Mindray Ventilator Cleaning and Disinfection Protocol

Protocol

1 Page

Ventilator Cleaning and Disinfection Protocol  Cleaning/Disinfectant Selections Name  Type  Ethanol 75%  Moderately efficient disinfectant  Isopropanol 75%  Moderately efficient disinfectant  Glutaraldehyde (2%)  Highly efficient disinfectant  Ortho-Phthalaldehyde disinfectant (such asCidex®OPA)(Suitable for SV series)  Highly efficient disinfectant  Soap water (pH value of 7.0~10.5)  Rinsing agent  Clean water  Rinsing agent  Steam autoclave(134ºC for 10 to 20 minutes recommended)  Highly efficient disinfection  Cleaning and Disinfection Cleaning Wipe: wipe with a damp cloth immersed in alkalescent detergent (soap water, etc.) or alcohol solution and then wipe off the remaining detergent with a dry lint free cloth. Immersion: flush with water first and then immerse it in alkalescent detergent (soap water, etc.) for approximately three minutes (recommended water temperature is 40ºC). Finally clean with water and dry completely.  Disinfection Wipe: wipe with a damp cloth immersed in a medium or high-efficiency detergent and then wipe off the remaining detergent with a dry lint free cloth. Immersion: immerse it in a medium or high-efficiency detergent for more than 30 minutes (recommended time). Then clean with water and dry completely. Steam autoclav: 134ºC for 10-20 minutes (recommended time) Ultraviolet radiation: 30-60 minutes (recommended time)  Recommended Frequency  Parts  Cleaning  Disinfection  Wipe  Wipe Ultraviolet radiation  Wipe  Wipe Ultraviolet radiation  Wipe  Wipe Ultraviolet radiation  Ventilator Housing Ventilator external surface (including housing, power cord, supply gas hose))  Trolley and support arm  Touchscreen  Fan dust filter Fan filter (filter at air intake vent) (E Series)  Each patient  Each patient  Each patient  Every four weeks/as Immersion Immersion necessary Every weeks/as necessary  Immersion  Ultraviolet radiation  Main unit air outlet dust filter (SV Series)  Every four weeks/as Immersion Immersion necessary  Air intake dust filter (SV Series)  Every four weeks/as Immersion Immersion necessary  Ventilator inspiration safety valve assembly Inspiration safety valve assembly  As necessary  Immersion Immersion Steam autoclave  Ventilator expiration valve assembly Expiration valve assembly  Immersion Each patient/weekly Immersion Steam autoclave  Expiration valve flow sensor (E series)  Each patient/weekly Immersion Immersion  Ventilator patient tubing (reusable) Patient tubing (including Each patient/weekly Immersion watertrap, Y piece, adapter)  Immersion Steam autoclave  Before cleaning and disinfecting your equipment, please read carefully the ‘Care and Cleaning’ chapter in the device's operators manual.  www.mindrayuk.com
File Type: PDF
File Size: 141 KB
File Name: Mindray Ventilator Cleaning and Disinfection Protocol.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions