CARESCAPE Modular Monitors Software Installation Instructions sw ver 2 2nd Edition June 2018
Software Installation Instructions
26 Pages
Preview
Page 1
CARESCAPE Modular Monitors Software Settings Transfer Guide Software version 2 host build v2.4
CARESCAPE Modular Monitors English 2nd edition 2106778-006 © 2018 General Electric Company. All rights reserved.
Due to continuing product innovation, specifications in this manual are subject to change without notice. For technical documentation purposes, the abbreviation GE is used for the legal entity names, GE Medical Systems Information Technologies, Inc. and GE Healthcare Finland Oy.
2
CARESCAPE Modular Monitors
2106778-006 2018-06-20
Contents Intended use of this guide ... 5 New settings in software version 2.4 ... 7 Extended setting: V Tach alarm priority ... 7 New QT/QTc alarm priority ... 7 New Ventilator disconnected alarm priority... 7 New settings: V Tach Criteria ... 8 Changed setting: Minimum HR/min... 8 New SpO2 option ... 9 New PR option ... 9 New impedance respiration option... 9 Extended setting: HR for SVT... 9 Extended WLAN security settings ... 10 Changed setting: Speed & Duplex ... 10 New ADT server connection option... 10 New E-LoFlo options ... 10 New settings in software version 2.3 ... 11 Change in non-invasive and invasive blood pressure alarm priorities ... 11 New option: Snapshot sweep speed ... 14 New option: 12 lead print speed ... 15 Extended settings: Allowed alarm priorities ... 15 New settings in software version 2.0.8 ... 17 New option: ECG lead off alarm priority ... 17 Extended settings: PVC, SVC, Impedance Resp, ST and Temp alarm priority... 17 Change in Art and ICP Invasive pressure labels ... 18 Change in other Invasive pressure labels ... 19 Extended settings: ECG Printout format ... 19 New settings in software version 2.0.7 ... 21 Extended settings: Pacemaker Detection with PDM... 21 Extended settings: 12SL Display Format ... 21
2106778-006
CARESCAPE Modular Monitors
3
Extended settings: New default value for Spirometry sensor type ... 21 Change in Menu structure (Diagnostic Tool)... 22 New global options ... 22 New options in NICU software package with CARESCAPE Respiratory module ... 22
4
CARESCAPE Modular Monitors
2106778-006
Intended use of this guide This guide provides information about new software settings that require manual configuration after a software upgrade for the following CARESCAPE Monitors: ●
CARESCAPE Monitor B450
●
CARESCAPE Monitor B650
●
CARESCAPE Monitor B850
In this document software version 2.0.6 refers to software version 2 with host build 2.0.6. In this document software version 2.0.7 refers to software version 2 with host build 2.0.7. In this document software version 2.0.8 refers to software version 2 with host build 2.0.8. In this document software version 2.3 refers to software version 2 with host build 2.3. In this document software version 2.4 refers to software version 2 with host build 2.4. The settings are divided in the tables as follows: ●
Settings listed in Chapter “New settings in software version 2.4” are available in the software version 2.4 only.
●
Settings listed in Chapter "New settings in software version 2.3" are available in the software versions 2.3 and 2.4.
●
Settings listed in Chapter “New settings in software version 2.0.8” are available in the software versions 2.0.8, 2.3, and 2.4.
●
Settings listed in Chapter “New settings in software version 2.0.7” are available in the software versions 2.0.7, 2.0.8, 2.3, and 2.4.
NOTE: If you upgrade software from version 1, refer to Software Setting Transfer Instruction, 2062973-011 for CARESCAPE B850 and B650 Monitor for new software settings that were introduced in version 2. For more detailed information regarding all settings, refer to the supplemental information manual. Use this guide together with the monitor’s service manual. The service manual has useful service information and important safety precautions.
2106778-006
CARESCAPE Modular Monitors
5
New settings in software version 2.4 Extended setting: V Tach alarm priority Settings
Selection in v2.0.8
Selection in v2.4
Default value in v2.4
Menu
Notes
V Tach alarm priority
High
High
High
Alarm Setup > Arrhytmia
Medium
Medium
Profile settings. Separate for each user profile.
Low
Low
Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > ECG.
Informational Off (NICU only)
NICU only: To enable the Off selection, the selection needs to be activated in Default Setup > Care Unit Settings > Parameters > ECG.
New QT/QTc alarm priority Setting
Selection in v2.4
Default value
Menu
Notes
QT/QTc alarm priority
Medium Low
Low
Alarm setup > Alarm Priorities
Profile setting. Separate for each user profile. Informational selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > ECG.
Informational
New Ventilator disconnected alarm priority Setting
Selection in v2.4
Default value
Menu
Notes
Ventilator disconnected alarm priority
Medium
Medium
Alarm setup > Alarm Priorities > Other Parameters
Profile setting. Separate for each user profile.
2106778-006
High
CARESCAPE Modular Monitors
7
New settings: V Tach Criteria Setting
Selection in version 2.4
Default value
Menu
Notes
Event Duration
Off
Off
Alarm Setup > Arrhytmia > Lethal Alarms > V Tach Criteria
Profile setting. Separate for each user profile.
Alarm Setup > Arrhytmia > Lethal Alarms > V Tach Criteria
Profile setting. Separate for each user profile.
Monitor Setup > Default Setup > Profile Settings > Alarm Delays
Profile setting. Separate for each user profile.
5s 10 s 15 s 20 s 25 s 30 s V Tach Criteria
AND
AND
OR
V Tach Event Duration Criteria
Allowed
Not allowed
Not allowed
This setting has been locked. To change the value, the setting must be unlocked in Default Setup > Locking settings > Parameters.
This setting has been locked. To change the value, the setting must be unlocked in Default Setup > Locking settings > Parameters.
Changed setting: Minimum HR/min Setting
Selection in version 2.0.8
Selection in version 2.4
Default value
Menu
Notes
Minimum HR/min
ICU, OR, PACU, and ED: 100 (not adjustable)
With PDM/PSM: 90-300 (adjustable) With combination monitoring:
100
Alarm Setup > Arrhytmia > Lethal Alarms > V Tach Criteria
Profile setting. Separate for each user profile.
NICU: 160 (not adjustable)
● For patients who
are 14 years or older: 100-300 (adjustable in steps of 10)
This setting has been locked. To change the value, the setting must be unlocked in Default Setup > Locking settings > Parameters.
● For patients
younger than 14 years: Not adjustable
8
CARESCAPE Modular Monitors
2106778-006
New SpO2 option Setting
Selection in v2.4
Default value
Menu
Notes
SpO2 alarm delay
0s
5s
Parameter Setup > SpO2 > Alarms > Priorities & Delays info
Profile setting. Separate for each user profile.
5s 10 s 15 s
Selection needs to be activated in Default Setup > Profile Settings > Alarm Delays.
New PR option Setting
Selection in v2.4
Default value
Menu
Notes
PR alarm delay
0s
5s
Parameter Setup > ECG > HR/PR Alarms > Priorities & Delays info
Profile setting. Separate for each user profile.
5s 10 s 15 s 20 s
Selection needs to be activated in Default Setup > Profile Settings > Alarm Delays.
New impedance respiration option Setting
Selection in v2.4
Default value
Menu
Notes
Impedance respiration alarm delay
0s
10 s
5s
15 s (NICU, Pediatric, or Infant)
Parameter Setup > Impedance Respiration > Alarms > Priorities & Delays info
Profile setting. Separate for each user profile.
10 s 15 s 20 s
Selection needs to be activated in Default Setup >Profile Settings > Alarm Delays.
25 s 30 s
Extended setting: HR for SVT Setting
Selection in v2.0.8
Selection in v2.4
Default value
Menu
Notes
HR for SVT /min
0 - 150
0 - 220
120
Alarm setup > Arrhytmia > Atrial Alarms
Profile setting. Separate for each user profile. NICU only.
2106778-006
CARESCAPE Modular Monitors
9
Extended WLAN security settings Setting
Selection in v2.0.8
Selection in v2.4
Default value in v2.4
Menu
Notes
WLAN security protocol
None
None
None
WEP
WEP
Monitor Setup > Service > Configuration > WLAN
Platform setting. For all user profiles.
WPA-PSK
WPA-PSK
WPA2-PSK
WPA2-PSK
Available for CARESCAPE Monitor B450 only.
802.1x AEP Custom
Changed setting: Speed & Duplex Setting
Default value in version 2.0.8
New default value in version 2.4
Menu
Notes
Speed & Duplex
auto
10 Mb half
Monitor Setup > Service > Configuration > Network > Wired Interfaces
Platform setting. For all user profiles. This setting is configurable in CARESCAPE Monitor B450 and CARESCAPE Monitor B850. In CARESCAPE Monitor B650 this is a non-configurable setting.
New ADT server connection option Setting
Selection in v2.4
Default value in v2.4
Menu
Notes
ADT server connection
Enabled Disabled
Disabled
Monitor setup > Service > Configuration > Network
Platform setting. For all user profiles.
New E-LoFlo options Setting
Selection in v2.4
Default value in v2.4
Menu
Notes
CO2 average
10 s
10 s
Parameter setup > Gases > Setup
Profile setting. Separate for each user profile.
20 s
Selection available only when E-LoFLo module connected. N2O noise level
0-30 30-80
0-30
Parameter setup > Gases > Setup
Profile setting. Separate for each user profile. Selection available only when E-LoFLo module connected.
10
CARESCAPE Modular Monitors
2106778-006
New settings in software version 2.3 Change in non-invasive and invasive blood pressure alarm priorities Settings
Default value in v2.0.8
Default value in v2.3
Menu
Notes
NIBP alarm priority
Escalating
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
Art alarm priority
Escalating
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
Fem alarm priority
Escalating
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Profile settings. Separate for each user profile.
FemV alarm priority
PA alarm priority
2106778-006
Escalating
Escalating
Medium
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Alarm Setup > Alarm Priorities > Other Parameters
CARESCAPE Modular Monitors
Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group. Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group. Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
11
Settings
Default value in v2.0.8
Default value in v2.3
Menu
Notes
CVP alarm priority
Escalating
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Profile settings. Separate for each user profile.
LAP alarm priority
RAP alarm priority
Escalating
Escalating
Medium
Medium
RVP alarm priority
12
Alarm Setup > Alarm Priorities > Other Parameters
Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group. Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
Escalating
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
Escalating
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Profile settings. Separate for each user profile.
ICP alarm priority
CPP alarm priority
Alarm Setup > Alarm Priorities > Other Parameters
Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
Escalating
Medium
Alarm Setup > Alarm Priorities > Other Parameters
CARESCAPE Modular Monitors
Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group. Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
2106778-006
Settings
Default value in v2.0.8
Default value in v2.3
Menu
Notes
UAC alarm priority
Escalating
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Profile settings. Separate for each user profile.
UVC alarm priority
P1 alarm priority
Escalating
Escalating
Medium
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Alarm Setup > Alarm Priorities > Other Parameters
Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group. Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group. Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
P2 alarm priority
Escalating
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
P3 alarm priority
Escalating
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Profile settings. Separate for each user profile.
P4 alarm priority
2106778-006
Escalating
Medium
Alarm Setup > Alarm Priorities > Other Parameters
CARESCAPE Modular Monitors
Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group. Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
13
Settings
Default value in v2.0.8
Default value in v2.3
Menu
Notes
P5 alarm priority
Escalating
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Profile settings. Separate for each user profile.
P6 alarm priority
P7 alarm priority
P8 alarm priority
Escalating
Escalating
Escalating
Medium
Medium
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Alarm Setup > Alarm Priorities > Other Parameters
Alarm Setup > Alarm Priorities > Other Parameters
Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group. Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group. Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group. Profile settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
New option: Snapshot sweep speed Settings
Available setting
Selection in v2.3
Default value in v2.3
Menu
Notes
Snapshot Setup
Sweep Speed
12.5 mm/s
12.5 mm/s
Trends > View > Snapshot > Snapshot Setup
Profile settings. Separate for each user profile.
14
25 mm/s 50 mm/s
CARESCAPE Modular Monitors
2106778-006
New option: 12 lead print speed Setting
Selection in v2.0.8
Selection in v2.3
Default value in v2.3
Menu
Notes
12 Lead Print Speed
50 mm/s
25 mm/s
25 mm/s
Not possible to change.
50 mm/s
Default Setup > Care Unit Settings > Parameters > 12 lead ECG
Care unit setting. Separate for each user profile.
Extended settings: Allowed alarm priorities Setting
Selection in version 2.08
Selection in version 2.3
Default value in version 2.3
Menu
Notes
NIBP alarm priority
Escalating
Escalating
High
High
High
Medium
Alarm Setup > Alarm Priorities > Other Parameters
Care unit settings. Separate for each user profile.
Medium
Medium
Low
Low or
Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
High Medium IP alarm priority
Escalating
Escalating
High
High
High Medium
Medium
Medium Low
Alarm Setup > Alarm Priorities > Invasive Pressures
Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
Low or High Medium
CO2 alarm priority
Escalating
Escalating
High
High
Medium
Medium
High Medium
Alarm Setup > Alarm Priorities > Other Parameters
Care unit settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > parameter group.
Alarm Setup > Alarm Priorities > Other Parameters
Care unit settings. Separate for each user profile. Selection needs to be activated in Default Setup > Care Unit Settings > Parameters
or High Medium FiAA alarm priority
Escalating
Escalating
High
High
High
Medium
Medium
Medium or High
2106778-006
Care unit settings. Separate for each user profile.
CARESCAPE Modular Monitors
15
Setting
Selection in version 2.08
Selection in version 2.3
Default value in version 2.3
Menu
> Parameter Alarms > Allowed Priorities > parameter group.
Medium
16
Notes
CARESCAPE Modular Monitors
2106778-006
New settings in software version 2.0.8 New option: ECG lead off alarm priority Settings
Selection in version 2.0.6 and version 2.0.7
Selection in version 2.0.8
Default value
Menu
Notes
ECG lead off alarm priority
Selection not available
Low
Low
Alarm Setup > Alarm Priorities > ECG
Profile setting. Separate for each user profile.
Informational
The new selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > ECG
Extended settings: PVC, SVC, Impedance Resp, ST and Temp alarm priority Settings
Selection in version 2.0.6 and version 2.0.7
Selection in version 2.0.8
Default value
Menu
Notes
PVC alarm priority
Escalating
Escalating
Escalating
High
High
Alarm Setup > Alarm Priorities > ECG
Profile setting. Separate for each user profile.
Medium Low
Medium Low
Low
Alarm Setup > Alarm Priorities > ECG
Profile setting. Separate for each user profile.
Escalating
Alarm Setup > Alarm Priorities > Other Parameters
Profile setting. Separate for each user profile.
Informational SVC alarm priority
Escalating
Escalating
High Medium
High Medium
Low
Low Informational
RR (Imped.) high/low alarm priority
Escalating
Escalating
High
High
Medium
Medium
Low
Low Informational
2106778-006
CARESCAPE Modular Monitors
The new selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > Other Parameters
17
Settings
Selection in version 2.0.6 and version 2.0.7
Selection in version 2.0.8
Default value
Menu
Notes
ST alarm priority
High
High
Medium
Medium
Medium
Alarm Setup > Alarm Priorities > ECG
Profile setting. Separate for each user profile.
Low
Low
The new selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > ECG Parameters
Informational
Temp alarm priority
Selection not available
Escalating
Escalating
Informational
Alarm Setup > Alarm Priorities > Other Parameters
Profile setting. Separate for each user profile. The new selection needs to be activated in Default Setup > Care Unit Settings > Parameters > Parameter Alarms > Allowed Priorities > Other Parameters
Change in Art and ICP Invasive pressure labels Settings
Selection in version 2.0.6 and version 2.0.7
Selection in version 2.0.8
Default value
Menu
Notes
Art
Art can be assigned to one InvBP channel at a time
Art can be assigned to multiple InvBP channel simultaneously
Multiple
Parameter Setup > Invasive Pressure > Px > Setup (Px = P1, P2, P3...)
Profile setting. Separate for each user profile.
ICP can be assigned to one InvBP channel at a time
ICP can be assigned to multiple InvBP channel simultaneously
Multiple
ICP
18
Parameter Setup > Invasive Pressure > Px > Setup (Px = P1, P2, P3...)
CARESCAPE Modular Monitors
If you assign additional Art channels, the settings are copied from the original Art channel. Profile setting. Separate for each user profile. If you assign additional ICP channels, the settings are copied from the original ICP channel.
2106778-006
Change in other Invasive pressure labels Settings
Selection in version 2.0.6 and version 2.0.7
Selection in version 2.0.8
Default value
Menu
Notes
ABP
available
not available
N/A
N/A
You need to check the settings manually after upgrading. *)
UAC
available only in NICU SW package
available in OR, ICU,NICU, ED and PACU SW packages
N/A
Parameter Setup > Invasive Pressure > Px > Setup
Profile setting. Separate for each user profile.
available only in NICU SW package
available in OR, ICU,NICU, ED and PACU SW packages
N/A
UVC
(Px = P1, P2, P3...)
Settings transferred in NICU SW package only.
Parameter Setup > Invasive Pressure > Px > Setup
Profile setting. Separate for each user profile.
(Px = P1, P2, P3...)
Settings transferred in NICU SW package only.
*) NOTE: Some of the ABP settings are transferred from the original Art and ABP settings. To ensure proper settings configuration, GE recommends that you check the settings manually after upgrading.
Extended settings: ECG Printout format Settings
Selection in version 2.0.6 and version 2.0.7
Selection in version 2.0.8
Default value
Menu
Notes
ECG Printout format
4x2.5 – 1 Rhythm
4x2.5 – 1 Rhythm
4x2.5 – 1 Rhythm
Profile setting. Separate for each user profile.
4x2.5 – 3 Rhythm
4x2.5 – 3 Rhythm
Default Setup > Care Unit Settings > Parameters > ECG
12 Rhythms 25mm/s 12 Rhythms 50mm/s Cabrera
12 Rhythms 25mm/s 12 Rhythms 50mm/s 2x6 Rhythms 50mm/s Cabrera 50mm/s
2106778-006
CARESCAPE Modular Monitors
19
New settings in software version 2.0.7 Extended settings: Pacemaker Detection with PDM Settings
Selection in version 2.0.6 with PDM
Selection in version 2.0.7 with PDM 2.4
Default value
Menu
Notes
Pacemaker detection
On
Sensitive
Off
Normal (New)
Sensitive (with PDM 2.4)
Parameter Setup > ECG > Advanced
Profile setting. Separate for each user profile.
Off
On (with PDM 2.3 or older) Sensitive (with E-PSM(P)) Pace2 (with TRAM)
New selection available only with PDM 2.4 and software version 2.0.7.
Extended settings: 12SL Display Format Settings
Selection in version 2.0.6
Selection in version 2.0.7
Default value
Menu
Notes
12SL Display Format
4x2.5 - 1 Rhythm 4x2.5 - 3 Rhythm 12 Rhythms Cabrera
4x2.5 - 1 Rhythm 4x2.5 - 3 Rhythm 12 Rhythms 2x6 Rhythms
4x2.5 - 3 Rhythm
Parameter Setup > ECG > 12 Lead Analysis > Settings
Profile setting. Separate for each user profile.
Cabrera
Extended settings: New default value for Spirometry sensor type Settings
Selection in version 2.0.6
New default value in version 2.0.7
Menu
Notes
Spirometry sensor type
Adult Pediatric
All user profiles in NICU software package: Pediatric
Parameter Setup > Spirometry/Gas Exchange
Profile setting. Separate for each user profile.
User profiles Infant and Pediatric in ED, ICU, OR, and PACU software package: Pediatric User profiles other than Infant and Pediatric in ED, ICU, OR, and PACU software package: Adult
2106778-006
CARESCAPE Modular Monitors
21
Change in Menu structure (Diagnostic Tool) Settings
Selection in version 2.0.6
Combined menu in version 2.0.7
Default value in version 2.0.7
Menu
Notes
ACITIPI
On/Off
ACITIPI
ACITIPI
On/Off
ACS
Parameter Setup > ECG > 12 Lead Analysis > Settings
Moved to other menu destination.
ACS
Off
Separate for each user profile.
New global options Settings
New available setting in version 2.0.7
Default value
Menu
Notes
Citrix Server version
Support for Citrix server versions 4.0/4.5 and 5.0.
Default server version: 4.0/4.5
Webmin > Configuration > Citrix
Platform setting. For all user profiles.
Citrix Server address 2–4
Support for 4 Citrix servers with 4 different IP addresses.
Default IP address: Empty
Webmin > Configuration > Citrix
Platform setting. For all user profiles.
Alarm Activation for PR
New available selection: Immediate or Delayed
Immediate
Default Setup > Care Unit Settings > Parameters > Allowed priorities > Other Parameters
Care Unit setting. For all user profiles.
Print Interpretations
New available selection: Enabled or Disabled
Enabled
Parameter Setup > ECG > 12 Lead Analysis > Settings
Profile setting. Separate for each user profile.
Low-Pass Filter
New available selections: 20Hz or 40Hz or 100Hz or 150Hz
150Hz
Parameter Setup > ECG > 12 Lead Analysis > Filter
Same default for all user profiles. Not possible to change and save as profile of unit settings.
New options in NICU software package with CARESCAPE Respiratory module NOTE: Not available in the US and its territories. NOTE: Not available in China.
22
CARESCAPE Modular Monitors