BeneVision N17/BeneVision N15/ BeneVision N12/BeneVision N12C Patient Monitor Service Manual
Intellectual Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray) owns the intellectual property rights to this product and this manual. This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability arising out of any infringements of patents or other rights of third parties.
,
, and
are the registered trademarks or trademarks owned by
Mindray in China and other countries.
Revision History This manual has a revision number. This revision number changes whenever the manual is updated due to software or technical specification change. Contents of this manual are subject to change without prior notice.
Version number
4.0
Release time:
December 2018
© 2016 -2017 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved.
I
Preface Manual Purpose This manual provides detailed information about the assembling, dissembling, testing and troubleshooting of the equipment to support effective troubleshooting and repair. It is not intended to be a comprehensive, in-depth explanation of the product architecture or technical implementation. Observance of the manual is a prerequisite for proper equipment maintenance and prevents equipment damage and personnel injury.
Intended Audience This manual is for biomedical engineers, authorized technicians or service representatives responsible for troubleshooting, repairing and maintaining the monitors
Passwords A password may be required to access different modes. The passwords are listed below:
User maintenance:
Manage Configuration: 315666
Factory maintenance:
332888
Demo mode:
2088
888888
II
Contents 1 Safety ... 1-1 1.1 Safety Information... 1-1 1.1.1 Dangers ... 1-1 1.1.2 Warnings ... 1-2 1.1.3 Cautions ... 1-2 1.1.4 Notes ... 1-3 1.2 Equipment Symbols ... 1-3 2 Operation Theory ... 2-1 2.1 Overview ... 2-1 2.2 Product System Architecture ... 2-1 2.2.1 Main Control Board ... 2-3 2.2.2 Internal Module Rack COM Board ... 2-3 2.2.3 Power Architecture ... 2-4 2.2.4 Independent Display Board (for the N17 Only) ... 2-5 2.2.5 iView Module (for the N17 Only) ... 2-5 2.2.6 Alarm LAMP Board ... 2-5 2.2.7 Power Switch Board ... 2-5 2.3 Data Logic Flow ... 2-5 3 WLAN Installation ... 3-1 3.1 Introduction... 3-1 3.2 Network Deployment Process ... 3-1 3.2.1 List of outputs ... 3-2 3.3 Network Requirements ... 3-3 3.4 Network Verification ... 3-4 3.4.1 Tools and Resources... 3-4 3.4.2 Wi-Fi Signal Calibration... 3-5 3.4.3 Network Verification Process ... 3-5 3.5 Network Coverage Assessment with Patient Monitors ... 3-8 3.6 Recommended Devices for WLAN ... 3-9 3.7 Setting Wireless Parameters for Patient Monitors ... 3-9 3.8 Troubleshooting ...3-11 4 Testing and Maintenance ... 4-1 4.1 Introduction... 4-1 4.1.1 Test Equipment ... 4-1 4.1.2 Test Report ... 4-1 4.1.3 Preventative Maintenance ... 4-1 4.1.4 Recommended Frequency ... 4-2 4.2 Preventative Maintenance Procedures ... 4-3
1
4.2.1 Visual Inspection ... 4-3 4.2.2 NIBP Tests... 4-4 4.2.3 Sidestream and Microstream CO2 Tests ... 4-6 4.2.4 AG Tests ... 4-9 4.3 Power On Test ...4-11 4.4 Module Performance Tests ... 4-11 4.4.1 ECG Tests ...4-11 4.4.2 Resp Performance Test... 4-12 4.4.3 SpO2 Test ...4-12 4.4.4 NIBP Tests...4-13 4.4.5 Temp Test ...4-13 4.4.6 IBP Tests ...4-13 4.4.7 C.O. Test ...4-15 4.4.8 Mainstream CO2 Tests ... 4-15 4.4.9 Sidestream and Microstream CO2 Tests ... 4-16 4.4.10 AG Tests ...4-16 4.4.11 EEG Test ...4-17 4.4.12 BIS Test ...4-18 4.4.13 RM Test ...4-19 4.4.14 CCO/SvO2 Tests ... 4-20 4.4.15 PiCCO Test ...4-21 4.4.16 NMT Tests ...4-23 4.5 Nurse Call Relay Performance Test ... 4-24 4.6 Analog Output Performance Test ... 4-25 4.7 Electrical Safety Tests...4-25 4.8 Recorder Check ...4-26 4.9 Network Print Test ...4-26 4.9.1 Device Connection and Setup ... 4-26 4.10 Device Integration Check ... 4-27 4.11 Battery Check...4-27 4.12 Mounting Check ...4-27 4.12.1 Safety check ...4-27 4.12.2 Overall Test and Check of Installed System ... 4-27 4.13 Factory Maintenance ... 4-28 4.13.1 Accessing Factory Maintenance Menu... 4-28 4.13.2 Monitor Information ... 4-28 4.13.3 Production Test ... 4-29 4.13.4 Setup...4-29 4.13.5 Debug...4-30 4.13.6 Power Info ...4-30 4.13.7 ClinicalData ...4-30 4.13.8 Transferring Clinical Data ... 4-31 4.13.9 Software Version ... 4-31 4.13.10 Monitor Information ... 4-31
2
5 Troubleshooting ... 5-1 5.1 Introduction... 5-1 5.2 Part Replacement ... 5-1 5.3 Check before Powering on the Monitor ... 5-1 5.4 Software Version Check ... 5-2 5.5 Technical Alarm Check ... 5-2 5.6 Troubleshooting Guide... 5-2 5.6.1 Power On/Off Failures ... 5-2 5.6.2 Display Failures ... 5-3 5.6.3 Module Rack Failures ... 5-4 5.6.4 Alarm Failures... 5-5 5.6.5 Recorder Failures... 5-6 5.6.6 Output Interface Failures ... 5-6 5.6.7 Power Supply Failures ... 5-7 5.6.8 Network Related Problems ... 5-8 5.6.9 Software Upgrade Problems ... 5-9 5.6.10 Device Integration Failures ... 5-10 6 Upgrade ... 6-1 6.1 Overview ... 6-1 6.2 Upgrade of Parameter Function Modules ... 6-2 6.3 Upgrade of Functional Assemblies ... 6-3 6.3.1 Upgrading SMR ... 6-4 6.3.2 Upgrading Wireless Network Functions ... 6-4 6.3.3 Upgrading Recorder ... 6-5 6.3.4 Upgrading iView System ... 6-5 6.3.5 Upgrading Independent Display Function ... 6-5 6.4 Upgrading Software ... 6-6 6.4.1 Description of Network Upgrade Tool ... 6-7 6.4.2 Guide to Software Upgrade Operations... 6-10 6.4.3 Guide to Upgrade through a USB Disk ... 6-11 6.5 Upgrading CAA license function ... 6-14 7 Repair and Disassembly ... 7-1 7.1 Tools ... 7-1 7.2 Preparations for Disassembly ... 7-1 7.3 Whole Unit Disassembly ... 7-2 7.3.1 Basic Disassembly ... 7-2 7.3.2 Disconnecting the Base ... 7-3 7.3.3 Separating the Front and Rear Half of the Monitor ... 7-4 7.4 Further Disassembly of the Front Case Assembly... 7-7 7.4.1 Removing the PowerSwitch Board ... 7-7 7.4.2 Removing the Alarm Lamp and Light Sensor Board ... 7-7 7.4.3 Removing the Screen Assembly Connection Line ... 7-8
3
7.5 Further Disassembly of the Rear Case Assembly ... 7-9 7.5.1 Removing the Recorder... 7-9 7.5.2 Further Disassembly of the Recorder ... 7-10 7.5.3 Removing the MPAN Board ... 7-11 7.5.4 Removing the Wi-Fi Module ... 7-12 7.5.5 Removing the Internal Module Rack Assembly... 7-13 7.5.6 Removing the Main Support Assembly... 7-17 7.5.7 Removing the iView Board Assembly (N17 Series)... 7-21 7.5.8 Removing the Independent Display Board Assembly (N17 Series) ... 7-23 7.5.9 Removing the Handle ... 7-24 7.6 Further Disassembly of the Base Assembly ... 7-25 7.7 Disassembling the Module Rack ... 7-26 7.8 Disassembling the M51C Module... 7-28 8 Parts ... 8-1 8.1 N12 Parts ... 8-1 8.1.1 N12 Whole Unit ... 8-1 8.1.2 N12-NLT Front Housing Assembly ... 8-2 8.1.3 N12-SHARP Front Housing Assembly... 8-4 8.1.4 N12 Rear Housing ... 8-5 8.1.5 N12 Rear Housing Assembly (FRU)(115-044499-00)... 8-7 8.1.6 N12 Battery Cavity Assembly (FRU)(115-044504-00)... 8-8 8.1.7 N12 Module Rack Body Assembly (FRU) (115-044507-00) ... 8-9 8.2 N 15 Parts ...8-10 8.2.1 N15 Whole Unit ... 8-10 8.2.2 N15-NLT Front Housing Assembly ... 8-11 8.2.3 N15-SHARP Front Housing ... 8-12 8.2.4 N15 Rear Housing ... 8-14 8.2.5 N15 Rear Housing Assembly (FRU) (115-044541-00)... 8-16 8.2.6 N15 Base Assembly (FRU)(115-044543-00) ... 8-17 8.2.7 N15N17 Module Rack Body Assembly (FRU) (115-044546-00) ... 8-18 8.3 N17 Parts ...8-19 8.3.1 N17 Whole Unit ... 8-19 8.3.2 N17-NLT Front Housing Assembly ... 8-20 8.3.3 N17-SHARP Front Housing ... 8-21 8.3.4 N17 Rear Housing ... 8-23 8.3.5 N17 Rear Housing Assembly (FRU)( 115-044548-00)... 8-25 8.3.6 N17 Base Assembly (FRU)( 115-044550-00) ... 8-26 8.3.7 Independent Display Upgrade Package (Without Mainboard) (115-044554-00) ... 8-27 8.3.8 Independent Display Upgrade Package (With Mainboard) (115-044579-00) ... 8-28 8.3.9 Independent Display Assembly ... 8-29 8.3.10 iView Module Assembly(Win7 system) ... 8-30 8.3.11 iView Module Assembly(Win10 system) ... 8-32 8.4 External Satellite Module Rack (SMR)... 8-34
4
8.5 M51C Module ...8-35 A Electrical Safety Inspection ... A-1
5
FOR YOUR NOTES
6
1 Safety 1.1 Safety Information
DANGER
Indicates an imminent hazard that, if not avoided, will result in death or serious injury.
WARNING
Indicates a potential hazard or unsafe practice that, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage.
NOTE
Provides application tips or other useful information to ensure that you get the most from your product.
1.1.1 Dangers There are no dangers that refer to the product in general. Specific “Danger” statements may be given in the respective sections of this manual.
1-1
1.1.2 Warnings WARNING
All installation operations, expansions, changes, modifications and repairs of this product are conducted by Mindray authorized personnel.
There is high voltage inside the equipment. Never disassemble the equipment before it is disconnected from the AC power source.
When you disassemble/reassemble a parameter module, a patient leakage current test must be performed before it is used again for monitoring.
The equipment must be connected to a properly installed power outlet with protective earth contacts only. If the installation does not provide for a protective earth conductor, disconnect it from the power line and operate it on battery power, if possible.
Dispose of the package material, observing the applicable waste control regulations and keeping it out of children’s reach.
1.1.3 Cautions CAUTION
Make sure that no electromagnetic radiation interferes with the performance of the equipment when preparing to carry out performance tests. Mobile phone, X-ray equipment or MRI devices are a possible source of interference as they may emit higher levels of electromagnetic radiation.
Before connecting the equipment to the power line, check that the voltage and frequency ratings of the power line are the same as those indicated on the equipment’s label or in this manual.
Protect the equipment from damage caused by drop, impact, strong vibration or other mechanical force during servicing.
1-2
1.1.4 Notes NOTE
Refer to Operation Manual for detailed operation and other information.
1.2 Equipment Symbols See the N series Operator’s Manual for information about the symbols used on this product and its packaging.
1-3
FOR YOUR NOTES
1-4
2 Operation Theory 2.1 Overview As a bedside workstation for multi-parameter monitoring, the N series can provide the complete patient management, abundant physiological parameter monitoring and physiological alarm functions, as well the powerful data review function and the flexible wired and wireless network configuration and application capabilities. The third-party application can be accessed easily through the iView application, meeting the increasingly common information requirements of hospitals. The provided series of CAA applications can help doctors to make auxiliary diagnosis for patients. Meanwhile, the N series provide the hospital management personnel with more excellent monitor management applications, rendering assistance in fixing the efficiency and quality problems during monitor equipment management of hospitals. The N series provide the product models with display screens of different sizes according to the demand of clinical application. In addition to touch screen operations, the user can use the mouse and keyboard to operate the monitor. The N series can connect to multiple display screens to function as mirror screens or extension screens. The series of products are compatible with the BeneView T series plug-in modules and related accessory products. They can work together with the TDS to implement the intra-hospital transfer application of patients. In comparison with the BeneView T series products, the N series boast better human-computer interaction design and clinical applicability, more complete IT solution capability of hospitals, and more abundant CDS applications.
2.2 Product System Architecture All the N17/N15/N12 monitors have only one main unit:
The N12 main unit uses the 12.1” TFT WXGA display screen.
The N15 main unit uses the 15.6” TFT FHD display screen.
The N17 main unit uses the 18.5”TFT FHD display screen.
All of them use the touch screen as an input device and can extend the mouse, keyboard and remote control.
An internal module rack is integrated, with 4 slots (N12) or 6 slots (N15/N17).
The MPAN and WiFi modules are optional.
The built-in recorder is optional.
2-1
The N15/N17 can connect to the external module rack and TDS; the N12 can connect to the TDS.
LCD TP control board
Front shell Back shell
Alarm lamp board
Speaker
DVI
Power switch board
WiFi module (Optional)
MPAN module (Optional)
Independent display board (N17 Optional) Internal module com board(4slots/6slots) Main control board
iView assembly (N17 Optional) DVI
RJ45
2*USB
SMR interface
Nurse call interface
4/6*Module communication interface
2*USB
AC/DC module
Recorder (Optional) 2*USB
2*USB
DVI
Bottom shell
AC/DC interface
Battery interface board (N12/N12C)
RJ45
Battery interface board (N15/N17)
Battery
Figure 2-1 System block diagram of the N17/N15/N12/N12C
2-2
Battery
2.2.1 Main Control Board There are the main control CPU, program memory, data memory, system configuration memory, system FPGA, WiFi module (optional), power management MCU, battery charging circuit, and DC-DC circuit on the main control board. The internal interface and external interfaces are also provided on the board. The internal interface is an interface between the recorder, internal module rack COM board, AC-DC, and the battery. The external interfaces refer to the DVI display interface, USB interface, and Ethernet interface.
MPAN Module
Internal Module Rack communication board
SPI
MMC
E2ROM(funnct ion cfg)
Nursecall interface
DDR3 I2C
RTC SMR Interface
MMC Alarm light controller
Touch panel
WiFi (optional)
Main Control board
Program Memory
Data Memory
Touchcontroller
I2C SPI
E2ROM
USB
USB hub
Main Processor
Photo senseor
Speaker
Audio codec/AMP
Power key board
Power M0
RJ-45
UART
RMII1
PHY
LCDC
FPGA (N15/N17) DVI transfer
PHY
USB 4*port Independent display board/iView LCD DVI interface
Back light driver
Figure 2-2 Diagram of the main control board
2.2.2 Internal Module Rack COM Board Two models of internal module rack COM boards are available. The N12 uses the 4-slot COM board, and the N15/N17 uses the 6-slot COM board. The internal module rack COM board is used to provide the interface for communication with the parameter module, the SMR interface and nurse call interface, and the MPAN module interface. Besides, the data forwarding FPGA and corresponding power circuit are also provided on the internal module rack COM board.
2-3
2.2.3 Power Architecture Power switch board
LCD&TP
+5V/+3.3V
Led driver
Photo Alarm lamp board senseor
+3.3V
+3.3V_1
M0
+12V
DC/DC
+3.3V_2
MPAN module
ON/OFF DC/DC
LDO 100~240V AC
Internal module rock COM board
ON/OFF AC/DC module
AC-DC in
DC/DC
switch ON/OFF BAT charger
VBUS: 9.6~11.1V, +15V
+12V
+5V DC/DC
+12V
iView module
+3.3V
Battery +5V
+5V
Independent display board
+12V
Recorder
Figure 2-3 Diagram of power architecture The AC/DC power module outputs 15V to the main control board, and 3.3V, 5V and 12V can be generated through the internal DC-DC conversion circuit in the main control board to provide a power supply to other modules or boards in the main unit. The battery charging circuit is powered by 15V, and the AC power supply and battery power supply can be switched according to AC on-line detection. The +12V power supply is provided to the power supply, including the external module rack, and the DC-DC isolation design is implemented at the module end. The iVIew assembly uses the power rail Vbus, which is the switching output between the AC-DC output and battery and aims to avoid abnormal power failure of the iView module and running exception of the Windows OS running on other modules due to an unexpected power failure of the AC power supply. The battery supports the main unit to stop the iView module in the normal power-off mode. In the case of battery power supply, the iView module cannot start.
2-4
2.2.4 Independent Display Board (for the N17 Only) It is used to connect the main unit to a display and extend the main screen display. It adopts the DVI interface. Moreover, the external display with a touch screen can be supported through the USB interface of the main unit. At present, the supported display with a touch screen is Elo 1919LM.
2.2.5 iView Module (for the N17 Only) As an embedded computer module, it provides the following external interfaces: the network interface, DVI interface, and the USB interface. It can connect to the keyboard, mouse, network cable, and display independently. The configuration of the iView module is mutually exclusive with that of the independent display module.
2.2.6 Alarm LAMP Board The LED alarm lamp and light sensor are provided on the board. The light sensor implements the ambient light detection and is used to adjust brightness of the LCD background light.
2.2.7 Power Switch Board There are the power switch and three indicators on the power switch board, which are the AC on-line indicator, battery indicator, and the power-on indicator.
2.3 Data Logic Flow Display and user interface
Data collection
Data forwarding
System application
Data output
Data storage
Figure 2-4 Data flow diagram The monitoring parameters are collected and analyzed through the module, and then forwarded to the system software through the internal or external module rack. The system software displays the waveform, numerical value and alarm information, and the data, alarm information and numerical value are also stored in the internal data memory at the same time. Meanwhile, they can also be sent to the central station or other monitors through the wired or wireless network.
2-5
FOR YOUR NOTES
2-6
3 WLAN Installation 3.1 Introduction This chapter describes how to install Mindray patient monitors using WLAN.
3.2 Network Deployment Process If the hospital has built its WLAN, the installation process is illustrated as follows:
Mindray/agent marketing or sales personnel
A0 Provide network requirements
Mindray headquarters
A1 Communicate with the hospital about intention to order
Mindray/agent service personnel
A2 Communicate with the hospital about network requirements and ask the hospital to perform corresponding setting and adjustment
A3 Do site survey
Pass or fail
Pass A4 Sign the contract
A5 Install the device and confirm system operation
Figure 3 -1 Network Deployment Process
3-1