care of sweden
Pressure Care Mattresses
CuroCell A4 -IQ Cirrus Air mattress systems Instructions for Use June 2022
Instructions for Use
25 Pages
Preview
Page 1
Instructions for use 2022-06-27
CuroCell® A4/IQ Cirrus Air mattress systems
Instructions for use item number: 95-001455-EN0000
EN
4.14 Maximum pressure notifications ... 9
Table of contents WARNINGS AND SAFETY PRECAUTIONS ... 1
4.15 Table of notifications ... 9 5
Operation CuroCell® IQ ... 11
1 Introduction ... 3
5.1 Function ... 12
1.1 General information ... 3
5.2 Program... 13
1.2 Intended purpose... 3
5.3 Maximum pressure (caring mode). 13
1.3 Intended user ... 3
5.4 Hand check (function control) ... 13
1.4 Intended use environment ... 3 1.5 Indications ... 3
5.5 CPR (Cardiopulmonary resuscitation) 13
1.6 Contraindications ... 3
5.6 Transport function ... 13
1.7 Clinical benefit ... 3
5.7 Pack&Go® ... 13
2 Assembly and installation ... 4
5.8 Restart ... 13
3 Common operations ... 5
5.9 Power failure ... 13
3.1 Sitting positioning in bed ... 5
5.10 Notifications ... 13
3.2 CPR (Cardipulmonary resuscitation) 5
5.11 Maximum pressure notifications. .. 13
3.3 Transport function ... 5
5.12 Table of notifications ... 14
3.4 Restart ... 5
6 Product description ... 15
3.5 Maximum pressure notification... 5
6.1 Control unit (A4/IQ) ... 15
3.6 Power failure ... 5
6.2 Mattress ... 16
3.7 Automatic Setting ... 5
7 Reuse and cleaning ... 17
4 Operation CuroCell® A4 Cirrus ... 6
7.1 Cleaning ... 17
4.1 Function ... 8
8 Storage ... 18
4.2 Programs ... 8
9 Maintenance ... 18
4.3 Maximum pressure (caring mode) ... 8
9.1 General ... 18
4.4 Panel lock... 8
9.2 Between users ... 18
4.5 Comfort settings ... 8
9.3 Replacing the air filter ... 18
4.6 Hand check (function control)... 9
10 Troubleshooting ... 19
4.7 Sitting positioning in bed ... 9
11 Technical specification ... 20
4.8 CPR (Cardiopulmonary resuscitation) 9
11.1 Standards ... 20
4.9 Transport function ... 9 4.10 Pack&Go® function ... 9
11.2 Symbolkey ... 21 12 Other information ... 22
4.11 Restart ... 9
12.1 Recommended lifetime of the product 22
4.12 Power failure ... 9
12.2 Disassembly and recycling ... 22
4.13 Notifications ... 9
12.3 Returns ... 22
WARNINGS AND SAFETY PRECAUTIONS Read all instructions before use or repair
WARNING! To minimize the risk of fire, personal injury and equipment/property damage adhere to the following instructions:
1. The product must only be installed and used for its intended purpose according to the instructions in this manual and/ or other documentation from Care of Sweden. The product must not be combined, assembled or repaired with parts (e.g. control unit and mattress), accessories or spare parts other than those described in this manual or other documentation from Care of Sweden. The product must not be modified in any way. 2. The product must be placed and used so that it does not become trapped or damaged. Be particularly aware of trapping damage when using side rails. Always make sure that the mattress is the correct size for the bed. 3. Regularly check product functionality by performing a hand check. 4. When the product is used for individuals needing special supervision, such as children, continuous monitoring is required. 5. The mattress is protected by a hygiene cover; avoid using multiple hygiene covers as this can affect the vapor permeability of the mattress. 6. The hygiene cover does not allow liquid or air to penetrate, but is vapour permeable. Make sure that the user is positioned correctly to avoid the risk of suffocation. 7. Be careful with sharp objects to prevent damage to the hygiene cover. 8. Do not open the control unit housing – risk for electric shock. Servicing and maintenance must be performed by Care of Sweden or one of its authorized service technicians. 9. Route the power cable to the control unit carefully to avoid tripping. Also make sure that the user is lying correctly on the mattress according to the instructions and use a cable holder if possible. 10. To avoid the risk of strangulation, make sure that the cable and tubes are routed to prevent someone getting caught up in them.
1
11. Do not use close to or in contact with fire sources/hot surfaces, such as fire, burning cigarettes, hot lamps, heating fans or heating stoves/open fires as this could damage the product.
12. Do not store or use the product in direct sunlight. The product may be damaged by the elevated temperature and UV light.
13. Strong magnetic fields or wireless communication equipment (e.g., wireless home network
products, mobile phones, walkie-talkies, cordless phones and their base stations, radio transmitters, etc.) may affect the product’s functionality and should be kept at a distance of at least 1 meters from the control unit.
14. Never use the product if the power cable, plug of the control unit or power supply housing is defective, if the control unit housing is damaged, or if it is not functioning properly. Contact an authorized service technician for examination and repair.
15. Never connect anything other than the Care of Sweden supplied power supply to the control unit power cable connector.
16. Never use the external communication input (3,5mm connector), this input should only be used by the manufacturer. 17. If the hygiene cover is equipped with side handles, these are intended for managing or
relocating the mattress. Do not use the handles to lift the mattress with the user lying on it. All other use takes place under your own liability and is not covered by the product warranty.
18. To prevent the power supply from being pulled out, exercise caution when there are children and pets in the environment around the equipment.
19. Use of this product adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this product and the other equipment should be observed to verify that they are operating normally.
20. To minimize the risk of wounds occurring on the feet, make sure that the user doesn’t come into contact with the hangers of the control unit.
2
1
1.4
Introduction
Intended use environment
The mattress system can be used in all kinds of health care environments, including home care.
These air mattress systems may be used as an aid to prevent and treat pressure ulcers/pressure injuries.
1.5 Indications
Suitable for a wide range of persons with CuroCell® A4 is a control unit that weighs the user increased risk for pressure ulcers/injuries, and adjust the pressure. It offers the operator the including those with superficial ulcers, up to option to adjust program and comfort. category 4 and unclassified PU/PI (unstageable CuroCell® IQ is a control unit that weighs the user and suspected deep tissue injury) in association and adjust the pressure without an option for the with an individualized plan of care. operator to choose program. Prescription shall be made by persons authorized Always read the instructions for use prior for prescriptions and with clinical education. Note that the user also may be the operator. to use.
1.1
General information
1.6
Contraindications
1.7
Clinical benefit
There are no known contraindications. The treating doctor assess and prescribe at each individual case.
The system is a medical device with CE marking in accordance with MDR (EU) 2017/745. According to this regulation the manufacturer is required to report all accidents or incidents involving the products. All information involving accidents or incidents relating to our products, shall be reported immediately to Care of Sweden.
The clinical benefits for M4, A4 and IQ together with any of the mattresses included in this instruction for use are: • Prevention and treatment of pressure 1.2 Intended purpose The mattress system consists of a control unit and ulcers/pressure injuries up to and including category 4 and unclassified PU/PI. a mattress and is intended to be used for • Reduction of shear forces. prevention and as an aid in the treatment of • Safety, comfort, pressure redistribution and pressure ulcers/pressure injuries (PU/PI). easy to handle. 1.3 Intended user • Silent running control units. The mattress system is intended to be used by all kind of users, including lay persons. Note that the Note! user and operator could be the same person. • For certain users, e.g., amputees, the The mattresses are intended for use by persons of recommended length measurement may not a recommended minimum length of 120 cm. be reached. Users in these groups may require other settings as the entire surface is not under load. • The mattress may be inappropriate for use during x-ray examinations because of the risk of blurred images or artefacts that may lead to diagnostic errors. • In the event of a power loss or similar, the mattress will retain air for at least 12 hours.
3
2 Assembly and installation
Check that no parts are damaged. If damage is found, contact Care of Sweden or your local distributor before using the product. Do not use sharp objects when unpacking as it might damage the product. 1. Place the mattress on the bed base. Secure the mattress to the bed using the fastening straps on the underside of the mattress.
jammed when raising or lowering the bed. 4. Open the lid on the air tube connector (marked CPR) and connect it to the side of the control unit.
If the mattress is used on an adjustable bed, the fastening straps at the middle of the mattress shall be fastened at the bed’s unmovable parts. The straps at the head end and foot end shall be fastened in the parts of the bed that may be raised or lowered.
5. A click is heard and felt when correctly connected. Secure that both sides of the connection are closed. 6. Check that the switch on the side of the control unit is set to ’0’ (off). Plug the power supply into an approved and easily accessible electrical socket (100– 240 V).
Note! • Make sure that the mattress is the correct size for the bed. • Make the bed with sheets. • Check the air cells to ensure they are correctly assembled. • The mattress should only be used lying in the lengthwise direction on the mattress with the feet at the end, marked with the feet symbol.
7. Check that the power cable has been correctly connected to the control unit and that the correct power supply has been used. (See Technical specification, section 11). The correct P/N is shown on the label on the power supply. The power supply is part of the equipment and may not be replaced. If the control unit has been stored in its minimum or maximum storage temperature (min -25°C, max 70°C), wait at least 1 hour before starting it. This time is based on an ambient temperature of 20°C.
2. Hang the control unit on the foot end of the bed or place it on a level, steady surface. 3. If the mattress is equipped with a cable holder, place the power cable in the cable holder by opening the press studs, placing the cable in the gap and closing the press studs again. Otherwise, place the power cable so there is no risk of stumbling over it, running over it with the bed wheels, or getting it
Note! • Do not hold the 12V plug on the power supply while touching the user.
4
3
Common operations
3.1
Sitting positioning in bed
Or Remove the power supply from the wall socket and leave the control unit hanging on the bed during transport. The mattress will retain air for at least 12 hours. We recommend using this function for short periods only.
Following operations apply to the Cirrus system regardless of which control unit that is used. When raising the head end of the bed into a sitting position, always secure the user´s position. To ensure the product functionality, we always recommend to perform a hand check (see section 4 or 5 depending on the pump and mattress of interest). This function is recommended to use for short periods only. For additional support, positioning pillows can be used.
3.5
Maximum pressure notification
3.6
Power failure
3.7
Automatic Setting
When the function Maximum pressure has been used for a long time, the Maximum pressure diode will blink. If the use is intentional, ignore the notification. In the event of a power failure, unplug the CPR connection, close the lid, place the CPR connection at the end of the bed and remove the control unit from the bed. The mattress will retain air for at least 12 hours. Perform a hand check to make sure the pressure of the mattress is not too hard or too soft.
• When the alternating or pulsating program is used and the head end of the bed is raised, make sure that the user and/or the mattress is not moving downwards due to the movement in the mattress. Also raise the foot end of the bed.
The mattress system regulates the mattress's internal pressure independently and without manual adjustment to different values based on the user's weight and position. No manual action is required to affect the internal pressure of the mattress. This function works in the following three ways: 1. The control unit carries out an automatic setting immediately after switching the system on. 2. During usage, if any significant change occurs, the control unit will perform an automatic setting. 3. When in use, the control unit will carry out automatic settings at fixed intervals to ensure correct control of the pressure in the mattress at all times. Once the automatic setting has completed, the control unit will start in the Pulsating mode.
CPR (Cardipulmonary resuscitation)
In case of an emergency where CPR (Cardiopulmonary resuscitation) is necessary, remove the connection (marked ‘CPR’) from the control unit and leave the lid open in order to empty the mattress of air quickly.
3.3
Restart
If a restart is required, set the On/Off switch on the side of the control unit to 0 (Off). Wait for approx. 10 seconds and restart the control unit.
Note! • When using a lift to place the user in the bed and the head end of the bed is raised, make sure that the user is not placed too high on the mattress. Otherwise, there is a risk of shear.
3.2
3.4
Transport function
If the user needs to be moved in bed, either: Unplug the CPR connection, close the lid, place the CPR connection at the end of the bed and remove the control unit from the bed. The mattress will retain air for at least 12 hours.
5
CuroCell® A4 Cirrus
4 Operation CuroCell® A4 Cirrus Operation CuroCell® A4 Cirrus (how to get started) 1. Set the On/Off switch on the side of the control unit to position 1 (On). See figure 1. 2. The mattress starts to inflate. This takes about 20 minutes depending on the size of the mattress. While the mattress is inflating, the ”mute the information signal” and ”information signal” diodes light up in orange. When these diodes goes out, the user can be placed on the mattress. 3. The pulsating mode setting is pre-set. If a different setting is desired, it can be selected when the user is placed on the mattress. 4. The control unit sets the inner pressure of the mattress according to the weight, length, and position of the user. This takes about 20 minutes. During this time, the diode above the selected program flashes. When the diode stops flashing, the mattress is set and the inner pressure has adjusted according to the user. 5. Perform a hand check to ensure that the settings are correct.
Note! • During the automatic setting, try to avoid larger movements on the mattress. Otherwise, you will get a notification on that the desired value could not be reached within the time limit and the weighing must start over. • The mattress must be inflated before the user can lie on the mattress.
Figure 1. On/Off switch.
• Once automatic setting is complete, the control unit switches to a basic setting of Pulsating mode (when used for the first time) or to the previous setting.
6
CuroCell® A4 Cirrus
Button
Function Mute the information signal Panel lock Pack & Go®. Function for deflating the system Alternating mode Pulsating mode CLP mode (Constant low pressure) Maximum pressure (caring mode) Information signal Incorrect connection of the air connector (CPR) Cycle time settings (10, 15, 20, 25 minutes). The diodes are also used for error notifications. Comfort settings
7
CuroCell® A4 Cirrus 4.1
Function
The mattress system independently and without manual adjustment controls the inner pressure of the mattress according to the weight, height, and position of the user. No manual action needs to be performed to adjust the inner pressure of the mattress. This function works as follows:
3. Pulsating mode. The cycle period can be changed according to the user needs and requirements. Choose between 10, 15, 20 or 25 minutes. The longer the cycle period, the slower the alternations. We recommend a basic setting of 10 minutes.
1. At start-up, automatic setting of the inner pressure of the mattress is always carried out according to the weight and height of the user. 2. If the user moves noticeably or changes their position, the system will independently control the inner pressure of the mattress. 3. The system performs an automatic setting at fixed intervals even if no significant changes have occurred. After automatic setting of the mattress inner pressure, the system returns to the previously selected program. At start-up, Pulsating Mode is always pre-set.
4.3
Maximum pressure (caring mode)
4.4
Panel lock
4.5
Comfort settings
4.2
With this function, the entire mattress is inflated and provides firmed support. This function reverts to the previous setting after 20 minutes. The function should be used when caring the user, shifting the user’s position or moving the user in or out of bed. Press the Panel Lock button to lock or unlock the control panel. The button indicates when the panel has been locked. The screen locks automatically if left untouched for five minutes. This is to prevent the settings being changed accidentally. To unlock, press the button for 2 seconds.
Programs
There are three programs to choose from: CLP means that the air pressure in all of the air cells is the same in the whole mattress. Alternating mode means that the air pressure in the air cells are different and alternates regularly after chosen cycle periods. Pulsating mode combinates CLP with the alternating function. Choose program by pushing the button for the program. We recommend the Pulsating mode which is also the preset mode.
The pressure can be increased in two steps depending on the user’s comfort requirements. This increase is made based on the automatic setting in 5.1. Note! • When only parts of the mattress are under load, for example, in the case of amputees – it may be necessary to raise the setting using the comfort settings.
1. Constant low pressure mode (CLP) No cycle period is needed. 2. Alternating mode. The cycle period can be changed according to the user needs and requirements. Choose between 10, 15, 20 or 25 minutes. The longer the cycle period, the slower the alternations. We recommend a basic setting of 10 minutes.
The selected setting is shown by a green light.
8
CuroCell® A4 Cirrus
4.6
The Pack&Go® diode will flash during deflation. The mattress will empty of air and be ready to be simply folded together in about 20 minutes. The control unit gives an audio signal once deflation is complete.
Hand check (function control)
Hand check is performed to ensure that the mattress system works properly. Hand check should be performed regularly; for CuroCell® A4 Cirrus, we recommend once per work shift as well as after installation of the system.
Carefully fold the mattress in half, place the control unit between the folds of the mattress and place the system in a transportation bag (accessory) or equivalent for protective storage. Ensure that the power supply is packed complete.
Note! • Make sure that the mattress system is filled, which is shown by a green light from the diode, before performing a hand check. 1. Open the cover and insert a hand, with the palm facing up, between the mattress and the bed base. The hand shall be inserted beneath the user’s sacrum (center of mattress).
4.8 CPR (Cardiopulmonary resuscitation) See 3.2
Transport function
4.10
Pack&Go® function
See 3.3
Power failure
4.13
Notifications
4.14
Maximum pressure notifications
4.15
Table of notifications
Different notifications exist based on how serious the warning is. With a malfunction or an error, a notification will be given by way of a flashing warning triangle. To mute the warning signal, press the mute button. When a notification occurs, the current cycle time diode will turn off and a notification code is shown on the four different cycle time diodes (10, 15, 20, 25). To read the cycle time during the error notification, unlock the control panel.
Sitting positioning in bed
4.9
4.12
See 3.6
3. If you can feel the user’s sacrum resting in the palm of your hand, the gap is too small. See section 8 ‘Troubleshooting’. See 3.1
Restart
See 3.4
2. Ensure there is a gap between the user and the bed base so that the user does not ‘bottom out’.
4.7
4.11
See 3.5
The product may be packed as follows: Ensure that no-one is lying on the mattress.
Information about each notification is shown in the notification table: Notifications from 1-10 are both audible and visual. The error code will be displayed until the error has been rectified. If the mute button is pressed the audible warning will cease for a period of 5 minutes and will return until the error has been rectified. The notifications 11-13 have no audible alarm. The error code is shown until the system is restarted.
• On the control panel, press the lock/unlock button. • Press the Pack&Go® button and hold it down for 2 seconds
9
CuroCell® A4 Cirrus Notification
Description and troubleshooting
1
High temperature. Valves and compressors are turned off. If the control unit is in direct sunlight, relocate it.
2
Default settings are not completed.
3
Incorrect input voltage. Make sure that the correct power supply is used, otherwise contact the technical support.
4
Low pressure. Secure the CPR, mattress, air tubes and air filter.
5
Automatic setting failure. The correct pressure has not been reached within the time limit.
6
Leakage under the automatic setting period. The mattress is leaking too much for the weighing to be accomplished. Control the mattress and the air connections.
7
High pressure. The pressure cannot be reduced to the desired value within the time limit.
8
The automatic setting has been restarted too many times during the automatic setting period.
9
The mattress control parameters have not been read. Connect the CPR.
10
The mattress control parameters have been changed during the use. Restart the system.
11
12
13
Leakage in one of the sections. Secure the CPR, mattress and connection tubes. Information to service technicians: this notification shows a leakage in the blue air cell section. More information in the Servicemanual for CuroCell® A4. Leakage in one of the sections. Secure the CPR, mattress and connection tubes. Information to service technicians: this notification shows a leakage in the green air cell section. More information in the Servicemanual for CuroCell® A4. Leakage in one of the sections. Secure the CPR, mattress and connection tubes. Information to service technicians: this notification shows a leakage in the red air cell section. More information in the Servicemanual for CuroCell® A4.
10
CuroCell® IQ Cirrus
5
Operation CuroCell® IQ Operation CuroCell® IQ (how to get started) 1. Set the On/Off switch on the side of the control unit to position 1 (On). See figure 1. 2. The mattress starts to inflate. This takes about 20 minutes depending on the size of the mattress. While the mattress is inflating, the “silent information signal” and “information signal” diodes light up in orange. When these diodes goes out, the user can be placed on the mattress. 3. The control unit sets the inner pressure of the mattress according to the weight, length, and position of the user. This takes about 20 minutes. During this time, the diode above the “check symbol” flashes. When the diode stops flashing, the mattress is set and the inner pressure has adjusted according to the user. 4. Perform a hand check to ensure that the settings are correct.
Note! • During the automatic setting, try to avoid larger movements on the mattress. Otherwise, you will get a notification on that the desired value could not be reached within the time limit and the weighing must start over. • The mattress must be inflated before the user can lie on the mattress.
Figure 1. On/Off switch.
• Once automatic setting is complete, the control unit switches to a basic setting of Pulsating mode (when used for the first time) or to the previous setting.
11
CuroCell® IQ Cirrus
Button
Function Mute the information signal Pack & Go®. Function for deflating the system Maximum pressure (caring mode) Information signal Incorrect connection of the air connector (CPR)
Check-symbol. System is ready to use Notification diodes
5.1
Function
The mattress system independently and without manual adjustment controls the inner pressure of the mattress according to the weight, length, and position of the user. No manual action needs to be performed to adjust the inner pressure of the mattress. This function works as follows: 1. At start-up, automatic setting of the inner pressure of the mattress is always carried out according to the weight and length of the user. 2. If the user moves noticeably or changes their position, the system will independently control the inner pressure of the mattress. 3. The system performs an automatic setting at fixed intervals even if no significant changes have occurred. After automatic setting, the system returns to Pulsating Mode.
12
CuroCell® IQ Cirrus 5.2
Program
Pulsating mode with a cycle period of 10 minutes is preset.
5.3
Maximum pressure (caring mode)
With this function, the entire mattress Is inflated and provides firmed support. This function reverts to the previous setting after 20 minutes. The function should be used when caring the user, shifting the user’s position or moving the user in or out of bed.
5.4
5.7
Hand check (function control)
Hand check is performed to ensure that the mattress system works properly. Hand check should be performed regularly; for CuroCell® IQ Cirrus, we recommend once per work shift as well as after installation of the system. Note! • Make sure that the mattress system is filled, which is shown by a green light from the diode, before performing a hand check. 1. Open the cover and insert a hand, with the palm facing up, between the mattress and the bed base. The hand shall be inserted beneath the user’s sacrum (center of mattress).
Pack&Go®
The product may be packed as follows: Ensure that no-one is lying on the mattress. On the control panel, press the lock/unlock button. Press the Pack&Go® button and hold it down for 2 seconds
The Pack&Go® diode will flash during deflation. The mattress will empty of air and be ready to be simply folded together in about 20 minutes. The control unit gives an audio signal once deflation is complete. Carefully fold the mattress in half, place the control unit between the folds of the mattress and place the system in a transportation bag (accessory) or equivalent for protective storage. Ensure that the power supply is packed complete.
5.8
Restart
5.9
Power failure
5.10
Notifications
5.11
Maximum pressure notifications.
See 3.4 See 3.6
2. Ensure there is a gap between the user and the bed base so that the user does not ‘bottom out’. 3. If you can feel the user’s sacrum resting in the palm of your hand, the gap is too small. See section 8 ‘Troubleshooting’.
5.5 CPR (Cardiopulmonary resuscitation)
Different notifications exist based on how serious the warning is. With a malfunction or an error, a notification will be given by way of a flashing warning triangle. To mute the warning signal, press the mute button. The notification code is shown on the four different diodes above the wrench symbol.
See 3.2
5.6
See 3.3
Transport function
See 3.5
13
CuroCell® IQ Cirrus 5.12
Table of notifications
Information about each notification is shown in the notification table: Notifications from 1-10 are both audible and visual. The error code will be displayed until the error has been rectified. If the mute button is pressed the audible warning will crease for a period of 5 minutes and will return until the error has been rectified. The notifications 11-13 have no audible alarm. The error code is shown until the system is restarted. Notification
Description and troubleshooting
1
High temperature. Valves and compressors are turned off. If the control unit is in direct sunlight, relocate it. Otherwise, contact the technical support.
2
Default settings are not completed. Contact the technical support.
3
Incorrect input voltage. Make sure that the correct power supply is used, otherwise contact the technical support.
4
Low pressure. Secure the CPR, mattress, air tubes and air filter. If the problem remains, contact the technical support.
5 6
Automatic setting failure. The correct pressure has not been reached within the time limit. If the problem remains, contact the technical support. Leakage under the automatic setting period. The mattress is leaking too much for the weighing to be accomplished. Control the mattress and the air connections. If the problem remains, contact the technical support.
7
High pressure. The pressure cannot be reduced to the desired value within the time limit. Contact the technical support.
8
The automatic setting has been restarted too many times during the automatic setting period. Contact the technical support.
9
The mattress control parameters have not been read. Connect the CPR or contact the technical support.
10 11
The mattress control parameters have been changed during the use. Restart the system. If the problem remains, contact the technical support. Leakage in one of the sections. Secure the CPR, mattress and connection tubes. Information to service technicians: this notification shows a leakage in the blue air cell section. More information in the Servicemanual for CuroCell® IQ Cirrus.
14
6
Product description
6.1 Control unit (A4/IQ) CuroCell® A4 and CuroCell® IQ are automatic air mattress systems used as an aid to prevent and treat pressure ulcers/pressure injuries. The automation means that the control unit’s built-in sensors use software to adjust the inner pressure of the mattress according to the user’s weight, height, position, and change in position. This means that no manual action needs to be performed to adjust the inner pressure of the mattress to conform to the user. With CuroCell® A4 there are three programs to choose from. CuroCell® IQ uses the pulsating mode which is a combination of the alternating mode and CLP mode. Any of these two control units are compatible with the CuroCell® Cirrus mattress. 1. 2. 3. 4. 5. 6. 7. 8.
Control panel Tube/CPR connection Power switch, On/Off 3,5 mm plug input (only for use by manufacturer) Connection power cable Air filter Hangers Power supply
1.
1.
CuroCell® A4
CuroCell® IQ
2.
2.
3.
8.
4. 5. 6 7.
15
6.2
Mattress
1.
1. Mattress 2. Hygiene cover (top part) 3. Air cells 4. Foam Core 5. Inner cover 6. Tubing set 7. CPR (quick deflation) 8. Hygiene cover (bottom part)
2. 3. 4. 5. 6. 7. 8.
16
7
Reuse and cleaning
The product is reusable. Follow the instructions below for cleaning and reconditioning. It is recommended to clean the mattress between users. Consult your hygiene manager or Care of Sweden for help and instructions if you are unsure. Cleaning and reconditioning should otherwise be performed as required. Cleaning CONTROL UNIT Wipe off the control unit with a damp cloth and mild cleaning solution for this purpose: such as alcohol with or without tenside surfactants, oxidizing solutions, chlorine max 1 % or hydrogen peroxide max 1,5 %. MATTRESS COVER Covers can be wiped off with a mild cleaning solution for this purpose such as: alcohol with or without tenside surfactants, Isopropanol 70 %, oxidizing solutions, chlorine max 1 % (max 10% for cover Stone) or hydrogen peroxide max 1,5 %. Excrement and blood stains must be removed as soon as possible using cold water. Carefully follow local instructions and the instructions for the detergent.
Multi-layer covers should be divided before washing. The parts should be washed with similar colours.
INNER COVER
Clean the affected area with a mild cleaning solution for this purpose such as: alcohol with or without tenside surfactants, Isopropanol 70 %, oxidizing solutions, chlorine max 1 % or hydrogen peroxide max 1,5 %. MATTRESS 1. Disconnect the tube connector from the control unit and remove the air from the mattress. 2. Remove the covers. 3. Move the mattress to a clean area that is suitable for cleaning. 4. Wipe off the air cells, all tubes and the CPR module with a mild cleaning solution for this purpose such as: alcohol with or without tenside surfactants, Isopropanol 70 %, oxidizing solutions, chlorine max 1 % or hydrogen peroxide max 1,5 %. Let it dry. 5. Wipe off the work surfaces using a suitable detergent and/ or disinfectant. 6. Put the mattress together. If air cells for any reason have become loose from the tubes, these must be replaced according to the drawing in section 6.2. FOAM CORE Clean the affected area with a mild detergent (such as washing-up liquid) and water or with an alcohol-based disinfectant (cleaner intended for this purpose. Gently squeeze out any water). Note! • Check the hygiene cover, air cells and hoses each time the product is cleaned. If damaged, it must be replaced or repaired. Also check the control unit, tube connectors and power cable during cleaning. Damaged parts must be replaced or repaired.
For chlorine, we recommend a mixture of max 1% (max. 10% for cover Stone).
• Do not wring or roll the foam core to extract the water. Let it dry in a warm, ventilated area (not in direct sunlight). The foam core must be completely dry before it is used again.
17
8
Storage
9
Maintenance
It is advisable to store the mattress and control unit in the product bag (accessory), original package or equivalent for protective storage. Handle the packaged product with caution. Do not place any heavy objects on top of it. For additional information about storage temperature, see section 11.
9.1
General
We recommend that the control unit will be regularly serviced and inspected to maintain functionality and performance. Service and maintenance must always be performed by Care of Sweden or one of its authorized technicians. Only use spare parts approved by Care of Sweden. For more information, see the service manual for CuroCell® A4 or IQ.
9.2
Between users
Between users, the Pack&Go function should be used to reset the system. When starting up the product again it will be set on the pulsating mode. Between users, also check that: • The power cable and power supply are undamaged. • The connecting tubes (marked CPR) on the side of the control unit are positioned correctly and not leaking. • The hygiene cover is intact and the cover and air cells are correctly assembled. • No tubes or connectors are damaged or jammed. Contact Care of Sweden or your local distributor if any spare parts are required.
9.3
Replacing the air filter
Before any maintenance is done, make sure that the control unit is turned off. Services shall not be done while using the product. To replace the air filter: 1. Loosen the small protective plate on the rear of the control unit using a size T10 Torx screwdriver.
2. Remove the filter from the holder. 3. Place the new filter in the holder with the pink side facing outwards. Put the protective plate back in place and secure using the screws. If the control unit is used in a dirty environment the filter should be checked regularly.
18