COSMED srl
Hemodynamic and Metabolic Monitors
Innocor Breath-by-Breath Method Guide Issue A Rev 2 July 2013
Guide
12 Pages
Preview
Page 1
Innovision
Innocor® Breath-by-Breath Method
TABLE OF CONTENTS 1 1.1 1.2 1.3 1.4 1.4.1 1.4.2 1.4.3 1.4.4 1.4.5 1.4.6 1.4.7 1.4.8 1.4.9 1.4.10 1.4.11 1.4.12 1.4.13 1.4.14 1.4.15 1.4.16 1.4.17 1.4.18 1.4.19 1.4.20
July 2013
BREATH-BY-BREATH METHOD ... 1 SCOPE ... 1 INTRODUCTION ... 1 BREATH-BY-BREATH PARAMETERS... 2 BREATH-BY-BREATH EQUATIONS ... 3 Oxygen uptake (Vo2)... 4 Carbon dioxide excretion (Vco2) ... 4 Expiratory minute ventilation (VE) ... 5 Oxygen uptake per kg (Vo2/kg) ... 5 Respiratory gas exchange ratio (R) ... 5 Alveolar ventilation (VA) ... 5 Anatomical dead space (VD) ... 6 Tidal volume (VT) ... 7 Respiratory rate (Resp.Freq.) ... 7 End-tidal concentration of oxygen (Fo2-et) ... 7 End-tidal concentration of carbon dioxide (Fco2-et) ... 7 Expiratory quotient for oxygen (VE/Vo2) ... 7 Expiratory quotient for carbon dioxide (VE/Vco2) ... 8 Anaerobic threshold (AT) ... 8 Respiratory compensation (RC) ... 8 Rest values ... 8 AT values ... 8 Max values ... 9 Gas delay ... 9 Conversion between ATP, STPD and BTPS ... 10
COR-MAN-0000-005-IN /EN, A/2
i
Innovision
Innocor® Breath-by-Breath Method
1
BREATH-BY-BREATH METHOD
1.1
SCOPE
The purpose of this document is to give an introduction to the Breath-by-Breath Method used in Innocor. This section applies to users of Innocor with only limited experience in gas exchange measurements. For more detailed information about the Breath-by-Breath method, please contact Innovision A/S or simply look in medical article databases.
1.2
INTRODUCTION
The breath-by-breath method is the fastest responding of the non-invasive gas exchange methods for determination of the metabolic rate. Gas composition and respiratory flow are measured continuously at a high frequency. The principle in the determination of the oxygen consumption and the carbon dioxide excretion is to estimate the difference between airway influx and efflux of oxygen and carbon dioxide during inspiration and expiration. This is done by integrating the product of oxygen or carbon dioxide concentration and flow in the respiratory gas over an interval, which covers a complete respiratory cycle. The results can be corrected partly for changes in oxygen and carbon dioxide contents of the lungs by using equations based on the change in functional residual capacity (FRC) from one breath to the next (Beaver et al., 1981) and by assuming constant F ETo2 and FETco2 from breath to breath. The change in FRC is determined from the nitrogen balance equation, which requires that the nitrogen flux across the alveolar membrane is practically zero. The respiratory (expiratory) minute ventilation is estimated by simply integrating the flow over one whole expiration and dividing by the length of the breath. This is also the case for the alveolar ventilation except that the product of anatomical dead space and respiratory rate is subtracted. The anatomical dead space is determined from the capnogram (CO2 curve) and the flow signal. The respiratory rate, the tidal volume and the time of inspiration are determined from the flow signal alone. The end-tidal concentrations are determined as the minimum/maximum concentrations of oxygen/carbon dioxide at the end of expiration. The respiratory and expiratory quotients are found as the ratio between carbon dioxide excretion and oxygen consumption, and as the ratio between respiratory minute ventilation and oxygen consumption or carbon dioxide excretion, respectively. This document will enable the reader to understand the cardiopulmonary parameters measured by Innocor and the way they are determined.
July 2013
COR-MAN-0000-005-IN /EN, A/2
1
Innocor® Breath-by-Breath Method
Innovision
1.3
BREATH-BY-BREATH PARAMETERS
The Breath-by-Breath parameters measured by the Innocor are: Primary parameters: Symbol Name Vo2 Oxygen uptake Vco2
Carbon dioxide excretion
VE
Expiratory minute ventilation
Secondary / derived parameters: Symbol Name Vo2/kg Oxygen uptake per kg R VA
Respiratory gas exchange ratio Alveolar ventilation
VD
Anatomical dead space
VT Resp.Freq. Fo2-et
Tidal volume Respiratory rate End-tidal concentration of oxygen End-tidal concentration of carbon dioxide Expiratory quotient / ventilatory equivalent for oxygen Expiratory quotient / ventilatory equivalent for carbon dioxide
Fco2-et VE/Vo2
VE/Vco2
Description & Unit The amount of oxygen extracted from the inspired gas in a given time, [l/min @ STPD]. The amount of carbon dioxide exhaled from the body into the atmosphere in a given time, [l/min @ STPD]. The volume of air exhaled from the body in one minute, [l/min @ BTPS].
Description & Unit The amount of oxygen extracted from the inspired gas in a given time per kg body weight, [ml/min/kg @ STPD] The ratio of the carbon dioxide excretion to the oxygen uptake. The volume of air exhaled from the alveoli per minute, [l/min @ BTPS]. The volume of the upper airways, trachea and bronchi. (Fowler dead space). [l @ BTPS]. Volume of expired air during a breath, [l @ BTPS]. Breathing frequency [/min]. The oxygen concentration in the expired gas at the end of the exhalation, [%]. The carbon dioxide concentration in the expired gas at the end of the exhalation, [%]. The ratio of ventilation to oxygen uptake.
The ratio of ventilation to carbon dioxide excretion.
The following parameters can be calculated after an incremental exercise test: Symbol Name Description & Unit AT Anaerobic threshold The exercise Vo2 above which anaerobic production supplements aerobic production. Exercise above AT is reflected by an increase in lactate concentration. The AT is measured by the V-slope* method on the plot of Vco2 against Vo2. RC Respiratory compensation The exercise Vo2 above which the ventilation starts to increase more than the Vco2 in order to compensate for the lactic acidosis. The RC is measured by the V-slope* method on the plot of VE against Vco2. @Rest Rest values The values of the BBB parameters at rest – an average of the last minute before start of exercise. @AT Values at AT point The values of the BBB parameters at the AT point – an average of the 30 seconds around the AT point. @Max Values at max exercise The values of the BBB parameters at the max exercise level – found as 30 seconds average around the maximum of Vo2. * Beaver WL, Wassermann K, Whipp BJ (1986) “A new method for detecting anaerobic threshold by gas exchange” J Appl Physiol 60:2020-2027 All the listed parameters will be explained in detail in the following. July 2013
COR-MAN-0000-005-IN /EN, A/2
2
Innovision
1.4
Innocor® Breath-by-Breath Method
BREATH-BY-BREATH EQUATIONS
In the following equations for the calculations of the Breath-by-Breath parameters the gas signals and flow signal are corrected for time delay. Refer to the figure below:
Figure 1.4-1 Illustration of a single breath O2 and CO2 concentrations and respiratory flow with definition of variables.
t1 = t2 = t3 = V= FMX = FAX = C1 = C2 =
start of an inspiration start of the following expiration start of the following inspiration flow at mouthpiece (here defined to be positive for inspiration) fractional concentration of gas X (O2, CO2 or N2) at mouthpiece fractional concentration of gas X (O2, CO2 or N2) in alveolar space, determined as the average value (FMX(t1)+FMX(t3))/2 conversion factor from ATP to STPD, see section 1.4.20 conversion factor from BTPS to STPD, see section 1.4.20
The gas analyser in the Innocor can not measure nitrogen directly. The nitrogen (incl. argon) concentration is determined as the rest: FM,N2 = 1 - FM,O2 - FM,CO2 - FM,N2O - FM,SF6
July 2013
COR-MAN-0000-005-IN /EN, A/2
3
Innocor® Breath-by-Breath Method
Innovision
1.4.1 Oxygen uptake (Vo2) The oxygen uptake is the amount of oxygen extracted from the inspired gas in a given time, [l/min @ STPD]. The oxygen uptake is the difference between the oxygen volume inspired and expired. When corrected for the change in lung volume from breath to breath the equation is:
t2 1 • V ATP ( ) • C1 • F MO 2 ( ) • d + ( t 3 - t1 ) t1
Vo2 =
t3
V
BTPS
( ) • C 2 • F MO2 ( ) • d -
t2
t2
F AO2 • [ V ATP ( ) • C 1 • F MN 2 ( ) • d + F AN 2 t1 ( ) • • ( ) • d ] V C F BTPS 2 MN 2 t2 t3
The oxygen uptake during rest is approximately 3.5 ml/min per kg or 250 ml/min for a 70 kg person. The oxygen uptake increases with exercise level, and during exercise on a bicycle ergometer Vo2 can be estimated as: Vo2 = 5.8∙weight+151+10.1*load [ml/min] Example: Weight = 70 kg Load = 100 watt Vo2 = 5.8∙70+151+10.1*100 ml/min = 1.567 l/min 1.4.2 Carbon dioxide excretion (Vco2) The carbon dioxide excretion is the amount of carbon dioxide exhaled from the body into the atmosphere in a given time, [l/min @ STPD]. The carbon dioxide excretion is the difference between the carbon dioxide volume expired and inspired. When corrected for the change in lung volume from breath to breath the equation is:
Vco2 =
t2 -1 • V ATP ( ) • C 1 • F MCO 2 ( ) • d + ( t 3 - t1 ) t1 t3
V
BTPS
( ) • C 2 • F MCO2 ( ) • d -
t2 t2
F ACO2 • [ V ATP ( ) • C 1 • F MN 2 ( ) • d + F AN 2 t1 t3
V
t2
July 2013
BTPS
( ) • C 2 • F MN 2 ( ) • d ]
COR-MAN-0000-005-IN /EN, A/2
4
Innovision
Innocor® Breath-by-Breath Method
1.4.3 Expiratory minute ventilation (VE) The expiratory minute ventilation is the volume of air exhaled from the body in one minute, [l/min @ BTPS]. The expiratory minute ventilation is the integrated flow during expiration divided by the breath length.
VE= -
t3 1 • V BTPS ( ) • d t 3 - t1 t 2
1.4.4 Oxygen uptake per kg (Vo2/kg) Oxygen uptake per kg is the amount of oxygen extracted from the inspired gas in a given time per kg body weight, [ml/min/kg @ STPD]. Vo2/kg = Vo2 / weight ∙ 1000 The Vo2/kg at rest is approximately 3.5 ml/min/kg. The maximum Vo2/kg during maximum exercise is a normalised parameter, which depends on the fitness of the subject. 1.4.5 Respiratory gas exchange ratio (R) Respiratory gas exchange ratio is the ratio of the carbon dioxide excretion to the oxygen uptake. R = Vco2 / Vo2 At rest and at steady state the R is in the range 0.7-1.0. During an exercise test the R increases – at the end above 1. During recovery the R increases further (<2). 1.4.6 Alveolar ventilation (VA) The alveolar ventilation is the volume of air exhaled from the alveoli per minute, [l/min @ BTPS]. The alveolar ventilation is the integrated flow during expiration subtracted the subject and instrument dead space divided by the breath length. V A=
t3 1 • −V BTPS ( ) • d −VD − Vtotal instrument dead space t 3 - t1 t 2
An alternative calculation of VA is: V A = VE − Resp.Freq • VD + Vtotal instrument dead space
Example: t3-t1 = 0.05 min (Resp.Freq = 20 /min) ∫ V∙dτ = -1 litre VD = 0.150 litre Vtotal instrument dead space = 0.136 litre VE = 1/0.05∙1 = 20.00 l/min VA = 1/0.05∙(1-0.150-0.136) = 14.28 l/min
July 2013
COR-MAN-0000-005-IN /EN, A/2
5
Innovision
Innocor® Breath-by-Breath Method
1.4.7 Anatomical dead space (VD) The anatomical dead space is the volume of the upper airways, trachea and bronchi. (Fowler dead space). [l @ BTPS]. The expirogram is normally divided into 3 phases: Phase I, where no CO2 is present, phase II, where the CO2 concentration rises rapidly during dead space wash-out, and phase III, identified by a slower and nearly linear increase in CO2 concentration, which represents delayed alveolar concentration. To estimate the anatomical dead space, a regression line is determined for phase III, and the dead space is defined as the expired volume where two areas are equal. One area is bounded by the zero line, the vertical line at the dead space and the CO2 curve, and the second area by the vertical line at the dead space, the alveolar regression line and the CO2 curve. The volume found in this way is subtracted the instrument dead space (from the inlet to the mouth): VD = V(To) – Vinstrument dead space, flowmeter
Figure 1.4.7-1 Illustration of an expirogram for the determination of the anatomical dead space.
The different dead spaces are defined in figure 1.4.7-2, and consist of VD Vinstrument dead space, flowmeter Vinstrument dead space, RVU Vtotal instrument dead space
July 2013
= Anatomical dead space. = gdFlowdeadspace, volume from mouth to inlet. = gdValvedeadspace, volume from inlet to ambient air port. = Vinstrument dead space, flowmeter + Vinstrument dead space, RVU
COR-MAN-0000-005-IN /EN, A/2
6
Innocor® Breath-by-Breath Method
Innovision
Figure 1.4.7-2 Illustration of the anatomical dead space and instrument dead spaces. A standard configured Innocor has the following dead spaces: Vinstrument dead space, flowmeter Vinstrument dead space, RVU Vtotal instrument dead space
= 116 ml (172 ml with flexible tube) = 20 ml = 136 ml (192 ml with flexible tube)
1.4.8 Tidal volume (VT) Tidal volume is the volume of air exhaled during a breath, [l @ BTPS]. t3
VT = − V BTPS ( ) • d t2
An alternative calculation of VT is: VT =
VE R esp.Freq
1.4.9 Respiratory rate (Resp.Freq.) The respiratory rate is the breathing frequency [/min].
Resp.Freq =
1 t 3 - t1
1.4.10 End-tidal concentration of oxygen (Fo2-et) The end-tidal concentration of oxygen is the oxygen concentration in the expired gas at the end of the exhalation, [%]. The end-tidal concentration of oxygen is found as a 250 ms average at the same time as the end-tidal concentration of carbon dioxide, see below. 1.4.11 End-tidal concentration of carbon dioxide (Fco2-et) The end-tidal concentration of carbon dioxide is the carbon dioxide concentration in the expired gas at the end of the exhalation, [%]. The end-tidal concentration of carbon dioxide is found as the maximum value of a 250 ms moving averaging. 1.4.12 Expiratory quotient for oxygen (VE/Vo2) The expiratory quotient / ventilatory equivalent for oxygen is the ratio of ventilation to oxygen uptake. The expiratory quotient is corrected for instrument dead space. VE/Vo2 = (VE – Resp.Freq∙Vtotal instrument dead space) / Vo2 Normal values of VE/Vo2 at AT is in the range 20 – 33.
July 2013
COR-MAN-0000-005-IN /EN, A/2
7
Innovision
Innocor® Breath-by-Breath Method
1.4.13 Expiratory quotient for carbon dioxide (VE/Vco2) The expiratory quotient / ventilatory equivalent for carbon dioxide is the ratio of ventilation to carbon dioxide excretion. The expiratory quotient is corrected for instrument dead space. VE/Vco2 = (VE – Resp.Freq∙Vtotal instrument dead space) / Vco2 Normal values of VE/Vco2 at AT is in the range 23 – 36.
1.4.14 Anaerobic threshold (AT) The anaerobic threshold is the exercise Vo2 above which anaerobic production supplements aerobic production. Exercise above AT is reflected by an increase in lactate concentration. The AT is measured by the V-slope* method on the plot of Vco2 against Vo2, see figure below. * Beaver WL, Wassermann K, Whipp BJ (1986) “A new method for detecting anaerobic threshold by gas exchange” J Appl Physiol 60:2020-2027
Figure 1.4.14-1 Illustration of the determination of the anaerobic threshold measured by the Vslope method. Normal values of AT is 50 to 65% of maximum Vo2. 1.4.15 Respiratory compensation (RC) The respiratory compensation is the exercise Vco2 above which the ventilation starts to increase more than the Vco2 in order to compensate for the lactic acidosis. The RC is measured by the Vslope method on the plot of VE against Vco2. 1.4.16 Rest values The rest values are the breath-by-breath parameters at rest – an average of the last minute before start of exercise (start exercise is defined as the time where a change from zero is found in the load, speed or slope. Start of unloaded pedalling is defined as start of exercise). 1.4.17 AT values The AT values are the breath-by-breath parameters at the anaerobic threshold – an average of 30
July 2013
COR-MAN-0000-005-IN /EN, A/2
8
Innovision
Innocor® Breath-by-Breath Method
seconds around the AT point. 1.4.18 Max values The max values are the breath-by-breath parameters at the max exercise level – a 30 seconds average. The time is found as the 30 seconds time period where the average of the Vo2 is highest. 1.4.19 Gas delay In order to calculate breath-by-breath parameters the flow and gas must be aligned time wise. Flow is measured instantaneously whereas the gas signals are delayed due to gas transport from inlet tip to measurement cell. Due to the delay in the gas signals, the gas signals must be shifted back in time. The accuracy of this delay is important – an error of 25 ms at rest gives an error of approx. 5% in the Vo2 and Vco2 calculations! The calculation of the delay is based on the respiratory flow and the fractional concentration of CO 2 / O2 at the change from expiration to inspiration, where the fastest change in CO2 concentration occurs. The time from a change in direction of the flow (to inspiration) to the corresponding change in CO2 / O2 concentration consists of two parts: The first part, which is flow dependent, is caused by the instrumental dead space from the breathing port to where the sample inlet is positioned (see figure1.4.7-2). The other part is the constant gas delay caused by the sample gas transit time and response time of the gas analyser. The delay between flow and gas signals is calculated from the time between the change in flow and the change in the CO2 / O2 signal, when this time is corrected for the time corresponding to the flushing of the instrumental dead space (Vinstrument dead space, RVU = 20 ml @ standard Innocor).
Figure 1.4.19-1 Illustration of the determination of the gas delay. Td = time where dead space volume is inspired. Ts = time where CO2 is dropped ½ based on an equal area determination Ts – Td = gas delay Flow is positive during inspiration.
In the Innocor the gas delay can be determined prior to a test, where the operator shall make 10 special breaths: Slow expirations followed by a fast inspiration in order to minimise the time to washout the dead space. The gas delay is also determined automatically during an exercise test, when the exercise level exceeds approx. 40-50% of max exercise level. The delay determination is based on the last 10 gas delay calculations using averaging and removal of any extreme values. Normal values for the gas delay on the Innocor is 1400 – 1700 ms.
July 2013
COR-MAN-0000-005-IN /EN, A/2
9
Innocor® Breath-by-Breath Method
Innovision
1.4.20 Conversion between ATP, STPD and BTPS The volume of a number of moles (n) of gas molecules depends on the thermodynamic temperature (T) and the ambient pressure (P). The following relationship holds for dry gas: V = n·R·T/P where R = gas constant, and T is expressed in Kelvin (T(K) = 273.2 + t(ºC)). Air and expired gas are made up of gas molecules and water vapour. In a gas mixture saturated with water vapour and in contact with water (such as occurs in the lung) the number of water molecules in the gas phase varies with temperature and pressure. As the number of molecules is not constant, the above gas law should be applied to dry gas. This also holds outside the lung when gas saturated with water vapour is compressed or cools down. BTPS:
In respiratory physiology lung volumes and flows are standardised to barometric pressure at sea level, body temperature, saturated with water vapour: body temperature and pressure, saturated. Measured at ambient temperature, pressure, saturated with water vapour (e.g. expired gas, which has cooled down): ambient temperature and pressure, saturated. Like ATPS, but not saturated with water vapour (e.g. room air). Like ATPS, but dry (e.g. from a gas bottle). Oxygen consumption and carbon dioxide excretion are standardised to standard temperature (0 ºC), barometric pressure at sea level (101.3 kPa / 760 mmHg) and dry gas: standard temperature and pressure, dry.
ATPS: ATP: ATPD: STPD:
Correction from ATP to STPD. Multiply the ATP-value by:
273 C1 = 273 + t a
PB −
RH PH 2O ( t a ) 100 760
Correction from BTPS to STPD. Multiply the BTPS-value by: C2 =
P − 47 273 B 273 + 37 760
where ta PB RH PH2O(ta) Temperature [°C] 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14
July 2013
= = = =
ambient temperature in °C barometric pressure in mmHg relative humidity in % saturated water vapour pressure in mmHg at temperature ta, see table below
Water vapour pressure [mmHg] 4.7 5.2 5.6 6.1 6.5 7.0 7.4 7.9 8.3 8.8 9.2 9.8 10.5 11.2 12.0
Temperature [°C] 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29
Water vapour pressure [mmHg] 12.8 13.6 14.5 15.5 16.5 17.5 18.7 19.8 21.1 22.4 23.8 25.2 26.7 28.3 30.0
Temperature [°C] 30 31 32 33 34 35 36 37 38 39 40
COR-MAN-0000-005-IN /EN, A/2
Water vapour pressure [mmHg] 31.8 33.7 35.7 37.7 39.9 42.2 44.6 47.1 49.7 52.4 55.3
10