Breas Medical Ltd

BREAS Nippy Systems

Nippy Clearway Functional Earth Testing Guide

Testing Guide

2 Pages

B & D Electromedical LLP Unit A2 The Bridge Business Centre Timothy’s Bridge Road Stratford-upon-Avon Warwickshire CV37 9HW  VENTILATORS & BREATHING SYSTEMS MD 558596  Nippy Clearway Functional Earth The Clearway is a class II device, equipped with a functional earth which is used for EMC screening. This is not a protective earth connection. Prevention of electric shock is provided by double insulation in accordance with EN60601-1: 2006 Sub clause 8.6.9 of EN60601-1-1 allows for a functional earth in class II equipment having the required isolation of the functional earth/screen and mains parts. The isolation of the screen in the Nippy Clearway complies with all relevant clauses of EN60601-1. Details of insulation and separation distances can be found in TUV test report 071-75917995-000 Extract from test report showing measured separation distances and insulation diagram:-.  Proprietors: P J Bachelor & N M Davis
File Type: PDF
File Size: 440 KB
File Name: Nippy Clearway Functional Earth Guide Testing.pdf