Service Manual
51 Pages
Preview
Page 1
Service Manual
E lectrosurgical G enerator
LIMITED WARRANTY For a period of two years following the date of delivery, CONMED Corporation warrants the CONMED Sabre Genesis™ Electrosurgical Generator against any defects in material or workmanship and will repair or replace (at CONMED’s option) the same without charge, provided that routine maintenance as specified in this manual has been performed using replacement parts approved by CONMED. This warranty is void if the product is used in a manner or for purposes other than intended.
STERILE EO © 2009 CONMED Corporation MdSS GmbH EC
REP
525 French Road Schiffgraben 41 STERILE D-30175 Hannover Utica, NY 13502-5994 USA Germany
R
STERILE A
U.S. Patent Numbers 6,830,569 - 6,875,210 - 6,948,503 D552241 and other patents pending.
2
0123 Authorization For Technical Service or Return LOTPhone: 303-699-7600 / 1-800-552-0138 Extension 5274 Fax 303-699-1628 CUT
COAG
For Customer Service or to order parts phone: 1-800-448-6506 / 315-797-8375 / Fax 315-735-6235 or contact your CONMED Representative. EC
REP
STERILE EO
European Authorized Representative MdSS GmbH Schiffgraben 41 MDSS GmbH STERILE D-30175 Hannover Schiffgraben 41 Germany D-30175 Hannover STERILE Germany
R A
9083
The revision level of this manual is specified by~ the highest revision letter found on either theLOT inside front cover 0123 or enclosed errata pages (if any).
2
Manual Number 60-8202-ENG Rev. - 10/09 CUT
COAG
Unit Serial Number_________________________________
9083
~
Table of Contents & List of Illustrations
Section
Title
Page
1.0 2.0 3.0
General Information... Refer to Operator’s Manual (60-8201) Specifications... Refer to Operator’s Manual (60-8201) Theory of Operation... 3-1
3.1
Mode Descriptions... 3-1
3.2
System Overview... 3-1
3.3
Optional System Configurations... 3-6
4.1 4.2 4.3 4.4 4.5
General Maintenance Information... 4-1 Maintenance Personnel... 4-1 Cleaning... 4-1 Periodic Inspection... 4-1 Periodic Performance Testing... 4-1
4.6
System Calibration... 4-6
4.7
Last Fault Code Retrieval & Clear... 4-9
4.8
Optional System Configuration ... 4-9
3.1.1 3.1.2 3.1.3 3.2.1 3.2.2 3.2.3 3.2.4 3.2.5 3.2.6 3.2.7 3.2.8 3.2.9 3.2.10
4.0
4.5.1 4.5.2 4.5.3 4.5.4 4.5.5 4.5.6 4.5.7 4.6.1 4.6.2 4.6.3 4.6.4 4.6.5 4.6.6 4.7.1 4.7.2
Cut Major Modes... 3-1 Coag Major Modes... 3-1 Bipolar Major Modes... 3-1 RF Power Supply (RFPS)... 3-4 RF Amplifier and Transformer... 3-4 Electrosurgical Outputs . ... 3-4 Activation Command Sensing ... 3-5 Automatic Return Monitor (A.R.M.)... 3-5 System Controllers and Monitor... 3-5 Low Voltage Power Monitoring... 3-5 Operator Control Panel... 3-5 Activation Tones... 3-6 Activation Relay Connector... 3-6
Maintenance... 4-1
Chassis Ground Integrity... 4-1 Displays, Alarms and Commands... 4-1 Output Power... 4-2 RF Leakage Measurement... 4-3 Line Frequency Leakage... 4-4 Automatic Return Monitor (A.R.M.) Check... 4-5 Output Coupling Capacitor Check... 4-5 Calibration Preliminaries ... 4-6 Selecting the Mode to Calibrate... 4-6 Calibrating a Monopolar Mode... 4-8 Calibrating Bipolar Mode... 4-8 Calibrating A.R.M... 4-8 Completing Calibration... 4-9 Last Fault Code Retrieval... 4-9 Clearing Last Fault Codes... 4-9
Section
Title
4.9 4.10
DACview ... 4-12 Troubleshooting... 4-12
4.11 4.12
Parts Ordering Information... 4-14 Assembly Breakdown/Parts Access... 4-15
4.13
Fault Codes... 4-18
4.10.1
4.12.1 4.12.2 4.12.3 4.12.4 4.12.5 4.12.6
Appx. A
Figure/Title
Page
HVPS Troubleshooting Hints... 4-14
Top Cover Removal & Replacement... 4-15 Bezel Removal & Replacement... 4-15 Control/Display Board Removal & Replacement... 4-16 Power Board Removal & Replacement... 4-16 RF Output Board Removal & Replacement... 4-17 Power Transistor Replacement... 4-17
Schematics & BOMs...A-1 Bill of Material: Top Assembly & Chassis... A-1 Bill of Material: Controller PCB Assembly... A-7 Bill of Material: Power PCB Assembly... A-13 Page
Figure 3.2 Figure 4.1 Figure 4.2 Figure A.1a Figure A.1b Figure A.3a
System Block Diagram... 3-3 Calibration Procedure Flow Chart... 4-7 DIP Switch Positions... 4-11 Controller PCB (Top)... A-7 Controller PCB (Bottom)... A-8 Power PCB... A-15
Table 4.1 Table 4.2 Table 4.3 Table 4.4 Table 4.5 Table 4.6 Table 4.7 Table 4.8 Table 4.9 Table 4.10 Table 4.11 Table 4.12
Monopolar Cut Mode RF Output Power Accuracy... 4-2 Monopolar Coag Mode RF Output Power Accuracy... 4-2 Bipolar Mode RF Output Power Accuracy... 4-2 Allowable RF Leakage Current to Ground... 4-3 Allowable RF Leakage Current - Inactive Monopolar Outputs ... 4-4 Allowable RF Leakage Current - Inactive Bipolar Outputs... 4-4 Line Frequency Allowable Leakage - Inactive... 4-4 Line Frequency Allowable Leakage - Active... 4-5 DIP Switch Settings... 4-10 DACview Channels... 4-12 Troubleshooting... 4-12 Fault Codes... 4-18
Schematic A.1a Controller PCB - Top Sheet... A-1 Schematic A.1b Controller PCB - Controller Sheet... A-2 Schematic A.1c Controller PCB - Monitor Sheet... A-3 Schematic A.1d Controller PCB - FPGA Sheet... A-4 Schematic A.1e Controller PCB - DAC Sheet... A-5 Schematic A.1f Controller PCB - Display Sheet... A-6 Schematic A.2a Power PCB - Power Amplifier... A-9 Schematic A.2b Power PCB - RF Output / Arm Sense... A-10 Schematic A.2c Power PCB - RF Output / RS-232/Relay Drive... A-11 Schematic A.2d Power PCB - RF Output... A-12 Schematic A.3a HV-Power factor Correction... A-13 Schematic A.3b Forward Converter... A-14
Theory of Operation Section 3.0
Sabre Genesis™ functions and essential circuit information are provided in this section. This section begins with a description of the key parameters for each mode. This is followed by an overview of how the system functions and some key operational information for the modules within the system.
3.1
Mode Descriptions
The key functional parameters for each mode are presented here. Nominal mode specifications are provided in section 1.2.11 of the Sabre Genesis™ Operators Manual. 3.1.1
Cut Major Modes
Major Mode
Minor Mode
Nominal RF Frequency
Modulation (Number of Modulation (Normal Pulses, Nominal Time On/Off Frequency & Period)
Cut
Pure
400 KHz
None
None
Pulsed
400 KHz
70µs / 600µs
670µs
Blend
400 KHz
9 pulses, 23µs / 17µs
25 KHz / 40µs
3.1.2
Coag Major Modes
Major Mode
Minor Mode
Nominal RF Frequency
Modulation (Number of Modulation (Normal Pulses, Nominal Time On/Off Frequency & Period)
Coag
Standard
495 KHz
Single pulse
40 KHz / 25µs
Standard mode is fundamentally different from the Cut mode in that the resonant circuit of the RF Amplifier and Transformer combination is excited by the energy of a single pulse, causing the resonant circuit to ring until the energy is dissipated. 3.1.3
Bipolar Major Modes
Major Mode
Minor Mode
Nominal RF Frequency
Modulation (Number of Modulation (Normal Pulses, Nominal Time On/Off Frequency & Period)
Bipolar
Macro
400 KHz
None
3.2
None
System Overview
Mains power is converted to electrosurgical output power through the RF Power Supply (RFPS), the RF Amplifier, and the Transformer and Output sections of the system. Mains power is converted to high voltage direct current power in the RFPS to supply the RF Amplifier. This is essentially a power transformer with a power factor corrected regulator. The power factor correction can be enabled or disabled under software control. Pulses generated in the RF Controller are amplified to electrosurgical power and voltage levels in the RF Amplifier and Transformer portions of the power train. Three high-voltage bipolar transistors and a single MOSFET make up the hybrid-cascode RF Amplifier. The hybrid-cascode amplifier is a fast, high-voltage amplifier that can be controlled by the combination of DC voltage (VBASE_PWM) and a fixed amplitude, variable pulse width signal (RFGATE). This amplifier is combined through a relay with either the Monopolar output transformer or the Bipolar output transformer to generate electrosurgical power. Electrosurgical power flows from the RF Amplifier / Transformer section to the Output section where the power is switched to the specific electrosurgical outputs. The Output section also has circuitry to detect
3-1
activations from accessories and the circuitry to perform the Automatic Return Monitor (A.R.M.) function to ensure the integrity of the dispersive electrode connection. The power section also includes a number of output voltage and current sensors that are used by the RF Controller for control of power delivery and by the Monitor to detect errant output conditions. Windings on the Monopolar output transformer and the Bipolar output transformer are the means for sensing output voltage. Separate primary-side current transformers are shared by the bipolar and monopolar channels for control and monitoring of the current. There are also separate current sensors on the monopolar outputs that are used to detect stuck output relays. The RF Controller is a Field Programmable Gate Array (FPGA) that generates the RFGATE and VBASE_PWM RF Amplifier drive signals based upon a comparison of measured parameters and settings-based parameters. The pulse train sequence is a settings-based parameter that is dependent on the selected mode. Target power, current limit, and voltage limit are all settingsbased parameters derived from a load curve that is specific to the selected mode and front panel power setting. The RF Controller samples electrosurgical output current and output voltage from sensors at a 20 Megahertz rate and uses these sampled values to calculate sensed current, sensed voltage, and output power. This high sample rate allows control of the real power delivered to the active accessory and also allows the Sabre Genesis™ to rapidly adapt to changing loads. The output current, output voltage, and output power are compared with corresponding settings based parameters of current limit, voltage limit, and target power, respectively, and the RF Controller adjusts value of VBASE_PWM in a closed-loop fashion to control these parameters. The RF Controller also provides fixed pulses for RFGATE within each mode-based pulse train sequence. The RF Monitor is a Digital Signal Processor (DSP) that is used to monitor the system for safety problems that can result from a variety of conditions. • The Monitor has independent sensors for output voltage and current, which it uses to calculate power for comparison with the power that the RF Controller senses and for comparison with the generator power setting.
3-2
• To ensure that the correct outputs are activated, the Monitor also independently senses current at each of the outputs, looking for current flow that would indicate electrosurgical power at outputs other than the selected output. • The Monitor senses the audio output to ensure that a tone occurs whenever electrosurgical outputs are active. • The RF Amplifier drive signal is sensed by the Monitor to detect improper frequencies or improper pulse sequences for the selected mode. • The Monitor independently compares the activation signal with that seen by the System Controller to ensure that the activation signal is consistent. The Monitor has the capability to independently disable the electrosurgical output if a problem is detected. The System Controller provides the primary control interface to the user and other outside systems, including the serial interface, the activation relay, tone generation, and displays. Finally, the Display accepts all user input and provides all user feedback. The Display is controlled by the System Controller through a serial interface and illuminates the LED display elements in a time division multiplexed fashion; the illuminated LED display elements are actually on less than half the time. The Display also provides for user input through the buttons on the control panel, including switch de-bouncing and conditioning. Figure 3.2 illustrates the key elements of the system in block diagram form.
Activation Request
Indicators: Activation & Mode
Keyboard Modes / Power
Displays
Activation Relay Connector
ACT RLY
Serial Interface Connector
RS232
SPI
AL TONE ACT TONE System Controller
A.R.M RF Output Relays
Patient
VARM RLY DRV
Mono
Bip
RF INH PFC EN HVEN RF Controller
RF MP VS RF BP VS
RF Amp
RF IS Host Bus
RFDRV
Tone Mon
RFHVSup VBPWM
HV Power Supply
Monitor
HFINH
M ISN MRF BP VSN MRF MP VSN MRF H1 SN MRF HF SN
Figure 3.2 System Block Diagram
3-3
3.2.1 High Voltage Power Supply (HVPS)
(Vbase_PWM).
The HVPS is comprised of a Power Factor Control (PFC) section and a Forward Converter (FC) section. The PFC converts Mains power to approximately 400 volts using techniques that ensure the mains current into the supply is sinusoidal and in phase with the mains voltage. By doing so, RMS current and harmonic distortion are reduced. The Forward Converter then converts the PFC output to an adjustable DC voltage for use by the RF amplifier.
The RFGATE drive pulses provide the basic pulse pattern that is used to form the electrosurgical waveform, and have a set pulse-pattern and pulsewidth for each mode. A drive of several pulses at a frequency that closely matches the resonant frequency of the amplifier characterize Cut and Blend modes, and the output pulses substantially correspond to the drive. Spray and Standard Coag modes, however, are characterized by pulses that occur less frequently where the amplifier is allowed to “ring” at its resonant frequency.
The System Controller can enable or disable the PFC section of the HVPS. The PFC is normally enabled during operation to ensure a resistive load is presented to the Mains. The Forward Converter is a switch-mode power converter that adjusts its operating frequency between 25KHz and 100KHz to ensure proper resolution for the commanded output voltage. Isolation between Mains power, the HVPS and the LVPS (+12V) output occurs in the Forward Converter. The RF Monitor enables the output of the HVPS. The forward converter includes current limiting on the output and has provisions to shutdown when the output of the Low Voltage Supply exceeds limits. The +12V output is then used to supply input voltage to a variety of low-voltage regulators on the controller board. 3.2.2
RF Amplifier and Transformer
The RF Amplifier and Transformer portions use a switch-mode resonant hybrid-cascode amplifier to convert the power from the RFPS to the RF energy necessary for electrosurgery. One may think of the amplifier as a high-speed switch that pulses current through a resonant circuit, which is formed by the monopolar or bipolar transformer together with capacitors that are connected to the transformer primary and secondary windings. The transformers are designed with a good deal of leakage inductance in order to provide inductance for resonating with the capacitors. One Metal-Oxide-Semiconductor Field Effect Transistor (MOSFET) is connected in series with three parallel bipolar transistors to provide the switching. The pulses to drive the gate on the MOSFET in this arrangement come from the RF Controller (RFGATE). The base connections of the three parallel bipolar transistors are also driven by a signal that originates from the RF Controller
3-4
Rapid regulation of the output power in this arrangement is provided by VBASE_PWM; as VBASE_PWM is increased, the output power increases. As noted in the RF Controller discussion, the RF Controller compares the output power with the desired power and adjusts the VBASE_PWM to minimize the difference. VBASE_PWM enters the amplifier as a 312 KHz Pulse-Width-Modulated (PWM) signal that is filtered to become a variable DC base drive signal. Finally, the RF Amplifier and Transformer provide capabilities for sensing RF output current and voltage. The voltage sensors that are used for power control and power monitoring are independent windings on the output transformers. The current is also measured on the primary side of the transformers. With proper characterization of the transformer, the controller obtains an accurate representation of the voltage, current, and thus the output power of the system. 3.2.3
Electrosurgical Outputs
Relays are provided to isolate electrosurgical outputs and select which outputs are active. The System Controller selects the appropriate output relays based upon activation command inputs. The Monitor utilizes current sensors implemented on each monopolar electrosurgical output to determine whether current is flowing only to the correct outputs. In the event that current flows in an output that is not selected, the Monitor can independently disable RF. The monitor uses the bipolar primary voltage to sense that the bipolar relay has been activated.
3.2.4
Activation Command Sensing
Each of the Hand Controlled Accessory receptacles incorporate inputs that are used to sense an activation command from the user. Each monopolar hand controlled accessory receptacle has an input for Cut and an input for Coag. The bipolar receptacle incorporates a single activation input. Each of these five inputs is isolated from the other electrosurgical outputs and from other low-level circuitry in the system. All are powered by a multiple output isolated power supply. The footswitch activation inputs on the back panel are configured in a similar way and share one of the isolated power supply outputs. 3.2.5
Automatic Return Monitor (A.R.M.)
The patient return connector interfaces to single and dual dispersive electrodes using a two-pin connector. A.R.M. circuitry uses an actively driven impedance measurement circuit, which allows the System Controller to detect the type of dispersive electrode connected and verify its integrity. 3.2.6
System Controllers and Monitor
Two processors and an FPGA are used for system interface & control, RF control, and system monitor functions. The RF control section consists of a dual-channel architecture with two independent channels where one is used exclusively for RF output control and the other is used for safety monitoring. All of these elements are located on the Control board, along with circuitry for interface with the user. • System Controller (System Microcontroller): A dedicated Microcontroller that handles the entire user interface, Serial Interface, and enables/disables the power factor control section of the RFPS using the PFC_EN signal. The System Controller can also disable the signal used to drive the RF Amplifier and can terminate RF drive at any time without interaction from either the RF Controller or the Monitor. The System Controller is comprised of a standard architecture microprocessor together with portions of the FPGA, which provides interface logic to a variety of signals, independent voltage regulators, a processor supervisory reset circuit, and other interface logic.
RFGATE outputs. To reduce the effects on the microprocessor circuits on the Control/ Display Board from RF noise at the output, VBASE_PWM and RFGATE are both differential mode signals running between the Control/Display Board and the power section. The RF Controller is capable of disabling RF output power and putting the system into a safe state without any interaction from the Monitor or the System Controller. The RF Controller independently monitors the RF output voltage and current for control purposes through several scaled inputs. The RF Controller is comprised of the major portion of the FPGA, together with circuitry necessary for converting the control signals between analog and digital form. • RF Monitor: A DSP that is dedicated to safety monitoring activities. The Monitor is capable of disabling RF output power and putting the system into a safe state without any interaction from the RF Controller or the System Controller. To ensure that the Monitor can correctly perform its function, the Monitor is resistively isolated from the System Controller and the RF Controller and has independent voltage regulation. The RF Monitor independently monitors a variety of inputs to detect safety problems and has control of disable signals for RF Amplifier drive. The Monitor is comprised of a DSP, together with circuitry necessary for converting the signals monitored between analog and digital form, an FPGA to provide interface logic, independent voltage regulators, isolation resistors and other interface logic. 3.2.7
Low Voltage Power Monitoring
The low voltage power supply is monitored in hardware and resets the processors if it is out of range. The microprocessor supervisory device on the Control/Display Board monitors +3.3V and +1.8V and will reset the system should the levels drop approximately 0.3V. The Control/Display
• RF Controller: An FPGA implementing digital signal processing elements for control of RF power using the VBASE_PWM and
3-5
Board has the circuit that will reset the system should the 3.3V supply exceed 3.6V. 3.2.8
Operator Control Panel
• Keyboard: The main operator input device for choosing operating modes and settings is the membrane keyboard panel. Tactile-feedback mechanical switches allow the operator to set modes and adjust power settings. • Display Panel: Consists of 7-segment displays and discrete dual colored LED’s that will display all controls and settings. LED display elements are illuminated in a time division multiplexed fashion; the illuminated LED display elements are actually on less than half the time. 3.2.9
Activation Tones
Tone is generated for all activation requests, fault detection and changes made on the Control Panel. The System Controller generates the tone signal (TONE_DRV), which is amplified by a driver. The activation tone is adjustable and controlled by an output from the System Controller, but alarm tones are not adjustable and are set to generate a tone greater than 65 dB. There is circuitry to permit the Monitor to verify the oscillation from voltage measured across the speaker, which provides confirmation that the speaker is indeed generating audible tones during activation. RF output is inhibited should the
3-6
speaker drive current be absent or too low. 3.2.10 Activation Relay Connector There is an Accessory Relay Connector, which provides a relay closure (SPST switch) that may be used for activating external accessories such as smoke evacuation units.
3.3
Optional System Configurations
An eight-position configuration dipswitch (S2), located on the Control/Display Board Assembly (A4) allows a qualified service technician to change some of the factory default settings. With the exception of the DACview switch, which is only effective in Test Mode, the configuration dipswitch settings are only detected when power is initialized, so any changes to the switch positions will not be detected until power is cycled. Each switch is OFF in the Down position and ON in the UP position. (The system detects changes in the DACview switch while power is on, so it is treated differently.) Relevant information for the configuration dipswitches appears in Section 4.8.
Maintenance Section 4.0
This section contains information useful in the maintenance and repair of the Sabre Genesis™.
Johnson Company. Formula 409® is a registered trademark of the Clorox Company.)
WARNING: High voltages are present at the connections and within the Sabre Genesis™. Maintenance personnel should take precautions to protect themselves. Read the safety summary in Section 1.1.4 of the Sabre Genesis™ Operators Manual before working on the ESU.
4.4
4.1
General Maintenance Information
Although the Sabre Genesis™ has been designed and manufactured to high industry standards, it is recommended that periodic inspection and performance testing be performed to ensure continual safe and effective operation. Ease of maintenance was a primary consideration in the design of the Sabre Genesis™. Maintenance features of this unit include microprocessor aided troubleshooting aids and push button calibration, built in fault detection, circuit protection, and easy access to circuitry while the unit is operational. These features, coupled with the warranty, local support, loaner equipment, factory support, toll free phone service to the factory and available factory training ensure a minimal maintenance effort with extensive support available.
4.2
Maintenance Personnel
Only Hospital Qualified Biomedical Technicians or ConMed factory technicians should perform service on the Sabre Genesis™. Refer all servicing to a Hospital Qualified Biomedical Technician. If necessary, your CONMED sales representative will be happy to assist you in getting your equipment serviced.
4.3
Cleaning
The interior of the unit may be vacuumed or blown out as required. The exterior of the unit may be cleaned by wiping it with a cloth that has been dampened (not dripping) with a mild detergent such as Windex® or Formula 409®. (Windex® is a registered trademark of the S.C.
Periodic Inspection
The Sabre Genesis™ should be visually inspected at least every six months. This inspection should include checks for the following: 1) Damage to the power cord and plug. 2) Proper mating and absence of damage to the accessory connectors. 3) Any obvious external or internal damage to the unit. 4) Any accumulation of lint or debris within the unit or heatsink. 5) Control Panel cuts, punctures, or dents.
4.5
Periodic Performance Testing
The Sabre Genesis™ should be tested for correct performance at least once every year. Every unit is supplied with a serialized Production Test Data Sheet that tabulates the results of the factory tests that were performed on the unit. This data is supplied so that it may be used as a reference for subsequent tests. Recommended periodic performance tests are listed in the following sections. 4.5.1
Chassis Ground Integrity
Connect a standard ohmmeter between the earth ground prong on the power plug and the Equipotential Ground Connection. Compensate for lead resistance. Confirm less than 0.2 ohms resistance is measured. 4.5.2
Displays, Alarms and Commands
Perform the Preliminary Functional Test procedure described in section 2.4.1 of the Sabre Genesis™ Operators Manual to verify proper operation of displays, alarms and commands.
4-1
4.5.3
Output Power
results.
1) Equipment Requirements: a) Monopolar Footswitch b) Bipolar Footswitch c) Commercial ESU Tester (e.g. Fluke 454A or equivalent) with non-inductive 50 load for bipolar modes and a non-inductive 500 ohm load for monopolar modes. Note: Micro Bipolar is particularly sensitive to the load resistance. A 50 ohm load should be used for checking power to obtain the best
2) Use test leads to connect the ESU tester to the unit’s return electrode output and the footswitch controlled active output. Set the Load resistance per mode as indicated in Tables 4.1 and 4.2. 3) Perform the monopolar power tests indicated in Tables 4.1 and 4.2. The acceptance range is given in both Watts and Amps to accommodate available test equipment. It is not necessary to test for both power and current. Table 4.1 Monopolar Cut Mode RF Output
Power Accuracy Mode
Load (ohms) Power Setting Watts (min) Watts (max)
Amps (min)
Amps (max)
Pure
500
20
17
23
0.184
0.214
500
100
90
110
0.424
0.469
500
200
180
220
0.600
0.663
500
20
17
23
0.184
0.214
500
100
90
110
0.424
0.469
200
180
220
0.600
0.663
Blend
500
Table 4.2 Monopolar Coag Mode RF Output Power Accuracy Mode
Load (ohms) Power Setting Watts (min) Watts (max)
Amps (min)
Amps (max)
Standard
500
20
17
23
0.184
0.214
500
50
45
55
0.300
0.332
500
80
72
88
0.379
0.420
4) Disconnect the ESU tester from the unit.
6) Perform the bipolar power tests indicated in Table 4.3. This table only provides the minimum number of points to be tested.
5) Use test leads to connect the ESU tester to the Bipolar Accessory outputs.
Table 4.3 Bipolar Mode RF Output Power Accuracy Mode
Load (ohms) Power Setting Watts (min) Watts (max)
Amps (min)
Amps (max)
Macro
50
20
17
23
0.238
0.277
50
50
45
55
0.387
0.428
4-2
4.5.4
RF Leakage Measurement
NOTE: To ensure accuracy when making leakage measurements, perform all leakage testing using methods and instruments that are compliant with the prcedures outlined in Section 19 of IEC60601-2-2 (Particular Requirements for the Safety of High Frequency Surgical Equipment). RF Leakage can present a hazard in the operating room because electrosurgical currents can flow
to the patient and operating room staff through unintended paths, which can cause injury. RF leakage occurs because the total energy in the output voltage waveform is provided with a conductive path through stray parasitic capacitance distributed within the generator and along the length of the leads. Table 4.4 presents the allowed RF leakage currents to ground.
Table 4.4 Allowable RF Leakage Current to Ground MEASURED TERMINAL
ACTIVATED ACCESSORY
MODE
RF LEAKAGE (Ma)
Dispersive Electrode
Coag Combination Monopolar
Standard Coag
< 100
Dispersive Electrode
Cut Combination Monopolar
Pure Cut
< 100
Dispersive Electrode
Hand Controlled
Standard Coag
< 100
Combination Monopolar Active
Coag Combination Monopolar
Standard Coag
< 100
Bipolar Right
Bipolar Footswitch
Bipolar Macro
< 67
Bipolar Left
Bipolar Footswitch
Bipolar Macro
< 67
Equipment: • ESU Tester with RF Leakage function -OR• 0-250 is RF Ammeter with a 200 ohm 10 W Non-inductive Resistor • Patient Plate Adapter Plug • 2 - Test leads, 1 m max. Length • 3 - Test leads, 10 cm max. Length • Wooden table approximately 1 m from floor. NOTE: Use a measuring device that meets IEC specification for RMS measured over one second. Procedure: 1) Ensure that the unit is fully assembled and all fasteners are tight. 2) Place the ESU tester or meter with resistor on the table so that they are at least 0.5m away from the unit under test and any other conductive surface. 3) Set the unit for full power for the modes noted in the table. Connect the ESU tester in accordance with the manufacturer’s instructions -OR- connect the 200-ohm noninductive resistor in series with the 250 mA RF ammeter to the Equipotential Ground Connection on the Rear Panel. Also make sure there are no connections to any output other than the one you are measuring. WARNING: HAND CONTROL ACTIVA-
TIONS SHOULD BE KEYED USING 3” OR LESS WELL-INSULATED JUMPER. USE OF AN INSULATING ROD TO INSERT THE JUMPER IS ADVISED TO PREVENT RF BURNS. 3) One at a time, connect test setup to each RF output terminal indicated in Table 4.4 and activate the unit using the corresponding command. Confirm no meter readings exceed the specified maximum. Hand controlled Coag activations are accomplished by connecting a jumper between the left jack and center jack of the desired hand switched accessory jack. RF leakage should also be measured between inactive outputs and the Dispersive Electrode connection. The procedure is as follows: 1) Set the unit for full power for the modes noted in Table 4.5. Connect the ESU tester according to manufacturer’s instructions - OR- the 200-ohm non-inductive resistor in series with the 250 mA RF ammeter to the Dispersive Electrode connection on the front panel. Also make sure there are no connections to any output other than the one you are measuring. 2) One at a time, connect this series combination to each RF output terminal indicated
4-3
in Table 4.5 and activate the unit using the corresponding command. Confirm that no meter readings exceed the specified maximum. Table 4.5 Allowable RF Leakage Current - Inactive Monopolar Outputs MEASURED TERMINAL
ACTIVATED ACCESSORY
MODE
RF LEAKAGE (Ma)
Combination Monopolar Active
Hand Controlled
Standard Coag
< 50
Combination Monopolar Active
Bipolar Footswitched
Macro
< 20
Hand Controlled Active
Combination Monopolar
Standard Coag
< 50
Hand Controlled Active
Bipolar Footswitched
Macro
< 20
Bipolar Left
Right Hand Controlled Standard
Coag
< 48
Finally, RF leakage should be measured between the inactive bipolar outputs while a monopolar accessory is activated. Do the following: 1) Set the unit for full power for the bipolar mode noted in Table 4.6. Connect ESU tester
according to manufacturer’s instructions -OR the 200-ohm non-inductive resistor in series with the 250 mA RF ammeter between the two bipolar output connections. 2) Activate and verify the limit in Table 4.6.
Table 4.6 Allowable RF Leakage Current - Inactive Bipolar Outputs MEASURED TERMINAL
ACTIVATED ACCESSORY
MODE
RF LEAKAGE (Ma)
Bipolar Right to Left
Hand Controlled
Standard Coag
< 48
4.5.5 Line Frequency Leakage CAUTION: To prevent RF current from destroying the test equipment and/or affecting leakage readings, set all power settings to zero. WARNING: ELECTROCUTION HAZARD. USE OF AN ISOLATED MAINS POWER SOURCE IS RECOMMENDED WHEN OPENING THE MAINS GROUND DURING THE FOLLOWING SAFETY TESTS.
Circuit ground and Neutral (Low Mains) must be connected together for Mains leakage testing. Equipment: These tests are performed most conveniently using any good quality biomedical electrical safety tester. Procedure: 1) Connect the electrical safety analyzer to make the measurements indicated in Table 4.7. 2) Mode: Measure leakage for Bipolar to Neutral and Chassis to Neutral.
Table 4.7 Line Frequency Allowable Leakage - Inactive RF output to Neutral
LINE
GND
LIMIT max
Equipotential Ground
Normal
Closed
30 µA
Equipotential Ground
Reversed
Closed
30 µA
Equipotential Ground
Normal
Open
270 µA
Equipotential Ground
Reversed
Open
270 µA
Dispersive Electrode
Normal
Closed
15 µA
Dispersive Electrode
Reversed
Closed
15 µA
Dispersive Electrode
Normal
Open
15 µA
Dispersive Electrode
Reversed
Open
15 µA
Bipolar Output*
Normal
Closed
15 µA
Bipolar Output*
Reversed
Closed
15 µA
Bipolar Output*
Normal
Open
15 µA
Open
15 µA
Bipolar Output* Reversed *Measure the Bipolar Output with Bipolar connections shorted together.
4-4
5) Since the Sabre Genesis™ monopolar active outputs are disconnected by relays when the unit is not activated, active-to-neutral leakage tests must be performed with the unit activated in order to be valid. 6) With all power controls set to zero, measure the leakage current as in step 1 from each of the three active output terminals to neutral;
see Table 4.8; while that output is activated in Cut by the appropriate footswitch or hand control jumper. Hand control cut activations are accomplished by connecting a jumper between the two outer jacks of where the handcontrolled accessory is plugged into the unit.
Table 4.8 Line Frequency Allowable Leakage - Active RF output to Neutral
LINE
GND
ACTIVATION
LIMIT max
Combination Monopolar Active
Normal
Closed
Combination Monopolar Cut
15 µA
Combination Monopolar Active
Reversed
Closed
Combination Monopolar Cut
15 µA
Combination Monopolar Active
Normal
Open
Combination Monopolar Cut
15 µA
Combination Monopolar Active
Reversed
Open
Combination Monopolar Cut
15 µA
Hand Controlled Active
Normal
Closed
Hand Controlled Cut
15 µA
Hand Controlled Active
Reversed
Closed
Hand Controlled Cut
15 µA
Hand Controlled Active
Normal
Open
Hand Controlled Cut
15 µA
Hand Controlled Active
Reversed
Open
Hand Controlled Cut
15 µA
4.5.6 Automatic Return Monitor (A.R.M.) Check A.R.M. has two specific ranges that will be tested initially and then the circuit will be tested to verify that the circuit measures dispersive electrode resistance correctly. For this testing, only a Decade Resistance Box (DRB) and a dispersive electrode cable adapter are required. Connect the DRB to the Dispersive Electrode Receptacle using the dispersive electrode cable adapter. A.R.M. may be reset by disconnecting the dispersive electrode connector or adjusting the DRB above 10K Ohms until the Single and Dual Dispersive Electrode Status/Alarm Indicators flash red in alternating fashion. Allow approximately two seconds after the DRB is changed before proceeding to the next step in the procedure. A.R.M. indicators not mentioned in the procedure must be off for each test. 1) Dual Electrode Alarm Limit: Set the DRB to 158 Ohms, then connect it to the Dispersive Electrode Receptacle and verify that the Single and Dual Dispersive Electrode Status/ Alarm Indicators flash red in alternating fashion. 2) Dual Electrode Upper Limit: Set DRB to 140 Ohms and verify that the Dual Dispersive Electrode Status/Alarm Indicator is Green.
3) Dual Electrode Lower Limit: Set the DRB to 15 Ohms and verify the Dual Dispersive Electrode Status/Alarm Indicator is Green. 4) Single Electrode Upper Limit: Set the DRB to 7 Ohms, then reset A.R.M. and verify the Single Dispersive Electrode Status/Alarm Indicator is Green. 4.5.7
Output Coupling Capacitor Check
WARNING: ENSURE ALL POWER SETTINGS ARE AT 0 WATTS BEFORE CONDUCTING THIS TEST TO PREVENT INJURY TO PERSONNEL AND DAMAGE TO TEST EQUIPMENT. NOTE: Not all capacitance meters will read properly for this test. The test frequency should be at or below 1 kHz for best accuracy. The following meters have been tried successfully: Fluke 189, Extech 285, Sencor LC75 and HP4284A (1 kHz setting or below). 1) Connect shorting plug to banana adapter to the two pin Dispersive Electrode Receptacle. Use 6” or shorter test leads to connect a capacitance meter between the shorting plug adapter and the footswitched Combo plug. 2) Measure capacitance and confirm it is less than 0.2 nF. 3) Confirm cut power is set to 0, then activate
4-5
and confirm capacitance is between 0.6 and 0.9 nF. 4) Do not activate for this bipolar test. Move test leads to Bipolar Output Accessory Receptacles. Confirm capacitance is between 2.2 and 2.5 nF.
4.6
System Calibration
The Sabre Genesis™ is calibrated during manufacture using equipment traceable to National Institute of Standards & Technology (NIST) standards and should retain its accuracy for a long period of time. Recalibrate the generator after repair or if it performs out of specification. Check the calibration in normal operating mode and only perform calibration if errors are identified. The Sabre Genesis™ stores its calibration in nonvolatile semiconductor memory, so the calibration will be retained without any action on the part of the user or maintenance staff. Calibration should be checked in normal operating mode during annual preventative maintenance to ensure there is no change. Calibration is required when: • “Err 138”, “Err 139”, or “Err 140” occurs: An error is detected with the stored calibration values. • “Err 143” or “Err 321” occurs: One or more modes require calibration. • “Err 135” occurs: An error is detected with stored ARM calibration values. • Either the Control/Display Board assembly (Conmed P/N 61-6991) or the output RF assembly (Conmed P/N 61-8102) is replaced.
system configuration DIP switch details. With this configuration set, turn on power while pressing and holding both Speaker Volume Up/ Down Keys. Release these Keys when CAL appears in the Monopolar Cut Power Digital Display and the software revision appears in the Monopolar Coag Power Digital Display. CAL and the software revision may persist in the displays for a few seconds after the Volume Adjust Keys are released. The display will then provide an indication of the calibration status: • “ALL” will appear in the Monopolar Cut Power Digital Display if the calibration memory is empty. • “nEr” will appear in the Monopolar Cut Power Digital Display, where “n” indicates how many major modes require calibration, will be displayed if only particular modes require calibration. All of the minor mode indicators will be illuminated and the minor modes needing calibration will flash. • “[U”, “[0A”, “bP”, or “Pad” will appear in the Monopolar Cut Power Digital Display to indicate the major mode when only minor modes under that major mode require calibration. All of the minor mode indicators will be illuminated and the minor modes needing calibration will flash. • “[U” will appear in the Monopolar Cut Power Digital Display with the Pure Cut Mode Indicator illuminated if all modes are calibrated.
Refer to Figure 4.1 for calibration process flow.
For all except the last of these, a single Press and release of the Tone Loudness Adjustment Down Key is required to proceed past this point on the menu. After pressing this key, “[U” will appear in the Monopolar Cut Power Digital Display with the Pure Cut Mode Indicator illuminated.
4.6.1
4.6.2
• Calibration differences are found during preventative maintenance. Calibration Preliminaries
Sabre Genesis™ calibration occurs in Calibration Operating Mode, which is entered by setting the system configuration DIP switches on the Control/Display Board. Set the Calibration system configuration DIP switch (Control/Display Board SW2.2) to the ON (UP) position and the Test system configuration DIP switch (Control/ Display Board SW2.1) to the OFF (DOWN) position. Other configuration DIP switch settings positions will not affect this. See Section 4.8 for
4-6
Selecting the Mode to Calibrate
Press the Monopolar Cut Power Adjustment Keys to select the major mode to calibrate as displayed in the Monopolar Cut Power Digital Display. The selections are “[U” for Cut, “[0A” for Coag, “Bp” for Bipolar, or “PAd” for the Dispersive Electrode A.R.M. connection. If any of the minor modes under these major modes are not calibrated, the displayed major mode will flash. Select the monopolar minor mode by pressing the appropriate Mode Select Key.
Calibration Set the Calibration System Configuration Dipswitch on the Controller to the ON position Turn main power switch on while pressing both Volume Adjust Keys. Release Volume Adjust Keys when the Monopolar Cut Power Digital Display indicates “ [AL” [AL Lxx
ALL
[u 500 P
2Er
3Er
[u
[0A
bP
PAd
Press Tone Loudness Adjustment Down Key
Press Monopolar Cut Power Adjustment Keys to select [u, [0A, bP and PAd. [u
[0A
bP
Pad 10
Press Cut Minor Mode Select Key for Pure or Blend. Do NOT select Pulsed Connect indicated load with meter to output
Connect 10 ohm load Two-Pin Dispersive Electrode Receptacle Press Bipolar Power Up Adjustment Key
Press Tone Loudness Adjustment Down Key Activate when target level displayed. Press Monopolar Coag Power Adjustment keys to match power/current to Calibration Target. Minimum activation 2 seconds, release. Press Tone Loudness Adjustment Down Key.
Press Up Arrow Key PAd 150 Connect 150 ohm load Two-Pin Dispersive Electrode Receptacle. Press Bipolar Power Up Adjustment Key.
Power off Set the Calibration System Configuration Dipswitch on the Controller to the OFF position End
Figure 4.1 Calibration Procedure Flow Chart
4-7
4.6.3
Calibrating a Monopolar Mode
4.6.4
Calibrating Bipolar Modes
This section applies to the Pure Cut, Blend, Standard Coag.
The Bipolar modes are calibrated using a method that is very similar to the Monopolar modes.
Calibration may be performed by measuring current or by measuring power. To select between calibration using measured current and measured power, press the Bipolar Power Adjustment Keys to set the calibration units to either “A” for current or “P” for power.
Calibration may be performed by measuring current or by measuring power. To select between calibration using measured current and measured power, press the Bipolar Power Adjustment Keys to set the calibration units to either “A” for current or “P” for power.
The resistance to be used for calibration will appear in the Monopolar Coag Power Digital Display. Connect a resistor of this value between the output connection that is being used for calibration and both pins on the Two-Pin Dispersive Electrode Receptacle.
The resistance to be used for calibration will appear in the Monopolar Coag Power Digital Display. Connect a resistor of this value between the two active connections in the Bipolar Accessory Receptacle.
Press and release the Tone Loudness Adjustment Down Key to begin calibration. After this key is pressed, the target level appears in the Monopolar Coag Power Digital Display. Activate using the appropriate Handswitch or Footswitch. Power will now flow to the resistor. While monitoring either the current or the power, adjust the power up or down using the Monopolar Coag Power Adjustment Keys until the measured value is as close to the target level as possible. The activation must be maintained for a minimum of 2 seconds to ensure the calibration is valid. After the power is properly adjusted, release the activation. Press and release the Tone Loudness Adjustment Down Key to complete the calibration sequence for the selected minor mode. To complete the Blend calibration, activate again using the appropriate Handswitch or Footswitch. Power will now flow to the resistor. While monitoring either the current or the power, adjust the power up or down using the Monopolar Coag Power Adjustment Keys until the measured value is as close to the target level as possible. The activation must be maintained for a minimum of 2 seconds to ensure the calibration is valid. After the power is properly adjusted, release the activation. Press and release the Tone Loudness Adjustment Down Key to complete the Blend calibration sequence. After a minor mode has been calibrated, the associated minor mode indicator will quit flashing. When all of the minor modes within a major mode have been calibrated, the major mode indicated in the Monopolar Cut Power Digital Display will quit flashing.
4-8
Press and release the Tone Loudness Adjustment Down Key to begin calibration. After this key is pressed, the target level appears in the Monopolar Coag Power Digital Display. Activate using the Bipolar Footswitch. Power will now flow to the resistor. While monitoring either the current or the power, adjust the power up or down using the Monopolar Coag Power Adjustment Keys until the measured value is as close to the target level as possible. The activation must be maintained for a minimum of 2 seconds to ensure the calibration is valid. After the power is properly adjusted, release the activation. Press and release the Tone Loudness Adjustment Down Key to complete the calibration sequence for the bipolar mode. After a Bipolar minor mode has been calibrated, the associated Bipolar minor mode indicator will quit flashing. When the Bipolar mode has been calibrated, the major mode indicated in the Monopolar Cut Power Digital Display will quit flashing. 4.6.5
Calibrating A.R.M.
A.R.M. is calibrated against a pair of known resistances. Press and release the Tone Loudness Adjustment Down Key to begin calibration. The resistance to be used for calibration will appear in the Monopolar Coag Power Digital Display. Connect a resistor of this value ±1% between the two active connections in the Two-Pin Dispersive Electrode Receptacle. Calibrate the particular value connected by pressing one of the Bipolar Power Adjustment Keys. When the value is accepted, a two-tone sequence
will sound and the resistance in the Monopolar Coag Power Digital Display will quit flashing. Now scroll to the other pair of known resistances using the Monopolar Coag Power Adjustment Keys. The resistance to be used for calibration will appear in the Monopolar Coag Power Digital Display. Connect a resistor of this value ±1% between the two active connections in the TwoPin Dispersive Electrode Receptacle. Calibrate the particular value connected by again pressing one of the Bipolar Power Up Adjustment Keys. When the value is accepted, a two-tone sequence will sound and the resistance in the Monopolar Coag Power Digital Display will quit flashing.
Window will next display “Err”; the Coag window will display the error code (a numeric value); and the Bipolar Window will display the storage location of that error code. Last Fault display example:
Err
381
1
3) Scroll through the stored error codes using the Bipolar Power Adjustment Keys. The error codes are stored Last in, First out. A “1” in the Bipolar Display shows the last error that occurred. Press the Bipolar Up key and a “2” will be displayed if more than one error occurred.
After A.R.M. has been calibrated, the major mode “PAd” indicated in the Monopolar Cut Power Digital Display will quit flashing.
4) To retrieve the settings when the error occurred, it is necessary to have a Handcontrol accessory connected. Press both Cut and Coag activation switches and the Display Panel will show the system settings when the error occurred.
4.6.6
4.7.2
Press and release the Tone Loudness Adjustment Down Key to complete A.R.M. calibration.
Completing Calibration
Turn power off and set the Calibration system configuration DIP switch (Control/Display Board SW2.2) to the OFF (DOWN) position. See Section 4.8 Displaying Optional System Configuration for system configuration DIP switch details. The ESU will be ready for normal operation the next time the power is turned on.
4.7
Last Fault Code Retrieval & Clear
Up to 50 error (Err) and accessory (A[[) codes can be stored in memory for retrieval. When retrieving the error codes, it is also possible to retrieve the system settings when the error occurred. 4.7.1
Last Fault Code Retrieval
1) Turn on power while pressing and holding both Volume Adjust Select Keys. Release these Keys when LF[ appears in the Monopolar Cut Power Digital Display and the software revision appears in the Monopolar Coag Power Digital Display. LF[ and the software revision may persist in the displays for a few seconds after the Volume Adjust Select Keys are released. This action will place the system in the Last Fault Code Mode (LFC). Electrosurgical outputs cannot be activated while the system is in LFC.
Clearing Last Fault Codes
As errors occur, fault codes from earlier errors are erased in a last-in-first-out fashion. While it is not absolutely necessary to clear the older codes, clearing the codes may be desirable in some situations. • Pressing the Monopolar Cut Power Adjustment down Key followed by the Tone Loudness Adjustment Down Key will clear the entire fault code memory. The cut window will display “[Lr” when codes are cleared.
4.8
Optional System Configuration
The eight-position configuration DIP switch (S2), located on the Control/Display Board Assembly allows a qualified service technician to change some of the factory default settings. The default switch is only read during Power on Self Test (POST) or when the system is powered on, so any changes to the switch positions should be made with the main power off. Each switch is OFF in the down position and ON in the up position. Relevant information for each switch is described in Table 4.9 and the positions are illustrated in Figure 4.2.
2) If any errors are stored in memory, the Cut
4-9
Table 4.9 DIP Switch Settings Config. Title / Display Switch Element Position
Default
Description for Off
Description for On
1
TEST / Cut 100’s
Off
Run Mode. Required position for surgery.
Activates Test Mode, which inhibits most of the system level monitoring for troubleshooting purposes. When this switch is ON, both Volume Adjust Select Keys on the Display Panel must be pressed until 888 appears in the Monopolar Cut Power Digital Display and the software revision appears in the Monopolar Coag Power Digital Display. 888 and the software revision may persist in the displays for a few seconds after the Bipolar Mode Select Keys are released. If both Bipolar Mode Select Keys are not pressed, and Err 100 is displayed and the power must be cycled.
2
CAL/ Cut 10’s
Off
Run Mode. Required position for surgery.
Required for calibration of output power and A.R.M. When this switch is ON, both Volume Adjust Select Keys on the Display Panel must be pressed until [AL appears in the Monopolar Cut Power Digital Display and the software revision appears in the Monopolar Coag Power Digital Display. [AL and the software revision may persist in the displays for a few seconds after the Bipolar Mode Select Keys are released. If both Bipolar Mode Select Keys are not pressed, and Err 100 is displayed and the power must be cycled.
3
Not Used / Cut 1’s
Off
Not Used
Not Used
4
Not Used / Coag 100’s
Off
Not Used
Not Used
5
LAST / Coag 10’s
On
Defaults to Pure Cut, Standard Coag, and Micro Bipolar and sets all power levels to zero (0W) each time the system is initialized.
Defaults all modes and power levels to the last activated settings from the last time the system was powered down.
4-10