Pocket Guide
Evita® V800 / Evita® V600 Software version 1.n
02 |
QUICK GUIDE EVITA® V800 / EVITA® V600
Content Product description
04
Operating concept
08
Getting started
12
Operation
16
Alarms
20
Monitoring
22
The Evita® V800 / V600 Pocket Guide is not a replacement or substitute for the Instructions for Use. It is for informational purposes only . Any use of the device requires full understanding and strict observation of the Instruction for Use. The availability of functionality depends on the country and availability. Options may not be included in a present device or may come at extra costs. The pocket guide may not reflect the latest software version of the device.
| 03
Intended use The intensive care ventilator Evita is intended for the ventilation of adults, adolescents, children, infants, and neonates. This device provides mandatory ventilation modes and ventilation modes for supporting spontaneous breathing as well as ventilation monitoring.
04 |
QUICK GUIDE EVITA® V800 / EVITA® V600
Product description Intensive care Ventilator
Display unit
Ventilation unit
Gas supply unit GS500 Power supply unit PS500 Trolley with 4 brakes
Evita® V800
| 05
Product description Display unit, front
Alarm bar
ON/OFF switch
Indicator for power and batteries
Touch screen
Rotary knob (touch turn confirm)
Alarm silence 2 minutes
06 |
QUICK GUIDE EVITA® V800 / EVITA® V600
Product description Display unit, rear
COM ports
USB ports
HDMI port
USB ports
USB port
LAN ports
Swiveling of display unit
| 07
Product description Ventilation unit, front
Flap
Operation display of ventilation Expiration valve
Expiratory flow sensor
Inspiration valve
Nebulizer port
Emergency air intake
Water trap
08 |
QUICK GUIDE EVITA® V800 / EVITA® V600
Operating concept Screen, general structure
Header bar Monitoring area Main menu bar
Therapy bar
| 09
Operating concept Main screen
Patient category
Ventilation mode / Therapy type
Power supply / charge status indicator of battery
Alarm messages
Button to open other modes
Buttons to open the - Alarms dialog - Views dialog - Trends/data dialog - Sensors dialog - Procedures dialog - System set up dialog
Button to - Start/Stop ventilation - Start/Stop O2 therapy
10 |
QUICK GUIDE EVITA® V800 / EVITA® V600
Operating concept Using the screen − Colours Alarms The following colors are used to identify the priority of alarms: Colour
Alarm priority
Action required
Red
High !!!
Immediate action is necessary in order to avert an acute danger
Yellow
Medium !!
Prompt action is necessary in order to avert a danger
Turquoise
Low !
Attention is necessary
| 11
Operating concept Using the screen − Colours Rotary knob If the device is turned on, the rotary knob is back lit with a color. Orange
The selected function or setting must be confirmed by pressing the rotary knob. 10 seconds before the time during which the confirmation may be performed ends, the background lighting flashes.
Blue
After the device is turned on, the background lighting shines for 2 seconds.
Touch – Turn – Confirm Selecting and setting parameters and functions Touch the control element. The control element turns orange.
Turn the rotary knob to the right or to the left to make the setting. The value is displayed in the control element.
Confirm by pressing the rotary knob. The setting is applied, and the control element turns gray or blue again
12 |
QUICK GUIDE EVITA® V800 / EVITA® V600
Getting started Turning on the device Prerequisites:
The accessories needed for operations on the patient are connected.
Ventilation unit, left side: Main switch
The mains power supply and the gas supply are connected. The main switch is turned on.
1. Press the ON/OFF key on the display unit. The system is started. The start dialog is displayed.
2. Touch the New patient button if a new patient is admitted or
1.
2.
3.
3. Touch the Current patient button to continue with existing settings.
| 13
Getting started System test Starting the system test 1. Touch the Start button in the System test line.
2. Confirm with the rotary knob. 3. Answer the questions in the information field by touching the Yes or No button.
After a successful system test has been performed, a message appears for continuing the breathing circuit test. Proceed with 'Yes'.
1.
3.
2.
Breathing circuit test Performing the Breathing circuit test 1. Touch the button in the Breathing circuit test line. 2. Select the breathing circuit used from the pick list and the type of humidification.
3. Confirm with the rotary knob.
4. Follow the instructions in the information field and touch the button OK when asked to do so.
1. 2.
4.
1.
14 |
QUICK GUIDE EVITA® V800 / EVITA® V600
Getting started Starting the therapy (for a new adult or paediatricpatient based on the patient height) 1. Touch the button for the relevant patient category.
2. Confirm with the rotary knob. 3. Touch the button for height.
4. Set the body height by turning the rotary knob and push to confirm.
1.
5.
3.
5. Touch the button for the relevant therapy type (Tube, NIV, O2 therapy).
7.
6. Confirm with the rotary knob.
2./4./6./8.
7. Check the alarm limits and touch the button Start ventilation.
8. Confirm with the rotary knob.
Starting the therapy (for a new adult or paediatric patient based on the patient category) 1. Touch the button for the relevant patient category.
3. Confirm with the rotary knob. 4. Check the therapy settings (ventilation mode, ventilation parameter, alarm limits) and touch the button Start ventilation.
5. Confirm with the rotary knob.
1. 2.
2. Touch the button for the relevant therapy type (Tube, NIV, O2 therapy).
4. 3./5.
| 15
Getting started Therapy Quick start ventilation In an emergency situation, therapy can be started immediately without operational readiness being checked. 1. Touch the button Quick start ventilation. The Patient and therapy step is displayed.
2. Press the rotary knob. The therapy is started. The main screen is displayed.
1. 2.
16 |
QUICK GUIDE EVITA® V800 / EVITA® V600
Operation Changing the ventilation mode 1. Touch the relevant tab.
2. The colour of the tab turns orange (if necessary, adjust the ventilation parameter).
3. Confirm with the rotary knob. The color of the tab turns white.
1./2. 3.
Setting ventilation parameters
1.
1. Touch the corresponding therapy control (colour turns orange). 2. Set the value by turning the rotary knob and push to confirm.
2.
| 17
Operation Advanced settings 1. Touch the button Advanced settings. The advanced settings of the selected ventilation mode are displayed and can be configured.
1.
2. The current settings are displayed in an information field next to the button. 2.
Changing the therapy type during therapy
3. Confirm with the rotary knob. The device is in standby mode.
4. Start the therapy by touching the Start ventilation button.
4.
2. Touch the button of the relevant therapy type – Tube – NIV – O2 therapy
1.
2.
1. Touch the field for the therapy type in the header bar.
3.
18 |
QUICK GUIDE EVITA® V800 / EVITA® V600
Operation Non-invasive ventilation (NIV) In the Adult and Pediatric patient patient categories, all the ventilation modes can be selected. In the Neonate patient category, only the SPN-CPAP and PC-CMV ventilation modes can be selected.
O2 therapy Patient-side accessories such as oxygen masks or nasal cannulas can be used for O2 therapy. During O2 therapy, only the O2 concentration, the inspiratory flow, and the inspiratory pressure are monitored. The alarm limits for the parameters VT, MVe, RR, Paw, and Tapn are not active. The alarm limits for O2 monitoring are automatically set by the device.
| 19
Operation Interrupting therapy 1. Touch the button Stop ventilation in the therapy bar. A pop-up window opens. 2. Touch the button Stop ventilation. 3. Confirm with the rotary knob. 4. The following message is displayed in the header bar: Standby mode activated
1. 4.
5. Touch the Alarm reset button in the header bar.
2.
6. Confirm with the rotary knob.
5.
3./6.
29 |
QUICK GUIDE EVITA® V800 / EVITA® V600
Alarms Display of alarms – Optical alarm signals Alarm bar Alarm message field More alarms Blinking parameter field
Display of alarms – Acoustic alarm signals The primary acoustic alarm system produces different alarm signals. The acoustic alarm signal can be suppressed for 2 minutes maximum.
1.
1. Press the alarm silence key.