Basic Service Manual
521 Pages
Preview
Page 1
Voluson® P8 / Voluson® P6/ Basic Service Manual • Voluson® P8 systems • Voluson® P6 systems
Document Number: 5450839-100 Revision: 5
0459
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
Important Precautions
WARNING WARNING (EN) (EN)
AVERTISSEMENT (FR)
WARNUNG (DE)
• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY. • IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE TRANSLATION SERVICES. • DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD. • FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL OR OTHER HAZARDS. • CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS. • SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE TRADUIRE. • NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS. • LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES. • DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER SPRACHE. • FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT, IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG ZU SORGEN. • WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN ANWWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND NACHVOLLZIEHEN KÖNNEN. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
-
i
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
AVISO (ES)
ATENÇÃO (PT)
AVVERTENZA (IT)
HOIATUS (ET)
ii
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
• ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, LA TRADUCCIÓN ES RESPONSABILIDAD DEL CLIENTE. • NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO. • LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE OTRA NATURALEZA. • ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL EM INGLÊS. • SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO. • NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA. • O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS, MECÂNICOS OU OUTROS. • IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN INGLESE. • SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE. • SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL CONTENUTO. • NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI. • KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES. • KUI KLIENDITEENINDUSE OSUTAJA NÕUAB JUHENDIT INGLISE KEELEST ERINEVAS KEELES, VASTUTAB KLIENT TÕLKETEENUSE OSUTAMISE EEST. • ÄRGE ÜRITAGE SEADMEID TEENINDADA ENNE EELNEVALT KÄESOLEVA TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST. • KÄESOLEVA HOIATUSE EIRAMINE VÕIB PÕHJUSTADA TEENUSEOSUTAJA, OPERAATORI VÕI PATSIENDI VIGASTAMIST ELEKTRILÖÖGI, MEHAANILISE VÕI MUU OHU TAGAJÄRJEL.
-
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
VAROITUS (FI)
ΠΡΟΕΙ∆ΟΠΟΙΗΣΗ (EL)
FIGYELMEZTETÉS (HU)
VIÐVÖRUN (IS)
• ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU. • REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG SKILIÐ ÞESSA ÞJÓNUSTUHANDBÓK. • EF EKKI ER FARIÐ AÐ ÞESSARI VIÐVÖRUN GETUR ÞAÐ VALDIÐ MEIÐSLUM ÞJÓNUSTUVEITANDA, STJÓRNANDA EÐA SJÚKLINGS VEGNA RAFLOSTS, VÉLRÆNNAR EÐA ANNARRAR HÆTTU.
-
iii
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
ĮSPĖJIMAS (LT)
ADVARSEL (NO)
• DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK. • HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE. • IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT. • MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ GRUNN AV ELEKTRISK STØT, MEKANISKE ELLER ANDRE FARER.
OSTRZEŻENIE (PL)
ATENŢIE (RO)
-
v
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
DAMAGE IN TRANSPORTATION All packages should be closely examined at time of delivery. If damage is apparent write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested within this 14 day period.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be performed by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall be performed by qualified GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes. The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
OMISSIONS & ERRORS If there are any omissions, errors or suggestions for improving this documentation, please contact the Voluson® P8 / Voluson® P6 Service Documentation Group with specific information listing the system type, manual title, part number, revision number, page number and suggestion details. Mail the information to:Service Documentation, 65-1, Sangdaewon-dong,Jungwon-gu, Seongnamsi,Gyeonggi-do 462-120, Korea. GE Healthcare employees should use TrackWise to report service documentation issues. These issues will then be in the internal problem reporting tool and communicated to the writer.
SERVICE SAFETY CONSIDERATIONS DANGER !! DANGER:
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING. TO PREVENT ELECTRIC SHOCK, SERVICE PERSONNEL TO AVOID CONTACT WITH HV AND MV CIRCUIT PARTS DURING SYSTEM ON CONDITION
WARNING WARNING Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety !! WARNING:
viii
glasses, and kneeling pad, to reduce the risk of injury.
-
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
LEGAL NOTES The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of Voluson® P8 / Voluson® P6. Voluson® P8 / Voluson® P6 may revise this publication from time to time without written notice.
TRADEMARKS All products and their name brands are trademarks of their respective holders.
COPYRIGHTS All Material Copyright© Voluson® P8 / Voluson® P6 by General Electric Inc. All Rights Reserved.
-
ix
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
Revision History
Revision
Date
Reason for change
1
August 2012
Initial release
2
September 2012
Updated Part numbers for Models
3
September 2012
Updated the rating plates and removed tilt indicator
4
September 2012
Upated isolation transformer information
5
January 2013
Updated Software version
List of Effected Pages
Pages Title
Revision 5
Pages
Revision
Chapter 3 - Setup Instructions pages 1-78 Chapter 4 - Functional Checks
Pages
Revision
5
Chapter 8 - Replacement Procedures pages 1-122
4
5
Chapter 9 - Renewal Parts pages 1-24
4
Important Precautions pages i-x
5
Table of Contents pages 1-24
5
Chapter 5 - Components and Functions (Theory) pages 1-78
5
Chapter 10 - Care & Maintenance pages 1-36
4
Chapter 1 - Introduction pages 1-18
5
Chapter 6 - Service Adjustments pages 1-4
5
Index pages 1-4
4
Chapter 2 - Site Preparation pages1-10
5
Chapter 7 - Diagnostics/ Truobleshooting pages 1-44
5
Rear Cover pages 1-2
4
x
pages 1-54
-
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
CHAPTER 1 Introduction Overview... 1 - 1 Purpose of Chapter 1... 1 - 1 Purpose of Service Manual... 1 - 1 Typical Users of the Basic Service Manual... 1 - 2 Models Covered by this Manual... 1 - 2 Purpose of Operator Manual(s)... 1 - 3 Important Conventions... 1 - 3 Conventions Used in this Manual... 1 - 3 Standard Hazard Icons... 1 - 4 Product Icons... 1 - 5 Safety Considerations... 1 - 7 Introduction... 1 - 7 Human Safety... 1 - 7 Mechanical Safety ... 1-7 Electrical Safety... 1 - 8 Safe Practices... 1 - 8 Probes... 1 - 8 Auxiliary Devices Safety... 1 - 8 Labels Locations... 1 - 10 Main Label... 1 - 11 UL Label... 1 - 12 Multi Caution Label... 1 - 13 Dangerous Procedure Warnings... 1 - 14 Lockout/Tagout Requirements (For USA Only)... 1 - 14 Returning/Shipping System, Probes and Repair Parts... 1 - 14 Electromagnetic Compatibility (EMC)... 1 - 15 What is EMC?... 1 - 15 Compliance... 1 - 15 Electrostatic Discharge (ESD) Prevention... 1 - 15 Customer Assistance... 1 - 16 Contact Information... 1 - 16 System Manufacturer... 1 - 16
Table of Contents
i
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
CHAPTER 2 Site Preparation Overview... 2 - 1 Purpose of Chapter 2... 2 - 1 General Console Requirements... 2 - 2 Environmental Requirements... 2 - 2 Cooling... 2 - 2 Lighting... 2 - 2 Electrical Requirements... 2 - 2 Voluson® P8 / Voluson® P6 Power Requirements... 2 - 3 Inrush Current... 2 - 3 Site Circuit Breaker... 2 - 3 Site Power Outlets... 2 - 3 Main Power Plug... 2 - 3 EMI Limitations... 2 - 4 Probe Environmental Requirements... 2 - 5 Time and Manpower Requirements... 2 - 5 System Specifications... 2 - 5 Physical Dimensions of Voluson® P8 / Voluson® P6... 2 - 5 Acoustic Noise Output... 2 - 5 Electrical Specifications... 2 - 5 Facility Needs... 2 - 6 Purchaser Responsibilities... 2 - 6 Mandatory Site Requirements... 2 - 7 Site Recommendations... 2 - 7 Recommended Ultrasound Room Layout... 2 - 8 Networking Setup Requirements... 2 - 9 Stand-alone Unit (without Network Connection)... 2 - 9 Unit Connected to Hospital’s Network... 2 - 9 Purpose of the DICOM Network Function... 2 - 9 DICOM Option Pre-installation Requirements... 2 - 9
ii
Table of Contents
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
CHAPTER 3 Setup Instructions Overview... 3 - 1 The Purpose of Chapter 3... 3 - 1 Set Up Reminders... 3 - 1 Average Installation Time... 3 - 1 Installation Warnings... 3 - 2 Moving/Lifting the System... 3 - 2 System Acclimation Time... 3 - 2 OPIO Position... 3 - 2 Brake Pedal Operation... 3 - 3 Safety Reminders... 3 - 3 Receiving and Unpacking the Equipment... 3 - 4 Preparing for Set Up... 3 - 8 Verify Customer Order... 3 - 8 EMI Protection... 3 - 8 Connection of Auxiliary Devices... 3 - 10 Connecting the LCD Monitor... 3 - 11 Connecting the Black & White Printer... 3 - 12 Connection Scheme: B&W Printer... 3 - 13 Connecting the Color Printer... 3 - 14 Connection Scheme: Color Printer... 3 - 15 Connecting the DeskJet Color Printer... 3 - 16 Connection via Bluetooth Adapter... 3 - 16 Connecting the Wireless Network Adapter... 3 - 18 Connecting the Footswitch... 3 - 19 Connecting the USB Flash Memory Stick... 3 - 20 Connecting the external USB Hard disk (Handydrive)... 3 - 20 General Remarks and Hints when using external USB-Devices... 3 - 21 External USB-Devices - Connection... 3 - 21 External USB-Devices - Disconnection... 3 - 21 Completing the Set Up... 3 - 23 Connecting the Unit to a Power Source... 3 - 23 Power On / Boot Up... 3 - 23 Scanner Power On... 3 - 23 Back End Processor Boot Up... 3 - 25 During a normal boot, you may observe... 3 - 26 Power Off / Shutdown... 3 - 27 Table of Contents
iii
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
Scanner Shutdown... 3 - 27 Transducer Connection... 3 - 29 Printer Installation... 3 - 30 Installing Digital Black & White Printer Sony UP-D897... 3 - 30 Installing Digital Color Printer Sony UP-D25MD... 3 - 30 Installing the Bluetooth Deskjet Printer... 3 - 31 Installing the USB Bluetooth Adapter and the Bluetooth Software... 3 - 31 Configuration of the Bluetooth Software... 3 - 33 Prepare the Printer and connect the Bluetooth Printer Adapter... 3 - 36 HP Officejet H470 Printer... 3 - 36 Installing the printer software/driver... 3 - 36 Configuration of Printer Properties... 3 - 40 HP Officejet H470 - Printer Properties... 3 - 41 Printer Installation manually... 3 - 42 Adjustment of Printer Settings... 3 - 46 UP-D897 - Printer Settings... 3 - 47 P95D - Printer Settings... 3 - 48 UP-D25MD - Printer Settings... 3 - 50 Remote Control Selection... 3 - 53 Report Printer Selection... 3 - 53 System Configuration... 3 - 54 System Setup... 3 - 54 To invoke the Setup procedure:... 3 - 54 How to enter Date and Time... 3 - 56 How to enter Hospital Name... 3 - 56 How to change Language and/or EUM Language... 3 - 56 How to activate Screen Lock... 3 - 57 How to adjust function of the Footswitch (Left/Middle/Right)... 3 - 57 How to change the Keyboard Layout... 3 - 58 How to configure Service Platform... 3 - 58 Measure Setup... 3 - 59 To invoke the Setup procedure:... 3 - 59 On-Board Optional Peripherals... 3 - 60 External I/O Connectors... 3 - 61 External I/O Pin Outs... 3 - 61 Available Probes... 3 - 63 Software/Option Configuration... 3 - 63 Connectivity Setup... 3 - 64 Connectivity Introduction... 3 - 64 The Dataflow Concept... 3 - 64 Dataflow Examples... 3 - 65 iv
Table of Contents
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
Stand-alone Voluson® P8 / Voluson® P6... 3 - 66 Voluson® P8 / Voluson® P6 + PC (with 4D View Software) within a “Sneaker Net”... 3 - 66 Connection between Voluson® P8 / Voluson® P6 and DICOM Server 3 - 66 Wireless Network Configuration... 3 - 67 Description of “Netgear” Configuration Parameters... 3 - 69 Network IP Address Configuration... 3 - 70 Map Network Drive... 3 - 71 Connectivity Setup Worksheet... 3 - 73 Paperwork... 3 - 75 Product Locator Installation... 3 - 75 User Manual(s)... 3 - 75
Table of Contents
v
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
CHAPTER 4 Functional Checks Overview... 4 - 1 Purpose of Chapter 4... 4 - 1 Required Equipment... 4 - 1 General Procedure... 4 - 2 Power On / Boot Up... 4 - 2 Scanner Power On... 4 - 2 Power Off / Shutdown... 4 - 4 Scanner Shutdown... 4 - 4 System Features... 4 - 5 Control Panel... 4 - 5 Monitor Display... 4 - 6 Functional Checks... 4 - 7 2D Mode Checks... 4 - 8 XTD-View Check... 4 - 11 M Mode Checks... 4 - 12 MCF Mode Check... 4 - 13 Spectral Doppler Mode Checks... 4 - 14 Color Doppler Mode Checks... 4 - 16 Volume Mode Checks... 4 - 18 Pre-Volume Mode Functions... 4 - 19 Functions after the Acquisition... 4 - 21 3D/4D Sub Menu... 4 - 23 Using Cine... 4 - 24 Activating Cine... 4 - 24 Cine-Split Function (Multiple Format)... 4 - 24 Activating 2D Cine... 4 - 24 Spectral Doppler- or M Cine Loop... 4 - 24 Activating 3D Rotation Cine... 4 - 24 Activating Volume Cine... 4 - 24 Activating Auto Cine... 4 - 24 Activating Cine Calc... 4 - 25 Generic Measurements... 4 - 26 Distance and Tissue Depth Measurements (2D and M Mode)... 4 - 26 Circumference/Area Measurements... 4 - 27 Volume Measurements... 4 - 27 Multiplane Measurements... 4 - 27 Measurements in Spectral Doppler Mode... 4 - 28 Auto Trace... 4 - 28 Manual Trace... 4 - 28 vi
Table of Contents
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
Heart Rate... 4 - 28 Calculations... 4 - 29 Worksheet (Report) Pages... 4 - 30 Probe/Connectors Usage... 4 - 31 Connecting a probe... 4 - 31 Activating the probe... 4 - 31 Deactivating the probe... 4 - 31 Disconnecting the probe... 4 - 31 Patient Archive (Image Management)... 4 - 32 Erasing DVD/CD... 4 - 33 Backup and Restore Database, Preset Configurations and Images... 4 - 34 Save Image Settings Only... 4 - 35 Load Image Settings Only... 4 - 36 Preparations... 4 - 36 Load “Complete Backup”... 4 - 37 Load only parts of the “Complete Backup”... 4 - 38 Save Full System Configuration (Full Backup)... 4 - 39 Load Full System Configuration (Full Backup)... 4 - 41 Delete Full System Configuration (Full Backup)... 4 - 43 Archiving Images... 4 - 44 Save Image Archive... 4 - 44 Load Image Archive... 4 - 46 Software Configuration Checks... 4 - 47 System Setup... 4 - 47 Measure Setup... 4 - 48 Peripheral Checks... 4 - 49 ECG Check Out... 4 - 49 Control Console Positioning... 4 - 49 Assembly procedure for converting Voluson P6/P8 from Low Version to High Version... 4 - 49 Brakes and Direction (Swivel) Locks... 4 - 65 Site Log... 4 - 66
Table of Contents
vii
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
CHAPTER 5 Components and Functions (Theory) Overview... 5 - 1 Purpose of Chapter 5... 5 - 1 General Information... 5 - 2 Description of Voluson® P8 / Voluson® P6 Operating Modes... 5 - 6 B-Mode or 2D-Mode... 5 - 6 Harmonic Imaging (HI)... 5 - 6 XTD-View... 5 - 6 M-Mode... 5 - 6 MCFM Mode (M Mode + Color Flow Mode)... 5 - 6 Color Doppler Modes... 5 - 7 Color Flow Mode... 5 - 7 Power Doppler... 5 - 7 Bi-Directional Angio (HD-Flow Mode)... 5 - 7 Pulsed (PW) Doppler... 5 - 8 3D Imaging... 5 - 8 3D Data Collection and Reconstruction... 5 - 8 3D Image Presentation... 5 - 8 3D Rendering... 5 - 9 Block diagram Voluson® P8 / Voluson® P6... 5 - 10 Front End Block diagram... 5 - 11 Back End Block diagram... 5 - 11 OPIO Block diagram... 5 - 11 Data Flow Control Description... 5 - 12 B-Mode... 5 - 12 Special B-Mode Techniques... 5 - 12 M-Mode... 5 - 13 D-Mode (Pulsed Wave- and Continuous Wave Doppler)... 5 - 13 D-Mode Autotrace (draws PC-calculated envelope to D-Spectrum) . . . 5 - 13 CFM-Mode (Color Flow Mode)... 5 - 14 3D-Mode (Freezes after 1 volume sweep)... 5 - 14 Real Time 4D-Mode (nonstop volume rendering)... 5 - 14 XBeam CRI-Mode (CrossBeam Compound Resolution Imaging)... 5 - 14 Archive write mode (store Image to Archive)... 5 - 14 Description of Software Options... 5 - 15 3D/4D Advance... 5 - 16 Real Time 4D... 5 - 16 3D/4D Basic... 5 - 16 Real Time 4D... 5 - 16 Real Time 4D Biopsy... 5 - 16 DICOM... 5 - 16 Anatomical M-Mode (AMM)... 5 - 16 Advanced SRI... 5 - 16 viii
Table of Contents
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
SonoNT... 5 - 17 Description of Hardware Options... 5 - 18 CW - Continuous Wave Doppler... 5 - 18 DVR Recorder... 5 - 18 ECG Module... 5 - 18 Wireless Network Adapter (WLAN - Wireless Local Area Network) . . . 5 - 18 Data Location... 5 - 18 FrontEnd Processor... 5 - 20 PIR - Probe Interface board... 5 - 21 BF64 - Beam former board... 5 - 22 RFS - Radio Frequency interface and System controller... 5 - 23 RFS board - FPGA... 5 - 23 BackEnd Processor... 5 - 25 Hard Disk Drive... 5 - 26 SOM (System On Module)... 5 - 26 Congatec... 5 - 26 DVR - Digital Video Recorder (option)... 5 - 27 Internal I/O... 5 - 28 OPIO (User Interface)... 5 - 29 OPIO Block Diagram... 5 - 29 OPIO layout... 5 - 30 Button Description... 5 - 30 Monitor... 5 - 35 External I/O... 5 - 36 Peripherals... 5 - 37 Recording Tools... 5 - 37 DVD Recorder... 5 - 37 Printers... 5 - 37 Black & White Digital Printer... 5 - 37 Color Digital Printer... 5 - 37 Color Deskjet Printer... 5 - 37 DVD+R/RW Drive (Writer)... 5 - 37 ECG Module... 5 - 37 Wireless Network Adapter... 5 - 37 Power Distribution... 5 - 38 ACFE - Primary Power Module... 5 - 38 CPS Block Diagram... 5 - 39 Table of Contents
ix
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
Power Distribution Block Diagram... 5 - 40 Power Up Sequence Description... 5 - 41 Power Up Sequence Description... 5 - 41 Power Down Sequence Description... 5 - 42 Mechanical Concept and Overview... 5 - 42 Functions of ACFE... 5 - 42 ACDC - primary DC Power Supply... 5 - 44 Function of ACDC... 5 - 44 DCDC - main DC Power Supply... 5 - 44 4D Module... 5 - 44 Rotation of the Control Console... 5 - 44 Height Adjustment of the OPIO Console... 5 - 44 Service Platform... 5 - 45 Introduction... 5 - 45 Access / Security... 5 - 45 Local Access... 5 - 45 Local(Extended) Access... 5 - 46 Remote Access... 5 - 47 Common Service Desktop (CSD)... 5 - 48 Internationalization... 5 - 48 Error Logs... 5 - 49 Diagnostics... 5 - 49 Image Quality... 5 - 50 Configuration... 5 - 50 Utilities... 5 - 51 Replacement... 5 - 51 PM... 5 - 51 Service Page... 5 - 52 Introduction... 5 - 52 Access / Security... 5 - 52 Service Login... 5 - 52 Auto Tester... 5 - 52 Update... 5 - 53 FMI from DVD... 5 - 53 ASI - Additional Software Installation... 5 - 53 Restore Software... 5 - 53 TCP/IP Buffersize... 5 - 53 Common Service Desktop (CSD)... 5 - 53 Request for Service... 5 - 53 Delete all Patients... 5 - 53 Export System Data... 5 - 53 Keyboard Layout... 5 - 54 x
Table of Contents
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
Monitor Test... 5 - 54 Load Application Settings... 5 - 54 Boot Screen Functions... 5 - 55 Overview... 5 - 55 Voluson... 5 - 55 GE-Service... 5 - 55 Rollback... 5 - 56 Memtest... 5 - 56 LINUX - Voluson Maintenance System... 5 - 57 Overview... 5 - 57 General Operation and Selection in LINUX (and BIOS) menu... 5 - 57 Access/Security... 5 - 57 Before Recovering or Initializing the Hard disk... 5 - 58 Voluson Maintenance System... 5 - 60 Media... 5 - 60 Advanced... 5 - 60 About... 5 - 60 Exit/Reboot... 5 - 60 Halt... 5 - 60 Media Menu... 5 - 61 Exit... 5 - 61 Restore... 5 - 61 Initialize... 5 - 64 NewDisk... 5 - 64 BIOS “Boot Priority Order” Setting... 5 - 65 New Hard Disk Setup... 5 - 68 Advanced Functions Menu... 5 - 72 Exit... 5 - 72 Network... 5 - 72 SSHD... 5 - 73 Restore... 5 - 73 Initialize Hard... 5 - 73 ServiceTools... 5 - 73 Exit... 5 - 73 Shell... 5 - 73 USBTree... 5 - 74 SmartHDDtest... 5 - 74 HDDWipeTool... 5 - 74 CollectLogs... 5 - 74 Bios Update... 5 - 74 After Recovering or Initializing the Hard disk... 5 - 75 Memory Check in LINUX... 5 - 77
Table of Contents
xi
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5
VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
CHAPTER 6 Service Adjustments Overview... 6 - 1 Purpose of Chapter 6... 6 - 1 Regulatory... 6 - 1 LCD Monitor Adjustment... 6 - 2 Modification of Keyboard Layout... 6 - 3 Setup the Voluson® P8 / Voluson® P6 Keyboard Language Layout... 6 - 3
xii
Table of Contents