care of sweden
Pressure Care Mattresses
Curoell M4, A4, IQ and IQ CX Series Air mattress systems Instructions for Use June 2022
Instructions for Use
36 Pages
Preview
Page 1
Instructions for use 2022-06-29
CuroCell® M4/A4/IQ Air mattress systems
Instructions for use item number: 95-001436-EN0000
EN
Table of contents
5.1 Function ... 13
1 Introduction... 3
5.3 Maximum pressure (caring mode)... 13
1.1 General information ... 3
5.4 Panel lock ... 13
1.2 Intended use ... 3
5.5 Comfort settings ... 13
1.3 Intended user ... 3
5.6 Hand check (function control) ... 13
1.4 Intended use environment ... 3
5.7 Sitting positioning in bed ... 14
1.5 Indications ... 3
5.8 CPR (Cardiopulmonary resuscitation) .. 14
1.6 Contraindications ... 3
5.9 Transport function ... 14
1.7 Clinical benefit ... 3
5.10
Pack&Go® function ... 14
2 Assembly and installation ... 4
5.11
Restart... 14
5.12
Power failure ... 14
5.13
Notifications ... 14
5.14
Maximum pressure notifications 14
5.15
Table of notifications... 14
3 Common operations ... 5 3.1 Sitting positioning in bed ... 5 3.2 CPR (Cardipulmonary resuscitation) ... 5 3.3 Transport function ... 5 3.4 Restart ... 5 3.5 Maximum pressure notification ... 5 3.6 Power failure... 5 4 Operation CuroCell® M4 ... 6 4.1 Function ... 8 4.2 Programs ... 8 4.3 Maximum pressure (caring mode) ... 8 4.4 Seating function ... 8 4.5 Panel lock ... 8 4.6 Hand check (function control) ... 9 4.7 Sitting positioning in bed ... 9 4.8 CPR (Cardiopulmonary resuscitation) ... 9 4.9 Transport function ... 9 4.10
Pack&Go® ... 9
4.11
Restart ... 9
4.12
Power failure ... 9
4.13
Notifications ... 9
4.14
Maximum pressure notification... 9
4.15
Table of notifications ... 10
5 Operation CuroCell® A4 ...11
5.2 Programs ... 13
6 Operation CuroCell® IQ ... 16 6.1 Function ... 18 6.2 Program... 18 6.3 Maximum pressure (caring mode)... 18 6.4 Hand check (function control) ... 18 6.5 CPR (Cardiopulmonary resuscitation) .. 18 6.6 Transport function ... 18 6.7 Pack&Go® ... 19 6.8 Restart ... 19 6.9 Power failure ... 19 6.10
Notifications ... 19
6.11
Maximum pressure notifications 19
6.12
Table of notifications... 20
7 Product description... 21 7.1 Control unit (M4/A4/IQ) ... 21 7.2 Mattresses ... 22 8 Reuse and cleaning ... 25 8.1 Cleaning ... 25 9 Storage ... 26 10 Maintenance... 26
10.1
General ... 26
10.2
Between users ... 26
10.3
Replacing the air filter ... 26
11 Troubleshooting ...27 12 Technical specification ...28 12.1
Standards ... 28
12.2
Symbolkey ... 29
13 Other information...30 13.1 Recommended lifetime of the product ... 30 13.2
Disassembly and recycling ... 30
13.3
Returns ... 30
WARNINGS AND SAFETY PRECAUTIONS Read all instructions before use or repair
WARNING! To minimize the risk of fire, personal injury and equipment/property damage adhere to the following instructions: 1. The product must only be installed and used for its intended purpose according to the instructions in this manual and/ or other documentation from Care of Sweden. The product must not be combined, assembled or repaired with parts (e.g. control unit and mattress), accessories or spare parts other than those described in this manual or other documentation from Care of Sweden. The product must not be modified in any way. 2.
The product must be placed and used so that it does not become trapped or damaged. Be particularly aware of trapping damage when using side rails. Always make sure that the mattress is the correct size for the bed.
3.
Regularly check product functionality by performing a hand check.
4.
When the product is used for individuals needing special supervision, such as children, continuous monitoring is required.
5.
The mattress is protected by a hygiene cover; avoid using multiple hygiene covers as this can affect the vapor permeability of the mattress.
6.
The hygiene cover does not allow liquid or air to penetrate but is vapour permeable. Make sure that the user is positioned correctly to avoid the risk of suffocation.
7.
Be careful with sharp objects to prevent damage to the hygiene cover.
8.
Do not open the control unit housing – risk for electric shock. Servicing and maintenance must be performed by Care of Sweden or one of its authorized service technicians.
9.
Route the power cable to the control unit carefully to avoid tripping. Also make sure that the user is lying correctly on the mattress according to the instructions and use a cable holder if possible.
10. To avoid the risk of strangulation, make sure that the cable and tubes are routed to prevent
someone getting caught up in them.
11. Do not use the product in bathrooms or other area where there is a risk of the control unit
coming into contact with water or other liquids. Except for specified cleaning, never handle a product that has come into contact with water/liquid. Pull the plug out of the electrical socket immediately and send the product to an authorized service technician for servicing
12. Do not use close to or in contact with fire sources/hot surfaces, such as fire, burning cigarettes,
hot lamps, heating fans or heating stoves/open fires as this could damage the product.
13. Do not store or use the product in direct sunlight. The product may be damaged by the
elevated temperature and UV light.
1
14. Strong magnetic fields or wireless communication equipment (e.g., wireless home network
products, mobile phones, walkie-talkies, cordless phones and their base stations, radio transmitters, etc.) may affect the product’s functionality and should be kept at a distance of at least 1 meters from the control unit.
15. Never use the product if the power cable, plug of the control unit or power supply housing is
defective, if the control unit housing is damaged, or if it is not functioning properly. Contact an authorized service technician for examination and repair.
16. Never connect anything other than the Care of Sweden supplied power supply to the control
unit power cable connector.
17. Never use the external communication input (3,5mm connector), this input should only be
used by the manufacturer.
18. If the hygiene cover is equipped with side handles, these are intended for managing or
relocating the mattress. Do not use the handles to lift the mattress with the user lying on it. All other use takes place under your own liability and is not covered by the product warranty.
19. To prevent the power supply from being pulled out, exercise caution when there are children
and pets in the environment around the equipment.
20. Use of this product adjacent to or stacked with other equipment should be avoided because it
could result in improper operation. If such use is necessary, this product and the other equipment should be observed to verify that they are operating normally.
21. To minimize the risk of wounds occurring on the feet, make sure that the user doesn’t come
2
1
Introduction
These air mattress systems may be used as an aid to prevent and treat pressure ulcers/pressure injuries. CuroCell® M4 is a control unit where the operator must manually adjust the pressure in the mattress based on the weight of the user. CuroCell® A4 is a control unit that sets the air pressure in the mattress based on the user’s length, weight and position without need for manual settings. It offers the operator the option to adjust program and comfort. CuroCell® IQ is a control unit that weights the user and adjust the pressure without an option for the operator to choose program. CuroCell® M4, A4 and IQ are compatible with five different mattresses: CuroCell® OP10 and CuroCell® CX10, CuroCell® CX15, CuroCell® CX16 and CuroCell® CX20.
General information
Up to 200 kg
CuroCell® CX15
Up to 220 kg
CuroCell® CX16
Up to 200 kg
CuroCell® CX20
Up to 250 kg
1.4
Intended use environment
1.5
Indications
1.6
Contraindications
1.7
Clinical benefit
The clinical benefits for M4, A4 and IQ together with any of the mattresses included in this instruction for use are: • Prevention and treatment of pressure ulcers/pressure injuries up to and including category 4 and unclassified PU/PI. • Reduction of shear forces. • Safety, comfort, pressure redistribution and easy to handle. • Silent running control units.
Intended use
Note!
Intended user
The mattress system is intended to be used by all kind of users, including lay persons. Note that the user and operator could be the same person. The mattresses are intended for use by persons of a recommended minimum length of 120 cm. The specifications for weight are listed in following table.
CuroCell® CX10
It is necessary for the prescriber of the mattress to make an individual assessment of the user and decide whether the characteristics of the mattress is beneficial for the user based on diagnosis.
The mattress system consists of a control unit and a mattress and is intended to be used for prevention and as an aid in the treatment of pressure ulcers/pressure injuries (PU/PI).
1.3
Up to 160 kg
Suitable for a wide range of persons with increased risk for pressure ulcers/injuries, including those with superficial ulcers, up to category 4 and unclassified PU/PI (unstageable and suspected deep tissue injury) in association with an individualized plan of care.
The system is a medical device with CE marking in accordance with MDR (EU) 2017/745. According to this regulation the manufacturer is required to report all accidents or incidents involving the products. All information involving accidents or incidents relating to our products, shall be reported immediately to Care of Sweden.
1.2
Recommended user weight
CuroCell® OP10
The mattress system can be used in all kinds of health care environments, including home care.
See more information about the mattresses in section 7.2. Always read the instructions for use prior to use.
1.1
Mattress
3
• For certain users, e.g., amputees, the recommended length measurement may not be reached. Users in these groups may require other settings as the entire surface is not under load. For function controls, see section 3. • The mattress may be inappropriate for use during x-ray examinations because of the risk of blurred images or artefacts that may lead to diagnostic errors. • In the event of a power loss or similar, the mattress will retain air for at least 12 hours.
2
Assembly and installation
Check that no parts are damaged. If damage is found, contact Care of Sweden or your local distributor before using the product. Do not use sharp objects when unpacking as it might damage the product. When using an overlay mattress (CuroCell® OP10 or CuroCell® CX10): 1a. Place the overlay mattress on the base mattress. Secure the mattress to the base using the 4 straps on the corners of the mattress.
3. If the mattress is equipped with a cable holder, place the power cable in the cable holder by opening the press studs, placing the cable in the gap and closing the press studs again. Otherwise, place the power cable so there is no risk of stumbling over it, running over it with the bed wheels, or getting it jammed when raising or lowering the bed. Put the power cable into the control unit.
When using a full replacement mattress (CuroCell® CX15, CuroCell® CX16 or CuroCell® CX20): 1b. Place the mattress on the bed base. Secure the mattress to the bed using the fastening straps on the underside of the mattress. If the mattress is used on an adjustable bed, the 2 straps in the middle shall be fastened to the non-movable parts of the bed. The 2 straps at the head end of the bed shall be fastened at the moving part of the head end. The 2 straps at the foot end of the bed shall be fastened at the moving part of the foot end.
4. Open the lid on the air tube connector (marked CPR) and connect it to the side of the control unit.
5. A click is heard and felt when correctly connected. Secure that both sides of the connection are closed. 6. Plug the power supply into an approved and easily accessible electrical socket (100– 240 V). 7. Check that the power cable has been correctly connected to the control unit and that the correct power supply has been used. (See section 9.1 Technical specification). The correct P/N is shown on the label on the power supply. The power supply is part of the equipment and may not be replaced.
Note!
• Make sure that the mattress is the correct size for the bed. • Check the cells and press studs to ensure they are correctly assembled. • Make the bed with sheets. • The mattress should be used lying in the lengthwise direction on the mattress, with feet at the end, market with the feet symbol.
If the control unit has been stored in its minimum or maximum storage temperature (25°C to 70°C), wait at least 1 hour before starting it. This time is based on an ambient temperature of 20°C.
2. Hang the control unit on the foot end of the bed or place it on a level, steady surface. Check that the switch on the side of the control unit it set to ‘0’ (off).
Note!
• Do not hold the 12V plug on the power supply while touching the user.
4
3
Common operations
3.1
Sitting positioning in bed
3.3
Following operations apply to CuroCell® M4, CuroCell® A4 and CuroCell® IQ regardless of which mattress that is used. When raising the head end of the bed into a sitting position, always secure the user´s position. To ensure the product functionality, we always recommend to perform a hand check (see section 4, 5 or 6 depending on the pump and mattress of interest). This function is recommended to use for short periods only. For additional support, positioning pillows can be used.
We recommend using this function for short periods only.
3.4
Restart
3.5
Maximum pressure notification
3.6
Power failure
If a restart is required, set the On/Off switch on the side of the control unit to 0 (Off). Wait for approx. 10 seconds and restart the control unit.
Note!
• When using a lift to place the user in the bed and the head end of the bed is raised, make sure that the user is not placed too high on the mattress. Otherwise, there is a risk of shear. • When the alternating or pulsating program is used and the head end of the bed is raised, make sure that the user and/or the mattress is not moving downwards due to the movement in the mattress. Also raise the foot end of the b d
3.2
Transport function
If the user needs to be moved in bed, either: • Unplug the CPR connection, close the lid, place the CPR connection at the end of the bed and remove the control unit from the bed. The mattress will retain air for at least 12 hours. Or • Remove the power supply from the wall socket and leave the control unit hanging on the bed during transport. The mattress will retain air for at least 12 hours.
When the function Maximum pressure has been used for a long time, the Maximum pressure diode will blink. If the use is intentional, ignore the notification.
CPR (Cardipulmonary resuscitation)
In case of an emergency where CPR (Cardiopulmonary resuscitation) is necessary, remove the connection (marked ‘CPR’) from the control unit and leave the lid open in order to empty the mattress of air quickly.
In the event of a power failure, unplug the CPR connection, close the lid, place the CPR connection at the end of the bed and remove the control unit from the bed. The mattress will retain air for at least 12 hours. Perform a hand check to make sure the pressure of the mattress is not too hard or too soft.
5
CuroCell® M4
4
Operation CuroCell® M4
Following operations apply only to CuroCell® M4 regardless of which mattress that is used. Read the label of the control unit carefully to make sure that you know which control unit you have. Operation CuroCell® M4 (how to get started) 1. Set the On/Off switch on the side of the control unit to position 1 (On). See figure 1. 2. Use the weight settings to set a suitable air pressure. 3. The mattress starts to inflate. This takes about 20-40 minutes depending on the size of the mattress. While the mattress is inflating, the “mute the information signal” and “information signal” diode light up in orange. When these diodes goes out, the user can be placed on the mattress. 4. The pulsating mode setting is pre-set. If a different setting is desired, it can be selected when the user is placed on the mattress. 5. When the user is placed on the mattress and the desired program is selected, the inner pressure of the mattress is adjusted according to the selected weight setting. This takes about 10 minutes. During this time, the diode above the selected program flashes. When the diode stops flashing, the mattress is set and the inner pressure has reached the selected value. 6. Perform a hand check to ensure that the settings are correct. 7. If the mattress is too soft or too hard, adjust the weight setting. 8. Perform another hand check.
Note! • Each time the system starts up, it will operate as follows: When using a mattress without an air safety mattress (CuroCell® OP10, CuroCell® CX10 or CuroCell® CX16), the control unit will adjust to the set weight. • When using a mattress with an air safety mattress (CuroCell® CX15 or CuroCell® CX20), the control unit always begins by completely inflating the safety mattress, followed by the other cells, and then perform an automatic setting. • When using a mattress with a safety mattress in foam (CuroCell® CX16), the user can lie on the mattress even when not inflated. • When using a mattress with an air safety mattress (CuroCell® CX15 or CuroCell® CX20), this must be inflated before the user can lie on the mattress. In both cases, the safety mattress will carry the load when the orange diode goes out.
6
Figure 1. On/Off switch.
CuroCell® M4
Button
Function Mute the information signal Panel lock Pack & Go®. Function for deflating the system Alternating mode Pulsating mode CLP mode (Constant low pressure) Maximum pressure (caring mode) User weight settings
Information signal Incorrect connection of the air connector (CPR)
Seating function Cycle time settings (10, 15, 20, 25 minutes). The diodes are also used for error notifications.
7
CuroCell® M4
4.1
Function
At start-up, the inner pressure of the mattress must be set manually based on the weight and height of the user. The mattress system maintains the pre-set inner pressure regardless of movement and position changes. This means that when the user changes position, for example, the mattress’s inner pressure must be adjusted manually. The weight settings on the control unit are used to change the inner pressure.
4.2
4.3
Maximum pressure (caring mode)
4.4
Seating function
4.5
Panel lock
With this function, the entire mattress is inflated and provides firmed support. This function reverts to the previous setting after approximately 20 minutes. The function should be used when caring the user, shifting the user’s position, or moving the user in or out of bed. When the function has been used for a long time, the diode will blink. If the use is intentional, ignore the notification.
Programs
There are three programs to choose from: Constant low pressure (CLP) means that the air pressure in all of the air cells is the same in the whole mattress.
The alternating mode switches to pulsating mode when the seating function is activated. A visible and audible notification is generated when the seating function has been active for two consecutive hours, notifying that repositioning of the individual may be necessary. The control unit will also notify if an attempt is made to switch on the system while seating function is active.
Alternating mode means that the air pressure in the air cells are different and alternates regularly after chosen cycle periods. Pulsating mode combinates CLP with the alternating function. Chose program by pushing the button for the program. We recommend the Pulsating mode which is also the preset mode.
Press the Panel Lock button to lock or unlock the control panel. The button indicates when the panel has been locked. The screen locks automatically if left untouched for five minutes. This is to prevent the settings being changed accidentally. To unlock, press the button for 2 seconds.
1. Constant low pressure mode (CLP) No cycle period is needed. 2. Alternating mode. The cycle period can be changed according to the user needs and requirements. Choose between 10, 15, 20 or 25 minutes. The longer the cycle period, the slower the alternations. We recommend a basic setting of 10 minutes. 3. Pulsating mode. The cycle period can be changed according to the user needs and requirements. The longer the cycle period, the slower the alternations. We recommend a basic setting of 10 minutes.
8
4.6
Hand check (function control)
CuroCell® M4
4.7
Sitting positioning in bed
4.8
CPR (Cardiopulmonary resuscitation)
See 3.1
Hand check is performed to ensure that the mattress system works properly and to ensure that the weight setting is correct. Hand check should be performed regularly; for CuroCell® M4, we recommend once per work shift as well as after installation of the system.
See 3.2
4.9
Transport function
4.10
Pack&Go®
See 3.3
For manual systems hand check should also be performed when moving from lying to seating position in bed, in case of positioning changes and when changing the comfort setting (and/or the weight setting).
The product may be packed as follows: • Ensure that no-one is lying on the mattress. • On the control panel, press the lock/unlock button. • Press the Pack&Go® button and hold it down for 2 seconds The Pack&Go® diode will flash during deflation. The control unit gives an audio signal once deflation is complete. Carefully fold the mattress together, place the control unit between the folds of the mattress and place the system in a transportation bag (accessory) or equivalent for protective storage. Ensure that the power supply is packed complete.
Note!
• Make sure that the mattress is filled, and that the system has adapted to the user before performing a hand check. • How to perform a hand check depends on which mattress is used – read the label of the mattress carefully to know which mattress you have.
When using an overlay mattress system (CuroCell® OP10 or CuroCell® CX10): 1a. Open the cover and identify a cell with less air at the user’s sacrum. Insert a hand, with the palm facing up, between the overlay mattress and the underlying mattress. The hand is inserted beneath the user’s sacrum (center of mattress).
4.11 Restart See 3.5
4.12 Power failure See 3.6
4.13
When using a full replacement mattress (CuroCell® CX15, CuroCell® CX16 or CuroCell® CX20)
Notifications
Different notifications exist based on how serious the warning is. With a malfunction or an error, a notification will be given by way of a flashing warning triangle. To mute the warning signal, press the mute button. When a notification occurs, the current cycle time diode will turn off and a notification code is shown on the four different cycle time diodes (10, 15, 20, 25). To read the cycle time during the error notification, unlock the control panel.
1b. Open the cover and identify a cell with less air at the user’s sacrum. Insert a hand, with the palm facing up, between the top cells and the underlying safety mattress. The hand is inserted beneath the user’s sacrum (center of mattress). 2. Ensure there is a gap between the user and the underlying mattress so that the user does not ‘bottom out’.
3. If you can feel the user’s sacrum resting in the 4.14 Maximum pressure notification palm of your hand, the gap is too small. See See 3.5 section 8 ‘Troubleshooting’.
9
CuroCell® M4
4.15
Table of notifications
Information about each notification is shown in the notification table: • Notifications from 1-7 are both audible and visual. The error code will be displayed until the error has been rectified. If the mute button is pressed the audible warning will crease for a period of 5 minutes and will return until the error has been rectified. • The notifications 8-10 have no audible alarm. The error code is shown until the system is restarted. Notifiering
Description and troubleshooting
1
High temperature. Valves and compressors are turned off. If the control unit is in direct sunlight, relocate it.
2
Default settings are not completed, contact technical support.
3
Incorrect input voltage. Make sure that the correct power supply is used.
4
Low pressure. Secure the CPR, mattress, air tubes and air filter.
5
High pressure. The pressure cannot be reduced to the desired value within the time limit.
6
The mattress control parameters have not been read. Connect the CPR.
7
The mattress control parameters have been changed during the use. Restart the system.
Leakage in one of the sections. Secure the CPR, mattress and connection tubes. Information to service technicians: this notification shows a 8 leakage in the blue cell section. More information in the Servicemanual for CuroCell® M4. Leakage in one of the sections. Secure the CPR, mattress and connection tubes. Information to service technicians: this notification shows a 9 leakage in the green cell section. More information in the Servicemanual for CuroCell® M4. Leakage in one of the sections. Secure the CPR, mattress and connection tubes. Information to service technicians: this notification shows a 10 leakage in the red cell section. More information in the Servicemanual for CuroCell® M4. If any of these problems remain, contact technical support.
10
CuroCell® A4
5
Operation CuroCell® A4
Following operations apply only to CuroCell A4 regardless of which mattress that is used. Read the label of the control unit carefully to make sure that you know which control unit you have.
Operation CuroCell® A4 (how to get started) 1. Set the On/Off switch on the side of the control unit to position 1 (On). See figure 1. 2. The mattress starts to inflate. This takes about 20-40 minutes depending on the size of the mattress. While the mattress is inflating, the ”mute the information signal” and ”information signal” diodes light up in orange. When these diodes goes out, the user can be placed on the mattress. 3. The pulsating mode setting is pre-set. If a different setting is desired, it can be selected when the user is placed on the mattress. 4. The control unit sets the inner pressure of the mattress according to the weight, length, and position of the user. This takes about 20-30 minutes. During this time, the diode above the selected program flashes. When the diode stops flashing, the mattress is set and the inner pressure has adjusted according to the user. 5. Perform a hand check to ensure that the settings are correct.
Note! Each time the system starts up, it will operate as follows: • When using a mattress without an air safety mattress (CuroCell® OP10, CuroCell® CX10 or CuroCell® CX16), the control unit will completely inflate the main cells and then perform an automatic setting. • When using a mattress with an air safety mattress (CuroCell® CX15 or CuroCell® CX20), the control unit always begins by completely inflating the safety mattress, followed by the other cells, and then perform an automatic setting. • During the automatic setting, try to avoid larger movements on the mattress. Otherwise, you will get a notification on that the desired value could not be reached within the time limit and the weighting must start over. • When using a mattress with a safety mattress in foam (CuroCell® CX16), the user can lie on the mattress even when not inflated. When using a mattress with an air safety mattress (CuroCell® CX15 or CuroCell® CX20), this must be inflated before the user can lie on the mattress. In both cases, the safety mattress will carry the load.
• Once automatic setting is complete, the control unit switches to a basic setting of Pulsating mode (when used for the first time) or to the previous setting. For more information see section 5.2.
11
Figure 1. On/Off switch.
CuroCell® A4
Button
Function Mute the information signal Panel lock Pack & Go®. Function for deflating the system Alternating mode Pulsating mode CLP mode (Constant low pressure) Maximum pressure (caring mode) Information signal Incorrect connection of the air connector (CPR) Cycle time settings (10, 15, 20, 25 minutes). The diodes are also used for error notifications. Comfort settings
12
CuroCell® A4
5.1
Function
5.2
Programs
3. Pulsating mode. The cycle period can be changed according to the user needs and requirements. The longer the cycle period, the slower the alternations. We recommend a basic setting of 10 minutes.
The mattress system independently and without manual adjustment controls the inner pressure of the mattress according to the weight, height, and position of the user. No manual action needs to be performed to adjust 5.3 Maximum pressure (caring mode) the inner pressure of the mattress. This With this function, the entire function works as follows: mattress is inflated and provides 1. At start-up, automatic setting of the inner firmed support. This function reverts pressure of the mattress is always carried out to the previous setting after 20 according to the weight and height of the user. minutes. The function should be used when 2. If the user moves noticeably or changes their caring the user, shifting the user’s position, or position, the system will independently control moving the user in or out of bed. the inner pressure of the mattress. 3. The system performs an automatic setting at 5.4 Panel lock fixed intervals even if no significant changes Press the Panel Lock button to lock or have occurred. unlock the control panel. The button indicates when the panel has been After automatic setting of the mattress inner locked. The screen locks automatically if left pressure, the system returns to the previously selected program. At start-up, Pulsating Mode is untouched for five minutes. This is to prevent the settings being changed accidentally. always pre-set. To unlock, press the button for 2 seconds. There are three programs to choose from:
5.5
Comfort settings
The pressure can be increased in two steps depending on the user’s comfort Alternating mode means that the air pressure in requirements. This increase is made based on the automatic setting in 5.1. the air cells are different and alternates regularly after chosen cycle periods. Note! Pulsating mode combinates CLP with the • When only parts of the mattress are alternating function. under load, for example, in the case of Constant low pressure (CLP) means that the air pressure in all of the air cells is the same in the whole mattress.
amputees – it may be necessary to raise the setting using the comfort settings.
Chose program by pushing the button for the program. We recommend the Pulsating mode which is also the preset mode.
The selected setting is shown by a green light. 1. Constant low pressure mode (CLP) No cycle period is needed. 5.6 Hand check (function control) Hand check is performed to ensure that the 2. Alternating mode. The cycle period mattress system works properly. Hand check can be changed according to the user should be performed regularly; for CuroCell® A4, we recommend once per work shift as well as needs and requirements. Choose after installation of the system. between 10, 15, 20 or 25 minutes. The longer the cycle period, the slower the alternations. We recommend a basic setting of 10 minutes.
13
CuroCell® A4 Press the Pack&Go® button and hold it down for 2 seconds
Note!
• Make sure that the mattress system is filled, which is shown by a green light from the diode, before performing a hand check. • How to perform a hand check depends on which mattress is used - read the label of the mattress carefully to know which mattress you have.
When using an overlay mattress system (CuroCell® OP10 or CuroCell® CX10) 1a. Open the cover and identify a cell with less air at the user’s sacrum. Insert a hand, with the palm facing up, between the overlay mattress and the underlying mattress. The hand is inserted beneath the user’s sacrum (center of mattress). When using a full replacement mattress (CuroCell® CX15, CuroCell® CX16 or CuroCell® CX20)
The Pack&Go® diode will flash during deflation. The mattress will empty of air and be ready to be simply folded together within 20 minutes. The control unit gives an audio signal once deflation is complete. Carefully fold the mattress together, place the control unit between the folds of the mattress and place the system in a transportation bag (accessory) or equivalent for protective storage. Ensure that the power supply is packed complete.
5.11
Restart
5.12
Power failure
5.13
Notifications
5.14
Maximum pressure notifications
5.15
Table of notifications
See 3.4 See 3.6
1b. Open the cover and identify a cell with less Different notifications exist based on air at the user’s sacrum. Insert a hand, with the how serious the warning is. With a palm facing up, between the top cells and the malfunction or an error, a notification underlying safety mattress. The hand is inserted will be given by way of a flashing warning beneath the user’s sacrum (center of mattress). triangle. To mute the warning signal, press the mute button. 2. Ensure there is a gap between the user and When a notification the underlying mattress so that the user does occurs, the current cycle not ‘bottom out’. time diode will turn off 3. If you can feel the user’s sacrum resting in the and a notification code is shown on the four palm of your hand, the gap is too small. See different cycle time diodes (10, 15, 20, 25). To section 8 ‘Troubleshooting’. read the cycle time during the error notification, unlock the control panel.
5.7
See 3.1
Sitting positioning in bed
5.8 CPR (Cardiopulmonary resuscitation) Se 3.2
See 3.5
Information about each notification is shown in 5.9 Transport function the notification table: Se 3.3 Notifications from 1-10 are both audible and visual. The error code will be displayed until the 5.10 Pack&Go® function The product may be packed as follows: error has been rectified. If the mute button is pressed the audible warning will crease for a Ensure that no-one is lying on the period of 5 minutes and will return until the mattress. error has been rectified. The notifications 11-13 On the control panel, press the have no audible alarm. The error code is shown lock/unlock button. until the system is restarted.
14
CuroCell® A4 Notification
Description and troubleshooting
1
High temperature. Valves and compressors are turned off. If the control unit is in direct sunlight, relocate it.
2
Default settings are not completed.
3
Incorrect input voltage. Make sure that the correct power supply is used, otherwise contact the technical support.
4
Low pressure. Secure the CPR, mattress, air tubes and air filter.
5
Automatic setting failure. The correct pressure has not been reached within the time limit.
6
Leakage under the automatic setting period. The mattress is leaking too much for the weighing to be accomplished. Control the mattress and the air connections.
7
High pressure. The pressure cannot be reduced to the desired value within the time limit.
8
The automatic setting has been restarted too many times during the automatic setting period.
9
The mattress control parameters have not been read. Connect the CPR.
10
The mattress control parameters have been changed during the use. Restart the system.
11
12
13
Leakage in one of the sections. Secure the CPR, mattress, and connection tubes. Information to service technicians: this notification shows a leakage in the blue cell section. More information in the Service manual for CuroCell® A4. Leakage in one of the sections. Secure the CPR, mattress, and connection tubes. Information to service technicians: this notification shows a leakage in the green cell section. More information in the Service manual for CuroCell® A4. Leakage in one of the sections. Secure the CPR, mattress, and connection tubes. Information to service technicians: this notification shows a leakage in the red cell section. More information in the Service manual for CuroCell® A4.
15
CuroCell® IQ
6
Operation CuroCell® IQ
Following operations apply only to CuroCell® IQ regardless of which mattress that is used. Read the label of the control unit carefully to make sure that you know which control unit you have. Operation CuroCell® IQ (how to get started) 1. Set the On/Off switch on the side of the control unit to position 1 (On). See figure 1. 2. The mattress starts to inflate. This takes about 20-40 minutes depending on the size of the mattress. While the mattress is inflating, the “silent information signal” and “information signal” diodes light up in orange. When these diodes goes out, the user can be placed on the mattress. 3. The control unit sets the inner pressure of the mattress according to the weight, length, and position of the user. This takes about 20-30 minutes. During this time, the diode above the “check symbol” flashes. When the diode stops flashing, the mattress is set and the inner pressure has adjusted according to the user. 4. Perform a hand check to ensure that the settings are correct.
Note! • Each time the system starts up, it will operate as follows: When using a mattress without an air safety mattress (CuroCell® OP10, CuroCell® CX10 or CuroCell® CX16), the control unit will completely inflate the main cells and then perform an automatic setting. • When using a mattress with an air safety mattress (CuroCell® CX15 or CuroCell® CX20), the control unit always begins by completely inflating the safety mattress, followed by the other cells, and then perform an automatic setting. • During the automatic setting, try to avoid larger movements on the mattress. Otherwise, you will get a notification on that the desired value could not be reached within the time limit and the weighting must start over. • When using a mattress with a safety mattress in foam (CuroCell® CX16), the user can lie on the mattress even when not inflated. When using a mattress with an air safety mattress (CuroCell® CX15 or CuroCell® CX20), this must be inflated before the user can lie on the mattress. In both cases, the safety mattress will carry the load.
16
Figure 1. On/Off switch.
CuroCell® IQ
Button
Function Mute the information signal
Pack & Go®. Function for deflating the system Maximum pressure (caring mode) Information signal Incorrect connection of the air connector (CPR)
Check-symbol. System is ready to use Notification diodes
17