GAYMAR
Thermacare REF TC3000 Operating Manual
Operating Manual
24 Pages
Preview
Page 1
www.gaymar.com P/N 09033-000 7/96
®
Thermacare TC3000 Convective Warming System Aparato de calentamiento convectivo
Operating Manual Manual del usuario
P/N 09033-000 7/96
Important
Symbols used within this manual:
The Thermacare TC3000 Convective Warming Unit is compatible with model TC1050, TC2050, TC2052, TC2054, and TC2060 Warming Quilts.
Attention: consult accompanying documents
The Thermacare Convective Warming System (figure 1, page 4) provides a continuous means of warming patients to help prevent and/or treat hypothermia. The system consists of a Convective Warming Unit (Power Unit) and a disposable Quilt. A connecting hose conducts heated air from the Power Unit to the Quilt.
Dangerous voltage
Protective earth (ground)
Read and understand the Operator's Manual and all precautions prior to using the TC3000 Convective Warming System.
Type BF applied equipment
We suggest that a biomedical engineer perform the FUNCTIONAL CHECK AND SAFETY INSPECTION regularly to assure the TC3000 Power Unit works properly (refer to TC3000 Service Manual).
Off-on switch
Review the SAFETY PRECAUTIONS in section 1.0 (pp. 2-3) prior to each application.
MIN (Fan Only)
Receiving Inspection
LOW (90°F, 32°C)
Upon receipt, unpack the Power Unit. Save all packing material. Perform a visual and mechanical inspection for concealed damage. If any damage is found, notify the carrier at once and ask for a written inspection. Photograph any damage and prepare a written record. Failure to do so within 15 days may result in loss of claim.
MED (100°F, 38°C)
HIGH (110°F, 43°C)
Contact your local dealer or contact Gaymar's Technical Service Department (see back cover for phone numbers) for assistance. NOTE: Extremely high storage temperatures (such as those found in rail cars or automobile trunks on hot summer days) can cause the overtemp thermostats within this device to actuate. Should this occur, the REMOVE FROM USE indicator will light when the Power Unit is turned on. If this happens, the thermostats must be manually reset by qualified personnel such as certified biomedical electronics technicians or certified clinical engineers according to the TC3000 Service Manual instructions.
© 1996. Gaymar Industries, Inc. All rights reserved.
MAX (115°F, 46°C)
REMOVE FROM USE (indicator light, amber color)
Do not use in operating room (OR) or Intensive Care Unit (ICU)
60°F - 85°F (16°C - 29°C)
Ambient operating temperature
www.gaymar.com P/N 09033-000 7/96
Before you begin . . .
www.gaymar.com P/N 09033-000 7/96
Table of Contents
Table of Contents Section 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0
Description Page Safety Precautions ... 2 Introduction ... 4 Operator Control Panel ... 5 Operating Instructions ... 6 Routine Maintenance ... 7 Troubleshooting ... 8 Quilts / Training / Warranty / Returns ... 8 Specifications ... 9
Illustrations Figure 1 2 3 4 5
Description Page Convective Warming System ... 4 Power Unit Mounting ... 4 Pole Mounting Height Limit ... 4 Operator Control Panel ... 5 Attaching Hose to Quilt ... 6
Tables Table 1
Description Page Setting the Prescribed Temperature ... 6
1
1.0
Safety Precautions
Review the following SAFETY PRECAUTIONS prior to using the Power Unit. These SAFETY PRECAUTIONS appear within this manual, the Power Unit label, or the Quilt Instructions for Use.
DANGER • Explosive hazard. Do not use in the presence of flammable anesthetics. • Risk of electric shock. Disconnect power before servicing the TC3000 Power Unit.
WARNING This device generates heated air flow. Excessive heat could cause thermal stress or skin lesions. Failure to follow these precautions could result in death or serious injury: • Use the Convective Warming System only under direction of a physician. A physician's order is required for setting temperature and for continued use of equipment. • If the patient's temperature is not responding or does not reach the prescribed temperature in the prescribed time, or deviates from the prescribed temperature range, notify the attending physician promptly. • Monitor core temperature. Neonates and pediatric patients of low weight will have a tendency to overheat more readily than adults. Failure to monitor core temperature could result in abnormal elevation of body temperature resulting in death or serious injury. • Monitor the patient's temperature, vital signs, and skin condition regularly (every 20 minutes or as prescribed by a physician). Pediatric, temperaturesensitive, and postoperative patients should be checked more frequently. If the patient's temperature is within the (continued next column) 2
WARNING prescribed range, select a lower temperature setting or discontinue therapy. If vital signs become unstable, notify the attending physician immediately. • When treating hemodynamically unstable patients, start therapy at LOW (90°F, 32°C) or MED (100°F, 38°C) temperature setting. • Do not apply heat to lower extremities during aortic cross-clamping. Ischemic injury may occur. • Do not use the MAX (115°F, 46°C) temperature setting in the operating room. • Do not use the TC3000 Power Unit in the MAX (115°F, 46°C) temperature setting when using the TC2050, TC2052, TC2054, or TC2060 Warming Quilts. • Do not use the MAX (115°F, 46°C) temperature setting for patients with any of the following conditions: 1. Low cardiac output 2. Poor peripheral perfusion or peripheral vascular disease 3. Total immobilization 4. Non-sensate patients • Do not apply heat to ischemic tissue. • Do not use the Power Unit unless the hose is connected to a warming quilt. Place your hand under the quilt to verify the air is warm. • Use this Power Unit only with those quilts recommended by the manufacturer (see inside cover of this manual for list). • Repairs should be performed only by qualified personnel such as certified biomedical electronics technicians or certified clinical engineers familiar with repair practices for servicing medical devices, and in accordance with the Service Manual. Damage to the Power Unit or malfunction could otherwise result.
www.gaymar.com P/N 09033-000 7/96
Section 1 - Safety Precautions
1.0
www.gaymar.com P/N 09033-000 7/96
Section 1 - Safety Precautions (cont’d)
Safety Precautions (cont’d) CAUTION
• U. S. Federal law restricts this device to sale by or on the order of a physician. • Evaporative cooling may result if the patient's skin is not dry. • Do not permit air from the quilt to blow onto an open wound. Convective airflow can cause airborne contamination to open wounds. Use tape on quilt to restrict movement of air toward the surgical site. • For grounding reliability, plug only into a grounded outlet labelled “Hospital Grade.” • When using an I. V. pole, do not mount the Power Unit higher than 3 feet (0.9 meter). Otherwise, the Power Unit could tip over.
3
2.0
Introduction
Please read the precautions found in section 1.0 and all operating instructions before attempting to use the Thermacare TC3000 Convective Warming System. In addition, we recommend you request in-service training from your local dealer.
2.1
Convective Warming System
The Convective Warming System (fig. 1) provides a continuous means of warming patients to help prevent and/or treat hypothermia. The system consists of a Convective Warming Unit (Power Unit) and a disposable Quilt. A connecting hose conducts heated air from the Power Unit to the Quilt.
2.2
Figure 1 - Convective Warming System I. V. Pole Clamp
TC3000 Power Unit
The TC3000 Power Unit provides a continuous source of air to the Quilt. The Power Unit consists of a HEPA filter, blower, heater, and temperature controller. Exiting air temperatures may be set to one of five temperature settings -- MIN (fan only), LOW (90°F, 32°C), MED (100°F, 38°C), HIGH (110°F, 43°C), and MAX (115°F, 46°C) -- as appropriate for the application. The Power Unit can be operated while hanging on the rails of a bed, stretcher, on an I. V. pole, or on an optional stand (figs. 2A/2B).
Figure 2A - I. V. Pole or Stand mounted
Bedrail Hook
CAUTION When using an I. V. pole, do not mount the Power Unit higher than 3 feet (0.9 meter) (see figure 3). Otherwise, the Power Unit could tip over.
2.3
Quilt
The Quilt distributes air over the covered areas of the patient through openings in the patient side of the Quilt. Quilts consist of layers of plastic and nonwoven material bonded together. On operating room model Quilts, a pressure-sensitive, nonirritating tape provides a seal between the air from the Quilt and the surgical site. Refer to the instructions enclosed with each Quilt. 4
Figure 2B - Bed rail mounted
CAUTION Do not mount higher than 3 ft. Figure 3 - Pole mounting height limit
www.gaymar.com P/N 09033-000 7/96
Section 2 - Introduction
www.gaymar.com P/N 09033-000 7/96
Section 3 - Operator Control Panel
®
®
Á
Â
Figure 4 - Operator Control Panel
WARNING Failure to follow the following precautions could result in death or serious injury: • Do not use the MAX (115°F, 46°C) setting in the operating room. • Do not use the MAX (115°F, 46°C) setting for patients with any of the following conditions: 1. Low cardiac output 2. Poor peripheral perfusion or peripheral vascular disease 3. Total immobilization 4. Non-sensate patients
À 3.0
Operator Control Panel
The operator controls are identified in figure 4.
3.1
POWER Switch À
The POWER switch controls power to the Power Unit. The POWER INDICATOR light in the switch will be lit when power is on.
3.2
TEMPERATURE SETTING Push Buttons Á
When the Power Unit is turned on, the system defaults to the LOW (90°F, 32°C) temperature setting. Temperature settings are selected by pressing the appropriate push button-- MIN (fan only), LOW (90°F, 32°C), MED (100°F, 38°C), HIGH (110°F, 43°C), or MAX (115°F, 46°C). If electrical power is momentarily interrupted, the existing temperature setting will remain. If electrical power remains off, the TC3000 will default to LOW (90°F, 32°C) when power is restored.
3.3
REMOVE FROM USE Indicator Â
The REMOVE FROM USE indicator will light: • if a patient limit thermostat or heater overtemp thermostat has tripped, or • if the HEPA filter has been installed incorrectly or is missing. The heating element and blower will automatically turn off when either of these conditions occur.
CAUTION If the REMOVE FROM USE indicator lights, turn the Power Unit off immediately. Discontinue use and refer equipment for servicing. 5
3.
Insert the end of the hose into the Quilt opening (fig. 5). Secure the hose tightly.
Read and understand these instructions and the precautions on the inside front cover before using the TC3000 Power Unit.
4.
Plug the Power Unit into a “Hospital Grade” outlet.
5.
Turn the power on.
Any application of the TC3000 Convective warming System must be under the direction of the attending physician.
6.
When the Power Unit is turned on, it defaults to the LOW (90°F, 32°C) setting. Select the prescribed temperature you desire for the patient (see table 1).
IMPORTANT
4.0
Temperature Setting
Operating Instructions
Perform the following start-up procedure each time the TC3000 Convective Warming System is placed into operation: 1.
Select the location for operation of the TC3000 Power Unit. Mount it using the bed rail hooks, pole clamp, or set it on a flat, hard surface. If mounting unit to an I. V. pole, insure unit is mounted no higher than 3 feet (0.9 meter). Keep the air inlets at the bottom of the unit clear of any obstructions. Do not set unit on a bed surface. NOTE: If Power Unit is placed on the floor, increased levels of dust could reduce filter life.
2.
Remove the Quilt from its packaging. Unfold the Quilt and place it over the patient. Refer to the Instructions for Use and Safety Precautions included with each Quilt.
Temperature
MIN
Fan only
LOW
90°F (32°C)
MED
100°F (38°C)
HIGH
110°F (43°C)
MAX
115°F (46°C)
Table 1 - Setting the prescribed temperature
collar
opening hose
tab strap strap collar Figure 5 - Attaching hose to Quilt
6
www.gaymar.com P/N 09033-000 7/96
Section 4 - Operating Instructions
NOTE: • The air temperature surrounding the patient will vary at each setting, depending upon room air temperature and the quality of the insulating blanket or sheet that is used over the Quilt.
4.10 Shutdown Procedure
• Operate the Power Unit only in the specified ambient temperature and voltage ranges (see Specifications, page 9).
5.0
WARNING Failure to follow the following precautions could result in death or serious injury: • Do not use the MAX (115°F, 46°C) setting in the operating room. • Do not use the MAX (115°F, 46°C) setting for patients with any of the following conditions: 1. Low cardiac output 2. Poor peripheral perfusion or peripheral vascular disease 3. Total immobilization 4. Non-sensate patients
7.
8.
Place your hand under the quilt to verify that the Power Unit is warming when set to the LOW (90°F, 32°C), MED (100°F, 38°C), HIGH (110°F, 43°C), or MAX (115°F, 46°C) temperature settings. Place a sheet or blanket over the inflated Quilt to maximize the efficiency and minimize the heat loss from the system.
WARNING Monitor the patient's temperature, vital signs, and skin condition regularly (every 20 minutes or as prescribed by a physician). If patient's temperature reaches the normal range, reduce air temperature or discontinue therapy. If vital signs become unstable, notify the attending physician immediately. Failure to follow these precautions could result in death or serious injury. 9.
If the REMOVE FROM USE indicator lights, discontinue use. Refer servicing to qualified personnel.
www.gaymar.com P/N 09033-000 7/96
Section 4 - Operating Instructions (cont’d) Section 5 - Routine Maintenance
When therapy has been completed, turn the unit off and disconnect the hose from the Quilt. Discard the Quilt.
Routine Maintenance
The following routine maintenance procedures will help insure dependable operation:
5.1
Cleaning the Power Unit
Clean the control panel, enclosure exterior, and hose with a soft cloth lightly dampened with a nonstaining hospital disinfectant. Clean accumulated dirt from the air vents with a vacuum cleaner.
5.2
Storing the Hose
Slide the hose under the bed rail hooks on the back of the Power Unit when not in use.
5.3
Storing the Power Cord
Coil the power cord. Hold it in place with the strap on the side of the Power Unit. Or, hang the power cord from the bracket on the back of the accessory stand.
5.4
Quilt Care
Refer to the instructions included with each Quilt for details on correct application and use. Small rips or tears in Quilt material may be repaired with adhesive tape. Quilts are not sterile. Quilts are intended for single patient use only. Quilts should be discarded after use.
5.5
Servicing the Equipment WARNING Repairs to the Power Unit should be performed by qualified personnel such as certified biomedical electronics technicians or clinical engineers familiar with repair practices for servicing medical devices, and in accordance with the TC3000 Service Manual. Damage to the Power Unit or malfunction may otherwise result. 7
5.6
Changing the Filter
Filters must be changed only by qualified service personnel. Refer to the TC3000 Service Manual for instructions on replacing the filter. Under normal use, replace the HEPA filter inside the Power Unit every 1000 operating hours or 12 months, whichever occurs first.
6.0
7.0
Miscellaneous
7.1
Quilts
Contact your dealer for information on the model TC1050, TC2050, TC2052, TC2054, and TC2060 Quilts available for use with the TC3000 Power Unit.
WARNING
Troubleshooting
Use only quilts recommended by manufacturer. Failure to do so may result in thermal injury.
Most Power Unit operating problems are minor and can be corrected by the operator. The following are troubleshooting hints. Perform corrective action in the sequence listed.
6.1 1. 2. 3. 4.
5.
Quilt Will Not Inflate Make sure the Power Unit is plugged in and turned on. Check both ends of hose for proper connection. Check hose and quilt inlet for kinks. Check Quilt for damage. If air is flowing from hose, try another Quilt. (Small rips or tears in the Quilt may be temporarily repaired with adhesive tape.) Have qualified service personnel check for a clogged or dirty HEPA filter.
6.2 REMOVE FROM USE Indicator is Lit This indicates that the heater and blower have shut down due to the air temperature exceeding the high temperature limit, or that the filter has been incorrectly installed or is missing. Remove the Power Unit from use immediately. Refer servicing to qualified personnel.
WARNING Repairs to the Power Unit should be performed by qualified personnel such as certified biomedical electronics technicians or clinical engineers familiar with repair practices for servicing medical devices, and in accordance with TC3000 Service Manual. Damage to the Power Unit or malfunction may otherwise result.
8
7.2
Accessory Stand
An optional accessory stand for mounting the TC3000 Power Unit is available from your local dealer.
7.3
Customer Training
In-service training is recommended and can be arranged through your dealer.
7.4
Limited Warranty
The Thermacare TC3000 Power Unit is warranted free of defects in material and workmanship under normal use and operation for a period of two years, under the terms and conditions of the Gaymar warranty in place at time of purchase. Consumable items such as filters are excluded. The full warranty is available from Gaymar upon request.
7.5
Return Authorization
Contact your local dealer or Gaymar's Technical Service Department at: Phone: 1 800 828-7341 (716) 662-2551 Fax:
(716) 662-8795
www.gaymar.com P/N 09033-000 7/96
Section 6 - Troubleshooting Section 7 - Miscellaneous
8.0
Specifications
8.3
8.1
Physical
The air temperatures are identified on the operator control panel and indicate the average air temperature at the hose end using a Quilt.
Dimensions
16 1/4"H x 10 3/4"W x 10 1/2"D 42 cm x 28 cm x 27 cm
Weight
15 lbs (6.8 kg)
Enclosure
Thermoplastic
Filter
HEPA filtration down to 0.2 micron particle size
8.2
Classification
Type BF, Class 1, grounded equipment suitable for continuous operation. Not classified for protection against harmful ingress of liquid.
Input
120 V nominal (+4%,-10%), 60 Hz, 12 amps max
Motor
1/25 HP, single phase
Heater
1200W heating element
Power Cord
15 ft, 14 AWG cord with hospital grade plug
Circuit Breaker 15 amp Current Leakage (Earth) 100 microamps max. Ground Resistance
0.15 ohms nominal; 0.50 ohms maximum
Ambient Operating Temperature
60°F to 85°F (16°C to 29°C)
Temperature Setting
Temperature
MIN
Fan only
LOW
90°F (32°C)
100°F (38°C)
Safety System
Dual Patient Safety Temperature Limit Thermostats Either of two independently operating thermostats will shut off the TC3000 at a preset high limit temperature. The heater element and blower will remain off until the thermostat is manually reset. Heater Overtemp Heater overtemp thermostat will shut off the TC3000 in the event of a blower failure or lack of air movement. The heater element and blower will remain off until the thermostat is manually reset. High Temp Indicator The REMOVE FROM USE indicator will light on the front panel when a patient safety thermostat or heater overtemp thermostat has tripped. Also, the heating element and blower will turn off. HEPA Filter Interlock Switch The REMOVE FROM USE indicator will light on the front panel when the HEPA filter has been installed incorrectly or is missing. Also, the heating element and blower will turn off.
8.5 MED
Temperature Settings
NOTE: The air temperature around the patient is affected both by the ambient room temperature and the use of an insulating blanket on top of the Quilt.
8.4
Electrical
www.gaymar.com P/N 09033-000 7/96
Section 8 - Specifications
Quilts
All Quilts are made of a nonwoven layer bonded to a plastic film. Quilt material meets U. S. flammability standards:
HIGH
110°F (43°C)
MAX
115°F (46°C)
• NFPA 702 “Normal Flammability” • CPSC Part 1632 • California Title 19, Subchapter 7 Clear drape material meets NFPA 702 “Normal Flammability.” 9