Electric Pen Drive Compact Drive With Specific Attachments For A Wide Spectrum Of Applications
User‘s Manual
Table of Contents
Introduction
Consoles
General Information
4
Explanation of Symbols
6
Standard and Basic Consoles
7
Set-up of Consoles
8
Speed Regulation
10
Irrigation11
Electric Pen Drive System
Attachments
Electric Pen Drive 60,000 rpm (05.001.010)
13
Hand Switch (05.001.012)
14
Foot Switch, 1 Pedal (05.001.016)
15
Foot Switch, 2 Pedals (05.001.017)
16
General Information
17
Drill Attachments
18
Screw Attachments
20
Kirschner Wire Attachment
20
Saw Attachments
21
Burr Attachments
23
Adapter for Intra Coupling
25
Perforators
26
Craniotome Attachment
27
Table of Contents
Cutting Tools
General Information
28
Cleaning and Disinfection
General Information
29
Preparation Prior to Cleaning
30
Manual Cleaning Instructions
31
Automated Cleaning Instructions with Manual Pre-Cleaning33
Care and Maintenance
Maintenance and Lubrication36 Function Control
38
Packaging, Sterilization and Storage
39
Repairs and Technical Service
40
Disposal
41
Troubleshooting
42
System Specifications
44
Electromagnetic Compatibility Accompanying Documents According to IEC 60601-1-2, 2014, ED. 4.0
50
Ordering Information
54
User‘s Manual
Electric Pen Drive
DePuy Synthes
3
Introduction
General Information Indications: The Electric Pen Drive (EPD) System is indicated for screw insertion, pin and wire placement, cutting of bone and metal, drilling, reaming, decorticating, shaping and smoothing of bones and teeth in a wide variety of surgical procedures, including but not limited to general orthopedic trauma, foot, hand, maxillofacial, neurosurgical, oral, otolaryngological, reconstructive and spine surgery. Safety Instructions The surgeon has to evaluate if the machine is suitable for an application based on several key factors, such as: • Power capability of the machine, attachment and cutting tool relative to the bone strength/anatomical situation • Handling of the machine, attachment and cutting tool relative to bone size In addition, contraindications of the implant must be considered and respected. If the Electric Pen Drive System is used in conjunction with an implant system make sure to consult the corresponding “Surgical Technique”. The Electric Pen Drive System is only to be used for patient treatment after careful consultation of the instructions for use. It is recommended that a backup EPD System is available to use during application, as technical problems can never be completely ruled out. The Electric Pen Drive System is designed for use by physicians and trained medical personnel. DO NOT use any component if damage is apparent. DO NOT use this equipment in presence of oxygen, nitrous oxide or a mixture consisting of flammable anesthetic and air. Never place the Electric Pen Drive in a magnetic environment since the machine might start unintentionally. To ensure proper operation of the tool, use only DePuy Synthes original accessories.
4
DePuy Synthes
Electric Pen Drive
User‘s Manual
Before first and every use, power tools and their accessories / attachments have to run through the complete reprocessing procedure. Protective covers and foils must be fully removed before sterilization. Check instruments for correct adjustment and functioning prior to every use. Always wear personal protective equipment (PPE) including safety goggles when handling with the Electric Pen Drive. To prevent overheating, always respect the duty cycles for each attachment listed on page 45. For the tool to function properly, DePuy Synthes recommends that it is cleaned and serviced after each use in accordance with the process recommended in the Cleaning and Disinfection section of this manual. It is mandatory to check used cutting tools after every use for wear and / or damage and to replace them if necessary. We recommend using new DePuy Synthes cutting tools for every surgery. Cutting tools must be cooled with irrigation liquid to prevent bone necrosis due to heat. The user of the product is responsible for proper use of the equipment during surgery. For important information regarding electromagnetic compatibility (EMC) please refer to the chapter “Electromagnetic Compatibility” in this manual. The tool is classified as type B against electrical shock and leakage current. The tool is suitable for use on patients in accordance with IEC 60601-1. This system requires regular maintenance service at least once a year, in order to maintain its functionality. This service has to be performed by the original manufacturer or an authorized site. Unusual Transmissible Pathogens Surgical patients identified as at-risk for Creutzfeldt-Jakob disease (CJD) and related infections should be treated with single-use instruments. Dispose of instruments used or suspected of use on a patient with CJD after surgery and / or follow current national recommendations.
Introduction
Precautions: • To avoid injuries, the locking mechanism of the tool has to be activated before every manipulation and before placing the tool back down, i.e. the mode switch has to be in the LOCK position ( ). • Should the machine drop on the floor and have visible defects, do not use it anymore and send it to the DePuy Synthes service center. • If a product drops on the floor, fragments may split off. This represents a danger for the patient and user as: – these fragments may be sharp – unsterile fragments may enter the sterile field or hit the patient Accessories / Scope of Delivery The main components of the Electric Pen Drive (EPD) System are the handpiece, hand switch, foot switch, console, electric cables as well as attachments and accessories. An overview of all components belonging to the Electric Pen Drive System can be found in the chapter “Ordering Information”. For using the EPD System the following components are a must: • Electric Pen Drive (05.001.010) • Hand Switch (05.001.012) or Foot Switch (05.001.016) with the Cable Foot Switch-Console (05.001.022) • Console (05.001.006 or 05.001.002) • Cable Electric Pen Drive – Console (05.001.021 or 05.001.025) • At least one attachment belonging to the system and a cutting tool fitting to the attachment. For an optimal function of the system, use only DePuy Synthes cutting tools. DePuy Synthes recommends the use of DePuy Synthes Graphic Cases and Washing Basket (68.001.800), which are designed specfically to sterilize and store the system.
such as cleaning brushes, DePuy Synthes Maintenance Oil for EPD and APD (05.001.095), Maintenance Spray (05.001.098), and a Maintenance Unit (05.001.099). No oils from other manufacturers may be used. Only DePuy Synthes oil may be used. Lubricants with other compositions can cause jamming, have a toxic effect, or have a negative impact on the sterilization results. Only lubricate the power tool and the attachments when clean. Locating of the Instrument or Fragments of Instruments DePuy Synthes instruments are designed and manufactured to perform within the scope of their intended use. However, if a Power Tool or accessory/ attachment breaks during use, a visual inspection or a medical imaging device (e.g. CT, Radiation Devices, etc.) can aid in locating the fragments and/or components of the instrument. Storage and Transport Only use the original packaging for dispatch and transport. If the packing material is no longer available, please contact the DePuy Synthes office. For storage and transport environmental conditions, see page 44. Warranty / Liability The warranty for the tools and accessories does not cover damage of any kind resulting from wear, improper use, improper reprocessing and maintenance, damaged seal, use of non-DePuy Synthes cutting tools and lubricants or improper storage and transport. The manufacturer excludes liability for damage resulting from repairs or maintenance carried out by unauthorized sites. For further information on the warranty please contact your local DePuy Synthes office.
For care and maintenance, special tools are available,
User‘s Manual
Electric Pen Drive
DePuy Synthes
5
Introduction
Explanation of Symbols Do not reuse
Caution
Products intended for single use must not be reused. Consult Instructions for Use
Reuse or reprocessing (e.g. cleaning and resterilization) may compromise the structural integrity of the device and / or lead to device failure which may result in patient injury, illness or death. Furthermore, reuse or reprocessing of single use devices may create a risk of contamination e.g. due to the transmission of infectious material from one patient to another. This could result in injury or death of the patient or user.
Indicates that the device complies with the classification type B against electrical shock and electrical leakage current. The device is suitable for use on patients as defined by IEC 60601-1 Do not immerse device in liquids
DePuy Synthes does not recommend reprocessing contaminated products. Any DePuy Synthes product that has been contaminated by blood, tissue and / or bodily fluids / matter should never be used again and should be handled according to hospital protocol. Even though they may appear undamaged, the products may have small defects and internal stress patterns that may cause material fatigue.
calibrate The marked device shall be re-calibrated annually 10PB
Product is UL classified to the requirements of both the United States and Canada
0123
The device meets the requirements of directive 93/42/EEC for medical devices. It is authorized by an independent notified body for which it bears the CE symbol.
IP X4
Ingress protection rating according to IEC 60529 Locked symbol. Drive Unit is off for safety
The European directive 2012/19/EC on waste electrical and electronic equipment (WEEE) applies to this device. This device contains materials that should be disposed of in accordance with environment protection requirements. Please observe national and local regulations.
Manufacturer
Indicates Environment Friendly Use Period of 10 years in China
Temperature limit
Indicates Environment Friendly Use Period of 50 years in China Humidity limitation
Date of manufacture Non-sterile
Catalog Number Lot number Serial number Packaging unit ISO 7000-2794 (2009-02)
Atmospheric pressure limitation Use-by-date Do not use if package is damaged 6
DePuy Synthes
Electric Pen Drive
User‘s Manual
INMETRO Ord. 350 certified
Consoles
Standard and Basic Consoles The Electric Pen Drive (05.001.010) can be supplied with the Standard Console (05.001.006), or the Basic Console without irrigation (05.001.002).
1
2
6
7
11
4
Standard Console with Irrigation, without Torque Limiting Function (05.001.006) 1.
Slide control for adjusting the maximum speed for 1 Electric Pen Drive ( )1
2. Slide control for adjusting the maximum speed for a second Electric Pen Drive 2 4. Selection switch for irrigation 5. Connection for Adapter for Small Battery Drive*( 6. Connection for Electric Pen Drive Electric Drive**( )1
and Small
7.
and Small
Connection for Electric Pen Drive Electric Drive** 2
)
5
8
9
10
8. Connection for Foot Switch 9. Twist lock for irrigation pump 10. Adjustment knob for irrigation flow rate 11. LED Irrigation Ready (Activation with Foot Switch) 1. Basic Console (05.001.002) 1.
Slide control for adjustment of the maximum speed for 1
1
5. Connection for Adapter for Small Battery Drive* 6. Connection for Electric Pen Drive Electric Drive ** 1 8. Connection for Foot Switch (
and Small )
5
6
8
*This connection is not of use anymore since the Adapter for Small Battery Drive is no longer available. **The Small Electric Drive is no longer available for purchase.
User‘s Manual
Electric Pen Drive
DePuy Synthes
7
Consoles
Set-up of Consoles Color Coding Set A color coding set is supplied with every console. Each set contains 3 text stickers, 8 silicone rings in 3 colors (red, blue and white) and 8 stickers of the colored rings. The stickers and silicone rings can be used to indicate on the console and the cables how the system has to be assembled.
Set-up Before the initial operation of the device, make sure that the power switch (12) is set to position 0 (OFF). Only connect the console to the power supply with the cable provided via the power plug socket (13) and set the power switch to position 1 (ON). The LED marked with on the front of the console signals the proper operation of the console. If the LED flashes, the console must be sent in for maintenance. A potential equilization connector (14) for an equipotential earth plug is integrated in the console. Existing equipotential earth plugs can be connected there. (15) Fuses: 2x3AF/250V, Breaking Capacity 1500A
8
DePuy Synthes
Electric Pen Drive
User‘s Manual
14 12 15 13
Consoles
Mounting Cables on Consoles To mount the respective cables on the consoles, position the nose on the plugs flush with the groove on the plug sockets and insert the plugs. For improved orientation, red dots are provided on the plug and socket that must face upward when connecting the cable. Removing Cables To remove the plug, grasp the release sleeve, pull back and remove. The connections 5–8 (page 8) are used to connect the following devices: 5
Connection for a Small Battery Drive handpiece*
6&7
Connection for an Electric Pen Drive and Small Electric Drive**
8
Connection for the Foot Switch
Connections not used can be sealed off with the protective caps provided. Precautions: • Do not lay textiles or objects under the console. These can be sucked up and block the ventilation inlet. • Do not block the ventilation opening on the back of the console with any object. • Only place consoles on smooth and flat surfaces. • Do not place the console in the sterile field. • Do not hang the irrigation liquid directly above the console to prevent liquid from dripping on the console. • Do not pull the cable! Always activate the release sleeve. • The use of HF (= high frequency) equipment for tissue coagulation can cause electromagnetic interferences – in this case cables should be separated as far as possible. • Ensure that the power cord can always be disconnected immediately from mains supply.
* This connection is not of use anymore since the Adapter for Small Battery Drive is no longer available. **The Small Electric Drive is no longer available for purchase.
User‘s Manual
Electric Pen Drive
DePuy Synthes
9
Consoles
Speed Regulation Functions of the Standard Console (05.001.006) and the Basic Console (05.001.002) Adjustment of the maximum speed (1, 2) Speed is automatically optimized and adjusted for each attachment; for certain attachments it is however recommended that you reduce the maximum speed of the connected handpiece. Adjustment can be performed in increments of 25 % with the slide controls for adjustment of maximum speed. It is recommended using this feature for high speed burrs. A letter code on each burr indicates the maximum speed specified by DePuy Synthes. Marking
Console Setting Pen 60k
A
100%
B
100%
C
75%
D
50%
E
25%
This is also explained by symbols on the supplied adhesive foil. The adhesive foil can be applied to the console as a reference aid.
11
DePuy Synthes
Electric Pen Drive
User‘s Manual
Consoles
Irrigation Function of the Standard Console (05.001.006)
4.2 4.3
4.1
4.5
4.4
The function of the Standard Consoles described in the following is not available on the Basic Console. Irrigation (4) With the integrated irrigation system, tools can be cooled to prevent tissue necrosis affected by excess heat. Irrigation nozzles for the attachments and sterile irrigation tubes are available for the irrigation system. In addition to the OFF position (position 4.1), there are two operating positions for Pen 1 and Pen 2: constant and variable irrigation. If constant irrigation for Pen 1/Pen 2 is chosen, a constant quantity of irrigation liquid will be released for Pen 1/Pen 2. A constant flow rate between 10 and 100 ml/min (0.34 and 3.4 oz/min) can be adjusted with the adjustment knob (10) for the irrigation flow rate (position 4.3/4.5).
4.1 Irrigation OFF 4.2 Variable irrigation for Pen 1 VAR 4.3 Constant irrigation for Pen 1 CON 4.4 Variable irrigation for Pen 2 VAR 4.5 Constant irrigation for Pen 2 CON
11
If variable irrigation for Pen 1/Pen 2 is chosen, the flow rate will be directly proportional to the speed selected on the Hand or Foot Switch, i.e. the higher the selected speed, the higher is the irrigation flow rate. The maximum flow rate can be set between 10 and 100 ml/min (0.34 and 3.4 oz/min) with the adjustment knob (10) for the irrigation flow rate (position 4.2/4.4). The LED lights up if the irrigation is activated (11) on the foot switch. Please refer to the chapter of the Foot Switch for a detailed explanation on how to activate or deactivate the irrigation.
User‘s Manual
Electric Pen Drive
DePuy Synthes
11
Consoles
Irrigation Tube Set (05.001.178.01S) and Clips for Irrigation Tube Set (05.001.179.05S), Inserting Irrigation Tube 1. Remove the sterile Irrigation Tube (Fig. 1) from the sterile package. 2. Secure the attachment-specific Irrigation Nozzle on the attachment (Fig. 2) used. 3. Fit the irrigation tube on the Irrigation Nozzle used first and then fix it with the clips (Fig 3) on the cable. 4. Route the Irrigation Tube end into the non-sterile area and open the twist lock for the irrigation pump in the direction of the arrow (see Fig. 4).
Fig. 1
5. Insert the Irrigation Tube in the pump in accordance with the marking (see Fig. 4) and close the twist lock. 6. Remove the protective cap from the cannula and connect cannula to irrigation bag. When doing so, make sure that the connection nipple of the cannula is not touched by non-sterile persons while securing irrigation nozzle to the attachment. The vent of the spike has to be open when the irrigation is used.
Fig. 2
Push the Irrigation Nozzles for the Burr Attachments and Reciprocating Saw Attachment as far as necessary over the attachment from the front. Fig. 3
Push the Irrigation Nozzle for the Sagittal Saw Attachment onto the attachment from the rear (from the attachment coupling side) before mounting the attachment on the handpiece.
Fig. 4
11
DePuy Synthes
Electric Pen Drive
User‘s Manual
Electric Pen Drive System
Electric Pen Drive 60,000 rpm (05.001.010) 1. 2. 3. 4. 5. 6. 7.
Adjustment sleeve Release sleeve for attachment Cable IN / OUT position LOCK position Forward position (clockwise) Reverse position counterclockwise Lock slide for adjustment sleeve
5
2
Turning the Adjustment Sleeve To avoid an unintentional change of the operating mode the lock slide for adjustment sleeve (7) locks the adjustment sleeve automatically. To be able to move the adjustment sleeve the lock slide has to be pushed in the direction indicated by the arrow. After having reached the desired position, release the lock slide (7) and the adjustment sleeve (1) is locked in the desired position.
3
6
1
4
Mounting Cable on Handpiece Move the adjustment sleeve (1) on the handpiece into the position cable IN / OUT (3). Align the groove on the plug with the notch in the sleeve and insert the plug. Turn the adjustment sleeve into the LOCK position (4). The cable is now firmly locked to the handpiece and the handpiece is in locked mode. To remove it again, move the adjustment sleeve (1) into the cable IN / OUT position (3) and remove the cable. FWD / REV Switchover By turning the adjustment sleeve into the forward position (5), the handpiece can be switched into a clockwise operating direction. In the reverse position (6) the handpiece can be operated in a counterclockwise direction. In addition to locking the cable, the LOCK position (4) is used for safety switch-off when changing attachments and tools to prevent accidental start-up of the unit. For instructions on mounting the attachments see page 18. Either a Hand Switch (page 15) or a Foot Switch (pages 16/17) can be used for speed control.
7 Precautions: • Do not place the handpiece on magnetic covers or in the immediate vicinity of other magnetic objects. This can activate the handpiece. • When two Electric Pen Drive handpieces are connected and the speed is controlled with the Foot Switch, one handpiece must be switched to LOCK . Otherwise both handpieces will be blocked for safety reasons. • In all other cases, the first activated device functions. As long as this device is activated, all others are deactivated. • HF (= high frequency) equipment for tissue coagulation can cause electromagnetic interferences and inadvertently activate the Electric Pen Drive. Cables should be separated as far as possible.
User‘s Manual
Electric Pen Drive
DePuy Synthes
11
Electric Pen Drive System
Hand Switch (05.001.012) 1. 2. 3. 4. 5.
Positioning arrow Pull-out finger rest Positioning arrow Guide groove Locking switch
5 1
Mounting the Hand Switch on the Handpiece (05.001.010) Position the Hand Switch on the handpiece so that both positioning arrows (1) of the Hand Switch cover the positioning arrows (3) over the guide grooves (4) of the pen. Then press downward vertically until the Hand Switch clicks into position. Removal To remove the Hand Switch, grasp the lever and pull it upwards, away from the handpiece. Operation The length of the Hand Switch can be individually adjusted with the pull-out finger rest (2). The speed can be continuously adjusted by operating the Hand Switch. The Hand Switch can be deactivated (LOCK position) or activated (ON position) with the locking switch (5).
11
DePuy Synthes
Electric Pen Drive
User‘s Manual
3
4
2
Electric Pen Drive System
Foot Switch, 1 Pedal (05.001.016) Connecting the Foot Switch to a Console The Foot Switch can be connected to the console with the Foot Switch Cable (05.001.022). To insert the plugs, align the red dots on the plugs with those on the sockets and insert the plug. A second foot pedal can be connected using the second socket on the back of the foot pedal. Use a foot pedal cable (05.001.022) to connect the two foot pedals. Note: this cannot be done with the basic console.
1
2
The second socket is covered with a protective cap on delivery; the cap may be removed if necessary. Disconnecting a Foot Switch Grasp the respective plug by the release sleeve, pull back and remove.
1 Irrigation ON / OFF button 2 Pedal
Operation The speed can be continuously adjusted with the pedal (2). Briefly pressing the irrigation button (1) connects or disconnects the irrigation. In the ON position, the setting preselected with the irrigation selection switch on the console is activated. The LED lights up if the irrigation is activated on the foot switch. If the user holds this button down, the irrigation quantity set with the irrigation flow-rate adjustment knob (page 8) is supplied continuously until the button is released. This function is independent of the position selected on the irrigation selection switch (page 8) and of the activation of the pedal (2) or Hand Switch (page 15). Precaution: When the cable is not connected properly to the foot switch, it is possible that the handpiece is activated without pressing the foot switch.
User‘s Manual
Electric Pen Drive
DePuy Synthes
11
Electric Pen Drive System
Foot Switch, 2 Pedals (05.001.017) Connecting the Foot Switch to a Console The Foot Switch can be connected to the console with the Foot Switch Cable (05.001.022). To insert the plugs, align the red dots on the plugs with those on the sockets and insert the plug. A second foot pedal can be connected using the second socket on the back of the foot pedal. Use a foot pedal cable (05.001.022) to connect the two foot pedals. Note: this cannot be done with the basic console.
3 2
The second socket is covered with a protective cap on delivery; the cap may be removed if necessary
1
Disconnecting the Foot Switch Grasp the respective plug by the release sleeve, pull back and remove. Operation Operating mode is forward FWD (1) when activating the right pedal and reverse REV (2) for the left pedal. The speed can be continuously adjusted with the pedals (1 and 2). Briefly pressing the irrigation button (3) connects or disconnects the irrigation. In the ON position the setting preselected with the irrigation selection switch on the console is activated. The LED lights up if the irrigation is activated on the foot switch. If the user holds this button down, the irrigation quantity set with the irrigation flowrate adjustment knob (page 8) is supplied continuously until the button is released. This function is independent of the position selected on the irrigation selection switch (page 8) and of the activation of the pedal (1 and 2). Precautions: • When operating the Electric Pen Drive with the Foot Switch, the operating mode is defined by the pedal used on the Foot Switch (i.e. right pedal for Forward mode and left pedal for Reverse mode) regardless of the operating mode (FWD or REV), defined by the adjustment sleeve of the handpiece. • When the cable is not connected properly to the foot switch, it is possible that the handpiece is activated without pressing the foot switch.
11
DePuy Synthes
Electric Pen Drive
User‘s Manual
1 FWD 2 REV 3 Irrigation ON / OFF button
Attachments
General Information Mounting the Attachments on the Handpieces (05.001.010) The attachments can be connected in 8 different positions (45° increments). To mount, turn the release sleeve for the attachments clockwise (see arrow on the release sleeve) until it engages. The release sleeve protrudes slightly from the black section of the handpiece towards the front. Insert the attachment into the attachment coupling from the front and press it lightly against the handpiece. The attachment automatically engages. If the release sleeve accidentally closes, turn the attachment clockwise while applying slight pressure against the handpiece until it engages without holding the release sleeve in place, or repeat the entire connection process. Check secure holding of the attachment on the handpiece by pulling on the attachment.
Release sleeve
Removing the Attachments from the Handpiece Turn the release sleeve for attachments clockwise until it disengages. Hold the attachment upward while doing so, then remove the attachment. Attachments and Accessories For easier change of burrs, without having the Burr Attachment or Craniotome Attachment connected to the handpiece, the Handle for Change of Instruments (05.001.074) can be used. Warranty: Only use DePuy Synthes saw blades, burrs and rasps while working with the Electric Pen Drive attachments. The use of other tools voids the device warranty.
User‘s Manual
Electric Pen Drive
DePuy Synthes
11
Attachments
Drill Attachments Speed Drill Attachments (05.001.030-05.001.032, 05.001.044): approx. 1,800 rpm The system includes straight Drill Attachments with Mini Quick Coupling, J-Latch Coupling and AO / ASIF Coupling, as well as a 45° Drill Attachment with AO/ASIF Coupling (see pg. 52).
1
The 45° Drill Attachment with AO / ASIF Coupling has a cannulation of 1.6 mm, which permits the use of this attachment for drilling and reaming over Kirschner Wire (e.g. for cannulated screws and for cup and cone technique).*
1
1
Mounting and Removing Tools Lock unit. Pull back the release sleeve (1) and insert / remove the tool. 1 Release sleeve
Drill Attachment 45°, Cannulated, with Jacobs Chuck (05.001.120) Speed drill attachments: approx. 1,800 rpm Clamping area: 0.5 mm–4.7 mm The cannulation of 1.6 mm permits the use of this attachment for drilling and reaming over Kirschner Wire (e.g. for cannulated screws and for cup and cone technique). Mounting and Removing Tools Lock unit. Open the chuck with the key provided (310.93K) or by hand by turning the two moving parts clockwise with respect to each other. Insert / remove the tool. Close the chuck by turning the moving parts counterclockwise and tighten it by turning the key clockwise.
* The cannulation was implemented from batch number 9232779.
11
DePuy Synthes
Electric Pen Drive
User‘s Manual
Attachments
1
Drill Attachment 90°, Short (05.001.035) and Long (05.001.036) with Mini Quick Coupling Speed: approx. 1,800 rpm Due to their very small angled head the 90° Drill Attachments enable good visibility during operations with narrow access (e.g. intra-oral, shoulder, etc.).
1
Mounting and Removing Tools Lock unit. Move the slide (1) to the side following the arrow on the slide and insert / remove the tool. To secure the tool, push the slide back again.
1 Slide
Oscillating Drill Attachment (05.001.033) Frequency: approx. 3,200 osc/min The oscillating drilling movement of the Oscillating Drill Attachment prevents tissue and nerves from wrapping around the drill. This can considerably improve the operating results. Mounting and Removing Tools
1
1 Release sleeve
Tools with a Mini Quick Coupling can be clamped with the Oscillating Drill Attachment. To do this, lock the unit, pull back the release sleeve and insert / remove the tool. Precaution: The handpiece must be in FWD position to use the oscillating drill attachment.
User‘s Manual
Electric Pen Drive
DePuy Synthes
11
Attachments
Screw Attachments Screw Attachments (05.001.028, 05.001.029, 05.001.034) Speed: approx. 400 rpm The system includes screw attachments with AO Coupling, Hexagonal and Mini Quick Coupling.
1 Release sleeve
Mounting and Removing Tools Lock unit. Pull back the release sleeve and insert / remove the tool. Precaution: Always use an appropriate Torque Limiting Device while inserting locking screws into a locking plate.
Kirschner Wire Attachment Kirschner Wire Attachment (05.001.037) 1
Speed: approx. 2,700 rpm With the Kirschner Wire Attachment, Kirschner Wires of any length with a diameter of 0.6 mm–1.6 mm can be tensioned. The tensioning lever (1) can be rotated by 300°, permitting individual adjustment (suitable for left and right-handed users). Mounting and Removing Kirschner Wires Lock unit. To insert and remove Kirschner Wires, press the tensioning lever (1). After the lever is released, the Kirschner Wire is automatically tensioned. To re-grasp, press the tensioning lever, pull back the unit along the Kirschner Wire and then release the tensioning lever again.
22
DePuy Synthes
Electric Pen Drive
User‘s Manual
1 Tensioning lever
1