Manual Cleaning Instructions. For Synthes battery-driven power equipment.
Small Battery Drive Trauma Recon System Battery Power Line
Synthes (USA) 1302 Wrights Lane East West Chester, PA 19380 Telephone: (610) 719-5000 To order: (800) 523-0322 Fax: (610) 251-9056 © 2009 Synthes, Inc. or its affiliates. All rights reserved.
Synthes (Canada) Ltd. 2566 Meadowpine Boulevard Mississauga, Ontario L5N 6P9 Telephone: (905) 567-0440 To order: (800) 668-1119 Fax: (905) 567-3185
Synthes is a trademark of Synthes, Inc. or its affiliates.
www.synthes.com Printed in U.S.A. 5/11 J9538-B
Manual Cleaning Instructions. For Synthes battery-driven power equipment. For Synthes Small Battery Drive, Trauma Recon System and Battery Power Line. This information is provided in accordance with ISO 17664 and ANSI/AAMI ST81.
Important: Do not sterilize batteries or power modules. Batteries/power modules will no longer function.
Important: Do not submerge or soak devices in aqueous solutions.
1 Disassemble Disassemble device if applicable, and remove the battery or power module from casing or handpiece.
2 Return batteries to charger Return batteries/power modules to charger after each use.
Batteries/power modules and charger may be wiped with a cloth. DO NOT disinfect, immerse, or sterilize atteries/power modules.
3 Remove debris Rinse device under running cold tap water for a minimum of three minutes. Use a sponge, soft lint-free cloth and/or soft-bristled brush to assist in the removal of gross soil and debris. Clean all cannulations, handpiece and attachments, with the Cleaning Brush (516.101). Manipulate all moving parts such as the triggers, release sleeves for attachments, mode switch, etc. under running cold tap water to loosen and remove gross debris. Note: Do not use pointed objects for cleaning.
Note: The screw coupling of the Sagittal Saw Attachment, long, 05.001.224, completely disassembles from saw for cleaning.
4
5
6
Spray with solution Spray and wipe device using a neutral pH enzymatic solution or foam spray for a minimum of three minutes. Follow the enzymatic solution or foam spray manufacturer’s instructions for use for correct temperature, water quality and concentration/dilution.
Rinse with tap water Rinse device with cold tap water for a minimum of two minutes. Use a syringe, pipette or water jet to flush lumens, channels and other hard to reach areas.
Clean with detergent Clean device manually under running warm water using a neutral enzymatic detergent for a minimum of five minutes. Manipulate all moving parts under running water. Use a soft-bristled brush and/or soft, lint-free cloth to remove all visible soil and debris. Follow the enzymatic detergent manufacturer’s instructions for use for correct temperature, water quality and concentration/dilution. Note: For Sagittal Saw Attachment, long, 05.001.224, process the Screw Coupling in an ultrasonic cleaner using a neutral pH enzymatic cleaner for a minimum of 15 minutes.
7
8
9
Rinse with tap water Rinse device using cool to lukewarm running tap water for a minimum of two minutes. Use a syringe, pipette or water jet to flush lumens and channels. Actuate joints, handles and other moveable device features in order to rinse thoroughly under running water.
Final rinse with deionized/purified water Perform a final rinse using deionized or purified water for a minimum of two minutes.
Dry Dry device using a soft, lint-free cloth or clean compressed air. Please refer to Synthes Battery Powered Equipment Lubrication Chart after cleaning and prior to sterilization.
Visually inspect device. Repeat Steps 3–7 until no visible soil remains on device.
Manual Cleaning Instructions. For Synthes battery-driven power equipment.
Small Battery Drive Trauma Recon System Battery Power Line
Synthes (USA) 1302 Wrights Lane East West Chester, PA 19380 Telephone: (610) 719-5000 To order: (800) 523-0322 Fax: (610) 251-9056 © 2009 Synthes, Inc. or its affiliates. All rights reserved.
Synthes (Canada) Ltd. 2566 Meadowpine Boulevard Mississauga, Ontario L5N 6P9 Telephone: (905) 567-0440 To order: (800) 668-1119 Fax: (905) 567-3185
Synthes is a trademark of Synthes, Inc. or its affiliates.
www.synthes.com Printed in U.S.A. 5/11 J9538-B