Assembly and Installation Instructions
19 Pages
Preview
Page 1
ALGO 5 Assembly and Installation Instructions
ASSEMBLY SAFETY WARNING These instructions assume familiarity with tools and above average mechanical aptitude. Read these assembly instructions in their entirety. Review and comply with all warning, cautions, and notes. Assembly of the ALGO 5 system should be performed by two (2) knowledgeable individuals using appropriate tools. If you experience problems with, or have any questions whatsoever about any portion of the assembly process or overall function of this system, call Natus Technical Support at 1-800-303-0306.
Preparation for Assembling the Cart The ALGO 5 system is packaged in 4 cartons. Examine the four shipping containers carefully upon receipt. If any of the boxes are damaged or the packing materials show signs of stress, notify the carrier and keep the shipping materials for the carrier’s inspection. Open all boxes to examine contents. Note: Save shipping boxes and packing foam in the event that any ALGO 5 component must be returned to the manufacturer.
Optional Accessories for ALGO 5 The following optional items are not included in the standard ALGO 5 system, and if ordered, will be in their individual cartons: • Barcode scanner • Page-size printer (laser or ink-jet) • LAN Isolator
Tools Not Supplied but Required for Assembly of ALGO 5 Cart A variety of hex wrenches are provided for assembly of the ALGO 5 system. These hex wrenches are listed on page 2 and are identified when required throughout this document. In addition to the supplied hex wrenches, the following common hand tools will be required to assemble the ALGO 5 Cart. • 7/16” standard wrench or adjustable wrench • #2 Phillips screwdriver • Scissors
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 1 of 19
Parts Reference The ALGO 5 arrives packaged in 4 separate cartons. The contents of each carton are listed on pages 2 – 3. Follow specific instructions on page 5 for unpacking the Algo 5 Column (Carton # 004691). Only major components are shown in Parts Reference. Specific hardware is shown throughout this document where required for assembly. If any parts are missing or damaged, call Natus Technical Support for replacement.
Carton # 004691 (Algo 5 Column) Item #
Description
Qty
1
Installation Guide
1
2
Column Assembly
1
3
#6-32 x 3/8'' Truss Head Phillips Screw
4
4
Mounting Adapter, VESA Plunger-plate
1
5
M4 x 20mm Pan Head Machine Screw (PHMS)
4
6
Spacer, 1/4'' Nylon
4
7
5/32'' Hex Wrench, T-Handle 9''
1
8
Power Cord Cleat
1
9
1/4-20 x 1-1/2'' Socket Head Cap Screw (SHCS)
2
10
1/4'' HI-Collar Lock Washer
2
11
Cable Clamp, Nylon
1
12
#6-32 x 3/8'' PHMS
1
13
Channel Cover, Rear (Green)
1
3 2 12
7
Carton # 000858 (Algo 5 Base) Item #
Description
Qty
1
Base
1
2
1/4-20 x KEP Nuts
2
3
1/4-20 x 1-1/4'' Socket Head Cap Screw (SHCS)
4
4
1/4'' Flat washer, 9/32'' (ID) x 5/8'' (OD)
4
5
1/4'' Spit Lock Washer
4
6
3/16'' T-handle Hex Wrench
1
7
Cable Tie, 11-3/8'' White Nylon
1
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
1
Page 2 of 19
Carton # 000910 (Accessories) Item #
Description
Qty
1
DSP Box
1
2
Preamp 5
1
3
ATA 5 Cable
1
4
Label Printer
1
5
PCA 5 Cable
2
6
Equipment Check Kit
1
7
Power Cord, AC
1
8
USB Cable
1
9
Sample Pack
1
10
CD
1
11
Documentation Package (only User Manual shown)
1
1 2
3 4
6
5
10 11 7
9
8
Carton # 004716 (PC/Display) Item # 1
Description
Qty
PC/Display Panel
1 1
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 3 of 19
Fully Assembled Cart
PC/Display
Column with Integrated Cable Management DSP Box Barcode Scanner (optional)
Writing Surface Preamp/Cable Holder Keyboard
Barcode Scanner Holder
Drawer for Disposables
Shelf Space
Label Printer
Power Cord Cleat
DSP Box Power Supply (preinstalled)
Label Printer Power Supply (preinstalled)
Ink Jet Printer Power Supply (not preinstalled) Page-size Printer (optional)
Isolation Transformer
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Main Power Switch
Page 4 of 19
Attaching the Column to the Base Tools Required • 7/16'' nut driver or socket wrench (not provided) • 3/16'' T-handle hex wrench (provided)
CAUTION: Two people are required to perform this task. 1. Remove materials (rear cover, etc) from carton, leaving only the protective packing at ends of Column in place inside the carton as shown. Installation Note: Retain packing material for use during assembly.
2. Unfold cardboard packing that was used to secure the upper Column and place on floor. Cardboard should be positioned to correspond with Keyboard Drawer and Shelves as shown.
3. Fold carton flaps to outside and carefully turn carton and Column upside down onto the cardboard. Do not allow the Column to fall out of the carton.
4. Lift carton off Column and carefully remove loose packing material from ends of Column. Do not remove packing material from Keyboard Drawer or Shelves. Column must be positioned as shown for attachment of Base.
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 5 of 19
5. Lay foam sheet or other packing material on floor at end of Column as shown. This material will protect the front edge of the Base during cable routing and attachment to Column.
6. Stand Base on its front edge at end of Column as shown. The opening at the rear of Base should be aligned as shown to allow routing of cables. Installation Note: Hardware and hex wrench required for attachment of Base to Column are taped to bottom of Base (hardware and Hex wrench shown below).
Opening aligned with Column
3/16'' Hex Wrench 1/4-20 Kep Nuts (2) 1/4'' Flat Washer (4) 1/4'' Split Lock Washer (4) 1/4-20 x 1-1/4'' SHCS (4)
7. Carefully guide cables (one at a time) through the hole in the Base as shown.
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 6 of 19
8. Align studs with corresponding mounting holes in opening in Base.
9. Ensure no cables are pinched between Column and Base and carefully guide Base onto two (2) studs. It may be necessary to lift the Base slightly to guide onto studs.
10. Thread two (2) KEP nuts onto the studs. Tighten KEP nuts with a 7/16'' nut driver or socket wrench.
1/4-20 Kep Nuts (2)
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 7 of 19
11. Locate four (4) remaining mounting holes in Base and assemble four (4) #1/4-20 x 1-1/4'' SHCS with lock washers and flat washers as shown below.
1/4'' Flat Washer (4) 1/4'' Split Lock Washer (4) 1/4-20 x 1-1/4'' SHCS (4)
12. Using the 3/16'' T-handle hex wrench provided, thread four (4) SHCS (as assembled in step 11) through the Base and into four (4) mounting holes in the Column. Tighten screws securely.
13. Connect all power cords to the isolation transformer (shortest cords to nearest outlets).
Right-angle power cord from Main Power Switch
Installation Note: Power cords are for (1) main power switch (shown left), (2) PC/Display power supply, (3) DSP box power supply, (4) label printer power supply, (5) laser printer, and (6) ink-jet printer power supply. The printers (laser or ink-jet) are both optional but their respective power cords are pre-installed and should remain connected to the isolation transformer even though neither printer is installed.
Power cords from: • PC/Display power supply • DSP box power supply • Label printer power supply • Laser printer (optional) • Ink-jet power supply (optional)
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 8 of 19
14. Insert Nylon cable tie (provided) through cable tie mount and loop around cables.
15. Fasten cable tie around cables and tighten. Cut off end of cable tie.
16. Ensure casters are locked as shown and carefully raise Cart to upright position (right).
17. Using a pair of scissors, cut ties holding keyboard tray and drawer in place.
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 9 of 19
Mounting DSP Box on Mounting Bracket Tools Required • #2 Phillips Screwdriver (not provided) • 5/32'' hex wrench (provided)
DSP Box
1. Remove power cord from Mounting Bracket.
#6-32 x 3/8'' Truss Head Phillips Srews (4)
2. Fasten Preamp to Mounting Bracket with four (4) #10-32 x 3/8'' truss head screws.
3. Using the 5/32'' hex wrench provided (shown below), slowly turn grounding screw clockwise until the screw breaks the outer coating of the DSP box, coming into contact with the aluminum housing. This will provide an electrical connection between the DSP Mounting Bracket and the DSP box.
5/32'' Hex Wrench
4. Connect power cord to DSP box.
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 10 of 19
Mounting the PC/Display at Top of Column Tools Required • #2 Phillips Screwdriver (not provided)
Mounting Adapter
M4 x 20mm PHMS (4) 1/4'' Nylon Spacer (4) M4 x 20mm PHMS (4)
1. Align spacers over four (4) mounting holes in back of PC/Display. Fasten Mounting Adapter to PC/Display with four (4)DSP M4 xBox 20mm PHMS as shown. Installation Note: Clearance holes must be oriented toward top and bottom of PC/Display to catch the plunger when Mounting Adapter slides into Mounting Plate.
Mounting Adapter
Nylon Spacers
Clearance (larger) Hole
2. Slide Mounting Adapter into Mounting Plate on PC/Display Mount. Pull spring-loaded plunger (photo below right) when Mounting Adapter makes contact. Resume sliding Mounting Adapter while releasing plunger. PC/Display is securely mounted when plunger snaps into clearance hole.
Plunger (spring-loaded)
3. Hand-tighten four (4) Nylon thumb screws on rear of Mounting Plate. Do not overtighten these screws. Installation Note: The Nylon screws should only be tightened enough to prevent looseness between Mounting Adapter and Mounting Plate.
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 11 of 19
Installing Power Cord Cleat Tools Required • 3/16'' hex wrench (provided)
1/4-20 x 1-1/2'' SHCS (2) 1/4'' HI-Collar Lock Washer (2) Power Cord Cleat
3/16'' Hex Wrench
1. Fasten the Power Cord Cleat to the rear left side of the Column with two (2) 1/4-20 x 1-1/2'' SHCS and HI-Collar lock washers.
Installing the Main Power Cord Tools Required • #2 Phillips screwdriver (not provided)
Power Cord
Cable Clamp #6-32 x 3/8'' PHMS (1)
1. Plug power cord into main power switch. Insert power cord in Cable Clamp and fasten Cable Clamp to Column with one (1) #6-32 x 3/8'' PHMS.
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 12 of 19
Installing Accessories Installing the Label Printer 1. Peel white backing off dual lock strips on bottom of label printer. This exposes the sticky side of the strips.
2. Place printer in location shown and press down until dual strips adhere to shelf. Connect power and USB cables.
Connecting Preamp, PCA 5 (Patient Cable Assembly), and ATA 5 (Acoustic Transducer Assembly) Cables The Preamp, PCA, and ATA cables are keyed (indicated by arrows). Align arrows and plug cables into sockets. Plug and unplug cables only by grasping the locking ring of the connector. Connect Preamp cable to DSP Box
Equipment Check Kit ATA 5 Cables
PCA 5 Cables
Connect PCA 5 and ATA 5 Cables to Preamp
Preamp
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 13 of 19
Connecting Cables to PC/Display Connect cables as shown below: Connections on Bottom of PC/Display
PC/Display Power
DSP Cable (COM Port 1)
USB Ports
Cable from DSP Box USB Cables from Printers and Keyboard
Power Cord for PC/Display
Installation Note: System will not function if DSP cable is connected to any port other than COM Port 1.
DSP Cable to (Com Port 1) Installation Note: System will not function if DSP cable is connected to any port other than COM Port 1.
PC/Display Power
On the back of the Panel PC, locate the threaded socket in the upper left corner and attach the wireless antenna
USB Ports • Keyboard • Label Printer • Page-Size Printer (optional) • Bar Code Scanner (optional)
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 14 of 19
Installing Optional Equipment Installation Note: Make and model of printers shown are for illustrative purposes only. Actual model may vary. Laser Printer
1. Remove Channel Cover from section of Column behind bottom shelf. Pull power cord out of channel. Installation Note: Channel Cover is removed by squeezing sides of Cover and pulling outward.
Power Cord
USB Cable (preinstalled)
2. Reinstall Channel Cover.
3. Place printer on shelf and connect power cord and USB cable to printer.
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 15 of 19
Ink Jet Printer Installation Note: The power supply for the ink jet printer must be installed in the rear channel.
1. Remove lower front Channel Cover. Installation Note: Channel Cover is removed by squeezing sides of Cover and pulling outward. USB Cable (preinstalled)
2. Open the hook-and-loop strap and remove the foam block from the lower power supply mount in the rear channel.
3. Remove backing from two (2) circles of double-backed tape.
Power Supply Cable
4. Connect power cord to bottom end of power supply and press power supply onto sticky side of double-backed tape. Installation Note: Power cord comes pre-installed in Column. Power Cord
5. Route power supply cable through opening in lower channel and onto the lower shelf.
6. Tuck excess power supply cable into channel below the power supply.
7. Reinstall channel cover. Place printer on shelf and connect power and USB cables.
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 16 of 19
Connecting the Barcode Scanner Connect the USB cord from the scanner to the PC/Display (see USB ports on page 13). Place the scanner onto the scanner holder on rear of cart.
Barcode Scanner in Barcode Scanner Holder on Rear of Cart
Barcodes for the LS2208 Barcode Scanner Scan the following barcodes into the scanner to produce the appropriate line feed, remove sound after good decode, and reduce its volume upon start-up. Acceptance of a valid scan is confirmed by a tone or tones emanating from the reader. Barcodes should be scanned in the following order: • This barcode sets the barcode scanner to default settings (factory settings). To ensure the scanner is set to default, scan this code.
• The following four barcodes must be scanned into the barcode scanner consecutively so that it will provide a line feed / tab to next field.
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 17 of 19
• This barcode disables the scanner of any sound after a successful decode.
• This barcode reduces the volume of the barcode scanner upon start-up.
Inspect Connections and Install Rear Channel Cover Ensure that all channel-mounted power supplies and cables are connected correctly, then install the green Channel Cover that protects the cable system. This is done by inserting one edge of the cover in place and squeezing the other side into place.
1. Ensure all power supplies are connected correctly and cables routed neatly inside channel.
2. Press Channel Cover into channel after inspecting connections and cable routing.
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B
Page 18 of 19
Adjusting Tilt Tension If the PC/Display does not maintain position when tilted, the tilt tension screws must be tightened. If the PC/Display does not tilt easily, the tilt tension screws must be loosened. See adjustment procedure below. 1. Using the 5/32'' hex wrench provided, equally tighten or loosen the two (2) tilt tension screws until desired tilt tension is achieved.
Tilt Tension Screws
Routine Maintenance Periodically inspect all fasteners associated with the mounting system. Tighten or adjust as necessary for optimal operation and safety. Cleaning the Mounting Assembly 1. The mounting assembly may be cleaned with most mild, non-abrasive solutions commonly used in the hospital environment (e.g. diluted bleach, ammonia, or alcohol solutions). 2. The surface finish will be permanently damaged by strong chemicals and solvents such as acetone and trichloroethylene. 3. Do not use steel wool or other abrasive material to clean the mounting assembly. 4. Damage caused by the use of unapproved substances or processes will not be covered by warranty. We recommend testing of any cleaning solution on a small area of the mounting assembly that is not visible to verify compatibility. 5. Never submerge the roll stand or allow liquids to enter the mounting assemblies. Wipe any cleaning agents off the mounting assembly immediately using a water-dampened cloth. Dry all mounting assemblies thoroughly after cleaning. CAUTION: Natus Medical Incorporated makes no claims regarding the efficacy of the listed chemicals or processes as a means for controlling infection. Consult your hospital’s infection control officer or epidemiologist. To clean or sterilize mounted instruments or accessory equipment, refer to the specific instructions delivered with those products.
ALGO 5 Newborn Hearing Screener Assembly Instructions P/N 004965 Rev B