 6.5 6.6
Using the control unit Operation Option 1: morcellation with protective tube Option 2: morcellation with Trocar Cannula Option 3: Myomectomy After use Disassembling Trocar Cannula Disassembly of motor Disassembly of gear unit and Morcellator with protective tube
20 22 24 25 27 28 29 30 30
7
Instrument preparation
31
7.1 7.2 7.3 7.4 7.5
Control unit and pedal 31 Preparation of the Trocar Cannula 32 Preparation of motor 33 Preparation of gear unit 35 Preparation of cutting tubes, protective tubes, filling rods, seals, seal unit and myoma drill 36
8
Maintenance and repair
38
8.1 8.2 8.3 8.4 8.5 8.6
General maintenance Replacing the control unit fuse Safety inspections Malfunction and troubleshooting Spare parts list with order numbers Information on disposal
38 38 39 39 40 40
9
Electromagnetic compatibility (EMC)
41
MN009-938-S0
10
ERGO 300
3
1
General information
1.1 Indications Morcellator set for morcellation and extraction of tissue and for removing myomas or the uterus during laparoscopic procedures in surgery and gynaecology.
1.2 Principle The ERGO 300 - Morcellator is a control unit for the BOWA Morcellator for use in morcellation and extraction of tissue. A rotating cylindrical tube with a blade at the distal end is inserted through an abdominal cavity that has been enlarged by an Insufflator. The tissue that is to be removed is guided to the cutting tube by a gripper, cut off and removed through the tube. This manual describes three different options for removing tissue by a laparoscopic procedure. Option 1: Tissue removal by morcellation with protective tube. Instruments (gripper) are inserted through a protective tube sealed by a valve system. The valve can be covered by a thumb for removing the gripper from the sealed protective tube to prevent gas loss. Option 2: Tissue removal by morcellation with Trocar Cannula. Instruments (cutter tube, gripper) are inserted through the Trocar Cannula. The Trocar Cannula contains a valve that prevents gas loss when withdrawing the instruments. Option 3: Tissue removal (myectomy) with myoma drill through the Trocar Cannula. The myoma drill is inserted through the Morcellator (protective tube or Trocar Cannula). The Trocar Cannula and the protective tube can be used with or without a cutting tube. No gas loss after removal of instruments with the Trocar Cannula, minimum gas loss during removal from the protective tube, sealing unit can be sealed with the thumb.
1.3 Contraindications Do not use the Morcellator for the following: •
Treatment of malignant tumors
•
Treatment of vascularized tissue
•
Preparation of tissue
1.4 Medical application This product is designed solely for use in medical facilities by medically qualified persons.
4
ERGO 300
MN009-938-S0
1.5 On receipt •
The shipping package must contain all parts listed in 1.6 Scope of delivery.
•
If any parts are damaged or missing, contact the BOWA agent from which the device was ordered or BOWA.
Immediate inspection of the ERGO 300 - Morcellator motor system (art. no. 905-000):
1. 2. 3. 4. 5. 6.
Morcellator control unit (905-001) including power cord 3 m Morcellator Gear unit complete including Seals (905-004) Cross seal (905-009) (10 pcs.) Seals for Ø 12-22 mm instruments (905-512) (10 pcs.) 7. Micromotor with cable 3 m (905-002) 8. Fuses (2 pcs.)
MN009-938-S0
9. Vario Foot switch with cable 3 m (905003) 10. Morcellator Spray nozzle gear unit for Nou Clean Spray (905-006) 11. Morcellator Spray nozzle motor for Nou Clean Spray (905-007) 12. Handle (905-005) 13. Manual on CD-ROM
ERGO 300
5
1.6 Scope of delivery, accessories and spare parts Art.-no.
Description
Quantity
Basic Set 905-000
Morcellator ERGO 300 basic set comprising the following products ---------------------------- 1 unit
905-001 905-002 905-003 905-004 905-005 905-006 905-007 MN031-578
Morcellator Control unit ERGO 300 including power cord 3 m ----------------------------------- 1 unit Morcellator Micromotor ERGO 300 with cable 3 m ------------------------------------------------- 1 unit Morcellator Vario Foot switch with cable 3 m --------------------------------------------------------- 1 unit Morcellator Gear unit complete including seals ------------------------------------------------------ 1 unit Morcellator Handle ------------------------------------------------------------------------------------------ 1 unit Morcellator Spray nozzle gear unit ---------------------------------------------------------------------- 1 unit Morcellator Spray nozzle motor -------------------------------------------------------------------------- 1 unit Manual on CD-ROM ---------------------------------------------------------------------------------------- 1 unit
Application-Sets 905-612
Morcellator application set Ø 12 mm ------------------------------------------------------------------- 1 unit
905-112 905-212 905-012
Morcellator Cutting tube Ø 12 mm ---------------------------------------------------------------------- 1 unit Morcellator Protective sleeve Ø 12 mm ---------------------------------------------------------------- 1 unit Obturator Ø 12 mm ----------------------------------------------------------------------------------------- 1 unit
905-615
Morcellator application set Ø 15 mm ------------------------------------------------------------------- 1 unit
905-115 905-215 905-015
Morcellator Cutting tube Ø 15 mm ---------------------------------------------------------------------- 1 unit Morcellator Protective sleeve Ø 15 mm ---------------------------------------------------------------- 1 unit Obturator Ø 15 mm ----------------------------------------------------------------------------------------- 1 unit
905-620
Morcellator application set Ø 20 mm ------------------------------------------------------------------- 1 unit
905-120 905-220 905-020
Morcellator Cutting tube Ø 20 mm ---------------------------------------------------------------------- 1 unit Morcellator Protective sleeve Ø 20 mm ---------------------------------------------------------------- 1 unit Obturator Ø 20 mm ----------------------------------------------------------------------------------------- 1 unit
Accessories and spare parts 905-312 905-315 905-008 905-010 905-011 905-009 905-512 905-013
Morcellator Trocar sleeve d12 mm ---------------------------------------------------------------------- 1 unit Morcellator Trocar sleeve d15 mm ---------------------------------------------------------------------- 1 unit Morcellator Carrier plate ERGO 300 -------------------------------------------------------------------- 1 unit Morcellator Myoma drill d10 mm ------------------------------------------------------------------------- 1 unit Morcellator Gasket holder ---------------------------------------------------------------------------- per 1 unit Morcellator Cross seal ------------------------------------------------------------------------------------ 10 unit Morcellator Seal d12-20 mm 12-20 mm--------------------------------------------------------------- 10 unit Morcellator Motor cable 3 m ------------------------------------------------------------------------------ 1 unit In line with regulations pertaining to hazardous materials, the following items are not delivered with the control unit but can be ordered separately from any official BOWA service centre. www.bowa.de
905-014
6
Morcellator Cleaning spray "Nou Clean" -------------------------------------------------------------------- 1 unit
ERGO 300
MN009-938-S0
1.7 Technical Data «ERGO 300 - Morcellator» Voltage -------------------------------------------------------------- variable: 100 V∼/ 115 V∼/ 230 V∼, 50–60 Hz Fuse, power supply ------------------------------------------------------------------------- 2 fuses T 1 AL 250 V AC Power consumption ------------------------------------------------------------------------------------------------- 60 VA Maximum motor speed -------------------------------------------------------------------------------------- 40,000 rpm Maximum torque, motor ------------------------------------------------------------------------------------------- 6 Ncm Torque at gear output -------------------------------------------------------------------------------------------- 80 Ncm Motor coupling ------------------------------------------------------------------------------ INTRA coupling ISO3964 Motor cable length ------------------------------------------------------------------------------------------------------ 3 m Pedal cable length ------------------------------------------------------------------------------------------------------ 3 m Morcellator speed ----------------------------------------------------------------------------------------- 50 – 1000 rpm Protection class ------------------------------------------------------------------------------------------------------class II Applied part type ------------------------------------------------------------------------------------------------- Typ BF * Foot switch -------------------------------------------------------------------------------------------------------------- IPX8 Dimensions (W x D x H) ------------------------------------------------------------------------ 120 x 180 x 107 mm Net weight control unit ---------------------------------------------------------------------------------------------- 1,8 kg * Applied part is the Morcellator
1.8 Ambient conditions Transport and storage:
Operation:
Relative humidity:
Max. 90 %
Max. 80 %
Temperature:
10°C to 60°C
10°C to 30°C
Atmospheric pressure:
700 hPa to 1060 hPa
800 hPa to 1060 hPa
1.9 Warranty coverage Purchasing the ERGO 300 - Morcellator entitles you to a 1-year warranty. If you return the warranty card for registration within four weeks of the date of purchase, warranty coverage will be extended for a further 6 months. Improper use or repair, or failure to observe these instructions, will relieve us from any obligation arising from warranty provisions or other claims.
1.10 Right to make technical changes reserved Illustrations, dimensions and weights and technical data may vary slightly as a result of continuing development. We reserve the right to make changes and we are not responsible for printing errors.
MN009-938-S0
ERGO 300
7
2
Explanation of symbols Important information
Autoclavable at 135°C
Warning
Suitable for thermal disinfection
Manufacturer Applied part of type BF is the Morcellator
3
Protection against continuous submersion in water. Protection class II
Certified by the Canadian Standards Association (CSA)
Electrical and electronic devices that have reached the end of their service life comprise hazardous waste and may not be disposed of together with household waste. Local disposal regulations apply.
CE mark with Notified Body
Symbol indicating the order number
Intermittent operation, 1 min on, 1 min off for 8 cycles Observe the instructions for use
Symbol indicating the serial number with the date of manufacture (year/month) Motor
Pedal
Biohazard
Safety information
Your safety, the safety of your team and of course that of your patients is very important to us. It is therefore essential to bear the following information in mind:
3.1 EMC Manufacturer’s Declaration of Conformity Please observe the information on electromagnetic compatibility provided with these operating instructions.
3.2 Modification and misuse • Modification or manipulation of the ERGO 300 - Morcellator and its accessories is prohibited. The manufacturer is not liable for any damages resulting from unauthorized modifications or manipulations. The warranty will be cancelled. • Use of the ERGO 300 - Morcellator outside the indications described in Section 1.1 is prohibited. The user or operator is solely responsible for any such use.
8
ERGO 300
MN009-938-S0
3.3 Essential requirements Before use read this manual and the manuals for all other equipment used in the procedure thoroughly.
The products described in this manual must be used by trained and qualified medical personnel only in accordance with the requirements and with reference to the manual.
For protection of the patient, the user and third parties the products used must be prepared as specified in the manual.
Improper use or repair of the device, or failure to observe these instructions, will relieve us from any obligation arising from warranty provisions or other claims.
The use of third-party products is the responsibility of the operator. Functionality and patient safety cannot be guaranteed with third-party accessories.
Prior to using the device, before startup, and before operation, the user must always ensure that the device and accessories are in good working order and are clean, sterile and operational.
Repairs may only be performed by authorized BOWA service technicians.
Use Nou-Clean spray to care for the motor and gear unit. Using other care products can result in malfunctions and/or cause the warranty to be revoked.
3.4 During use The device is not sterile on delivery. All sterilizable parts must be sterilized before use (see Chapter 7 Instrument preparation).
Never operate the clamping mechanisms of the cutting tubes while the system is running. This could result in instrument damage.
Operate the device only outside the danger zone of explosive, flammable mixtures or gases.
Keep hands clear of the cutting tubes when operating the device. Danger of serious injury.
The motor ventilation slots must be kept clear in order to prevent the motor from overheating.
The ERGO 300 - Morcellator must be operated under the continuous supervision of medically qualified persons only.
Disregarding the guidelines on intermittent operation (1 min on, 1 min off for 8 cycles) may cause burns at contact with the gear unit.
Make sure that the operating voltage to the mains voltage.
MN009-938-S0
ERGO 300
9
4
Device overview
Front view
Control panel
Rear view
1. Display
1. Mode button (variable speed range) 1. Mains connection
2. Motor LED
2. Low button (low speed range)
2. Power switch
3. Pedal connection
3. Medium button (medium speed
3. Fuse module 4. Window with national
range) 4. Motor connection
4. High button (high speed range)
voltage display
5. Change rotational direction button 5. Type plate with type 6. Up button (increase value)
designation, reference
7. Down button (reduce value)
number, serial number, information on power supply and device fuse
Rear view
Handle
1. Flange for protective tube con- 1. Release lever 2. Slots for motor ventilanect 2. Thread for seal holder
tion
Motor
1. Device connection plug 2. Motor cable 3. Motor cap 4. Motor
3. Cutting tube holder
5. Handpiece holder
4. Seal holder
6. Motor connection
5. Thread for membrane screw
plug
6. Membrane screw 7. Seal holder thread 8. Motor connection coupling 10
ERGO 300
MN009-938-S0
Protective tube with filling rod
Cutting tube
1. Cutting edge 2. Locking pins
5. Fingerstütze 1. Protective tube
6. Unterer Schen-
2. Lock nut
kel
3. Tissue positioner
7. Feder
4. Filling rod
8. Flexibler Arm der Raste
Trocar Cannula
6
1. Protective tube
4. Valve body
2. Seal cap holder
5. Main body
3. Valve actuator
6. Nose
6. Unterer Schenkel 7. Feder 8. Flexibler Arm der Raste
MN009-938-S0
ERGO 300
11
4.1 ERGO 300 Morcellator system chart
12
ERGO 300
MN009-938-S0
5
Before use The ERGO 300 Morcellator,and all required accessories and instruments must be placed on a level nonslip surface. Do not allow the operating range of the device with cable to be restricted by limiting factors. The display, keypad and indicator lights must be fully visible at all times. It is very important to ensure that no objects can fall on the pedal. The power plug at the rear of the device must be accessible at all times.
5.1 Connection to the power supply •
Before plugging the power cable into the power socket for the first time, you must check the supply voltage setting next to the power switch.
•
In order to prevent the risk of electric shock, the device may only be connected to a power network with a
PE protective earth conductor. If the voltage shown does not correspond to the local mains voltage, the grey fuse holder must be set to the correct voltage:
1.
Swich the device off.
2. Unplug the power cable. 3. Use a screwdriver to open the fuse slot. 4. Remove the fuse holder. 5. Remove the grey fuse holder and reinsert it so that the local mains voltage setting is shown in the small window. 6. Slide the grey fuse holder back in and close the fuse slot. 7. Check the mains voltage shown on the fuse slot. 8. Plug the power cable back into the device.
5.2 Device preparation The device is not sterile on delivery. All sterilizable parts must be sterilized before use (see Chapter 7 Instrument preparation).
If the accessories have already been sterilized: when removing them from the sterile packaging, ensure that the sterile packaging is not damaged and that the sterility indicator confirms sterility and shelf life of the sterile item has not expired.
The Morcellator (as per option 1 to option 3) must be assembled under sterile conditions (wear gloves and a mask, place Morcellator and accessories on a sterile surface).
MN009-938-S0
ERGO 300
13
5.2.1 Assembly of the Morcellator set with protective tube (option 1)
There is a risk of injury to tissue by the sharp edge at the distal end of the cutting tube. • Handle the cutting tube carefully. • If the cutting tube is not used, rotate the protective tube to the NO CUT position. • If the cutting tube is not used, it must be placed on a flat, sterile surface so it cannot fall and cause injury
• Check the cutting tube. • Make sure that the cutting edge of the blade is sharp and undamaged (e.g. no cracks, deformation or burrs). • Use only cutting tubes in perfect condition.
1.
Inserting the cutting tube through the gear unit Remove the seal unit from the gear unit.
Insert the cutting tube through the gear unit. The two guide pins on the cutting tube must fit into the slot on the flange of the gear unit.
2.
Fixing the cutting tube to the gear unit Hold the cutting tube with one hand. Connect the sealing unit to the gear unit with the bayone closure with the other hand.
3.
Mounting protective tube Select a suitable protective tube. Push the protective tube over the distal end of the cutting tube. The guide pin of the protective tube must be in the distal slot of the gear unit. Lock the protective tube in position with a click.
4.
Protecting the cutting tube The distal end of the cutting tube is very sharp. To prevent injury, the cutting tube must be covered at all times when it is not in use. To cover the cutting tube, rotate the protective tube until the words “NO CUT” are visible in the distal slot of the gear unit. The cutting tube is completely covered by the protective tube.
14
ERGO 300
MN009-938-S0
5. Replacing seals Remove the seal unit from the gear unit by rotating it slightly. Unscrew the membrane screw from the seal unit. Check seals before use. Do not use damaged seals (frayed, discolored, yellowed, porous, non-elastic). Position the edge of the cross seal (1) to point to the inside stop (2) of the seal holder (3) and press the cross seal in. Position the edge of the membrane seal (4) to point to the slot (5) of the membrane screw (6) and press the membrane seal into the slot. Screw the membrane screw (6) into the thread of the seal holder (3) to the stop and connect the seal unit to the gear unit.
6.
Inserting filling rod Insert the filling rod through the seal unit, gear unit and protective tube to the stop.
7. Attaching handle Lift and hold the release lever (1) of the handle (2). Push the handle (2) over the motor connection of the gear unit. Release the handle release lever. The handle release lever must lock in position.
Do not open the handle release lever to remove the handle during the procedure..
8.
Connecting motor Insert the motor coupling into the bottom end of the handle. Continue to insert the motor coupling until it locks into the gear unit motor connection.
MN009-938-S0
ERGO 300
15
Align the red dot on the device connection plug of the motor with the red dot on the motor connection of the control unit. Plug the motor device connection plug into the motor connection of the control unit. The plug must lock in position.
9.
Connecting pedal Align the red dot on the device connection plug of the pedal with the red dot on the pedal connection of the control unit. Plug the pedal device connection plug into the pedal connection of the control unit. The plug must lock in position.
10. Functional check • • • • • • • • •
Check Morcellator for loose parts Check that Morcellator is correctly assembled Are all labels easily readable? Check cutting tube: cutting tube evenly ground, no broken edges or bends? Switch on control unit Switch on motor (press foot switch). Run motor with Morcellator for about 20 seconds; it must reach the set speed. If it does not run evenly, switch off motor with Morcellator.. Switch off control unit.
The control unit will only operate if the foot switch has been connected to the control unit. •
Check that Morcellator is correctly assembled.
•
Check that the gear unit runs smoothly; if necessary treat it with Nou-Clean care spray.
•
Check that the motor is clean and runs smoothly; if necessary clean it and treat with Nou-Clean care spray.
•
Repeat the function check. Risk of injury by motor failure Problem: the motor becomes very hot. Cause: the motor was damaged during preparation. Remedy: do not continue to use the motor. Always have a spare motor available.
16
ERGO 300
MN009-938-S0
5.2.2 Assembly of the Morcellator set with Trocar Cannula (option 2) Using the Trocar Cannula The Trocar Cannula is used instead of the protective tube. The integrated valve flap prevents the escape of gas when instruments are moved in and out. The nose of the Trocar Cannula partially extends beyond the cutter tube to protect organs in the abdominal cavity from the sharp blade. This requires the Morcellator to be rotated under visual supervision (endoscopy) so the nose of the Trocar Cannula covers the cutter tube blade to protect the organ. 1. Visual check •
Check the product for sealing, corrosion and sterility.
•
Ensure that it is free from rust, dents and scratches.
•
The distal end of the filling rod must not be damaged.
•
The sealing flap and the seal must not be damaged. Risk of injury Select blade. Make sure that the cutting edge of the blade is sharp and undamaged (e.g. no cracks or burrs).
2.
Insert the cutting tube through the gear unit Remove the seal unit from the gear unit. Insert the cutting tube through the gear unit. The two guide pins on the cutting tube must fit into the slot on the flange of the gear unit.
3.
Fixing the cutting tube to the gear unit Hold the cutting tube with one hand. Connect the seal unit to the gear unit with the bayonet closure with the other hand.
4.
Mounting Trocar Cannula Select a suitable Trocar Cannula. Open the Trocar Cannula valve by actuating the lever to prevent damage to the cutting tube by pushing the Trocar too far. Push the Trocar Cannula over the distal end of the cutting tube. The guide pin of the Trocar Cannula must be in the distal slot of the gear unit. Lock the Trocar Cannula in position with a click.
MN009-938-S0
ERGO 300
17
5. Replacing seals Remove the seal unit from the gear unit by rotating it slightly. Unscrew the membrane screw from the seal unit. Position the edge of the cross seal (1) to point to the inside contact (2) of the seal holder (3) and press the cross seal in. Position the edge of the membrane seal (4) to point to the slot (5) of the membrane screw (6) and press the membrane seal into the slot.
Screw the membrane screw (6) into the thread of the seal holder (3) to the stop and connect the seal unit to the gear unit.
6. Inserting filling rod Insert the filling rod through the seal unit, gear unit and Trocar Cannula. 7. Attaching handle Lift and hold the release lever (1) of the handle (2). Push the handle (2) over the motor connection of the gear unit. Release the handle release lever. The handle release lever must lock in position.
8. Connecting motor Insert the motor coupling into the bottom end of the handle. Continue to insert the motor coupling until it locks into the gear unit motor connection. Align the red dot on the device connection plug of the motor with the red dot on the motor connection of the control unit. Plug the motor device connection plug into the motor connection of the control unit. The plug must lock in position. 9. Connecting pedal Align the red dot on the device connection plug of the pedal with the red dot on the pedal connection of the control unit. Plug the pedal device connection plug into the pedal connection of the control unit. The plug must lock in position.
18
ERGO 300
MN009-938-S0
Functional check • • • • • • • • •
Check Morcellator for loose parts Check that Morcellator is correctly assembled Are all labels easily readable? Check cutting tube: cutting tube evenly ground, no broken edges or bends? Switch on control unit Switch on motor (press foot switch). Run motor with Morcellator for about 20 seconds; it must reach the set speed. If it does not run evenly, switch off motor with Morcellator. Switch off control unit. The control unit will only operate if the foot switch has been connected to the control unit. •
Check that Morcellator is correctly assembled.
•
Check that the gear unit runs smoothly; if necessary treat it with Nou-Clean care spray.
•
Check that the motor is clean and runs smoothly; if necessary clean it and treat with Nou-Clean care spray.
•
Repeat the function check.
Problem: the motor becomes very hot. Cause: the motor was damaged during preparation. Remedy: do not continue to use the motor. Always have a spare motor available.
5.3 Mains connection Risk of electric shock Connect the device to an earthed AC mains system that conforms to IEC guidelines only. Connect the device to a power supply that corresponds to the connection values on the type plate on the back of the device. Risk of damage Do not switch on the control unit and the other equipment until all cables have been connected. Otherwise the equipment may be damaged.
The device is not water-proof. Keep the device away from liquids. Keep the mains connection socket dry. Do not use the device if liquid penetrates it.
Connecting the power cable Switch off the power switch on the back of the device (“O” position). Plug the included power cable into the socket on the back of the device. Plug the other end of the power cable into the power socket.
MN009-938-S0
ERGO 300
19
6
Use Before use Make sure that the product has been correctly prepared and inspected. Risk of injury The ERGO 300 - Morcellator control unit has been designed for operation with the morcellation blades as a set. The user is solely responsible if different products are used with the set. Incompatibility may risk injury to patients. Switching on and off Switch on the control unit. The display lights up. A short signal tone sounds. The control unit displays the value “200 rpm”. Operating mode (motor) The motor is designed for intermittent operation (1 minute on, 1 minute off for 8 cycles).
6.1 Using the control unit Setting the speed The speed can be set manually in stages (within a range of 50 to 100 rpm). Press «
» to increase the speed.
Press «
» to reduce the speed.
The set speed is shown on the display. Starting motor Press foot switch. The blade rotates at the set speed. The motor LED lights up. To stop the motor remove foot from foot switch.
Risk of injury and damage due to heat! The Morcellator may generate heat if operated for a long period. Remedy: Use Morcellator only in intermittent mode of “1 minute on, 1 minute off for 8 cycles”. Stop the Morcellator immediately if one or more components of the Morcellator set become very warm.
Precaution when changing parameters. Unusual behaviour of instruments while operating may provoke false reactions and jeopardize the patient. Every setting and the new behaviour of the instrument has to be verified and tested.
20
ERGO 300
MN009-938-S0
Available rotary modes «
» (100 – 400 rpm)
«
» (300 – 700 rpm)
«
» (500 – 1000 rpm)
Activating mode setting Press «
» to activate mode setting.
The Mode LED lights up. The Low LED lights up. (Default setting) The display shows “VARIO LOW, 100 – 400 rpm”. Selecting mode Press « », « corresponding mode.
» or «
» to select the
The display shows the mode and the corresponding speed range that can be selected with the pedal. The corresponding LED lights up.
Starting motor Press foot switch. The motor LED lights. The blade rotates at a speed within the range set by how far down the foot switch is pressed. To stop the motor remove foot from pedal. Exit mode setting Press «
»
Changing direction of rotation To change the direction of rotation, press«
»
The display shows the mode and the corresponding speed range that can be selected with the pedal. The corresponding LED lights up.
MN009-938-S0
ERGO 300
21