Supplement to Solar SpO2 Module with Masimo SET Service Manual for the CARESCAPE Monitor B850 Chapter 1. Introduction
This document supplements the Solar SpO2 Module with Masimo SET Service Manual (PN 2004407-001). The purpose of this supplement is to identify the differences from the Solar SpO2 Module with Masimo SET Service Manual when the module is used with the CARESCAPE Monitor B850. The information is presented in chapter order of the Solar SpO2 Module with Masimo SET Service Manual. NOTE If you do not have a copy of the Solar SpO2 Module with Masimo SET Service Manual, you can order one from GE.
Related documents This supplement is intended to be used alongside the following documents:
Chapter 3. Maintenance
Solar SpO2 Module with Masimo SET Service Manual (PN 2004407-001)
CARESCAPE Monitor B850 V2 Technical Manual (PN 2062973-001)
CARESCAPE Modular Monitors Supplemental Information Manual (2062971-002 or 2062971-004 510k)
CARESCAPE Modular Monitors User's Manual (PN 2062971-001 or 2062971-003 510k)
For maintenance information besides that listed in this supplement, refer to the Solar SpO2 Module with Masimo SET Service Manual.
Visual inspection Refer to the Solar SpO2 Module with Masimo SET Service Manual.
Cleaning Refer to the patient monitor’s user’s manual for cleaning procedures.
2068969-005 17 January 2013
CARESCAPE Monitor B850
1
Checkout procedure Required tools/equipment The following lists the test equipment, adapters, and cables necessary to complete the checkout procedures.
CARESCAPE Monitor B850.
Simulators and cables listed below provide waveforms and patient vital signs. Item
Manufacturer and Part Number/Model
SpO2 Simulator
408610-001 or equivalent
SpO2 Simulator Cable
2006011-001 or equivalent
Any patient cable or leadwire set that you would usually use on patients.
Preparation 1. Install the module in a Tram-rac 4A housing. 2. Apply power to the monitor by turning the rear panel power switch to the ON position. 3. Ensure that the power indicator of the monitor is on. 4. Ensure that the power indicator of the module is illuminated green.
SpO2 test Follow the SpO2 test procedure described in the Solar SpO2 Module with Masimo SET Service Manual. NOTE On the CARESCAPE Monitor B850, PR indicates pulse rate.
Electrical safety tests Refer to the electrical safety tests found in the CARESCAPE Monitor B850 Technical Manual. This information supersedes the instructions found in the Solar SpO2 Module with Masimo SET Service Manual.
AC Hi-Pot Test The AC Hi-Pot test only needs to be performed when the module is disassembled to a board level.
2
CARESCAPE Monitor B850
2068969-005
Check form 1. Complete one sheet per unit. 2. Prior to testing, verify all equipment is calibrated via calibration labeling, and record calibration due date. Test configuration, conditions and test equipment for unit under test Product/Model
Customer asset tag
Maintenance type: Planned maintenance
Serial number
Comments
Corrective maintenance Measuring equipment used
ID number
Manufacturer
Model number
Description
Results
Maintenance procedure
Pass
Visual inspection:
Serial number
Checkout procedure
Fail
Cal due date
Results Pass
Fail
Preparation
Case
SpO2 test
Cables and connectors Exterior cleaning Electrical safety tests Refer to the CARESCAPE Monitor B850 Technical Manual
Test results: Pass Fail
2068969-005
Signature of tester:
CARESCAPE Monitor B850
Date:
3
GE Medical Systems Information Technologies Inc. 8200 W. Tower Ave., Milwaukee, WI 53223 U.S.A. © 2013 General Electric Company. All rights reserved.
Chapter 4. Troubleshooting
With the exception of the troubleshooting procedure below, refer to the Solar SpO2 Module with Masimo SET Service Manual.
Troubleshooting procedure Is LED on front panel green?
Yes
No
Is patient Yes data displayed No at the monitor? Yes
Try one or more of the following solutions: 1. Verify the Tram-Rac 4A housing is functional. Verify the Tram-Rac power LED is ON. Check for a loose or faulty cable from the patient monitor to the Tram-Rac housing. Power cycle the Tram-Rac 4A housing if it has an external power supply. Swap the Tram-Rac 4A housing with a known good one. 2. If module LED is off or PROBE OR MODULE MALFUNCTION displays on the screen, there is a hardware problem with the module. Reseat the Solar SpO2 module in the Tram-Rac 4A housing. Remove the sensor from the module, then reseat the Masimo module in the Tram-Rac housing. Verify the Tram-Rac housing is a Tram-Rac 4A housing and is turned ON. Verify the host is a patient monitor and is turned ON. Verify the patient monitor is not resetting itself. Swap the module with a known good one. Try one of the following solutions: 1. Verify correct patient or simulator leadwire connections to the module. 2. Contact GE Technical Support.
No
Is the parameter box displayed at the monitor? Yes
Verify that the patient cable is connected.
No
End
This is the extent of troubleshooting steps.
Chapter 5. Parts lists and drawings
No SpO Module with Masimo SET Service Manual. Refer Yes to the Solar 2
Yes 4
No
CARESCAPE Monitor B850
2068969-005