Test instructions / Service Card IPM-L
413
Workstation Critical Care
Warning All servicing and/or test procedures on the device require detailed knowledge of this documentation. Use of the device requires detailed knowledge and observance of the relevant Instructions for Use.
Revision 21.0 9500553
Workstation Critical Care Table of contents
1 1.1
2 2.1 2.2
2.3 3 3.1
3.2
4 4.1
4.2
4.3 2 / 70
Important notes 6 Device configuration 6 Device configuration ... 6 1.1.1 Serial numbers (if not otherwise recorded)... 6 1.1.2 Software versions ... 8 Maintenance parts 8 Maintenance intervals, overview... 8 2.1.1 Maintenance intervals and required sets and parts... 8 Maintenance parts by specified interval... 9 2.2.1 V-unit, battery M7 NM2,5 ... 9 2.2.2 V-unit, particle filter (O2 filter) ... 9 2.2.3 V-unit, air filter... 10 2.2.4 V-unit, Spirolog sensor cable harness ... 10 2.2.5 (Option) GS500, room air filter (filter mat) ... 10 2.2.6 (Option) GS500, air filter blower unit (microfilter) ... 10 2.2.7 (Option) PS500, battery set PS500 (Service) 15Ah ... 10 2.2.8 (Option) PS500, battery set PS500 (Service) 24Ah ... 11 2.2.9 (Option) Medication nebulizer, pressure regulator (O2-side)... 11 Maintenance parts as required ... 11 Electrical safety 12 Electrical safety of V-unit or ME system according to DIN EN 62353 (IEC 62353) ... 12 3.1.1 Visual inspection... 13 3.1.2 Protective earth resistance ... 13 3.1.3 Protective earth resistance measuring points... 14 3.1.4 Applied parts for device leakage current ... 14 3.1.5 Device leakage current... 14 3.1.6 Application components for measurement of leakage current from application 16 component... 3.1.7 Applied part leakage current, mains on applied part, type BF ... 16 Electrical safety according to IEC 60601-1 ... 17 3.2.1 Visual inspection... 17 3.2.2 Protective earth resistance ... 18 3.2.3 Earth leakage current medical device ... 18 3.2.4 Patient leakage current... 20 Function and condition test 21 General condition check ... 21 4.1.1 Medical device... 21 4.1.2 Accompanying documents ... 21 4.1.3 Rating plates, labels and notices on the medical device ... 21 4.1.4 Rating plates, labels and notices on the installed options ... 24 C500 and V-unit condition test ... 25 4.2.1 C500 ... 25 4.2.2 V-unit ... 26 Condition test, options ... 26 Test instructions / Service Card IPM-L
|
Revision 21.0
|
9500553
No.0806_0000007435
Table of contents
Workstation Critical Care Table of contents
4.4
4.5 4.6
No.0806_0000007435
4.7
4.8
4.3.1 (Option) MCable Mainstream CO2 module M11.1 (CO2 sensor)... 4.3.2 (Option) Neonatal flow measurement... 4.3.3 (Option) Drug nebulizer ... 4.3.4 (Option) Trolley... 4.3.5 (Option) Gas cylinder holder ... 4.3.6 (Option) Hose holder ... 4.3.7 (Option) GS500 ... 4.3.8 (Option) PS500... 4.3.9 (Option) Cable duct ... Fuse rating and readout of software version and operating data ... 4.4.1 Fuse value... 4.4.2 Software version of medical device... 4.4.3 Battery operating data of the medical device with M7.3... 4.4.4 Battery operating data of the medical device with M7.3 Plus... Preparations for the function tests ... 4.5.1 Preparations ... C500, function tests ... 4.6.1 Pixel test... 4.6.2 Acoustic alarm... 4.6.3 "Red" and "yellow" visual alarms... 4.6.4 Rotary encoder lighting ... V-unit, function test of module M1.3 (gas dosage module)... 4.7.1 V-unit self-tests... 4.7.2 AIR proportional valve step response time... 4.7.3 O2 proportional valve step response time... 4.7.4 Zeroing all flow sensors... 4.7.5 Relative flow difference S2/S3 ... 4.7.6 Relative flow difference S1/S3 ... 4.7.7 Leakage LPSV V3 ... 4.7.8 Valve calibration for workstations with gas supply unit GS500 ... 4.7.9 Comparator shut-off pressure ... 4.7.10 Comparator switch-on pressure ... 4.7.11 Proportional valve AIR 120 L/min ... 4.7.12 Proportional valve AIR 5 L/min ... 4.7.13 Proportional valve O2 120 L/min ... 4.7.14 Proportional valve O2 5 L/min ... 4.7.15 "System Leakage Test (M1V3)"... 4.7.16 Leakage, AIR proportional valve ... 4.7.17 Leakage, O2 pipeline supply non-return valve ... 4.7.18 Leakage, O2 proportional valve ... 4.7.19 Leakage, AIR pipeline supply non-return valve ... 4.7.20 Comparative pressure measurement, Ptank and inspiratory pressure sensor ... 4.7.21 Comparative flow measurement M1.3... V-unit, function test of module M12.4 (Pato standalone) ... 4.8.1 Calibration ... 4.8.2 Measurement accuracy ...
9500553
|
Revision 21.0
|
Test instructions / Service Card IPM-L
26 26 27 28 29 29 30 30 30 30 30 31 31 32 33 33 33 34 34 34 34 35 35 35 35 35 35 36 36 36 36 37 37 37 37 37 38 38 39 39 39 40 40 41 41 41 3 / 70
Workstation Critical Care Table of contents
V-unit, function test of module M4.1 (pressure/flow temperature monitoring)... 42 4.9.1 Calibration of pressure sensors... 42 4.9.2 Offset of the inspiratory pressure sensor... 42 4.9.3 Offset of the expiratory pressure sensor... 43 4.9.4 Temperature measurement of module M4.1... 43 4.9.5 Calibration of expiratory flow sensor ... 43 4.9.6 Reference voltage of module M4.1... 43 4.10 V-unit, function test of module M21 East ... 44 4.10.1 Function check ... 44 4.11 V-unit, function test of safety devices... 44 4.11.1 Emergency expiratory valve SV2 ... 45 4.11.2 Emergency respiratory valve SV3 ... 46 4.11.3 Non-return valve in expiratory valve ... 47 4.11.4 Expiratory resistance of adult expiratory valve ... 47 4.12 V-unit, function test of the power supply unit module M7.3 / M7.3 Plus... 48 4.12.1 Fan ... 48 4.12.2 Supply voltage ... 48 4.12.3 Internal temperature ... 48 4.13 V-unit, function test of internal power failure backup battery, part 1 ... 49 4.13.1 Powerfail mode function test ... 49 4.14 V-unit, "PI self-tests" function tests ... 50 4.14.1 V-unit self-tests ... 50 4.14.2 Flush flow M12.4 ... 50 4.14.3 "IST Airway Pressure Measurement"... 51 4.14.4 "Pneumatic Release Logic Test"... 51 4.14.5 Pressure sensors S2 and S8... 51 4.14.6 Expiratory flow measurement ... 52 4.15 V-unit, function test of internal power failure backup battery, part 2 ... 52 4.15.1 Battery parameters ... 53 4.16 V-unit, test of device internal temperature ... 53 4.16.1 Testing ... 53 4.17 V-unit, function test of piezo... 54 4.17.1 Function check ... 54 4.18 V-unit, function test of nurse call ... 54 4.18.1 Nurse call with nurse call box (8417370)... 54 4.18.2 Nurse call with test adapter (7911671) ... 55 4.19 V-unit, function test of expiratory valve heater ... 55 4.19.1 Function check ... 55 4.20 Option, V-unit, Function test of nebulizer ... 56 4.20.1 Function check ... 56 4.21 Option, V-unit, function test of drug nebulizer (white housing)... 56 4.21.1 Function check ... 57 4.22 Option, V-unit, function test of Mcable-Mainstream CO2 module M11.1 (CO2 sensor)... 57 4.22.1 Function check ... 57 4.23 Option, V-unit, Function test of neonatal expiratory valve... 58 4.23.1 Function check ... 58
4 / 70
Test instructions / Service Card IPM-L
|
Revision 21.0
|
9500553
No.0806_0000007435
4.9
Workstation Critical Care Table of contents
58 59 59 59 60 60 60 61 61 62 62 63 63 63 65 65 65 65 66 66 68
9500553
5 / 70
No.0806_0000007435
4.24 Option, V-unit, Function test of neonatal flow sensor ... 4.24.1 Function check ... 4.25 Option, V-unit, Function test of GS500 ... 4.25.1 Function check ... 4.26 Medical device, function tests in operation mode ... 4.26.1 Self-test ... 4.26.2 Device checks ... 4.26.3 Expiratory valves leak test... 4.26.4 Function tests in ventilation mode ... 4.26.5 V-unit display function test... 4.26.6 Function test of acoustic alarm, only for SW2.40 and SW2.41 ... 4.26.7 Function test of Silence key ... 4.27 Option, PS500 function test ... 4.27.1 Function check ... 4.28 Medical device, final action... 4.28.1 Device status report ... 4.28.2 Error messages and remedies ... 4.28.3 Device handover with test label... 5 Test equipment 5.1 Test equipment subject to mandatory calibration... 5.2 Test equipment not subject to mandatory calibration...
|
Revision 21.0
|
Test instructions / Service Card IPM-L
Workstation Critical Care
Important notes
Important notes Configuration of Workstation Critical Care (medical device): - Infinity C500 operating and display unit (C500) - Evita Infinity V500 ventilation unit (V-unit) with expiratory valves Options: - Nebulizer-ejector - Mcable-Mainstream CO2 module M11.1 (CO2 sensor) - Medication nebulizer - Neonatal flow measurement - Neonatal expiratory valve - Gas supply unit (GS500) - Power supply unit (PS 500) - Trolley 2
NOTE Prior to using these test instructions, check that they are the latest revision (compare revision with latest service documentation). All results and inputs must be documented in the "Test Report" and "Result Sheet". These test instructions apply to software version 2.0 and higher. Conversion table: 1 bar = 14.504 PSI / 1 mbar = 1.01973 cm H2O.
1
Device configuration This chapter contains test steps to document the configuration.
1.1
Device configuration Introduction
The testing covers the following components: – C500 – V-unit with expiratory valves – All options as per cover sheet
Serial numbers (if not otherwise recorded) Prerequisites
6 / 70
The medical device is switched off.
Action
• Read off serial number of C500. Note: The serial number is located on the rear of the C500.
Result
Serial number of C500
Action
• Read the serial number of the ventilator.
Result
Ventilator serial number
Test instructions / Service Card IPM-L
[________txt]
[________txt]
|
Revision 21.0
|
9500553
No.0806_0000007435
1.1.1
Workstation Critical Care
Device configuration Action
• Remove the inspiratory safety valve (inspiration unit) from the V-unit and read off its serial number. Note: The serial number is located on the side of the inspiratory safety valve.
Result
Serial number of inspiratory safety valve (inspiration unit)
Action
• Document the expiratory valve used.
Result
Document the use of a disposable expiratory valve (YES/NO) with "Yes" or "No".
[________txt]
[________txt]
NOTE
No.0806_0000007435
Disposable expiratory valves have no serial number
9500553
|
Action
• Read the serial number of the adult expiratory valve (reusable). Note: The serial number is located on the side of the expiratory valve.
Result
Serial number of the adult expiratory valve (reusable)
Action
• Read off the serial number of the additional reusable adult expiratory valve, if fitted.
Result
Serial number of additional reusable adult expiratory valve
Action
• Read off the serial number of the MCable-Mainstream CO2 option (CO2 sensor), if fitted. Note: The serial number is located on the sensor housing.
Result
Serial number of MCable-Mainstream CO2 (CO2 sensor) option
Action
• Read off the serial number of the neonatal expiratory valve option, if fitted. Note: The serial number is located on the neonatal expiratory valve.
Result
Serial number of neonatal expiratory valve option
Action
• If the additional optional neonatal expiratory valve is fitted, read off its serial number. Note: The serial number is located on the neonatal expiratory valve.
Result
Serial number of additional neonatal expiratory valve option
Action
• Read off the serial number of the GS500 option, if fitted. Note: The serial number is located on the rating plate, above the power supply socket (remove left side cover).
Result
Serial number of GS500 option
Action
• Read off the serial number of the PS500 option, if fitted. Note: The serial number is indicated on the rating plate on the rear of the PS500.
Result
Serial number of PS500 option
Action
• Read off the serial number of the trolley option, if installed. Note: The serial number is indicated on the rating plate on the rear of the trolley.
Result
Serial number of trolley option
Action
• On the inside of the housing flap, read whether the nebulizer/ejector additional hardware option (without serial number) is fitted and enter "Yes" (fitted) or "No" (not fitted) in the text box.
Result
Nebulizer ejector additional hardware option (without serial number) (Yes or No)
Revision 21.0
|
Test instructions / Service Card IPM-L
[________txt]
[________txt]
[________txt]
[________txt]
[________txt]
[________txt]
[________txt]
[________txt]
[________txt]
7 / 70
Workstation Critical Care
Maintenance parts
1.1.2
Action
• On the inside of the housing flap, read whether the neonatal flowmeter additional hardware option (without serial number) is fitted and enter "Yes" (fitted) or "No" (not fitted) in the text box.
Result
Neonatal flowmeter additional hardware option (without serial number) (Yes or No)
[________txt]
Software versions
NOTE The software versions are read under "Readout of software version and operating data".
2
Maintenance parts This chapter contains information on maintenance parts, measures and tests.
2.1
Maintenance intervals, overview
2.1.1
Maintenance intervals and required sets and parts The following table presents an overview of the required sets and maintenance parts over 6 years. After 6 years the table is worked through again starting with the first year. Designation Part number V-unit, battery M7 NM2.5 8415290 V-unit, particle filter (O2 filter) 8415127 V-unit, air filter (Air filter) 8421479 V-unit, Spirolog sensor cable harness 8416012 (Option) GS500, room air filter (filter 8419758 mat) (Option) GS500, air filter blower unit 6737545 (microfilter) (Option) PS500, battery set PS500 8422969 (Service) 15Ah – LFP batteries (Option) PS500, battery set PS500 (Service) 24Ah
1
2 x
3
x
6 x x x x*
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
8420500
|
Revision 21.0
|
9500553
x
No.0806_0000007435
* The start of the maintenance interval is reset after a part has been replaced, if required.
Test instructions / Service Card IPM-L
5
x
– VRLA batteries Only for devices with (optional) medication nebulizer: V-unit, pressure regulator (O2-side) 8417029
8 / 70
4 x
Workstation Critical Care
Maintenance parts
2.2
Maintenance parts by specified interval
2.2.1
V-unit, battery M7 NM2,5
NOTE After replacing the batteries, the battery wear data must be reset in the "Operating Data" service menu. In addition, for a new battery the material number with revision index and the serial number of the new battery must be entered.
NOTE Batteries are special waste. Dispose of batteries in accordance with local waste disposal regulations.
NOTE – In connection with a M7.3 power supply unit (8415328): If the PS500 option is present, there is no internal battery. – In connection with a M7.3 Plus power supply unit (8421200): The internal battery is always installed. The PS500 option may also be present. An installed M7.3 Plus power supply unit can be identified on the additional connection box behind the left side door.
NOTE After replacing the battery it must be fully charged once. The battery is fully charged when none of the bars in the on-screen battery icon is flashing any more. Quantity 1
Result
Designation
Number
Location/Remark
Battery M7 NM2.5
8415290
Refer to the repair instructions, the data sheet and the safety data sheet of the battery
Battery M7 NM2.5 Next replacement: [________dat]
2.2.2
V-unit, particle filter (O2 filter) Quantity 1
No.0806_0000007435
Result
Designation
Number
Location/Remark
Particle filter (O2 filter)
8415127
In O2 gas inlet/Replacement by specialist.
Particle filter (O2 filter) Next replacement: [________dat]
9500553
|
Revision 21.0
|
Test instructions / Service Card IPM-L
9 / 70
Workstation Critical Care
Maintenance parts
2.2.3
V-unit, air filter Quantity 1 Result
Designation
Number
Location/Remark
Air filter
8421479
In Air gas inlet/Replacement by specialist.
Air filter Next replacement: [________dat]
2.2.4
V-unit, Spirolog sensor cable harness Quantity 1 Result
Designation
Number
Spirolog sensor cable 8416012 harness
Location/Remark Behind the connection housing / replacement by specialist
Spirolog sensor cable harness Next replacement: [________dat]
2.2.5
(Option) GS500, room air filter (filter mat) Quantity 1 Result
Designation
Number
Location/Remark
Room air filter (filter mat)
8419758
Refer to the maintenance instructions
Room air filter (filter mat) Next replacement: [________dat]
2.2.6
(Option) GS500, air filter blower unit (microfilter) Quantity 1 Result
Designation
Number
Location/Remark
Air filter blower unit (microfilter)
6737545
Refer to the maintenance instructions
Air filter blower unit (microfilter) Next replacement: [________dat]
2.2.7
(Option) PS500, battery set PS500 (Service) 15Ah
NOTE
No.0806_0000007435
The battery must be fully charged after replacement. The charging process does not need to be monitored. The battery is fully charged when no bar is flashing near the battery symbol on the screen.
10 / 70
Test instructions / Service Card IPM-L
|
Revision 21.0
|
9500553
Workstation Critical Care
Maintenance parts
Quantity 1
Designation
Number
Location/Remark
Battery set PS500 (Service) 15Ah
8422969
Refer to the repair instructions, the safety data sheet and the battery's data sheet
– LFP batteries Result
Battery set PS500 (Service) 15Ah Next replacement: [________dat]
2.2.8
(Option) PS500, battery set PS500 (Service) 24Ah
NOTE After replacing the battery it must be fully charged once. The battery is fully charged when none of the bars in the on-screen battery icon is flashing any more. Quantity 1
Designation
Number
Location/Remark
Battery set PS500 (Service) 24Ah
8420500
Refer to the repair instructions, the safety data sheet and the battery's data sheet
– VRLA batteries Result
Battery set PS500 (Service) 24Ah Next replacement: [________dat]
2.2.9
(Option) Medication nebulizer, pressure regulator (O2-side) Quantity 1 Result
Designation
Number
Pressure regulator (O2- 8417029 side)
Location/Remark Medication nebulizer / replacement by specialist
Pressure regulator (O2-side) Next replacement: [________dat]
Maintenance parts as required
No.0806_0000007435
2.3
9500553
|
Designation V-unit Expiratory valve diaphragm
Part number
Location/Remark
8413661
Room air filter
8416753
Diaphragm in the expiratory valve Check condition Replacement by user Filter in the housing fan Check condition Replacement by user
Expiratory valve Lip seal (gray)
8416201
Revision 21.0
|
Test instructions / Service Card IPM-L
Expiratory valve Replacement by maintenance personnel
11 / 70
Workstation Critical Care
Electrical safety
Designation Part number Flow sensor 8416203 sleeve, complete (gray) Neonatal expiratory valve (if fitted) Lip seal 8415261
Location/Remark Expiratory valve Replacement by maintenance personnel Neonatal expiratory valve Replacement by maintenance personnel
NOTE The following part must be replaced every 6 years (see the maintenance intervals, overview). The following applies every 2 years before replacement, i.e., in the second and fourth year: At the end of the test, check the device status report for specific error messages, see the final procedure Error messages and remedies. Replace the following part, if required (in case of an error): Designation V-unit Spirolog sensor cable harness
3
Part number
Location/Remark
8416012
Behind the connection housing Replacement by specialist
Electrical safety This chapter contains tests on the electrical safety. Perform the tests according to IEC 62353 or IEC 60601-1 (depending on the national guidelines. In Germany, for example, standard DIN EN 62353 applies).
3.1
Electrical safety of V-unit or ME system according to DIN EN 62353 (IEC 62353) NOTE The medical device under test conforms to the conditions of protection class I. With the MCable-Mainstream CO2 module M11.1 (CO2 sensor) (sensor connected) or the neonatal flow sensor the medical device conforms to type BF.
NOTE
No.0806_0000007435
The tester, e.g. SECUTEST, must be correctly configured for all measurements. If implausible measurement results are obtained, such as a leakage current of 0.0 µA, check the tester configuration in addition to the test setup!
12 / 70
Test instructions / Service Card IPM-L
|
Revision 21.0
|
9500553
Workstation Critical Care
Electrical safety
NOTE In testing to IEC 62353, the medical electrical device (ME device) or the medical electrical system (ME system) must be tested. ME systems must be treated like ME devices. An ME system is a combination of several devices, as specified by the manufacturer, of which at least one must be an ME device, which are interconnected by a functional connection or by means of a multiple socket outlet.
NOTE With devices that are connected to other devices by means of a data cable, this connection must be disconnected prior to performing the electrical safety check, in order to avoid incorrect measurements.
3.1.1
Visual inspection Prerequisites
The tester and the device under test are switched off.
WARNING Hazardous voltage. Touching live components can lead to serious injury or death. ► Disconnect the power cord from the AC outlet before checking the power fuse-links. Action Test
• Disconnect the power plug from the mains socket. – The power fuse-links of the device under test match the specifications on the rating plate. – The power cable and plug are not dirty or damaged.
Result
3.1.2
Condition checked.
[________OK]
Protective earth resistance
1936
Test set-up
Fig. 1
Protective earth resistance
No.0806_0000007435
Item Designation 1 2 3
9500553
|
Revision 21.0
Tester Device under test Tester probe cable
|
Test instructions / Service Card IPM-L
13 / 70
Workstation Critical Care
Electrical safety
Item L N PE Action
Designation Conductor Neutral conductor Protective earth
• Prepare the test setup. • Switch the tester on. • Configure the tester appropriately and follow the instructions on the tester. • Using the tip of the probe cable, scan each of the measuring points on the device under test listed under "Protective earth resistance measuring points" one after the other, moving the mains power supply cord along the entire length during the measurement. The resistance must not change when you do so.
Test
Result
The protective earth resistance of single devices (including the permanently installed power supply cord, or power supply cord detachable only by use of tools) must not exceed 0.3 Ohm in each case. Maximum measured value.
[________Ω]
or Test Result
3.1.3
The protective earth resistance of ME systems (including mains power cable) must not exceed 0.5 ohms in each case. Maximum measured value of ME system.
[________Ω]
Protective earth resistance measuring points Action
• Scan the following measuring points for protective earth resistance measurement one after the other using the tip of the probe cable: – Potential equalization pin on the power supply unit of the V-unit
Result
3.1.4
Measuring points scanned.
[________OK]
Applied parts for device leakage current
3.1.5
Action
• Use the following device-specific test adapters for the device under test:
Result
7901772, VDE Babylog test plug (if Neo-Flow option installed)
[________OK]
Device leakage current
NOTE The equipment leakage current can be tested by the differential measurement method or the direct measurement method. In direct measurement, set up the device under test with insulation and scan all accessible conductive components using the probe (the protective earth is internally interrupted in the tester). The tester is switched on. No.0806_0000007435
Prerequisites
14 / 70
Test instructions / Service Card IPM-L
|
Revision 21.0
|
9500553
Workstation Critical Care
Electrical safety
1932
Test set-up
Fig. 2
Equipment leakage current
Item Designation 1 2 3 4 L N PE Action
Tester Device under test (Applied part) Device-specific test adapter for tester (Applied part) Configurable ports for applied part Conductor Neutral conductor Protective earth
• Prepare the test setup. • (Applied part) Connect the device-specific test adapter on one end to the device under test and on the other end to the tester's configurable port "A" for applied parts (paying attention to the configuration!). • Follow the instructions on the tester.
NOTE Perform the test twice. In the second test attach the connector in the socket rotated 180°. In many test devices the mains plug rotation is simulated by means of a built-in selector switch. Document the higher measured value.
NOTE Always enter the reference value (initial value measured) in the "Test Report" or "Result Sheet" document.
NOTE If the measured values are between 90% and 100% of the permissible limit value, apply the reference value and the previously measured values of the recurrent test to assess electrical safety! Test Result Test
Reference value
9500553
|
[________µA]
The recurrent test value must not exceed 500 µA. Recurrent test
No.0806_0000007435
Result
The reference value must not exceed 500 µA.
Revision 21.0
|
Test instructions / Service Card IPM-L
[________µA]
15 / 70
Workstation Critical Care
Electrical safety
3.1.6
Application components for measurement of leakage current from application component
NOTE Only conduct the leakage current test from the applied part if the Neoflow measurement option is available. With the optional MCable Mainstream CO2 sensor, no measurement of the patient leakage current is performed due to the galvanic isolation in the sensor head and the sensor having no metallic contacts. As a substitute, a visual inspection is performed in the Condition test chapter.
3.1.7
Action
• Use the following device-specific test adapter for the device under test:
Result
VDE Babylog test plug (with the Neoflow option)
[________OK]
Applied part leakage current, mains on applied part, type BF Prerequisites
The tester is switched on.
2267
Test set-up
Fig. 3
Leakage current on applied part
Item Designation 1 2 3 4 L N PE Action
Tester Device under test Applied part, device-specific test adapters for tester Configurable connection sockets for applied parts (A - K) Conductor Neutral conductor Protective earth
• Prepare the test setup. • (Applied part) Connect the device-specific test adapter on one end to the device under test and on the other end to the tester's configurable port "A" for applied parts (paying attention to the configuration!). • Follow the instructions on the tester.
The test must be performed twice! The second test is performed with the plug rotated 180° in the socket. In many test devices the mains plug rotation is simulated by means of a built-in selector switch. The higher measured value must be documented.
16 / 70
Test instructions / Service Card IPM-L
|
Revision 21.0
|
9500553
No.0806_0000007435
NOTE
Workstation Critical Care
Electrical safety
NOTE The reference value (initial measured value) must always be transmitted!
NOTE If the measured values are between 90% and 100% of the permissible limit value, the reference value and the previously measured values of the recurrent test should be applied to assess electrical safety! Test Result Test Result
3.2
The reference value must not exceed 5000 µA. Reference value
[________µA]
The recurrent test value must not exceed 5000 µA. Recurrent test
[________µA]
Electrical safety according to IEC 60601-1 NOTE The medical device under test conforms to the conditions of protection class I. With the MCable-Mainstream CO2 module M11.1 (CO2 sensor) (sensor connected) or the neonatal flow sensor the medical device conforms to type BF.
NOTE An optional multiple socket-outlet, if any, must be included in the individual tests (medical electrical system).
NOTE With devices that are connected to other devices by means of a data cable, this connection must be disconnected prior to performing the electrical safety check, in order to avoid incorrect measurements.
3.2.1
Visual inspection Prerequisites
The following preconditions must be met for the visual check: – The medical device is switched off. – Make sure the power plug is unplugged. – The pipeline supply tubing has been removed.
Test
– The ratings of the V-unit power fuse links conform to the specifications imprinted on the power supply unit label. – The mains power cable and cable retainer of the V-Unit are not dirty or damaged. Condition checked.
[________OK]
No.0806_0000007435
Result
9500553
|
Revision 21.0
|
Test instructions / Service Card IPM-L
17 / 70
Workstation Critical Care
Electrical safety
3.2.2
Protective earth resistance
769
Test set-up
Fig. 4
Test setup for protective conductor resistance
Item Designation 1 2 3 L N PE Action
Tester, e.g. SECUTEST, 7910594 Device under test Test probe with tip (e.g. SECUTEST accessories) Conductor Neutral conductor (Protective Earth) conductor
• Prepare the test setup. • Switch the tester on. • Configure the tester appropriately, and follow the instructions on the tester. • Using the tip of the test probe, scan the following points on the V-unit, moving the power cable section-by-section while doing so: – Power supply unit potential equalization pin
Test Result
3.2.3
The protective conductor resistance must not exceed 0.2 Ohm (including mains power cable) in each case. Protective conductor resistance of V-unit
[________Ω]
Earth leakage current medical device
NOTE
No.0806_0000007435
In order to avoid incorrect measurement, set up the V-unit so that it is insulated.
18 / 70
Test instructions / Service Card IPM-L
|
Revision 21.0
|
9500553
Workstation Critical Care
Electrical safety
218
Test set-up
Fig. 5
Earth leakage current test setup
Item Designation 1 2 L N PE Action
Tester, e.g. SECUTEST, 7910594 Device under test Conductor Neutral conductor (Protective earth) conductor
• Prepare the test setup. • Switch on medical device. • Follow the instructions on the tester.
NOTE For symmetrical mains plugs that have no preferential position in the socketoutlet, the earth leakage current test must be performed twice! The second test is performed with the plug rotated 180° in the socket. In many test devices the mains plug rotation is simulated by means of a built-in selector switch. The higher measured value must be documented. IEC 60-601-1 Test Result Test Result
Normal condition (N.C.): The value must not exceed 500 µA. Normal condition (N.C.) IEC 60-601-1
[________µA]
Single fault condition (S.F.C.): The value must not exceed 1000 µA. Single fault condition (S.F.C.) IEC 60-601-1
[________µA]
UL 60-601-1 Test Result Test
Normal condition (N.C.) UL 60-601-1
[________µA]
Single fault condition (S.F.C.): The value must not exceed 500 µA. Single fault condition (S.F.C.) UL 60-601-1
[________µA]
No.0806_0000007435
Result
Normal condition (N.C.): The value must not exceed 300 µA.
9500553
|
Revision 21.0
|
Test instructions / Service Card IPM-L
19 / 70
Workstation Critical Care
Electrical safety
3.2.4
Patient leakage current
NOTE Perform the patient leakage current test only if the Neo-Flow measurement option is fitted. With the MCable-Mainstream CO2 sensor option no patient leakage current measurement is performed, as the electrical isolation is in the sensor head and the sensor has no metallic contacts. A visual check is performed alternately in the "Condition test" section.
425
Test set-up
Fig. 6
Patient leakage current test setup
Item Designation 1 2 3 4 L N PE Action
Tester, e.g. SECUTEST, 7910594 Device under test Test plug, 7901772 Ports for applied parts Conductor Neutral conductor (Protective Earth) conductor
• Prepare the test setup. • Connect the Neo-Flow sensor cable to the V-unit. • Remove the Neo-Flow sensor. • Connect the test plug (7901772) to the Neo-Flow sensor socket and to the tester. • Start up the test specimen and perform patient leakage current measurements. • Follow the instructions on the tester.
For symmetrical mains plugs that have no preferential position in the socketoutlet, the patient leakage current test must be performed twice! The second test is performed with the plug rotated 180° in the socket. In many test devices the mains plug rotation is simulated by means of a built-in selector switch. The higher measured value must be documented. IEC 60-601-1
20 / 70
Test instructions / Service Card IPM-L
|
Revision 21.0
|
9500553
No.0806_0000007435
NOTE