airvo_2_technical_in_service_notes_oct_2012.pdf
Page 1
7/11/2012
AIRVO 2
AIRVO 2
Flow Range
•
•
NEW:
More flow
Was 15 to 45 L/min
Now 15 to 50 L/min
AIRVO 2
Temperature Range
•
•
Dew point temp was 37°C only
Now 37°C, 34°C or 31°C
NEW:
34°C Dew Point
setting
1
Page 2
7/11/2012
AIRVO 2
O2% Range
•
•
•
Was 21 to 45% at 45 L/min
Now 21 to 74% at 45 L/min
– will do up to 90% at lower flow rates
Need a new way for reading oxygen
NEW:
Maximum
recommended
O2 input:
now 30L/min
AIRVO 2
Setting O2%
•
•
•
O2% Chart was niggly and confused people. It’s gone!
O2% is now displayed on screen
O2sensor range 25%-80%
NEW:
Ultra-sound
Oxygen sensor
built into flow
path
AIRVO 2
New display
•
Vastly improved user interface
NEW:
Full colour
OLED display
2
Page 3
7/11/2012
AIRVO 2
Patient settings
•
Customisation for the patient
NEW:
3 lockable
settings
AIRVO 2
Environment Settings
•
•
Customisation for the fleet
Top level control for technical staff and reps
NEW:
Mode control
Lang. control
Limit control
AIRVO 2
Errors
•
On-screen troubleshooting animations
NEW:
Full colour
OLED display
3
Page 4
7/11/2012
AIRVO 2
Paediatric Patients
•
Junior mode for Optiflow Junior Cannula
– 34°C Dew Point
– 5 to 20 L/min
– 21 to 80% O2
NEW:
Junior Mode
Junior Circuit
Junior Interfaces
AIRVO 2
Filter
•
NEW:
Filter
Larger, sealed air filter and filter cover
– Same material as the RT019/RT020
– O2 inlet extension now located on pole side
– Change required every 1000 hours of use
AIRVO 2
Reprocessing
•
Better support for the disinfection process
– Information at start-up
– Disinfection progress chart
NEW:
Press power
button to start
4
Page 5
7/11/2012
AIRVO 2
Other changes
•
•
•
•
•
•
14 day change out 900PT501
> 2 minute power down alarm
Updated pole tray
Oxygen adapter now on the side closest to the
pole
Non-return valve
Blue lower cover with molded rubber feet
AIRVO 2
Spec Sheet
Project KRAMER
Objectives – Better mobile pole stand
•
•
•
•
•
Stability
Mobility
Versatility
applications
Shipping
Assembly
•
Portability
-Pass the ISO tilt test
-Good “push feel”
-Be usable for all FPH
-One box, flat packed
-Low packaging
-Low no. of steps
-Carry on shoulder, air travel
5
Page 6
7/11/2012
Project KRAMER
Objectives – Better mobile pole stand
•
•
•
•
•
•
•
Pass the ISO tilt test
Good “push feel”
Be usable for all FPH applications
One box, flat packed
Low packaging
Low no. of steps
Carry on shoulder, air travel
NEW:
4 x mounting
points
Questions
?
6