Technical Documentation Fabius MRI Anesthesia System
WARNING! Each servicing and/or testing of the device requires full understanding of this Technical Documentation. Carefully read this Technical Documentation and any applicable Instructions for Use prior to any use of the device.
Revision 2.0 5330.660 9036337
Inhaltsverzeichnis General 1
Symbols and Definitions
6
2
Notes
6
1
Cautions and Warnings
8
1.1
Patient safety ... 8
1.2
Warnings ... 8
Function Description 1
Abbreviations
12
2
General
12
3
Basics of magnetic resonance tomography
12
4
Safety instructions
12
5
General Information about the Fabius MRI
14
6
Battery backup
17
7
Fabius MRI Piping Diagram
18
8
Function description of the gas box
18
9
SORC (Sensitive Oxygen Ratio Controller)
19
10
Cosy 2.6 breathing system
21
10.1 Ventilation mode ... 25
Copyright reserved.
K5330660IVZ.fm 09.01.08
10.2 Manual ventilation ... 26 10.3 Spontaneous breathing ... 27 10.4 Volume/pressure control ventilation mode ... 29 10.5 Cosy 2.6 absorber ... 31 3
Inhaltsverzeichnis 11
Lung ventilator
31
11.1 Pressure limiting valve ... 34 11.2 Auxiliary-air valve ... 34 12
Pneumatic assembly
35
12.1 PEEP/Pmax valve control ... 35 12.2 APL bypass valve control ... 36 13
Control PCB
36
14
Function Description: Control PCB
37
15
Control panel assembly
38
16
FiO2 measurement
40
17
Respiratory flow measurement
41
18
Gas flow rate measurement
42
19
Anesthetic vaporizer(s)
43
Maintenance Procedures 1
Diagnostics
46
Annex Parts catalog
4
K5330660IVZ.fm 09.01.08
Copyright reserved.
Test List
General
5
General 1
Symbols and Definitions
Fabius MRI
WARNING A WARNING statement provides important information about a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION A CAUTION statement provides important information about a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or in damage to the equipment or other property.
NOTE A NOTE provides additional information intended to avoid inconvenience during operation or servicing of the equipment.
Definitions:
2
Notes
Inspection
= examination of actual condition
Maintenance
= measures to maintain specified condition
Repair
= measures to restore specified condition
Servicing
= inspection, maintenance, and repair
This Technical Documentation conforms to the IEC 60601-1 standard. Read each step in every procedure thoroughly before beginning any test. Always use the proper tools and specified test equipment. If you deviate from the instructions and/or recommendations in this Technical Documentation, the equipment may operate improperly or unsafely, or the equipment could be damaged. Dräger recommends that only Dräger supplied repair parts be used for maintenance. Otherwise the correct functioning of the device may be compromised.
NOTE Unless otherwise stated, reference is made to laws, regulations or standards (as amended) applicable in the Federal Republic of Germany for equipment used or serviced in Germany. Users or technicians in all other countries must verify compliance with local laws or applicable international standards.
6
5330.660
Copyright reserved.
This Technical Documentation is for the purpose of information only. Product descriptions found in this Technical Documentation are in no way a substitute for reading and studying the Instructions for Use.
4.0_Printed on_09.01.08_General_Technical_Documentation.fm
The maintenance procedures described in this Technical Documentation may be performed by properly trained service personnel only. These maintenance procedures do not replace inspections and servicing by the manufacturer.
Fabius MRI
General
Copyright reserved.
4.0_Printed on_09.01.08_General_Technical_Documentation.fm
NOTE If the test values are not met, please contact your local service organization.
5330.660
7
General
Fabius MRI
1
Cautions and Warnings
1.1
Patient safety
The design of the medical device, the accompanying literature, and the labeling on the medical device take into consideration that the purchase and use of the medical device are restricted to trained professionals, and that certain inherent characteristics of the medical device are known to the trained operator.
1.2
Warnings
The following WARNINGS and CAUTIONS apply to general operation of the device. WARNINGS and CAUTIONS specific to subsystems or particular features appear with those topics in later sections of the manual. WARNING Any person involved with the setup, operation, or maintenance of the Fabius MRI anesthesia system must be thoroughly familiar with the instruction manual.
CAUTION Only the accessories indicated on the list of accessories 8607185 en (1st edition or higher) have been tested and approved to be used with the medical device. Accordingly it is strongly recommended that only these accessories be used in conjunction with the specific medical device. Otherwise the correct functioning of the medical device may be compromised.
WARNING This MR conditional anesthesia machine has been tested with magnets with field strengths of 1.5 tesla and 3 tesla by a fringe field strength of 40 mtesla. Use of the machine with higher strengths could result in ventilator and device malfunction. Additionally, unmanageable attractive forces could lead to serious injury.
WARNING Always lock the caster brakes after the Fabius MRI has been positioned in the MRI scanner room. Magnetic attractive forces between the magnet and the anesthesia machine may cause unintentional movement of the anesthesia machine if the casters are unlocked.
8
5330.660
Copyright reserved.
CAUTION Only the combinations approved by Dräger Medical, with monitoring may be used. Otherwise the correct functioning of the device maybe compromised.
1.0_Printed on_09.01.08_L5330660_Warnhinweise.fm
CAUTION No third-party components shall be attached to the anesthesia machine, ventilator, or breathing system (except for certain approved components), otherwise the correct functioning of the medical device may be compromised. For more information, contact DrägerService or your local authorized service organization.
Fabius MRI
General
WARNING Do not place any object on this machine unless it is specifically labeled to be used in an MR scanning room and on a Fabius MRI anesthesia system. Objects placed on this machine that are not designed for use with this anesthesia system may be strongly attracted to the magnet and may cause serious injury or death when the machine is used in an MR scanning room.
WARNING Do not bring any ferromagnetic tools or equipment into the scanning room. Ferromagnetic objects (made of steel, iron, or stainless steel) are strongly attracted to the magnet and can become harmful projectiles.
WARNING Be careful in handling the power cord and main power plug. These parts still contain minor magnetic components. The power cord can be attracted to MRI system.
WARNING The Fabius MRI and its patient connections must be carefully positioned so that the patient cannot be disconnected when being removed from the MRT system.
CAUTION Do not use any type of Desflurane vaporizer in the MR environment. In an MR environment functionality of the Desflurane vaporizer will be compromised.
Copyright reserved.
1.0_Printed on_09.01.08_L5330660_Warnhinweise.fm
CAUTION Only Vapor 2000 vaporizers can by used on the Fabius MRI in MRT scanner rooms.
5330.660
9
10
5330.660
1.0_Printed on_09.01.08_L5330660_Warnhinweise.fm
Copyright reserved.
General Fabius MRI
Function Description
11
MRI 1
Fabius MRI Abbreviations
MRI →Magnetic Resonance Imaging MRT →Magnetic Resonance Tomography Tesla (T) →Magnetic flux density/induction HF →High Frequency
2
General
Fabius MRI is a variant of the Fabius GS/Tiro which operates in a MRI environment.
3
Basics of magnetic resonance tomography
Magnetic resonance (MR) or nuclear magnetic resonance imaging is a diagnostic technique which produces high-resolution pictures (images) of the human body without the use of dangerous X-rays. Signals are generated and received using a strong magnet and a radio-frequency antenna. The resulting images are evaluated and displayed by a computer.
4
Safety instructions
As the MR scanner generates very strong magnetic fields, special safety precautions must be taken. Strong, high-frequency magnetic fields can heat metal. There is a risk of burns caused by metal objects on the person or implants. Ferromagnetic (metal) objects can also be accelerated with great force and could cause injury. The data content of cards featuring magnetic strips (such as credit cards) may be wiped.
–
Coins
–
Key
–
Watches
–
Items of jewelry
–
Tools
–
Cards with magnetic strips
Copyright reserved.
Persons with implants should consult a specialist doctor before entering the room.
12
5330.660
1.0_Printed on_09.01.08_F5330660_Magnetresonanztomographie.fm
Before entering the MR scanner room remove all metal objects from your person. This includes:
Fabius MRI
MRI Metal implants inside the body may cause pain and injuries when they are introduced into the magnetic field. Therefore, patients with certain implants are contraindicated from MR imaging. Prior to MR examination, the MR physician or technologist will ask the patients about the following risk factors: –
cardiac pacemakers
–
metal plates, nails, or metal implants
–
artery clamps
–
artificial cardiac valves
–
intrauterine contraceptive devices
–
body jewelry
–
cosmetic decorations or tattoos (the dyes used may contain metal flakes or slivers)
–
shrapnel
–
pregnancy (should be reported)
Copyright reserved.
1.0_Printed on_09.01.08_F5330660_Magnetresonanztomographie.fm
According to current knowledge, damaging effects are unlikely at the magnetic field strength used (up to 3.0 tesla).
5330.660
13
Function description 5
General Information about the Fabius MRI
Fabius MRI The Fabius MRI comprises the following assemblies: –
Display and control panel
–
Flowmeter assembly
–
Gas box: Gas inlet assembly and related items
–
Breathing system
–
Pneumatic assembly
–
Lung ventilator
–
Anesthetic vaporizer(s)
–
Trolley
–
Cover with additional alarm
Fig. 1
Front view of Fabius MRI anesthesia system, for legend see Table 1
Table 1
Legend to Fig. 1
14
Name
1
Breathing system Cosy 2.6
2
Lung ventilator
3
Oxygen flowmeter (auxiliary)
4
Additional alarm lights
5
Display
Copyright reserved.
No.
5330.660
1.0_Printed on_09.01.08_F5330660_Function_Description.fm
Monitoring, electrical connections and gas connections as shown in Fig. 1, Fig. 2, Fig. 3, Fig. 4 and Fig. 5.
Fabius MRI
Function description
Name
6
Control panel
7
Anesthetic vaporizer mount
8
Pipeline supply manometers
9
Flow control valves
10
Total fresh gas flowmeter
11
Cylinder Manometer
12
Trolley
13
O2 flush key
Fig. 2
Rear view with interface plate; legend, see Table 2
Table 2
Legend to Fig. 2
No. Copyright reserved.
1.0_Printed on_09.01.08_F5330660_Function_Description.fm
No.
5330.660
Name
1
Serial communication ports
2
O2 sensor connection
3
Spirolog sensor cable
4
ON/OFF switch
15
Function description
Fabius MRI
No.
Name
5
Battery fuse
6
Airway pressure connection
7
Tube connection for PEEP valve
8
Tube connection for APL bypass valve
Fig. 3
Rear view showing gas pipeline and PIN index cylinder connections, for legend see Table 3
Table 3
Legend to Fig. 3 Name
1
N2O PIN index cylinder connection
2
O2 or AIR PIN index cylinder connection
3
O2 PIN index cylinder connection
Central tube connections and high-pressure connections
Copyright reserved.
Fig. 4
16
5330.660
1.0_Printed on_09.01.08_F5330660_Function_Description.fm
No.
Fabius MRI
Function description
Fig. 5
6
Battery backup
Multiple socket strip
Fabius MRI battery power is provided by two rechargeable series-connected 12 V batteries. These batteries remain on charge as long as the machine is plugged into an active AC outlet. Should power supply fail while the machine is in operation, the batteries will allow the machine to continue operating for a minimum of 45 minutes, provided that the batteries are fully charged.
Fig. 6
Battery backup arrangement
Copyright reserved.
1.0_Printed on_09.01.08_F5330660_Function_Description.fm
The batteries are accessible by opening the ventilator compartment. The 3.15A battery fuse is located at the back of the control box.
5330.660
17
Function description 7
Fig. 7
8
Fabius MRI
Fabius MRI Piping Diagram
Fabius MRI piping diagram, variant US 2-Gas
Function description of the gas box
The supply gases flow through the filters and non-return valves in the gas inlet assembly. Pipeline supply pressures are indicated on pipeline pressure gauges located on the flowmeter assembly. Cylinder pressure gauges are located on the trolley assembly. The pressures of O2 and N2O delivered to the flowmeter assembly are set by regulators on the gas inlet assembly.
18
5330.660
1.0_Printed on_09.01.08_F5330660_Function_Description.fm
Copyright reserved.
If the O2 supply fails or its pressure decreases below a certain limit, the O2 low alarm switch generates an alarm.
Fabius MRI
Function description
Fig. 8
Gas Box Function Diagram, part 1
Fig. 9 Copyright reserved.
1.0_Printed on_09.01.08_F5330660_Function_Description.fm
If the O2 flush button is pressed, oxygen is delivered to the fresh-gas outlet. The fresh-gas ejector prevents the fresh gas from flowing back into the anesthetic vaporizer. This avoids an increase in anesthetic gas concentration.
9
5330.660
SORC (Sensitive Oxygen Ratio Controller)
Gas Box Function Diagram, part 2
The SORC is a control element that functions like an N2O shut-off device and ensures a vital O2 concentration in the fresh gas. In the event of an O2 shortage, the SORC limits the N2O flow such that the O2 concentration in the fresh gas does not decrease below 21 vol.%.
19
Function description
Fabius MRI If the O2 flow control valve is closed or if the O2 flow is lower than or equal to 200 mL/min, the SORC interrupts the N2O flow. N2O can be added as of an O2 flow of approx. 300 mL/min. In this case, the SORC also prevents O2 concentrations below 21 vol.%. The SORC bypass allows the oxygen to bypass the resistor in the SORC when O2 flows above 10 L/min are needed.
Fig. 10
SORC function diagram, part 1
The O2 and N2O flows are adjusted with the flow control valves. Resistors located at the outlets of the SORC generate back-pressures. These back-pressures exert a force on the control diaphragms of the SORC. The O2 back-pressure opens the SORC. The N2O back-pressure closes the SORC. The pressure ratio at the control diaphragm affects the N2O flow.
20
5330.660
1.0_Printed on_09.01.08_F5330660_Function_Description.fm
Copyright reserved.
The resistors and the spring force are dimensioned such that a minimum concentration of 21 vol.% of O2 is always ensured. The maximum O2 flow is approx. 12 L/min.
Fabius MRI
Function description
Fig. 11
SORC function diagram, part 2, for legend see Table 4
Table 4
Legend to Fig. 11
No.
10
Cosy 2.6 breathing system
Name
1
Control diaphragms
2
Resistors
3
N2O non-return valve
4
Operating-point adjusting screw
5
Flow control valves
The Cosy 2.6 breathing system allows three modes of patient ventilation: –
Manual ventilation and spontaneous breathing
–
Volume controlled ventilation
–
Pressure controlled ventilation
In the “MAN” position, the breathing system is closed to atmosphere. This position is used for manual ventilation of the patient. The APL valve opening pressure can be adjusted from 5 to 70 cmH2O (mbar). In the “SPONT” switch position the APL valve is open to atmosphere. This position is used for spontaneous breathing. Using the control box and the PEEP/Pmax valve, the pressure limit (Pmax) can also be adjusted during volume control from 15 cmH2O (mbar) to 70 cmH2O (mbar) via the membrane keypad.
Copyright reserved.
1.0_Printed on_09.01.08_F5330660_Function_Description.fm
On APL valves with control knob, switching from “IPPV/SPONT” to “MAN” is carried out by turning the knob.
5330.660
21