Innovision CathLab Step-by-step Guide May 2018

Guide

1 Page

Step-by-Step procedure CathLab  18/05/2018/kp  Gas exchange CathLab •  Power Innocor on.  •  Press Measurement on Screen.  •  Select appropriate configuration.  •  Press Prepare.  •  Press change on Ambient data and enter actual data.  •  Press calibrate on Flowmeter and follow calibration guide using a syringe.  •  Press calibrate on Flow-gas delay and follow calibration guide using a subject who performs fast inspirations/slow expirations.  •  Press calibrate on Oxygen, expose tip of inlet to air and wait for “Current signal” to become stable. Press Start. Wait approximately 30 seconds. Press Save and Exit.  •  Press OK.  •  Press OK.  •  Press New Pt. and enter patient data.  •  Move cursor to the patient you want to study and press Select.  •  Press Test.  •  Connect patient to breathing assembly.  •  Press OK when patient ready.  •  You are now measuring. If scaling is not optimal then press Graph Setup and optimize axis settings.  •  A measurement can be stopped manually by pressing Stop test.  •  Go to Results, Details or Graph to evaluate data.  •  Go to Print Prev. to print data.  Step-by-Step procedures - CathLab.doc  page 1 of 1
File Type: PDF
File Size: 66.2 KB
File Name: Step-by-Step procedures - CathLab.pdf