A Series Disassembly of Breathing System for Cleaning Disinfection Autoclaving Guide Rev A Feb 2020

Guide

3 Pages

Disassembly of A-Series Breathing System for Cleaning, Disinfection and Autoclaving  1 Disassembly  2  3  Remove the bellows dome by turning it counterclockwise and lifting it away from the breathing system.  Detach the bellows from the base plate.  Detach the top plate from the bellows.  4  5  6  Remove the bellows dome O-ring.  Remove the valve dome by turning it counterclockwise. Remove the valve from the breathing system.  7  8  9  Remove flow sensors by turning the locking nuts counterclockwise. Pull out the inspiration and expiration connectors together with their locking nuts. And then pull out the flow sensors horizontally.  Disconnect the oxygen sensor cable from the main unit. Then hold the oxygen sensor and pull straight out firmly from the breathing system block.  Loosen the base collar of the APL valve by turning the collar (not the valve knob) counterclockwise until it is no longer threaded. Then, firmly pull the APL valve upward to remove.  Remove the bellows adapter ring from inside the bellows. Note the orientation of the bellows adapter ring as it is being removed to ensure that it is properly inserted during reassembly. (If the ring contains grooves, the ring should be oriented so that the grooves are facing downward in the final reassembly.)  Page 1 of 3