Step-by-Step procedure CathLab
18/05/2018/kp
Gas exchange CathLab •
Power Innocor on.
•
Press Measurement on Screen.
•
Select appropriate configuration.
•
Press Prepare.
•
Press change on Ambient data and enter actual data.
•
Press calibrate on Flowmeter and follow calibration guide using a syringe.
•
Press calibrate on Flow-gas delay and follow calibration guide using a subject who performs fast inspirations/slow expirations.
•
Press calibrate on Oxygen, expose tip of inlet to air and wait for “Current signal” to become stable. Press Start. Wait approximately 30 seconds. Press Save and Exit.
•
Press OK.
•
Press OK.
•
Press New Pt. and enter patient data.
•
Move cursor to the patient you want to study and press Select.
•
Press Test.
•
Connect patient to breathing assembly.
•
Press OK when patient ready.
•
You are now measuring. If scaling is not optimal then press Graph Setup and optimize axis settings.
•
A measurement can be stopped manually by pressing Stop test.
•
Go to Results, Details or Graph to evaluate data.
•
Go to Print Prev. to print data.