Galaxy Ultra HD-LED® Standard Models Installation Guide
IG-GLEDSTD Issue B3 – 03 December 2013
Galaxy Ultra HD-LED Standard Models
© Brandon Medical Company Ltd 2009 All rights are reserved. No part of the content may be reproduced, adapted, stored in a retrieval system or transmitted in any form or by any means, electronic, mechanical, photocopying, recording or otherwise without the prior permission of the Copyright owner. Whilst every effort is made to ensure the accuracy and completeness of this guide, we do not warrant that the content is error free. The brand names or product names mentioned or referred to throughout this guide are fully recognised as the trademark or registered trademark of their respective owners.
Page 2 of 34
IG-GLEDSTD
Contents
Contents 1
Before starting the installation ...4 1.1 1.2 1.3 1.4 1.5 1.6
2
Electrical installation ...8 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8
3
Supply requirements ...8 48V power supplies ...8 Switching and intensity control ...9 Emergency power switch-over ...10 Circuit protection and indication...11 Cable size ...11 Cable rating table...11 Wiring schematic ...12
Installation, wiring and adjustment instructions...14 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 3.10 3.11
Table of instructions...16 Selecting the correct ceiling anchor ring ...17 Mounting the flange plate (stem) ...17 Fixing the articulation arm assembly (structure)...20 Fixing the articulation spring arms ...21 Fixing the horizontal arm ...22 Fitting secondary arm and yoke arm ...23 Fitting a single lamp head...24 Fitting the main lamp head ...24 Fitting the sterilisable handle systemError! Bookmark not defined. Releasing spring arm transit lock...27
4
Commissioning the unit...28
5
Additional facilities and options ...29
6
Installation Checklist...30
7
Warranty...32 7.1 7.2
8
IG-GLEDSTD
Symbols used ...4 Models covered...4 Installer qualifications ...5 Classifications ...6 Environmental conditions...6 Related guides ...7
Terms of the Warranty ...32 Conditions of the Warranty ...32
Declaration of Conformity ...33
Page 3 of 34
Galaxy Ultra HD-LED Standard Models
1
Before starting the installation For your full satisfaction with this equipment it is recommended that you read this section carefully before starting the installation.
1.1
Symbols used The symbols used in this guide and on the product are explained in the following table. Symbol
Description
!
WARNING/DANGER – Pay special attention. Degree of protection from electric shock.
1.2
Note this point. Refer to another section/guide.
Models covered Brandon Galaxy Ultra lamps are a modular design and are available in virtually any combination of 1, 2, 3, 4 or 5 lamp heads. Within the specific model numbers, eg Galaxy Ultra 54, the first digit defines the number of LED modules in the main lamp head and the subsequent digit defines the satellites. A low ceiling model has the suffix LC.
Throughout this guide reference is typically made to GLED5, GLED4, etc and this relates equally to the number of modules a main lamp head or a satellite lamp head. Up to three lamp heads can be mounted on a single central spindle. Virtually any combination can be configured and some degree of customisation is also possible. For example, a GLED54 will be made up of a 5-module main lamp and a 4-module satellite lamp and is a standard version. A GLED54LC is the same but is configured for low ceiling installations.
Page 4 of 34
IG-GLEDSTD
Before starting the installation
1.3
Main Lamp
No. of LED modules
Intensity (K Lux)
Lamp Power Rating (W)
GLED50
5
160
135
GLED40
4
140
108
GLED30
3
120
80
GLED20
2
85
54
GLED10
1
50
27
Satellite Lamp
No. of LED modules
Intensity (K Lux)
Lamp Power Rating (W)
GLED05
5
160
135
GLED04
4
140
108
GLED03
3
120
80
GLED02
2
85
54
GLED01
1
50
27
Installer qualifications Installation should only be undertaken by an installer who has undergone an accredited Brandon Medical training course.
IG-GLEDSTD
Each installation requires two people, one of whom is an approved installer.
Page 5 of 34
Galaxy Ultra HD-LED Standard Models
1.4
Classifications The classifications are explained in the following table.
1.5
Classification
Description
Electrical classification
Class 1
Degree of protection against electric shock
Class B
Classification according to EU Medical Device Directive 93/42/EEC
Class 1
Lamp head degree of protection from ingress of dust and water
IPXO
Degree of protection against the presence of flammable anaesthetic mixtures
Not for use in a flammable atmosphere
Environmental conditions The environmental conditions relate to transportation/storage and operating.
1.5.1
Transportation and storage The environmental conditions relating to transportation and storage are explained in the following table.
1.5.2
Condition
Range
Temperature
-40oC to 70oC
Relative humidity
10% to 95% non-condensing
Pressure
500hPa to 1060hPa
Operating The environmental conditions relating to operating are explained in the following table.
Page 6 of 34
Condition
Range
Temperature
15oC to 30oC ambient temperature
Relative humidity
10% to 80% non-condensing
IG-GLEDSTD
Before starting the installation
1.6
Related guides The other guides that relate to the Galaxy Ultra HD-LED operating table luminaires are:
IG-GLEDSTD
Galaxy Ultra HD-LED Quick Start Guide
Galaxy Ultra HD-LED Instructions for Use
Galaxy Ultra HD-LED Low Ceiling Models Installation Guide
Galaxy Ultra HD-LED Tandem Models Installation Guide
Galaxy Ultra HD-LED Routine Maintenance Guide.
Page 7 of 34
Galaxy Ultra HD-LED Standard Models
2
Electrical installation The instructions in this section apply to all Galaxy Ultra HD-LED Standard and Low Ceiling models and should be used in conjunction with the instructions later in this guide that are specific to the model you are installing.
2.1
Supply requirements
!
All standard models are 48V and should be supplied by a Brandon Medical switched-mode power supply which is separately purchased. The location of the power supplies and the associated wiring should be carried out before fitting the lamp. Check that the electrical requirements as detailed on the serial number label of the lamp correspond with your supply voltage before connecting.
A single-phased fused supply is required for each of the lamp power supplies. It is recommended that the power supplies to the main lamp and the satellites are kept separate so that failure of any one component will only affect one lamp head. A mains supply isolator is recommended to isolate the live and neutral supply for maintenance work. The isolator should be rated at a minimum of 20A and be 2- or 4pole if an emergency DC supply is present.
2.2
Refer to Section 2.8 for the wiring schematic. Brandon Medical can supply electrical accessories as required for any electrical supply.
48V power supplies The specification is shown in the following table.
Page 8 of 34
Classification
Description
Model
EPS15023048
Power
150W
Supply Voltage
85 – 265VAC (47 – 63Hz)
Output Voltage
38.4 – 52.8VDC
Size (W x H x D)
42 x 82 x 160
Weight
550g
IG-GLEDSTD
Electrical installation
!
2.3
The power supplies can only be mounted in certain orientations as shown in the following drawing.
The above specification is only for a stand-alone power supply and does not include any DC/DC change-over system or enclosure.
Switching and intensity control A remote switch to put the equipment in hibernation mode can be fitted.
Refer to the wiring schematic in Section 2.8. All on/off and intensity controls are controlled via the lamp head-mounted control pod.
IG-GLEDSTD
Refer to Instructions for Use for details. Where requested, alternative dimming arrangements can also be accommodated.
Page 9 of 34
Galaxy Ultra HD-LED Standard Models
2.4
Emergency power switch-over An emergency power switch-over module is available which provides all the necessary relays to switch over from mains supply to 24V emergency power supply and which incorporates the power supply. The specification is shown in the following table.
Page 10 of 34
Classification
Description
Model
13508
Supply Voltage (DC)
19.6 – 36VDC (Max Cable Size 6 mm)
Supply Voltage (AC)
85 – 265VAC (47 – 63Hz) (Max Cable Size 2.5 mm)
Supply Current (DC)
20A Max
Output Voltage
48VDC (Max Cable Size 6 mm)
Output Current
7.3A
Size (W x H x D)
200 x 100 x 70
Weight
150g
IG-GLEDSTD
Electrical installation
!
The power supplies can only be mounted in certain orientations as shown in the following drawing.
The above module can be supplied together with the power supply inside a pre-wired enclosure.
2.5
Circuit protection and indication Brandon Medical recommends the fitting of circuit breakers and indicator lamps as shown in the wiring schematic to facilitate maintenance and to protect the equipment from fault conditions.
Refer to the wiring schematic in Section 2.8.
2.6
Cable size It is recommended the distance between the power supply/emergency power switchover module and the lamp is kept to a minimum. It is recommended that the correct cable size is selected for the low voltage feed from the power supply to the lamp, to minimise the volt drop.
Refer to the cable rating table in Section 2.7.
2.7
Cable rating table Cabel size current ratings and volt drops for unsheathed single core PVC-insulated cables taken from Tables 4D1A and 4D1B of BS7671: 1992.
Cross-
In
In
In
In conduit
Clipped
Clipped
Volt drop
Volt drop
sectional
conduit
conduit
conduit
on wall
direct (A)
direct (A)
(mV/A/m)
(mV/A/m)
area
in
in
on wall
(A) 3 or 4
2 cables
3 or 4
2 cables
3 or 4
(mm2)
thermal
thermal
(A) 2
cables
insulation
insulation
cables
(A) 2
(A) 3 or 4
cables
cables
1.0
11.0
10.5
13.5
12.0
15.5
14.0
44.0
38.0
1.5
14.5
13.5
17.5
15.5
20.0
18.0
29.0
25.0
2.5
19.5
18.0
24.0
21.0
27.0
25.0
18.0
15.0
4.0
26.0
24.0
32.0
28.0
37.0
33.0
11.0
9.5
6.0
34.0
31.0
41.0
36.0
47.0
43.0
7.3
6.4
10.0
46.0
42.0
57.0
50.0
65.0
59.0
4.4
3.8
16.0
61.0
56.0
76.0
68.0
87.0
79.0
2.8
2.4
IG-GLEDSTD
cables
cables
Page 11 of 34
Galaxy Ultra HD-LED Standard Models
2.8
Page 12 of 34
Wiring schematic
IG-GLEDSTD
Electrical installation
IG-GLEDSTD
Page 13 of 34
Galaxy Ultra HD-LED Standard Models
3
Installation, wiring and adjustment instructions Unpack the lamp and check that all the parts listed in the following table are present. Part
Quantity
Anchor ring (if ordered)
1
Flange plate (stem) – made-to-measure or adjustable version
1
Ceiling cover and ceiling cover support ring
1
MEIGaN isolation washers
12-off
Main articulation arm assembly (structure)
1
Spring balance arms (spring arm)
1 per lamp head
Secondary arm and yoke arm assembly
1 per lamp head (not GLED10 or GLED1
These are supplied as one assembly except on GLED1 where the yoke/lamp head are one. Sterilisable handle carrier assemblies
1 per lamp head
Sterilisable handle covers
1 per lamp head
Power supply
1 per lamp head
Accessory items as ordered
-
Examine all the parts carefully for any damage in transit. Report any damaged or missing items to the supplier within 5 working days. The instructions you have to follow are determined by the size of light you are installing.
Page 14 of 34
Refer to Section 3.1.
IG-GLEDSTD
Installation, wiring and adjustment instructions
A typical mechanical installation is shown in the following illustration.
IG-GLEDSTD
Page 15 of 34
Galaxy Ultra HD-LED Standard Models
3.1
Table of instructions Please check which sub-sections you must refer to, to ensure the correct installation.
Page 16 of 34
Model
3.2
3.3
3.4
3.5
3.6
3.7
3.8
3.9
3.10
3.11
GLED55
-
-
GLED54
-
-
GLED53
-
-
GLED52
-
-
GLED51
-
GLED50
-
GLED44
-
-
GLED43
-
-
GLED42
-
-
GLED41
-
GLED40
-
-
-
GLED33
-
-
GLED32
-
-
GLED31
-
GLED30
-
-
-
GLED22
-
-
GLED21
-
GLED20
-
-
-
GLED11
-
-
-
GLED10
-
-
-
-
IG-GLEDSTD
Installation, wiring and adjustment instructions
3.2
Selecting the correct ceiling anchor ring
!
The ceiling anchor ring is used to adjust the levelling of the lamp stem to ensure that the lamp is perfectly vertical. It is essential that the mounting is both secure and perpendicular to ensure safe and trouble-free operation as the consequences are potentially serious.
It is recommended you use ceiling anchor ring, ie Part No 12981, shown in the illustration.The anchor ring is secured in position during building works so that the lamp can be fitted easily at a later stage and without risk of damage.
3.3
Refer to Section 3.3.
Mounting the flange plate (stem) Before mounting the stem tube, plan for a safe routing and connection of the 48V supply cables and run in any draw wires that might be required. Where the room has a false ceiling, the usual method is to fasten the mounting flange to the structural surface above the false ceiling so that the bottom of the stem tube is the only partly protruding into the room as shown in the illustration.
FALSE CEILING
FLANGE PLATE (STEM)
IG-GLEDSTD
Page 17 of 34
Galaxy Ultra HD-LED Standard Models
3.3.1
Identifying the flange plate (stem) Identity whether the stem is adjustable – mounting flange and mounting stem are separate – or made-to-measure.
3.3.2
If the stem is adjustable refer to Adjustable Stem Instructions for Use.
Fixing the flange plate (stem) and electrical isolation The light installation must be electrically isolated using the MEIGaN insulation washers provided in order to meet MEIGaN, IEE Guidance Note (Chapter 10) and IEE 60364-7-710 for Electrical Installation in Medical Locations. The lamp mounting plate/stem must be securely fastened to either the ceiling anchor ring or directly to the structural surfaces using 6-off M16, or equivalent fixings.
!
It is essential that the mounting is both secure and perpendicular to ensure safe and trouble-free operation as the consequences are potentially serious.
To adjust the level of the stem, adjust the half nuts up or down the threads and resecure using the M16 full nuts. The typical MEIGaN fixing is shown below.
Page 18 of 34
IG-GLEDSTD
Installation, wiring and adjustment instructions
3.3.3
Assembling the ceiling cover Start by sliding the ceiling cover support ring over the stem and secure it temporarily in place using the 3 x M4 grub screws supplied. Next, join the two halves of the ceiling cover together around the stem and secure using the M5 bolts provided. Slide the ceiling plate cover over the flange plate followed by the ceiling cover support ring. Secure with M4 grub screws.
!
Do not overtighten the M4 grub screws. You are able to fit the ceiling cover later, but you must install the ceiling cover support ring now.
SPLIT CEILING
M5 HEX DRIVE FIXING BOLTS (Qty 4) M4 HEX DRIVE GRUB SCREWS (Qty 3) CEILING COVER SUPPORT RING
IG-GLEDSTD
Page 19 of 34
Galaxy Ultra HD-LED Standard Models
3.4
Fixing the articulation arm assembly (structure)
This instruction applies to all models except GLED40, GLED30, GLED20 and GLED10. Refer to Section 3.6 for the relevant instruction. Offer up the articulation arm assembly spindle to the flange plate stem tube and draw the lamp supply cables through the flange plate to the point of connection. Insert the lamp mechanism spindle into the bottom of the stem tube. Ensure the fixing holes align with the threaded holes in the lamp spindle. Secure with the 8 x M8 hex drive countersunk head screws supplied.
!
Page 20 of 34
The 8 screws must be thread-locked using a proprietary threadlock adhesive, such as Loctite 243 or PermaBond A1042.
IG-GLEDSTD