Dräger Medical

Fabius Family Disassembly and Assembly Guide Jan 2013

Guide

1 Page

D-58067-2012  Dräger Fabius Family Compact Breathing System Disassembly/Assembly  Disassembly  Remove inspiratory and expiratory valve.  Remove APL-valve.  D-58140-2012  D-58129-2012  Attach flow-sensor guard. Insert flow sensor, insert and tighten expiratory port.  Mount APL-valve.  Mount inspiratory and expiratory valve.  Remove hook for gas sample line (optional).  Remove cover of ventilator chamber, remove ventilator diaphragm.  Remove waste-gas port.  D-58178-2012  D-58147-2012  D-58143-2012  Remove cleaning disc from the adapter. Close adapter until it snaps.  Make sure all seals are in place. Screw CLIC adapter (optional) or absorber canister.  Loosen and pull out expiratory port; pull out flow sensor. Remove flow-sensor guard.  D-58124-2012  D-58106-2012  D-58104-2012  D-58098-2012  Mount the breathing system, connect all hoses, cables and the absorber.  Mount hook for gas sample line (optional).  Insert cleaning disc into the adapter. Close adapter until it snaps.  Unscrew CLIC adapter (optional) or absorber canister. Visually inspect the sealing diaphragm of the CLIC adapter.  Mount waste-gas port.  Insert ventilator diaphragm and mount cover of ventilator chamber.  Assembly  Reprocessing Instruction* Sterilization 134 °C (273.2 °F)  *No replacement or substitute for the instructions for use.  Machine cleaning and disinfection 93 °C (200 °F)  Immersion Disinfection  Recommendation: Position components so that all interior spaces and surfaces are completely flushed and water can drain off freely.  Breathing System: After the machine cleaning and disinfection, a hot steam sterilization is required to dry the breathing system. Valves, ventilator cover, and AGS sleeve: Make sure that the cleaning and rinsing liquids can flow in the opening direction of the valve. Remove any water which may have accumulated in the ventilator diaphragm. For further instructions and explanations please see the respective instructions for use.  D-62063-2012 & 58163 & 41883 & 46343  01.13-1 | Marketing Communications | PP | PR | LE | Printed in Germany | Chlorine-free – environmentally compatible | Subject to modifications | © 2013 Drägerwerk AG & Co. KGaA  Remove all hoses, cables and the absorber, lift off the breathing system.
File Type: PDF
File Size: 1.46 MB
File Name: Dräger Medical - Fabius Family Disassembly and Assembly Guide Jan 2013 - 2013-01.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions