Multi-Patient Cleaning Instructions
SleepStyle™ 200, 600 & Auto Series CPAP Accessories
Step 1: Soil Removal
• Before disinfection, disassemble and clean the breathing tube and chamber with a mild detergent solution.
• Rinse thoroughly under running tap water to remove detergent.
• Allow to thoroughly air dry before proceeding with high-level disinfection.
Failure to properly clean these items may result in inadequate disinfection.
Step 2: High-Level Disinfection
Not
Recommended
For High-Level
Disinfection
These parts are not
recommended for high-level
disinfection and should be
replaced if the CPAP is used
between multiple patients.
900HC221
SleepStyle™ 200 Series
standard chamber
SleepStyle™ 600 Series
standard chamber
900HC522
SleepStyle™ 600 Series
heated breathing tube
Soak the chamber in a hot
water bath for 30 minutes at
158°F (70°C).
Follow the manufacturer’s
instructions for the
Sterrad® 100S
sterilization system.
Immerse the chamber in
Cidex® OPA for 12 minutes.
Rinse all parts thoroughly
in clean water.
✔
✔
✔
SleepStyle™ 200 Series
long-life chamber
HC365
CIDEX® OPA
30 minutes at
167°F (75°C)
HC355
FILTER
STERRAD® 100S
✔
SleepStyle™ 200 Series
breathing tube
HC385
Pasteurization
✔
✔
✔
✔
✔
Refer to the Healthcare
Provider Manual
(185045167)
for detailed instructions
Refer to the Healthcare
Provider Manual
(185045167)
for detailed instructions
✔
185045068 REV C EN © 2012 Fisher & Paykel Healthcare Limited
OUTLET PORT
SleepStyle™ 600 Series
CPAP
HC360
SleepStyle™ 600 Series
long-life chamber
HC300
Extended Life
Humidification Chamber Kit.
Suitable for use with the
HC150 Heated Humidifer
✔
Refer to the Operating
Manual (185042488)
for detailed instructions
Refer to the Outlet
Cleaning Kit (185043557)
for detailed instructions
✔
✔
✔
✔
✔
Refer to the Operating
Manual (185043383)
for detailed instructions
✔
Refer to the User
Instructions (185041054)
for detailed instructions.
ITEMS RECOMMENDED FOR HIGH-LEVEL DISINFECTION HAVE BEEN VALIDATED TO WITHSTAND 20 CYCLES OF THE ABOVE PARAMETERS
Do not use the items if damaged. Note that it is the user’s responsibility to qualify any deviations from the recommended methods. The information provided in this document is an indication only.
Please refer to the User Instructions provided with each item for the most up-to-date information.
www.fphcare.com