Service Manual
916 Pages
Preview
Page 1
P R E L I M I N A R Y
(OCTOBER 7, 2019) COVER.FM
GE Healthcare
LOGIQ S8/LOGIQ E8/ LOGIQ S7 XDclear2.0 Service Manual
Document Number: 5394227 Revision: 13
This manual is a reference for the LOGIQ S8/LOGIQ E8/LOGIQ S7 XDclear2.0 and LOGIQ S8 Vet ultrasound systems (Hereafter listed as LOGIQ S8). All information provided in this manual is relevant for all four systems unless otherwise specified.
GE Healthcare Direction 5394227, 13
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
Important Precautions TRANSLATION POLICY
WARNING (EN)
AVERTISSEMENT (FR)
WARNUNG (DE)
THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY. • IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE TRANSLATION SERVICES. • DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD. • FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL OR OTHER HAZARDS. CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS. • SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE TRADUIRE. • NE PAS TENTER D’INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE MANUEL SERVICE N’A PAS ÉTÉ CONSULTÉ ET COMPRIS. • LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE TECHNICIEN, L’OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES. DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER SPRACHE. • FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT, IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG ZU SORGEN. • VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN WURDE. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
-
i
GE Healthcare Direction 5394227, 13
AVISO (ES)
ATENÇÃO (PT-Br)
AVISO (PT-pt)
AVVERTENZA (IT)
ii
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN. • NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO. • LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSAS ELÉCTRICAS, MECÁNICAS O DE OTRA NATURALEZA. ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL EM INGLÊS. • SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO. • NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA. • O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELÉTRICOS, MECÃNICOS OU OUTROS. ESTE MANUAL DE ASSISTÊNCIA ESTÁ DISPONÍVEL APENAS EM INGLÊS. • SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO. • NÃO TENTE EFECTUAR REPARAÇÕES NO EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO PREVIAMENTE ESTE MANUAL. • A INOBSERVÂNCIA DESTE AVISO PODE RESULTAR EM FERIMENTOS NO TÉCNICO DE ASSISTÊNCIA, OPERADOR OU PACIENTE EM CONSEQUÊNCIA DE CHOQUE ELÉCTRICO, PERIGOS DE ORIGEM MECÂNICA, BEM COMO DE OUTROS TIPOS. IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN INGLESE. • SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE È TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE. • SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL CONTENUTO. • NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI. -
GE Healthcare Direction 5394227, 13
VIÐVÖRUN (IS)
VÝSTRAHA (CS)
ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU. • REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG SKILIÐ ÞESSA ÞJÓNUSTUHANDBÓK. • EF EKKI ER FARIÐ AÐ ÞESSARI VIÐVÖRUN GETUR ÞAÐ VALDIÐ MEIÐSLUM ÞJÓNUSTUVEITANDA, STJÓRNANDA EÐA SJÚKLINGS VEGNA RAFLOSTS, VÉLRÆNNAR EÐA ANNARRAR HÆTTU. TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKÉM JAZYCE. • V PŘÍPADĚ, ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTŘEBUJE NÁVOD V JINÉM JAZYCE, JE ZAJIŠTĚNÍ PŘEKLADU DO ODPOVÍDAJÍCÍHO JAZYKA ÚKOLEM ZÁKAZNÍKA. • NEPROVÁDĚJTE ÚDRŽBU TOHOTO ZAŘÍZENÍ, ANIŽ BYSTE SI PŘEČETLI TENTO SERVISNÍ NÁVOD A POCHOPILI JEHO OBSAH. • V PŘÍPADĚ NEDODRŽOVÁNÍ TÉTO VÝSTRAHY MŮŽE DOJÍT ÚRAZU ELEKTRICKÁM PROUDEM PRACOVNÍKA POSKYTOVATELE SLUŽEB, OBSLUŽNÉHO PERSONÁLU NEBO PACIENTŮ VLIVEM ELEKTRICKÉHOP PROUDU, RESPEKTIVE VLIVEM K RIZIKU MECHANICKÉHO POŠKOZENÍ NEBO JINÉMU RIZIKU.
ADVARSEL
DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK. • HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE. • FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE DENNE SERVICEMANUAL ER BLEVET LÆST OG FORSTÅET. • MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFØRE SKADE PÅ GRUND AF ELEKTRISK, MEKANISK ELLER ANDEN FARE FOR TEKNIKEREN, OPERATØREN ELLER PATIENTEN.
WAARSCHUWING (NL)
DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS VERKRIJGBAAR. • ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST, DAN IS DE KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN. • PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS. • INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, ZOU HET ONDERHOUDSPERSONEEL, DE OPERATOR OF EEN PATIËNT GEWOND KUNNEN RAKEN ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK, MECHANISCHE OF ANDERE GEVAREN.
(DA)
iv
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
-
GE Healthcare Direction 5394227, 13
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
BRĪDINĀJUMS (LV)
ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ. • JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU. • NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS IZLASĪŠANAS UN SAPRAŠANAS. • ŠĪ BRĪDINĀJUMA NEIEVĒROŠANA VAR RADĪT ELEKTRISKĀS STRĀVAS TRIECIENA, MEHĀNISKU VAI CITU RISKU IZRAISĪTU TRAUMU APKALPES SNIEDZĒJAM, OPERATORAM VAI PACIENTAM.
ĮSPĖJIMAS (LT)
ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA. • JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ, VERTIMU PASIRŪPINTI TURI KLIENTAS. • NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT VADOVAUTUMĖTĖS ŠIUO EKSPLOATAVIMO VADOVU IR JĮ SUPRASTUMĖTE • NEPAISANT ŠIO PERSPĖJIMO, PASLAUGŲ TEIKĖJAS, OPERATORIUS AR PACIENTAS GALI BŪTI SUŽEISTAS DĖL ELEKTROS SMŪGIO, MECHANINIŲ AR KITŲ PAVOJŲ.
ADVARSEL (NO)
DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK. • HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE. • IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT. • MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ GRUNN AV ELEKTRISK STØT, MEKANISKE ELLER ANDRE FARER.
OSTRZEŻENIE (PL)
NINIEJSZY PODRĘCZNIK SERWISOWY DOSTĘPNY JEST JEDYNIE W JĘZYKU ANGIELSKIM. • JEŚLI FIRMA ŚWIADCZĄCA KLIENTOWI USłUGI SERWISOWE WYMAGA UDOSTĘPNIENIA PODRĘCZNIKA W JĘZYKU INNYM NIŻ ANGIELSKI, OBOWIĄZEK ZAPEWNIENIA STOSOWNEGO TłUMACZENIA SPOCZYWA NA KLIENCIE. • NIE PRÓBOWAĆ SERWISOWAĆ NINIEJSZEGO SPRZĘTU BEZ UPRZEDNIEGO ZAPOZNANIA SIĘ Z PODRĘCZNIKIEM SERWISOWYM. • NIEZASTOSOWANIE SIĘ DO TEGO OSTRZEŻENIA MOŻE GROZIĆ OBRAŻENIAMI CIAłA SERWISANTA, OPERATORA LUB PACJENTA W WYNIKU PORAŻENIA PRĄDEM, URAZU MECHANICZNEGO LUB INNEGO RODZAJU ZAGROŻEUŃ.
-
v
GE Healthcare Direction 5394227, 13
OPOZORILO (SL)
UPOZORENJE (HR)
UPOZORNENIE (SK)
VARNING (SV)
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
TA SERVISNI PRIROČNIK JE NA VOLJO SAMO V ANGLEŠČINI. • ČE PONUDNIK SERVISNIH STORITEV ZA STRANKO POTREBUJE NAVODILA V DRUGEM JEZIKU, JE ZA PREVOD ODGOVORNA STRANKA SAMA. • NE POSKUŠAJTE SERVISIRATI OPREME, NE DA BI PREJ PREBRALI IN RAZUMELI SERVISNI PRIROČNIK. • ČE TEGA OPOZORILA NE UPOŠTEVATE, OBSTAJA NEVARNOST ELEKTRIČNEGA UDARA, MEHANSKIH ALI DRUGIH NEVARNOSTI IN POSLEDIČNIH POŠKODB PONUDNIKA SERVISNIH STORITEV, UPORABNIKA OPREME ALI PACIENTA. OVAJ SERVISNI PRIRUČNIK DOSTUPAN JE SAMO NA ENGLESKOM JEZIKU. • AKO KLIJENTOV SERVISER ZAHTIJEVA JEZIK KOJI NIJE ENGLESKI, ODGOVORNOST KLIJENTA JE PRUŽITI USLUGE PREVOĐENJA. • NEMOJTE POKUŠAVATI SERVISIRATI OPREMU AKO NISTE PROČITALI I RAZUMJELI SERVISNI PRIRUČNIK. • AKO NE POŠTUJETE OVO UPOZORENJE, MOŽE DOĆI DO OZLJEDE SERVISERA, OPERATERA ILI PACIJENTA PROUZROČENE STRUJNIM UDAROM, MEHANIČKIM I DRUGIM OPASNOSTIMA. TÁTO SERVISNÁ PRÍRUČKA JE K DISPOZÍCII LEN V ANGLIČTINE. • AK ZÁKAZNÍKOV POSKYTOVATEĽ SLUŽIEB VYŽADUJE INÝ JAZYK AKO ANGLIČTINU, POSKYTNUTIE PREKLADATEĽSKÝCH SLUŽIEB JE ZODPOVEDNOSŤOU ZÁKAZNÍKA. • NEPOKÚŠAJTE SA VYKONÁVAŤ SERVIS ZARIADENIA SKÔR, AKO SI NEPREČÍTATE SERVISNÚ PRÍRUČKU A NEPOROZUMIETE JEJ. • ZANEDBANIE TOHTO UPOZORNENIA MÔŽE VYÚSTIŤ DO ZRANENIA POSKYTOVATEĽA SLUŽIEB, OBSLUHUJÚCEJ OSOBY ALEBO PACIENTA ELEKTRICKÝM PRÚDOM, PRÍPADNE DO MECHANICKÉHO ALEBO INÉHO NEBEZPEČENSTVA. DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA. • OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA ÖVERSÄTTNINGSTJÄNSTER. • FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN. • OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA I SKADOR PÅ SERVICETEKNIKERN, OPERATÖREN ELLER PATIENTEN TILL FÖLJD AV ELEKTRISKA STÖTAR, MEKANISKA FAROR ELLER ANDRA FAROR.
-
vii
GE Healthcare Direction 5394227, 13
DİKKAT (TR)
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR. • EĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ BİR DİLDE OLMASINI İSTERSE, KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN SORUMLULUĞUNDADIR. • SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE ETMEYİNİZ. • BU UYARININ GÖZ ARDI EDİLMESİ, ELEKTRİK ÇARPMASI YA DA MEKANİK VEYA DİĞER TÜRDEN KAZALAR SONUCUNDA TEKNİSYENİN, OPERATÖRÜN YA DA HASTANIN YARALANMASINA YOL AÇABİLİR.
(JA)
Traditional Chinese
viii
-
GE Healthcare Direction 5394227, 13
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
(ZH-CN)
(KO)
-
ix
GE Healthcare Direction 5394227, 13
x
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
-
GE Healthcare Direction 5394227, 13
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
DAMAGE IN TRANSPORTATION All packages should be closely examined at time of delivery. If damage is apparent write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested within this 14 day period.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be performed by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall be performed by qualified GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes. The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
OMISSIONS & ERRORS If there are any omissions, errors or suggestions for improving this documentation, please contact the GE Healthcare Global Documentation Group with specific information listing the system type, manual title, part number, revision number, page number and suggestion details. Mail the information to: Service Documentation, 9900 Innovation Drive (RP-2123), Wauwatosa, WI53226, USA. GE Healthcare employees should use Application Lifecycle Management (ALM) to report service documentation issues. These issues will then be in the internal problem reporting tool and communicated to the writer.
SERVICE SAFETY CONSIDERATIONS DANGER DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING. WARNING WARNING USE ALL PERSONAL PROTECTION EQUIPMENT (PPE) SUCH AS GLOVES, SAFETY
SHOES, SAFETY GLASSES, AND KNEELING PAD, TO REDUCE THE RISK OF INJURY. For a complete review of all safety requirements, see the Chapter 1, Safety Considerations section in the Service Manual.
-
xi
GE Healthcare Direction 5394227, 13
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
LEGAL NOTES The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare. GE Healthcare may revise this publication from time to time without written notice.
PROPRIETARY TO GE HEALTHCARE Permission to use this Advanced Service Software and related documentation (herein called the material) by persons other than GE Healthcare employees is provided only under an Advanced Service Package License relating specifically to this Proprietary Material. This is a different agreement from the one under which operating and basic service software is licensed. A license to use operating or basic service software does not extend to or cover this software or related documentation. If you are a GE Healthcare employee or a customer who has entered into such a license agreement with GE Healthcare to use this proprietary software, you are authorized to use this Material according to the conditions stated in your license agreement. However, you do not have the permission of GE Healthcare to alter, decompose or reverse-assemble the software, and unless you are a GE employee, you MAY NOT COPY the Material. The Material is protected by Copyright and Trade Secret laws; the violation of which can result in civil damages and criminal prosecution. If you are not party to such a license agreement or a GE Healthcare Employee, you must exit this Material now.
TRADEMARKS All products and their name brands are trademarks of their respective holders.
COPYRIGHTS All Material Copyright© 2011 by General Electric Company Inc. All Rights Reserved.
xii
-
GE Healthcare Direction 5394227, 13
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
Revision History Revision
Date
Reason for change
1
Jun. 2011
Initial Release
2
Jul. 2011
Correction
3
Mar. 2012
Correction
4
May. 2013
Initial release for R2.x
5
March.. 2014
Minor correctionfor R2.x
6
July, 2015
Release for R3
7
September, 2015
Update
8
November, 2015
Update BW Printer setup
9
Februry, 2016
Add UDI label information
10
Februry, 2017
Update adding R4 information
11
June, 2017
Revise with Service Note information
12
March, 2019
Update for R4.2.5x release
13
October, 2019
Add LOGIQ S7 XDclear2.0.
List of Effected Pages (LOEP) Pages
Revision
Pages
Revision
Pages
Revision
Title Page
13
Chapter 3
13
Chapter 8
13
Warnings i to xii
13
Chapter 4
13
Chapter 9
13
TOC
13
Chapter 5
13
Chapter 10
13
Chapter 1
13
Chapter 6
13
Back Cover
N/A
Chapter 2
13
Chapter 7
13
-
xiii
GE Healthcare Direction 5394227, 13
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
This page was intentionally left blank.
xiv
-
GE Healthcare Direction 5394227, 13
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
Table of Contents CHAPTER 1 Introduction Overview...1 - 1 Purpose of this chapter...1 - 1 Contents in this chapter...1 - 1 Service manual overview...1 - 2 Contents in this service manual...1 - 2 Typical users of the Service Manual...1 - 3 Models covered by this manual...1 - 4 Product description...1 - 6 Important conventions...1 - 7 Conventions used in this manual...1 - 7 Standard Hazard Icons...1 - 8 Product Icons...1 - 10 Safety Considerations...1 - 15 Contents in this section...1 - 15 Introduction...1 - 15 Human Safety...1 - 15 Mechanical Safety...1 - 17 Electrical Safety...1 - 19 Auxiliary Devices Safety...1 - 20 Battery Safety...1 - 22 Labels Locations...1 - 23 The labels on the cover...1 - 23 Rating plate...1 - 26 Dangerous Procedure Warnings...1 - 27 Lockout/Tagout (LOTO) Requirements...1 - 28 Returning/Shipping System, Probes and Repair Parts...1 - 29 For LOGIQ S8 Vet system...1 - 29 Electromagnetic Compatibility (EMC) and Electrostatic Discharge (ESD)...1 - 30 What is EMC?...1 - 30 Compliance...1 - 30 Table of Contents
1
GE Healthcare Direction 5394227, 13
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
Electrostatic Discharge (ESD) Prevention...1 - 30 Customer Assistance...1 - 31 Contact Information...1 - 31 System Manufacturer...1 - 32 Authorized EU Representative...1 - 32
2
-
GE Healthcare Direction 5394227, 13
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
CHAPTER 2 Site preparations Overview... 2 - 1 Purpose of this chapter... 2 - 1 Contents in this chapter... 2 - 1 General console requirements... 2 - 2 Environmental Requirements... 2 - 2 Electrical Requirements... 2 - 3 EMI Limitations... 2 - 4 Probe Environmental Requirements... 2 - 6 Time and Manpower Requirements... 2 - 6 Facility needs... 2 - 7 Purchaser Responsibilities... 2 - 7 Required Facility Needs... 2 - 8 Privacy and Security... 2 - 8 Desirable features... 2 - 9 Minimal floor plan suggestions... 2 - 9 Networking setup requirements... 2 - 10 Environmental Dangers... 2 - 12 Patient Vicinity UL60601-1 (USA)... 2 - 12 Patient Environment IEC60601-1 (IEC60601-1-1) and ANSI AAMI ES60601-1 2 - 13
-
3
GE Healthcare Direction 5394227, 13
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
CHAPTER 3 System Setup Overview...3 - 1 Purpose of this chapter...3 - 1 Contents in this chapter...3 - 1 Setup reminders...3 - 2 Average Setup Time...3 - 2 Setup warnings...3 - 3 Receiving and unpacking the equipment...3 - 5 Receiving and unpacking warnings...3 - 5 Receiving the LOGIQ S8...3 - 6 Transportation Box Label...3 - 8 Unpacking the LOGIQ S8...3 - 9 Packing materials - recycling information...3 - 10 Preparing for setup...3 - 11 Verify Customer Order...3 - 11 Physical inspection...3 - 11 EMI Protection...3 - 11 Completing the Set Up...3 - 12 System Specifications...3 - 12 Electrical specifications...3 - 13 Connection the I/O Rear Panel...3 - 14 Connecting Probes...3 - 15 Power On/Off...3 - 17 Connect AC (mains) Power to the LOGIQ S8...3 - 18 Power ON...3 - 19 Login...3 - 20 Logoff/Shutdown/Exit...3 - 21 Complete shutdown...3 - 22 Sleep Mode (R1 to R4)...3 - 23 To change your password...3 - 24 Password policies...3 - 25 Disk encryption...3 - 29 System Configuration...3 - 32 Purpose of this section...3 - 32 4
-
GE Healthcare Direction 5394227, 13
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
System Configuration... 3 - 32 Software Option installation... 3 - 35 Optional Peripherals/Peripheral Connection... 3 - 37 Connecting the Secondary “Patient” LCD Monitor... 3 - 38 Connecting the Off-Board/On-Board UVC... 3 - 42 Connecting the Footswitch... 3 - 46 ECG connector... 3 - 47 Connecting the USB Flash Drive and USB External Hard disk drive... 3 - 48 Printer Installation... 3 - 49 Installing Digital Black & White Printer Sony UP-D897/D898... 3 - 49 Installing Digital Color Printer Sony UP-D25MD... 3 - 58 Adding the printer to the system... 3 - 64 Setting the printer to print reports... 3 - 65 PC printer setup... 3 - 65 Connectivity Setup... 3 - 66 Connectivity Introduction... 3 - 66 The Dataflow Concept... 3 - 66 Stand-alone LOGIQ S8... 3 - 68 LOGIQ S8 + PC within a “Sneaker Net”... 3 - 68 Connection between LOGIQ S8 and DICOM Server... 3 - 68 Wired Ethernet from LOGIQ S8 to a Workstation... 3 - 69 Connection using Wireless Option (R2.x.x or later)... 3 - 69 Network icon... 3 - 70 Configuring Connectivity... 3 - 72 Overview... 3 - 72 Contents in this section... 3 - 72 Structured Reporting... 3 - 72 Connectivity Functions... 3 - 73 TCPIP... 3 - 74 Device... 3 - 76 Service... 3 - 77 Instructions for setting up Network Storage on LOGIQ S8 R4... 3 - 79 Setting up InSite Connection (R1 to R4)... 3 - 82 Overview... 3 - 82 Contents in this section... 3 - 82 Prerequisites for InSite Setup... 3 - 82 Configuration Steps... 3 - 83 Configuring InSite... 3 - 83 -
5
GE Healthcare Direction 5394227, 13
LOGIQ S8/E8/S7 XDclear2.0 Service Manual
Configuring Agent Configuration...3 - 83 Configuring Advanced Configuration...3 - 86 Configuring Request for Service (RFS)...3 - 90 Verifying InSite Connection - CSD...3 - 91 Verifying InSite Connection - VCO...3 - 92 Agent Configuration (R4.2.5x and later)...3 - 93 Overview...3 - 93 Verifying back office connection - CSD...3 - 93 Verifying Back Office Connection - VOC...3 - 94 Connectivity Setup Worksheet...3 - 95 Paperwork...3 - 97 Product Locator Installation Card...3 - 97 User Manual(s)...3 - 97
6
-