Dräger Medical

Savina Alarm Messages Sw 2.0

Alarm Messages

98 Pages

Alarm Messages  Ventilation monitoring  Alarm Messages Ventilation monitoring Upper airway pressure limit First pressure limit  For internal use only. Copyright reserved.  R5664900FL2.fm 03.12.01  Cause: 1.  The patient fights the ventilator or is coughing.  2.  The inspiration tube is kinked or obstructed.  3.  One or both pressure sensor(s) deliver(s) too high a measured value.  4.  The pressure source delivers a pressure that deviates from the set pressure.  Detection: 1.  The measured value of the inspiratory pressure sensor Pinsp(t) exceeds the value of the upper airway pressure limit Paw-high + offset.  Dräger Medizintechnik GmbH  Alarm Messages Savina Version 2.0  Page 1