CME McKinley

T and T34 Syringe Pumps

T34 Quick User Guide v7 Sept 2016

Quick User Guide

2 Pages

Alerts, Alarms and Troubleshooting Screen prompts  Result/cause  Keypad Locked  Only the STOP, START and INFO keys are accessible.  Disengage keypad lock if further access required.  The infusion has been stopped.  Press YES to Resume the infusion or press NO to continue stopped state.  Alert: Program is about to end/syringe is almost empty.  Prepare to change syringe or discontinue pump use.  Low Battery  Alert: Battery is almost depleted.  Prepare to change battery.  Pump Paused Too Long  Alarm: The pump has been stopped/paused for more than 2 minutes without any key presses.  Press YES to resume the infusion, press NO to continue pause for another two minutes or power off.  Syringe Empty, Remove Syringe  Alarm: Current infusion program has completed/ syringe is empty.  Prepare to change syringe or discontinue pump use.  Alarm: Battery will fail imminently.  Change battery.  Syringe Displaced, Check Syringe  Alarm: One or more of the syringe detection sensors is not detecting.  Check the syringe and re-seat as necessary. Check screen messages for assistance.  Occlusion/Empty Syringe, Check Line  Alarm: Clamped line, occluded or kinked. Actuator has reached the minimum travel position.  Release the clamp, flush/replace the access device or clear the occlusion.  Pump Stopped, Press YES to Resume Program Nearly Complete  Troubleshooting  End Battery  System Error. Press & Hold INFO for Details. If problem persists send pump for service. ERROR Startup MotMov Fail If problem persists send pump for service.  Possible actions  Alarm: An internal system error has occurred. Two examples of system failure screen messages are shown here, refer to the pump service manual for a full list of error codes. The user may be prompted to power off and restart, which may rectify the error. If the error recurs, take pump out of use. Press INFO to obtain error message, record error code and summary of fault and return pump to designated service centre.  Program protection and "Resume" • •  Pre-Loading and syringe empty alarm clears a program from the pump memory. If the option is available to resume a program, a screen prompt displays.  20ml BD Plastipak Select /, Press YES  Press YES to Resume, NO for New Syringe  T34™ Ambulatory Syringe Pump LOCK ON DURATION (Prime and Load) Quick User Guide  •  The information contained in this guide is a summary only, based on default settings. Refer to Operating Manual for full operating instructions.  •  Users must have undertaken training before operating this device.  •  Screen information is representative only and some text wording/ screen information may vary slightly with different software versions.  •  You must refer to local policy and procedures for specific guidance on pump settings/set-up and use of accessories (e.g. cannula and administration lines).  •  Always follow screen prompts. Before pressing keys to proceed, ensure selections made correspond with what is required.  If the syringe size/brand displayed matches the one used, press YES. Press YES to retain the current program (mL/hr rate is protected). Press NO to delete the current program which allows a new infusion to be programmed.  “Resume” protects the infusion rate for the current program, so: • If the syringe volume is increased and the infusion is resumed, the duration of delivery will increase • If the syringe volume is decreased and the infusion is resumed, the duration of delivery will decrease NOTE: Follow local policy/procedure for the option to press when this screen displays following purge.  Doc Ref: T34 QUG01/Lock ON Duration P-L/September 2016/v7 Tel: 01253 206700 | Fax: 01253 896648 | Email: customersupport@cmemedical.co.uk | cmemedical.co.uk
File Type: PDF
File Size: 288 KB
File Name: T34 Quick User Guide v7 Sept 2016.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions