CareFusion

Alaris Gateway Workstation and DS Docking Station

Alaris Docking Station Technical Guide Dec 2010 - Replacing DS PSU assembly

Technical Guide

2 Pages

CareFusion Switzerland 317 Sàrl CH-1180 Rolle  carefusion.com  Publication: Issue: Date Issued: Status:  1000PB01938 Subject: 1 December 2010 Replacing DS PSU assembly  Alaris® Docking Station removal and installation PSU procedure  OVERVIEW This Technical Publication shows the removal and installation of the Alaris® Docking station PSU. REMOVAL PROCEDURE REQUIRED EQUIPMENT •  Posi screwdriver or Torx T20 depending on the version of Docking Station  1. Remove the bottom Tile assembly from the DS unit. 2. Disconnect the DC, AC and Earth cables from the tile assembly and separate from the base. 3. Unscrew the Earth cable(s) from PSU Heat sink and assembly from the base. 4. Unscrew the Power tile from the base and remove. 5. Unplug the switch wiring loom from PL1 on the PSU. 6. Unscrew the PSU from the front cover  1  2  3  4  5  6  © 2010 CareFusion Corporation or one of its subsidiaries. All rights reserved. For additional information, please contact your local office or distributor.  1000PB01938 Issue 1  Page 1 of 2
File Type: PDF
File Size: 121 KB
File Name: alaris_docking_station_technical_guide_dec_2010_-_r.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions