Technical Data Sheet
8 Pages
Preview
Page 1
BRUMABA | Technical Data | Operating Table CL GENIUS 02 // 08
Contents
1. TECHNICAL SPECIFICATIONS 1.1. Dimension sheet 1.2. Dimensions and weight 1.3. Connected loads 1.4. Ambient conditions 1.5. Performance data 1.6. Type plate 1.7. FDA - Establishment Registration
Page
2 2 2 3 3 4 5
2. INFORMATION ON ELECTROMAGNETIC COMPATIBILITY (EMC) 5
1. Technical data
Dimension sheet BRUMABA CL GENIUS DTA, BRUMABA CL GENIUS DTN
1.1. Dimension sheet
Fig. 1.1.A: Dimension sheet BRUMABA CL GENIUS DTA, BRUMABA CL GENIUS DTN
1.2. Dimensions and weights
Fig. 1.1.A: Dimension sheet BRUMABA CL GENIUS STA, BRUMABA CL GENIUS STN
SPECIFICATION
VALUE
Weight
352.7 lbs
160 kg
Weight
308.6 lbs
140 kg
Weight (BRUMABA CL GENIUS STN)
264.6 lbs
120 kg
Length
71 – 78.41 in
1803,5 – 1991,5 mm
Width
22.24 in
565 mm
Height (BRUMABA CL GENIUS STA, BRUMABA CL GENIUS STN) excluding padding
20.67 – 32.09 in
525 – 815 mm
Height (BRUMABA CL GENIUS DTA, BRUMABA CL GENIUS DTN) excluding padding
18.31 – 35.24 in
465 – 895 mm
Height padding (Visco padding)
2.17 in 2,56 in
55 mm 65 mm
SPECIFICATION
VALUE
Nominal voltage
100 – 240 V AC
Nominal frequency
50 – 60 Hz
Rated power consumption
Max. 350 W
Protection class
I
Applied part
Type B
(BRUMABA CL GENIUS DTA)
(BRUMABA CL GENIUS STA,BRUMABA CL GENIUS DTN)
1.3. Connected loads Electric (BRUMABA CL GENIUS STN, BRUMABA CL GENIUS DTN)
02
Power supply unit for plug-in version
Operating Table CL GENIUS | Technical Data | BRUMABA
Operating Table CL GENIUS | Technical Data | BRUMABA 03 // 08
Electric (for battery version) (BRUMABA CL GENIUS STA, BRUMABA CL GENIUS DTA)
Battery
SPECIFICATION
VALUE
Nominal voltage (input)
100 – 240 V AC
Nominal frequency
50 – 60 Hz
Max. power consumption (input) Nominal power consumption
1.6 A
Protection class
II
SPECIFICATION
VALUE
160 W
Battery capacity
22 Ah
Operating time of a new battery fully charged (depends on the type of use)
1 – 2 weeks
Nominal voltage
24 V DC
1.4. Ambient conditions In service
Storage and transport
Load capacity
1.5. Performance data Adjustment range (± 10 %)
SPECIFICATION
VALUE
Temperature
+10 °C to + 40 °C
Relative humidity
30 % – 75 %
Air pressure
700 hPA – 1060 hPa
SPECIFICATION
VALUE
Temperature
-20 °C to +70 °C
Relative humidity
30 % – 95 %
Air pressure
500 hPA – 1060 hPa
10.15 psi – 15.37 psi
-4 °F to +158 °F 7.25 psi – 15.37 psi
PARAMETERS
VALUE
Maximum load capacity of the operating table (maximum patient weight)
250 kg, 551 lbs
Maximum load capacity of the head section
50 kg, 110 lbs
Maximum load capacity of the foot section
80 kg, 176 lbs
PARAMETERS
VALUE
Inclination of the head section
from +30° to -45°
Inclination of the chest support
from +80° to -22°
Inclination of the chest support
from +80° to -11°
Inclination of the seat section
from +25° to -9°
Inclination of the seat section
from +25° to -9°
Inclination of the lower leg section
from +28° to -90°
Inclination of the lower leg section
from +28° to -90°
(BRUMABA CL GENIUS STA, BRUMABA CL GENIUS STN)
(BRUMABA CL GENIUS DTA, BRUMABA CL GENIUS DTN)
(BRUMABA CL GENIUS STA, BRUMABA CL GENIUS STN)
(BRUMABA CL GENIUS DTA, BRUMABA CL GENIUS DTN)
(BRUMABA CL GENIUS STA, BRUMABA CL GENIUS STN)
(BRUMABA CL GENIUS DTA, BRUMABA CL GENIUS DTN)
Speed modes
+50 °F to +104 °F
PARAMETERS
VALUE
Head section
4 – 6 °/s
Chest support
1,7 – 7,6 °/s
Seat section
2,1 – 8,4 °/s
Foot section
2,7 – 11,27 °/s
Height
9,6 – 12,1 mm/s
BRUMABA | Technical Data | Operating Table CL GENIUS
03
BRUMABA | Technical Data | Operating Table CL GENIUS 04 // 08
1.6. Type plate
The type plate is located on the telescopic column between the battery case or power supply housing and control casing and contains the following information: • Manufacturer’s con• Maximum load capacity tact information • Country of manufacture • Type • Production year • Connected loads • CE marking
Fig. 1.6.A: Type plate B RUMABA CL GENIUS STA
Fig. 1.6.B: Type plate BRUMABA CL GENIUS STN
Fig. 1.6.C: Type plate B RUMABA CL GENIUS DTA
Fig. 1.6.D: Type plate B RUMABA CL GENIUS DTN
The type plate is located on the telescopic column between the battery case or power supply housing and control casing and contains the following information:
SYMBOL/ABBREVIATION
MEANING This symbol indicates the position of the product's manufacturer address.
SN
Serial number
REF
Order number
Protection rating IP X4 ED 20 %
(l=2 min/0=8 min)
Device is resistant to water splashes from any direction The duty cycle is 20 % of the cycle duration. This means that the device could be operated for approx. 2 minutes under the design load and must then pause for approx. 8 minutes. Type B applied part. This applied part complies with the standard requirements to protect against electric shock, particularly taking the permissible patient leakage current and patient auxiliary current into consideration. Attention! Type B applied parts are not suitable for direct usage on the heart. This symbol indicates that the instruction manual must be observed. This symbol indicates the production date of the operating table. This symbol indicates that the operating table or some of its parts may not be disposed of with household waste. Pb: Battery contains more than 0.004 percent lead by weight IEC protection class II
04
Operating Table CL GENIUS | Technical Data | BRUMABA
Operating Table CL GENIUS | Technical Data | BRUMABA 05 // 08
SYMBOL/ABBREVIATION
MEANING
MD
Medical Device The symbol indicates the tare weight of the operating table and the max. working load EU conformity marking
The medical device must not be disposed of with household waste
(11) Date of manufacture (21) Serial number (01) UDI-DI (product identification number)
1.7. FDA – Establishment The FDA – Establishment Registration Number for BRUMABA Ltd is: 3005607354 Registration
2. Information on electromagnetic compatibility (EMC) EMC – Electromagnetic emissions
Avoid using this device in the immediate vicinity of other devices or stacked with other devices because this could cause faulty operation. However, if usage in the above-mentioned manner is necessary, observe this device and the other devices, to ensure that they are working properly. Electromagnetic faults can cause malfunctions that can lead to device defects. Please direct your questions regarding operation within a valid EMC environment to BRUMABA.
EMISSIONS TEST
COMPLIANCE
RF emissions as per CISPR 11
Group 1
RF emissions as per CISPR 11
Class B
ELECTROMAGNETIC ENVIRONMENT - GUIDELINE The “BRUMABA CL GENIUS” exclusively uses RF energy for its internal functions. Therefore, its RF emissions are very low and it is unlikely that it will cause interference with other devices in its vicinity.
Guidelines and Manufacturer’s Declaration – Electromagnetic Interference Immunity
The “BRUMABA CL GENIUS” is intended for use in an environment as specified below. The customer or user of the “BRUMABA CL GENIUS” should ensure that it is operated in such an environment.
INTERFERENCE IMMUNITY CHECKS
IEC 60601 TEST LEVEL
Conducted high frequency interferences according to IEC 61000-4-6
3V 0,15 MHz – 80 MHz 6 V in ISM between 0.15 MHz and 0 MHz 80% AM at 1 kHz
Radiated high frequency interferences according to IEC 61000-4-3
3 V/m 80 MHz to 2,5 GHz
INTERFERENCE IMMUNITY FOR AMBIENT FIELDS OF DEVICES WITH HIGH FREQUENCY RADIO COMMUNICATION 28 V/m 450 MHz, ±5 kHz FM, 1 kHz sinus 810 MHz, 50% PM at 18 Hz 870 MHz, 50% PM at 18 Hz 930 MHz, 50% PM at 18 Hz 1720 MHz, 50% PM at 217 Hz 1845 MHz, 50% PM at 217 Hz 1970 MHz, 50% PM at 217 Hz 2450 MHz, 50% PM at 217 Hz 27 V/m 385 MHz, 50% PM at 18 Hz
9 V/m 710 MHz, 50% PM at 217 Hz 745 MHz, 50% PM at 217 Hz 780 MHz, 50% PM at 217 Hz 5240 MHz, 50% PM at 217 Hz 5500 MHz, 50% PM at 217 Hz 5785 MHz, 50% PM at 217 Hz
With P as the rated output of the transmitter in Watts (W) according to the specifications of the transmitter's manufacturer and at the recommended safety distance in meters (m). The field strength of stationary radio transmitters should be lower than the compliance level for all frequencies based on an on-site examination. Interference is possible in the vicinity of devices that have the following pictogram. REMARK 1: The higher frequency range applies at 80 MHz and 800 MHz. REMARK 2: These guidelines may not be applicable in all cases. The propagation of electromagnetic factors is impacted by the absorption and reflection of buildings, objects, and humans. The field strength of stationary transmitters such as base stations for radio telephones and handheld radio devices, amateur radio stations, AM and FM radio and TV stations can theoretically not be precisely pre-determined. To determine the electromagnetic environment regarding the stationary transmitters, an examination of the installation site should be considered. If the measured field strength in the place in which the “BRUMABA GENIUS” is used exceeds the compatibility levels named above, then the “BRUMABA GENIUS” should be observed, to verify proper functioning. If unusual performance characteristics are exhibited, additional measures may be required, e.g. changing the orientation or a different location of the “BRUMABA CL GENIUS”
BRUMABA | Technical Data | Operating Table CL GENIUS
05
BRUMABA | Technical Data | Operating Table CL GENIUS 06 // 08
Guidelines and Manufacturer’s Declaration - Electromagnetic Interference Immunity The “BRUMABA GENIUS” is intended for use in an environment as specified below. The customer or operator of the “BRUMABA GENIUS” should ensure that it is operated in such an environment.
INTERFERENCE IMMUNITY CHECKS
COMPLIANCE
Electrostatic discharge (ESD) according to IEC 61000-4-2
Contact: ±8 kV Air: ±2 kV; ±4 kV; ±8 kV; ±15 kV
Magnetic field of the supply frequency (50/60 Hz) according to IEC 61000-4-8
30 A/m 50 & 60 Hz
Electrical fast transient interferences /bursts according to 61000-4-4
± 2 kV ±1 kV for signal cable 100 kHz repetition frequency
Surges according to IEC 61000-4-5
L-N (sym) ±0.5 kV; ±1 kV L-PE and N-PE (unsym.) ±2 kV
Voltage drops, short interruptions and fluctuations of the supply voltage according to IEC 61000-4-11
0 % UT, 0.5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315° 0 % UT, 1 cycle and 70% UT; 25/30 cycles Individual phase: at 0° and 0 % UT; 250/300 cycles
ELECTROMAGNETIC ENVIRONMENT - GUIDELINES
Flooring should be made of wood, concrete, or tile. If the floor is covered with synthetic material, the relative humidity must be at least 30 %. Magnetic fields of the grid frequency should correspond to the standard values found in industrial and hospital environments.
Recommended safety distances between portable and handheld RF telecommunication devices and the “BRUMABA GENIUS” The “BRUMABA GENIUS” is intended for use in an electromagnetic environment in which the high frequency interferences are controlled. The customer or operator of the “BRUMABA GENIUS” can help in preventing electromagnetic interferences by maintaining the minimum distance of 30 cm between portable and handheld RF telecommunication devices (transmitters) and the “BRUMABA GENIUS” –depending on the output of the communication device, as described below.
INSTRUCTIONS AND MANUFACTURER’S DECLARATION – EMC The proper operation of the “BRUMABA CL GENIUS” is ensured within the ambient conditions described below. The customer or operator should ascertain that the required ambient conditions are maintained. Emissions test
Compliance
RF emissions as per CISPR 11
Group 1
RF emissions as per CISPR 11
Class B
Harmonic emissions according to IEC 61000-3-2
Class A
Voltage fluctuations/ flicker emissions according to IEC 610001-3-3
complies
06
Operating Table CL GENIUS | Technical Data | BRUMABA
Electromagnetic Environment - Guidelines RF energy is only needed for the internal functions of the operating table “BRUMABA CL GENIUS”. Therefore, its RF emissions are very low and it is unlikely that it will cause interference with other devices in its vicinity. Please use the device as intended. The device can be connected to the public low-voltage grid.
Operating Table CL GENIUS | Technical Data | BRUMABA 07 // 08
BRUMABA | Technical Data | Operating Table CL GENIUS
07