Spare Parts Instruction
Pneumatic Spring G5009701
Replacing the Pneumatic Spring Tools needed: •
8mm Hex wrench
Note: • • • •
The installer is required to use an ESD (electro-static discharge) Grounding Wrist strap during this procedure to avoid unintentional damage to the electronic devices in the machine. Do not remove any cards or lC chips from their antistatic containers until you are ready to install them. When removing cards or chips from a system, immediately place them in an antistatic bag or container. When handling cards or lC's, hold them by their edges. Avoid touching the components and connector leads on the card. Avoid touching the leads on the le. Do not slide cards or lC's over any surface. Avoid plastic, vinyl , and Styrofoam in your work area
Setup instructions
1. 2.
Check that the machine is switched off. Open the cabinet using a 8mm Hex wrench!
3.
Remove the locking washers and replace the pneumatic spring Close the cabinet
4.
Thanks for providing us with information he I ping us improving our service.
5.
To continuously improve our customer support, please send the following information to the below address. • Running Time: • Serial number: • Reason for change: • Changed component: • Country/Hospital l"tlrlla~'m I hen.com E-mail - ric .customer
Final check: •
Check that the break foot stays up, and also that it is possible to press it down to break and to pull it up again to un-break.
Working time necessary: Replacement: Final check:
10 minutes. 2 minutes.
,.,
Template-2235
GAMBRO Renal Products
Page 1/1
Revision: 4 Reference: POP·2279
MDREC-1 001 05
Version: 2.0
Effective date: 2006-09-04