MAC™ 5500 Resting ECG Analysis System Service Manual 2020299-020 Revision G
The information in this manual only applies to MAC 5500 Resting ECG Analysis Systems with product code SCD. It does not apply to earlier software versions. Due to continuing product innovation, specifications in this manual are subject to change without notice. MUSECASE, MAC, MARS, and 12SLare trademarks owned by GE Medical Systems Information Technologies, a General Electric Company going to market as GE Healthcare. All other trademarks are owned by their respective owners. © 2005-2010, 2012 General Electric Company. All rights reserved.
T-2
MAC™ 5500 Resting ECG Analysis System 2020299-020
Revision G 14 August 2012
Contents Manual Information... 1-3 Revision History... 1-3 Manual Purpose... 1-3 Intended Audience... 1-3 Warnings, Cautions, and Notes... 1-4 Safety Messages... 1-5 Responsibility of the Manufacturer... 1-5 General... 1-6 Equipment Symbols... 1-7 Service Information... 1-8 Service Requirements... 1-8 Equipment Identification... 1-8 Serial Number Format... 1-9 Label Format... 1-9 General Description... 2-3 Front View... 2-3 Back View... 2-4 Connector Identification... 2-5 MAC 5500 Block Diagram... 2-6 Theory of Operation... 2-7 Overview / Block Diagram... 2-7 Power Supplies... 2-10 +3V-C... 2-10 +3V-M... 2-10 +3V-EMI... 2-10 +5V-M... 2-10 +5V-EMI... 2-10 +18V... 2-11 +1.8V... 2-11 +12V... 2-11 REF2V5... 2-11 VAna+, VAna-... 2-12 Clocks... 2-12 Super I/O and FPGA... 2-12 CPU ATMEL AT91RM9200... 2-12 CPU (Stooges)... 2-13 RTC... 2-13
Revision G
MAC™ 5500 Resting ECG Analysis System 2020299-020
i
CPU... 2-13 External Bus Interface... 2-13 Boot Loader... 2-14 FPGA Internal Logic... 2-15 Board ID Register... 2-16 XBus Controller... 2-16 Video Interface... 2-16 Interrupt Controller... 2-21 System Interrupt Timer... 2-22 Acquisition Module Interface... 2-22 Thermal Printhead Interface... 2-23 Serial EEPROM Interface... 2-23 BBus Interface... 2-24 PWM Analog Outputs... 2-24 Beep Generator... 2-24 PC Card Logic... 2-25 SDRAM... 2-25 NAND Flash... 2-25 Secure Digital Card Interface... 2-25 Serial EEPROM... 2-25 Daughter Board Interface... 2-26 VGA LCD/CRT Interface... 2-26 LCD Panel EMI Reduction Components... 2-26 CRT Video DAC / Sync / Buffers... 2-26 Acquisition Module Transceiver / Power Switch... 2-27 Transceiver... 2-27 Acquisition Power Regulator / Switch... 2-27 COMM Port Power Switch / Current Limiter / KISS Power... 2-27 Thermal Printhead Power / Pixel Test Hardware... 2-28 Super I/O Peripheral Controller... 2-28 RS-232 Serial Ports (One Dual Mode RS-232 / IrDA)... 2-28 Clock/Calendar... 2-29 PS2 Keyboard Port... 2-29 The Three Stooges... 2-29 Startup Self Identification... 2-29 BBus... 2-30 Shemp... 2-30 Larry... 2-30 Moe... 2-32 Preparation for Use... 3-3 General... 3-3 Trolley Height Adjustment... 3-4 Installing the MAC 5500 resting ECG analysis system... 3-6 Installing the Optional External Modem Kit... 3-8 Magnetic Card Reader Installation... 3-10 Bar Code Reader Installation... 3-12 Type-S Trolley Assembly... 3-14
ii
MAC™ 5500 Resting ECG Analysis System 2020299-020
Revision G
MAC 5500 ST Requirements and Configuration... 3-15 Compatible Blood Pressure Units... 3-15 Compatible GE Medical Systems Information Technologies Treadmills... 3-16 Analog Treadmills... 3-16 Bicycle Ergometers... 3-17 Assembly Descriptions... 4-3 Introduction... 4-3 Assembly Block Diagram... 4-3 General Fault Isolation... 4-4 Power-up Self-test... 4-4 Power-up Flow Chart... 4-5 Poor Quality ECGs... 4-6 Visual Inspection... 4-7 Diagnostic Tests... 4-8 Introduction... 4-8 Loading the System Diagnostics... 4-8 Substitute Master Password... 4-8 System Diagnostics Main Menu... 4-9 Display Tests... 4-9 Pixel Verification Test... 4-9 Grey Scale Test Patterns... 4-9 Speaker Test... 4-10 Keyboard Test... 4-10 Writer Tests... 4-11 Battery Tests... 4-13 Battery Status... 4-13 Battery Discharge Test... 4-13 Battery Charge Test... 4-14 Print Charge/Discharge Test Results... 4-14 Communication Tests... 4-14 COM Port Loopback Test... 4-14 External Modem Test... 4-15 Internal Modem Test... 4-15 Ethernet Module Test... 4-15 Acquisition Module Test... 4-15 Analog I/O Tests... 4-15 Analog Output Test... 4-15 Analog Input Test... 4-16 DCOut Loopback Test... 4-16 ECGOut/QRSTrigger Loopback Test... 4-16 Floppy Drive Tests... 4-16 Internal Memory Tests... 4-16 SD Card Tests... 4-17 Equipment Problems... 4-18 ECG Data Noise... 4-18 Missing ACI-TIPI Report... 4-18
Revision G
MAC™ 5500 Resting ECG Analysis System 2020299-020
iii
No BP from External Device... 4-18 Treadmill/Ergometer Does Not Move... 4-18 System Errors... 4-19 Frequently Asked Questions... 4-20 Maintenance... 4-20 Save Setups... 4-20 Storing ECGs... 4-20 Format an SD Card... 4-20 Cleaning... 4-20 Battery Capacity... 4-21 System Setup... 4-21 Location Number... 4-21 Patient Questions... 4-21 Passwords... 4-22 Clinical... 4-22 Report Format... 4-22 Hi-Res and Phi-Res... 4-22 Editing... 4-22 Entering Patient Data... 4-23 Transmission... 4-23 Losing Fields When Transmitting... 4-23 Input and Output Connectors... 4-24 A Pins (J1)... 4-24 COM1 (COM3/4) Pins (J3)... 4-24 COM2 Pins (J5)... 4-25 Analog Pins (J6)... 4-25 EXT. VID. Pins (J7)... 4-26 CPU PCB Input/Output Signals... 4-27 Battery Pack/Monitor, J2... 4-27 LCD Backlight, J4... 4-27 Keyboard, J8... 4-27 LCD, J10... 4-28 Power Supply/Motor, J11... 4-29 Thermal Printer, J12... 4-30 Floppy Disk Drive, J13... 4-31 Acquisition Module, J14... 4-32 Introduction... 5-3 Recommended Maintenance... 5-3 Required Tools and Supplies... 5-3 Inspection and Cleaning... 5-4 Visual Inspection... 5-4 Exterior Cleaning... 5-4 Interior Cleaning... 5-4 General... 5-4
iv
MAC™ 5500 Resting ECG Analysis System 2020299-020
Revision G
Thermal Printhead... 5-4 Battery and Patient Cable Replacement... 5-6 Battery Replacement... 5-6 Patient Cable Replacement... 5-6 Disassembly Guidelines... 5-7 Preliminary Steps... 5-7 Trolley Disassembly... 5-7 Type-S Trolley Disassembly... 5-8 Power Supply... 5-8 Removal... 5-8 Reassembly... 5-9 Top Cover... 5-9 Removal... 5-9 Reassembly... 5-10 Display/Keyboard Assembly... 5-11 Removal... 5-11 Display/Keyboard Reassembly... 5-12 Reassembly... 5-12 Main CPU Board... 5-13 Removal of CPU Board... 5-13 Reassembly of CPU Board... 5-13 Software... 5-15 Service Only Setups... 5-16 Restore System Setups... 5-17 Restore Options... 5-17 Disable Options... 5-18 Printhead Replacement... 5-19 Removal... 5-19 Reassembly... 5-19 COMM Board Replacement... 5-19 Setting up LAN Communications... 5-20 Writer Roller/Carriage Assembly... 5-22 Removal... 5-22 Reassembly... 5-22 Trolley Casters... 5-22 Functional Checkout Procedures... 5-25 Tools... 5-26 Visual Inspection... 5-27 Checkout Procedures... 5-27 Operational Checks... 5-27 Diagnostic Tests... 5-28 Electrical Safety Checks... 5-29 Ordering Parts... 6-3 General Information... 6-3
Revision G
MAC™ 5500 Resting ECG Analysis System 2020299-020
v
Field Replaceable Units... 6-4
Appendix A –Abbreviations Standard Abbreviations... A-3
Appendix B – Technical Specifications Technical Specifications... B-3 Display...B-3 Computerized Electrocardiograph...B-3 Writer...B-4 Keyboard...B-4 Electrical...B-4 Vectorcardiography...B-4 Hi-Res and PHi-Res Signal-Averaged Electrocardiography...B-5 Physical...B-5 Environmental...B-5 Safety...B-7
Appendix C – Electromagnetic Compatibility Electromagnetic Compatibility (EMC)... C-3 Guidance and Manufacturer's Declaration - Electromagnetic Emissions...C-3 Guidance and Manufacturer's Declaration - Electromagnetic Immunity...C-4 Guidance and Manufacturer's Declaration - Electromagnetic Immunity...C-5 Recommended Separation Distances...C-6 Compliant Cables and Accessories...C-7
Index
vi
MAC™ 5500 Resting ECG Analysis System 2020299-020
Revision G
Introduction:
1
Revision G
Introduction
MAC™ 5500 Resting ECG Analysis System 2020299-020
1-1
Introduction:
For your notes
1-2
MAC™ 5500 Resting ECG Analysis System 2020299-020
Revision G
Introduction: Manual Information
Manual Information Revision History Each page of the document has the document part number and revision letter at the bottom of the page. The revision letter identifies the document’s update level. The revision history of this document is summarized in the table below. Table 1. Revision History, PN 2020299-020 Revision
Date
Comment
A
25 July 2005
B
14 December 2006
Added FRUs for MobileLink Silex serial server and HHP Model 4600 barcode reader.
C
18 October 2007
Items added to the FRU list.
D
10 October 2008
Added -007 board, 009C SD Card and -002 Comm PC board. Included Functional Checkout Procedure.
E
08 January 2010
Added CAM 14 Assembly V2 in Appendix C
F
20 June 2012
Internal release.
G
14 August 2012
Added 2022332-003 to Parts List.
Initial release of this document.
Manual Purpose This manual supplies technical information for service representative and technical personnel so they can maintain the equipment to the assembly level. Use it as a guide for maintenance and electrical repairs considered field repairable. Where necessary the manual identifies additional sources of relevant information and or technical assistance. See the operator’s manual for the instructions necessary to operate the equipment safely in accordance with its function and intended use.
Intended Audience This manual is intended for the person who uses, maintains, or troubleshoots this equipment.
Revision G
MAC™ 5500 Resting ECG Analysis System 2020299-020
1-3
Introduction: Warnings, Cautions, and Notes
Warnings, Cautions, and Notes The terms danger, warning, and caution are used throughout this manual to point out hazards and to designate a degree or level or seriousness. Familiarize yourself with their definitions and significance. Hazard is defined as a source of potential injury to a person.
1-4
Term
Definition
DANGER
Indicates an imminent hazard which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potential hazard or unsafe practice which, if not avoided, could result in death or serious injury.
CAUTION
Indicates a potential hazard or unsafe practice which, if not avoided, could result in minor personal injury or product/property damage.
NOTE
Provides application tips or other useful information to assure that you get the most from your equipment.
MAC™ 5500 Resting ECG Analysis System 2020299-020
Revision G
Introduction: Safety Messages
Safety Messages Additional safety messages may be found throughout this manual that provide appropriate safe operation information. DANGER Do not use in the presence of flammable anesthetics.
WARNINGS This is Class 1 equipment. The mains plug must be connected to an appropriate power supply. Operate the unit from its battery if the integrity of the protective earth conductor is in doubt.
CAUTIONS This equipment contains no serviceable parts. Refer servicing to qualified service personnel. U.S. Federal law restricts this device to the sale by or on the order of a physician.
Responsibility of the Manufacturer GE Medical Systems Information Technologies is responsible for the effects of safety, reliability, and performance only if:
Revision G
Assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized by us.
The electrical installation of the relevant room complies with the requirements of the appropriate regulations.
The equipment is used in accordance with the instructions for use.
MAC™ 5500 Resting ECG Analysis System 2020299-020
1-5
Introduction: Safety Messages
General The intended use of this device is to record ECG signals from surface ECG electrodes. This device can analyze, record, and store electrocardiographic information from adult and pediatric populations. This data can then be computer analyzed with various algorithms such as interpretive ECG and signal averaging for presentation to the user. This device is intended for use under the direct supervision of a licensed health care practitioner. Failure on the part of the responsible individual, hospital, or institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards. To ensure patient safety, use only parts and accessories manufactured or recommended by GE Medical Systems Information Technologies. Contact GE Medical Systems Information Technologies for information before connecting any devices to this equipment that are not recommended in this manual. If the installation of this equipment, in the USA, will use 240 V rather than 120 V, the source must be a center-tapped, 240 V, single-phase circuit. Parts and accessories used must meet the requirements of the applicable IEC 60601 series safety standards, and/or the system configuration must meet the requirements of the IEC 60601-1-1 medical electrical systems standard. The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system. Consideration relating to the choice shall include:
1-6
use of the accessory in the PATIENT VICINITY; and
evidence that the safety certification of the ACCESSORY has been performed in accordance to the appropriate IEC 60601-1 and/or IEC 60601-1-1 harmonized national standard.
MAC™ 5500 Resting ECG Analysis System 2020299-020
Revision G
Introduction: Equipment Symbols
Equipment Symbols For a list and examples of the symbols that appear on the equipment and packaging, refer to the MAC™ 5500 Resting ECG Analysis System Operator’s Manual (PN 2020299-153).
Revision G
MAC™ 5500 Resting ECG Analysis System 2020299-020
1-7
Introduction: Service Information
Service Information Service Requirements Refer equipment servicing to GE Medical Systems Information Technologies authorized service personnel only. Any unauthorized attempt to repair equipment under warranty voids that warranty. It is the user’s responsibility to report the need for service to GE or to one of their authorized agents.
Equipment Identification The serial number label is located inside the device where shown below.
12A
Every GE Medical Systems Information Technologies device has a unique serial number for identification. The serial number is formatted as shown in “Serial Number Format” on page 1-9. NOTE The examples shown are representative only. Your product label may differ.
1-8
MAC™ 5500 Resting ECG Analysis System 2020299-020
Revision G
Introduction: Service Information
Serial Number Format
### ## ## #### # #
A
B
C
D
E
F
Table 2. Serial Number Format A
1
B
Year manufactured (00-99) 00 = 2000 01 = 2001 02 = 2002 (and so on)
C
Fiscal week manufactured
D
Production sequence number
E
Manufacturing site
F
Miscellaneous characteristic
Product code
1. This manual applies to MAC 5500 with product code SCD.
Label Format B A C D E
Table 3. Equipment Identification Label
Revision G
A
Date of manufacture in YYYY-MM format
B
Part number of product
C
Product code description
D
Serial number (described above)
E
Manufacturing site
MAC™ 5500 Resting ECG Analysis System 2020299-020
1-9
Introduction: Service Information
1-10
MAC™ 5500 Resting ECG Analysis System 2020299-020
Revision G
Equipment Overview:
2
Revision G
Equipment Overview
MAC™ 5500 Resting ECG Analysis System 2020299-020
2-1
Equipment Overview:
For your notes
2-2
MAC™ 5500 Resting ECG Analysis System 2020299-020
Revision G