neopuff_technical_manual_rev_m_2011.pdf
Page 1
|
Infant Resuscitation | Technical Manual
Page 2
TABLE OF CONTENTS
1. Specifications......................................................................................................................................................................2
1.1 Warnings...........................................................................................................................................................................2
1.2 Performance Specifications...............................................................................................................................................2
1.3 Technical Specifications....................................................................................................................................................2
1.4 Symbol Definitions............................................................................................................................................................3
1.5 Label Identification . .........................................................................................................................................................3
2. Cleaning ..............................................................................................................................................................................4
2.1 Cleaning Products to Avoid................................................................................................................................................4
2.2 Cleaning of the Test Lung...................................................................................................................................................4
2.2.1 Latex (black) Test Lung (discontinued May 2007) ..................................................................................................4
2.2.2 Silicone (blue) Test Lung RD020-01........................................................................................................................5
2.3 Cleaning Reusable T-Piece (discontinued September 2008)...............................................................................................5
2.4 Sterilization.......................................................................................................................................................................5
2.5 After Cleaning....................................................................................................................................................................5
3. Service Information..............................................................................................................................................................6
3.1 Functional Schematic........................................................................................................................................................6
3.2 Installation Checks and Preventative Maintenance............................................................................................................6
3.2.1 Installation Checks.................................................................................................................................................6
3.2.2 Preventative Maintenance.......................................................................................................................................6
3.3 Testing of the F&P Neopuff™/ Perivent™ Performance (Manometer and Valve System)......................................................6
3.3.1 Testing of the Manometer.......................................................................................................................................7
3.3.2 Testing of the Valve System....................................................................................................................................7
3.3.3 Setting Max Pressure Relief to 40 cmH2O................................................................................................................7
3.3.4 Resetting the Manometer to Zero............................................................................................................................7
3.3.5 Manometer Replacement.......................................................................................................................................7
3.3.6 Valve System Replacement.....................................................................................................................................7
3.4 Set-up and Maintenance Checklist ....................................................................................................................................8
4. Assembly Diagrams..............................................................................................................................................................9
4.1 F&P Neopuff™/Perivent™ Infant Resuscitator (Post June 2010).........................................................................................9
4.1.1 Assembly Diagram.................................................................................................................................................9
4.1.2 Parts List................................................................................................................................................................9
4.1.3 Spares Kit............................................................................................................................................................10
5. Mounting Options..............................................................................................................................................................11
5.1 900RD301 Side-Mounting Block......................................................................................................................................11
5.2 900MR088 Rail Mount Bracket........................................................................................................................................11
5.3 900RD302 Wall Mount Bracket........................................................................................................................................11
5.4 RD050-01 Pole and Rail Central Mount............................................................................................................................12
N E O P U F F ™ t e c h n i c a l m a nu a l
1
Page 3
1. Specifications
1.1 Warnings
Input gas flow rate 5 to 15 L/min. Recommended operating gas
flow rate 8 L/min. Input flow ranges are circuit specific, refer to
circuit User Instructions.
Do not use oil, grease or other substances that are incompatible
with oxygen on any part of the F&P Neopuff / Perivent
T-Piece Circuit.
Do not attempt to use a higher gas flow rate than 15 L/min.
Incorrect use can be hazardous.
Factory setting of pressure limit valve is at 40 cmH2O. This setting
is user adjustable up to 80 cmH2O. Do not attempt to set the
pressure relief valve above 80 cmH2O. Reset pressure relief valve
to 40 cmH2O after use (refer section 3.3.3 of this manual).
The device must not be used on unattended patients.
Use only recommended F&P Neopuff™ / Perivent™ Infant
Resuscitator accessories.
To be used for resuscitation only.
The operation of the F&P Neopuff / Perivent must be checked prior
to first use.
Refer to section 3.3 “Testing of the F&P Neopuff / Perivent
Performance”.
Use only a F&P gas supply line or approved equivalent.
1.2 Performance Specifications
Peak Inspiratory Pressure (PIP) Range
1.3 Technical Specifications
Height
250 mm (9.8”)
@ 5 L/min
approx. 2 to 70 cmH2O [mbar]
Width
200 mm (7.9”)
@ 8 L/min
approx. 3 to 72 cmH2O [mbar]
Depth
104 mm (4.1”)
@ 10 L/min
approx. 4 to 73 cmH2O [mbar]
Weight
1.9 kg (4.2 lb)
@ 15 L/min
approx. 8 to 75 cmH2O [mbar]
Manometer Range
-10 to 80 cmH2O [mbar]
Manometer Accuracy
+/-2.0% of Full Scale Deflection
Maximum Pressure Setting
65 to 80 cmH2O [mbar]
(dependent on flow rate)
Positive End Expiratory Pressure (PEEP) Range
@ 5 L/min
approx. 1 to 6 cmH2O [mbar]
@ 8 L/min
approx. 1 to 10 cmH2O [mbar]
900RD010 / 900RD110 Dead Space
6 ml
@ 10 L/min
approx. 2 to 15 cmH2O [mbar]
RD1300-10 Dead Space
3.3 ml
@ 15 L/min
approx. 4 to 17 cmH2O [mbar]
Storage Temperature Range
-10 to 50 °C (+14 to +122 °F),
up to 95% humidity
Input Gas Flow Range
Minimum
5 L/min
Maximum
15 L/min
Operating Time (400 L Cylinder)
5 L/min
80 minutes
10 L/min
40 minutes
15 L/min
26 minutes
Operating Temperature Ranges
Humidified Circuit
+18 to 26 °C (+64 to +78 °F),
30 - 75% humidity
Non-Humidified Circuit
-18 to 50 °C (-0.4 to +122 °F),
up to 95% humidity
Recommended Patient
Body Weight
Delivered Oxygen Concentration
0 to 10 kg (22 lb)
Up to 100% depending on gas
supply
NOTE: All performance figures listed above are representative only.
PEEP values stated are based on typical clinical PIP settings.
Higher PEEP values can be achieved if higher PIP values are set.
N E O P U F F ™ t e c h n i c a l m a nu a l
2
Page 4
1. Specifications continued
1.4 Symbol Definitions
SYMBOL
DEFINITION
1.5 Label Identification
The label pictured is typical of the information contained on a
F&P Neopuff™ / Perivent™ Infant Resuscitator.
Attention: Consult the Operating
Instructions
MAX-P
PiP
Sets the Max Pressure Relief that may
be delivered to the patient
(factory set at 40 cmH2O)
Controls the Peak Inspiratory Pressure
delivered to the patient
5-15 L/min
Gas inlet connection from gas supply
(5 to 15 litres per minute)
REF Gas outlet connection to patient
LOT SN -
This details the product code which identifies the operating
instruction language and fascia type
The production date: (10) = 2010, (06) = June,
(15) = fifteenth day
The serial number is specific to a particular F&P Neopuff /
Perivent Infant Resuscitator. The first six digits match the LOT
number and last six digits identify the specific product serial
number.
The bar code consists of the following predefined application identifiers:
01
- Not used
0 - Sequence start
94 - Country of manufacture - New Zealand
200124 - Company Identifier - Fisher & Paykel Healthcare Ltd
1093 - Product Code
1 - Check Sum
REF - RD900XXX
RD900
XXX
Operating Instruction Language
Fascia Type
900 Series
AEU
English
English Text
ADU
Danish
AFU
French
Symbols
Only
ALU
Italian
AMU
Finnish
ANU
Dutch
ASU
Spanish
ATU
Portuguese
AVU
Norwegian
AWU
Swedish
AZU
Chinese
AGU
German (Perivent)
T-Piece
Resuscitator
N E O P U F F ™ t e c h n i c a l m a nu a l
3
Page 5
2. Cleaning
• Comply with hospital, local and national guidelines for product
cleaning frequencies.
• Ensure all oxygen and air supplies are turned off and disconnected
from the F&P Neopuff™ / Perivent™ Infant Resuscitator before
performing cleaning procedures. Explosion and fire hazards can
exist when performing cleaning procedures in an oxygen-enriched
environment.
• Cleaning shall be performed at ambient conditions.
• Before cleaning, remove and discard all used disposable products
using the recommended method of disposal.
• Dust all surfaces with a clean damp soft cloth.
• Clean all plastic surfaces with detergent-based solution (maximum
2% in water) ensuring the manufacturer’s directions for use of the
cleaning agent are followed.
• The following proprietary chemical cleaning wipes are
recommended if the F&P Neopuff / Perivent fascia requires cleaning
for infection control purposes.
Caviwipes (Metrex Research Corporation)
Sani-cloth HB (Professional Disposables, Inc.)
Asepti-Wipes II (Ecolab, Inc.)
Spartan’s TB-Cide Quat Wipes
Vernacare Tuffie Wipes Alcohol Free
• Dry all surfaces after cleaning with a clean soft cloth or
paper towel.
CAUTION: Ensure no part of the F&P Neopuff / Perivent Infant
Resuscitator or related accessories is immersed in any cleaning liquid
or cleaning solution. For Test Lung cleaning instructions, see section
2.2.2.
CAUTION: Do not clean the F&P Neopuff / Perivent Infant Resuscitator
fascia with proprietary cleaning products containing either
hydroxides, hypochlorites, peroxides, gluteraldehyde or cleaning
products with a greater than 30% alcohol base.
NOTE: The recommended chemical cleaning wipes listed above have
been checked for long-term compatibility with the IW900 Series Infant
Warmers and the F&P Neopuff / Perivent Infant Resuscitator.
2.1 Cleaning Products to Avoid
CAUTION: The chemicals used in these proprietary cleaning products
may lead to discoloration, crazing and eventual cracking of the fascia.
Examples of proprietary cleaning products which contain such
cleaning chemicals include but are not limited to:
Asepti-Wipes (Ecolab, Inc)
Clorox (The Clorox Company)
Endbac 256 (Johnson Wax Professional)
Quat (3M)
Sporiciden (Liberty Industries, Inc)
Sporox II (Sultan Healthcare)
Steris (Steris Corporation)
Terralin (Schülke & Mayr)
Virtek (A Virtek Company)
Virox (Virox Technologies)
CAUTION: Do not use abrasive cleaning solutions.
CAUTION: Ensure F&P Neopuff/ Perivent Infant Resuscitator and
accessories are checked before returning the Infant Resuscitator to
service.
2.2 Cleaning of the Test Lung
If required, the Test Lung can be disinfected. There are two different
models of Test Lung and each has specific cleaning methods.
• T he Test Lung is a consumable item. Irrespective of model, the
Test Lung should regularly be inspected for signs of damage such
as discoloration, perishing or cracking. Replace the Test Lung if
damage is observed.
2.2.1 Latex (black) Test Lung (discontinued May 2007)
WARNING The Test Lung contains natural rubber latex that may
cause allergic reactions.
CAUTION: Latex is susceptible to attack by solvents. Ensure no solvents
are used to clean the Test Lung.
This Test Lung consists of two parts, the flexible latex test lung and
a rigid black plastic connector. The Test Lung can be disassembled
for disinfection. It can be disinfected using ethylene oxide gas only.
Some carrier gases can cause stress cracking and are not suitable. If
in doubt, check with the chemical supplier.
N E O P U F F ™ t e c h n i c a l m a nu a l
4
Page 6
2. Cleaning continued
2.2.2 Silicone (blue) Test Lung RD020-01
The Test Lung must be cleaned before every use or in
accordance with hospital cleaning protocols. This test lung consists of
two parts, the flexible silicone test lung and a rigid plastic connector.
The two parts of the Test Lung may be disassembled
for disinfection. It can be cleaned using disinfectants containing
either peracetic acid (e.g. Perasafe®) or orthopthaldehyde
(e.g. Cidex® OPA). Other disinfection methods can cause damage
to the Test Lung and are not recommended.
Please follow the manufacturer’s instructions for storage, handling
and usage of these chemicals.
2.4 Sterilization
CAUTION:
• Do not autoclave or gas-sterilize any part of the
F&P Neopuff™ / Perivent™ Infant Resuscitator.
• Do not autoclave any of the Neopuff / Perivent accessories,
e.g. Test Lung and Gas Supply Line.
• For cold sterilization: Ensure the cold sterilization agents are safe
for use with the relevant surfaces.
2.5 After Cleaning
CAUTION:
• Ensure all F&P Neopuff / Perivent Infant Resuscitator parts and
accessories are checked before returning the device to service.
Silicone Test Lung
Plastic Connector
WARNING Ensure the silicone Test Lung and plastic connector are
both removed from the T-Piece before use on a patient.
2.3 Cleaning Reusable T-Piece
(discontinued September 2008)
The reusable T-Piece 500RD104 consists of two parts and can be
disassembled for disinfection (refer diagram). The T-Piece should
be disinfected by autoclaving at up to 136 °C, 220 kPa for 4
minutes. Following reassembly, the T-Piece should be tested prior
to use to ensure that it is functioning correctly. Refer to the F&P
Neopuff™ / Perivent™ Operating Manual (Fisher & Paykel Healthcare
Part No. 185041726) for set-up instructions.
Unscrew PEEP cap for disinfection
N E O P U F F ™ t e c h n i c a l m a nu a l
5
Page 7
3. Service Information
3.1 Functional Schematic
3.3 Testing of the F&P Neopuff™ /Perivent™
Performance (Manometer and Valve System)
Circuit Pressure Manometer
Gas Inlet Port
Gas Outlet Port
Manometer
Maximum
Pressure
Relief Valve
Peak
Inspiratory
Pressure Valve
3.2 Installation Checks and Preventative
Maintenance
WARNING Dropping the F&P Neopuff™ / Perivent™ Infant
Maximum
Pressure
Relief Cover
Peak
Inspiratory
Pressure
Control
Inlet Port
Outlet Port
Resuscitator or other similar forms of impact may cause damage
resulting in incorrect operation of the unit. If you suspect damage
to have occurred, please perform checks as outlined in section 3.3
before connection to a patient.
3.2.1 Installation Checks
Prior to first use:
– Remove manometer cover
– Complete Performance Testing (section 3.3)
3.2.2 Preventative Maintenance
The integrity of the F&P Neopuff / Perivent manometer and valve
system can be tested using the following guidelines. The inlet port
must be connected to a gas supply capable of generating constant
flow at 5, 10 and 15 L/min, and a certified gauge and bleed valve
should be available to check the manometer accuracy. The
F&P Neopuff / Perivent resuscitation circuit and T-Piece can
be used in place of a bleed valve as shown below.
• The integrity of the system and manometer should be checked prior
to first use, annually and after servicing, using the
“Testing of the F&P Neopuff / Perivent Performance” procedure
(section 3.3).
Use T-Piece with
Test Lung attached
as a bleed valve
• The use of a mounting bracket to help prevent the
F&P Neopuff / Perivent being damaged or dropped is recommended.
Mounting solutions available can be found in section 5 of this
manual.
50
60 70
40
• All maintenance and service procedures must be performed by
qualified personnel using only Fisher & Paykel Healthcare parts.
• Always ensure gas passages are free from contaminants, especially
hydrocarbons, oils and grease, prior to reassembly.
30
20
10
80
–
cm
H 2O
+
0 -10
Certified Gauge
50
60 70
40
30
• Please contact an authorized Fisher & Paykel Healthcare
representative for further assistance with any servicing or
maintenance requirements.
• The Test Lung is a consumable item; it should be monitored for
signs of wear and material degradation and replaced as required.
20
10
80
–
0 -10
cm
H 2O
+
-20
Connect pressure gauge
between F&P Neopuff / Perivent
outlet and resuscitation circuit
N E O P U F F ™ t e c h n i c a l m a nu a l
6
Page 8
3. Service Information continued
3.3.1 Testing of the Manometer
1. L ift the cover off the Max Pressure Relief Valve slightly and turn out
of the way.
2. D
isconnect all devices from the F&P Neopuff™ / Perivent™ outlet
port. Check that the manometer needle is within +/-2 cmH2O of
zero on the manometer gauge. If the manometer does not read
zero, the resetting of the manometer to zero procedure (section
3.3.4) should be followed.
3. C
onnect the outlet of the F&P Neopuff / Perivent Infant Resuscitator
to a bleed valve and a certified gauge (e.g. Mensor Digital Pressure
Gauge Series 2400). Set the gas supply to 10 L/min. Completely
close the maximum pressure limit valve by turning the left-hand
knob completely clockwise. With the bleed valve closed, adjust
the Peak Inspiratory Pressure knob to set the pressure so that
the certified gauge reads 10, 20 and 40 cmH2O. Check that the
manometer reads within +/-2 cmH2O of these values at each set
point.
4. W
ith the pressure set to 40 cmH2O, open and close the bleed valve
three times and check the manometer needle rises and falls
smoothly. If the F&P Neopuff / Perivent Infant Resuscitator
fails any of these tests, the manometer should be regarded
as inaccurate and replaced with a new manometer (Part No.
043040841).
Follow the manometer replacement guidelines in section 3.3.5 of
this manual, or contact your Fisher & Paykel Healthcare service
representative for further guidance.
Alternative settings for the Max Pressure Relief should be made in
accordance with hospital protocol.
1. Adjust gas flow to 10 L/min.
2. C
lose the Peak Inspiratory Pressure valve by turning the knob fully
clockwise.
3. A
djust the Max Pressure Relief knob clockwise or counterclockwise
until the manometer reads 40 cmH2O.
4. T urn the Peak Inspiratory Pressure knob counterclockwise so the
manometer reads 20 cmH2O and shut off the gas flow.
3.3.4 Resetting the Manometer to Zero
To set the manometer to zero:
1. D
isconnect the F&P Neopuff™ / Perivent™ Infant Resuscitator from
any other equipment.
2. Remove the opaque plastic plug in the lens of the manometer.
3. U
sing a suitable slot screwdriver, carefully adjust the screw in
the manometer face clockwise or counterclockwise to reset the
manometer to zero. Care must be taken when doing this, as overrotation of the screw can damage the manometer internals.
4. Replace the plastic plug in the lens of manometer.
5. V
erify that the manometer needle is now within +/-2.0 cmH2O of
zero. If not, repeat the above procedure.
3.3.5 Manometer Replacement
The manometer is not a serviceable item and must be replaced by
Manometer Kit 043040841.
1. Remove the back cover, fixed by four screws.
3.3.2 Testing of the Valve System
2. Disconnect the tube from the manometer.
1. S
et the gas supply to 5 L/min. Completely close the Peak
Inspiratory Pressure control and the Max Pressure control by
turning both knobs completely clockwise. Close the bleed valve
and check that the gauge reads at least 60 cmH2O.
3. Remove the manometer by unscrewing the two retaining nuts.
2. S
et the gas supply to 15 L/min. Close the bleed valve and check
that the gauge reads no higher than 80 cmH2O.
3. S
et the gas supply to 10 L/min. Close the bleed valve and turn the
Max Pressure Relief valve until the manometer reads 40 cmH2O.
Check that the manometer needle rises and falls smoothly. Gently
rotate the Max Pressure Relief cover over the Max Pressure Relief
knob.
4. R
eset the peak inspiratory pressure to 20 cmH2O and turn off the
gas flow. Testing is now complete.
If the F&P Neopuff / Perivent Infant Resuscitator fails any of these
tests, the valve assembly should be regarded as faulty and replaced
with a new valve assembly (see the parts list (section 4.1.2) for
part numbers). Follow the valve replacement guidelines in section
3.3.6 of this manual, or call your Fisher & Paykel Healthcare service
representative for further information.
3.3.3 Setting Max Pressure Relief to 40 cmH2O
This is required if the Max Pressure Relief has been changed. The
factory setting for the Max Pressure Relief is 40 cmH2O.
4. F it the new manometer into the front panel, tighten the retaining
nuts and reconnect the manometer tube.
5. Refit the front panel to the back cover with the four screws.
6. C
arry out the manometer performance test as per section 3.3.1. It
is recommended to record the lot number from the box label of the
replacement manometer on the maintenance checklist.
3.3.6 Valve System Replacement
NOTE: The valves are an integral part of the valve, panel and
manifold assembly and are not able to be serviced. Please specify
the model number from the parts list (section 4.1.2) when ordering a
replacement valve assembly.
1. Remove the back cover, fixed by four screws.
2. Disconnect the tube from manometer.
3. Remove the manometer by unscrewing the two retaining nuts.
4. F it the existing manometer into the new valve assembly panel,
tighten the retaining nuts and reconnect the manometer tube.
5. Refit the front panel to the back cover with the four screws from
step 1.
6. Carry out the valve system performance test as per section 3.2.2.
It is recommended to record the lot number of the new valve
assembly onto the Set-up and Maintenance Checklist.
N E O P U F F ™ t e c h n i c a l m a nu a l
7
Page 9
3. Service Information continued
3.4 Set-up and Maintenance Checklist
The following table is provided to record the results of the performance tests described in section 3.3. Any components replaced should be
recorded also, as appropriate. The table may be photocopied or otherwise reproduced as required.
Upon receipt of the product and prior to use on a patient, please complete the tests and fill in the table below.
Name:
F&P Neopuff™/Perivent™
Performance Testing Record
Customer Details
Address:
Country:
Phone Number:
Email:
Test Date:
Serial Number:
For 900IW130 only:
Infant Warmer Serial Number:
Tested By:
Next Test Due Date:
Check
Ref
or x
Comments
Manometer Checks:
Replacement Manometer Serial Number:
Accuracy @ 0 cm (+/-2 cmH2O)
3.3.1-2
Accuracy @ 10 cm (+/-2 cmH2O)
3.3.1-3
Accuracy @ 20 cm (+/-2 cmH2O)
3.3.1-3
Accuracy @ 40 cm (+/-2 cmH2O)
3.3.1-3
Needle movement smooth?
3.3.1-4
Valve System Checks:
Pressure greater than 60 cmH2O?
3.3.2-1
Pressure less than 80 cmH2O?
3.3.2-2
Maximum pressure set to 40 cmH2O
3.3.2-3
PIP reset to 20 cmH2O
3.3.2-4
Additional Comments:
N E O P U F F ™ t e c h n i c a l m a nu a l
8
Page 10
4. Assembly Diagrams
4.1 F&P Neopuff™ / Perivent™ Infant
Resuscitator (Post June 2010)
4.1.2 Parts List
Units manufactured Post June 2010
This F&P Neopuff / Perivent module contains a sealed valve assembly
that can not be serviced in the field. If a faulty valve occurs, the entire
panel and valve assembly must be replaced.
#
Description
Part Number
Reqd
1
Handle Neopuff Spare
043043976
1
4.1.1 Assembly Diagram
2
Plug (set of four)
693040706
2
3
Screw M8x20
614040309
8
4
End cap (upper)
043042565
1
5
End cap (lower)
043042564
1
6
Back cover
641040816
1
Model specific panel and
valve assembly
(see Spares Kit section 4.1.3)
(see table in
section 4.1.3)
1
7
8
Manometer kit
043040841
1
9
Cap Valve Neopuff Spare (Blue)
043043977
1
10
Cover, maximum pressure relief valve
043041057
1
11
Foot
693041436
4
12
Screw #8x1” Csk
616050011
4
13
Screw M4x8 Pan hd
(handle attachment)
614040117
2
NOTE: White valve cap (pre-June 2010) can be ordered using 043042345
(1 / bag).
N E O P U F F ™ t e c h n i c a l m a nu a l
9
Page 11
4. Assembly Diagrams continued
4.1.3 Spares Kit
NOTE: All RD900 units manufactured from May 1999 can
accommodate the above spares kit.
Model specific panel and valve assembly Blue (kit includes fascia)
043043595
Fascia & Valve Assy Blue English
043043596
Fascia & Valve Assy Blue German
043043597
Fascia & Valve Assy Blue Italian
043043598
Fascia & Valve Assy Blue Spanish
043043599
Fascia & Valve Assy Blue French
043043600
Fascia & Valve Assy Blue Norwegian
043043601
Fascia & Valve Assy Blue Dutch
043043602
Fascia & Valve Assy Blue Swedish
043043603
Fascia & Valve Assy Blue Portuguese
043043604
Fascia & Valve Assy Blue Danish
043043605
Fascia & Valve Assy Blue Finnish
Model specific panel and valve assembly Purple (kit includes fascia)
043042347
Fascia & Valve Assy English
043042348
Fascia & Valve Assy German
043042349
Fascia & Valve Assy Italian
043042350
Fascia & Valve Assy Spanish
043042351
Fascia & Valve Assy French
043042353
Fascia & Valve Assy Dutch
043042354
Fascia & Valve Assy Swedish
043042355
Fascia & Valve Assy Portuguese
043042356
Fascia & Valve Assy Danish
043042357
Fascia & Valve Assy Finnish
043042358
Fascia & Valve Assy Norwegian
N E O P U F F ™ t e c h n i c a l m a nu a l
10
Page 12
5. Mounting Options
Impact to the F&P Neopuff™ / Perivent™ Infant Resuscitator,
caused by rough handling or the unit being dropped, can damage the
valve system and produce irregular resuscitation pressures.
To help prevent any impact to the device, Fisher & Paykel Healthcare
recommends the use of one of the mounting systems shown below.
5.2 900MR088 Rail Bracket
For mounting the F&P Neopuff / Perivent Infant Resuscitator centrally
on standard rails (2.5 to 5.5 cm x 1 cm / 0.98 to 2.17” x 0.39”).
Also requires RD050-01.
5.1 900RD301 Side-Mounting Block,
900MR170 C-Clamp, 900MR030 Pole Bracket
The 900RD301 Side Mounting Block fits into the dovetail slot on the
side of the Infant Resuscitator. The 900RD301 may then be connected
to a 17 to 40 mm pole using either the 900MR170 C-Clamp or
900MR030 Pole Bracket.
Mounting Option
Parts to order
Quick disconnect (C-Clamp)
900RD301 & 900MR170
Permanent Pole Bracket
900RD301 & 900MR030
5.3 900RD302 Wall Mount Bracket
The 900RD302 Wall Mount Bracket has two tongue fittings that
may be screwed to the back of the F&P Neopuff™ / Perivent™ Infant
Resuscitator by removing the four plastic plugs from the rear panel,
and a T-shaped bracket that should be screwed to the wall.
This allows the infant resuscitator to be quickly and easily removed
from and replaced on the bracket as required.
900RD301
900MR170
C-Clamp
TLET
GAS OU
GAS INL
ET
in
5-15 L/m
900MR030
Pole Bracket
N E O P U F F ™ t e c h n i c a l m a nu a l
11
Page 13
5. Mounting Options continued
5.4 RD050-01 Pole and Rail Central Mount
For mounting the F&P Neopuff™ / Perivent™ Infant Resuscitator
centrally on a pole or on standard rails (2.5 to 5.5 cm x 1 cm /
0.98 to 2.17” x 0.39”), a central mounting block can be affixed to
the back of the unit by removing the four plastic plugs from the rear
panel. The unit will then fit the F&P bracket and clamps.
N E O P U F F ™ t e c h n i c a l m a nu a l
12
Page 14
For more information please contact
your local Fisher & Paykel Healthcare representative
Manufacturer
Fisher & Paykel Healthcare Ltd
15 Maurice Paykel Place
East Tamaki, Auckland 2013
PO Box 14 348 Panmure
Auckland 1741
New Zealand
Tel:
Fax:
Email:
Web:
+64 9 574 0100
+64 9 574 0158
info@fphcare.com
www.fphcare.com
Australia
Tel: +61 3 9879 5022
Fax: +61 3 9879 5232
Austria
Tel: 0800 29 31 23
Fax: 0800 29 31 22
Benelux
Tel: +31 40 216 3555
Fax: +31 40 216 3554
China
Tel: +86 20 3205 3486
Fax: +86 20 3205 2132
France
Tel: +33 1 6446 5201
Fax: +33 1 6446 5221
Germany
Tel: +49 7181 98599 0
Fax: +49 7181 98599 66
185041597 RevM EN © 2011 Fisher & Paykel Healthcare Limited
India
Tel: +91 80 4284 4000
Fax: +91 80 4123 6044
Irish Republic
Tel: 1800 409 011
Italy
Tel: +39 06 7839 2939
Fax: +39 06 7814 7709
Japan
Tel: +81 3 3661 7205
Fax: +81 3 3661 7206
www.fphcare.com
Northern Ireland
Tel: 0800 132 189
Spain
Tel: +34 902 013 346
Fax: +34 902 013 379
Sweden
Tel: +46 8 564 76 680
Fax: +46 8 36 63 10
Switzerland
Tel: 0800 83 47 63
Fax: 0800 83 47 54
Taiwan
Tel: +886 2 8751 1739
Fax: +886 2 8751 5625
Turkey
Tel: +90 312 354 34 12
Fax: +90 312 354 31 01
UK
(EU Authorised Representative)
Fisher & Paykel Healthcare Ltd
Unit 16, Cordwallis Park
Clivemont Road, Maidenhead
Berkshire SL6 7BU, UK
Tel: +44 1628 626 136
Fax: +44 1628 626 146
USA/Canada
Tel: +1 800 446 3908
or
+1 949 453 4000
Fax: +1 949 453 4001