Technical Publication
VIVID™ S60N AND VIVID™ S70N Basic Service Manual Direction Number: FR091521 Revision: Rev. 1
BASIC SERVICE DOCUMENTATION. COPYRIGHT GENERAL ELECTRIC COMPANY.
© 2016 General Electric Company.
P R E L I M I N A R Y
© 2016 General Electric Company. Manufacturer: GE Vingmed Ultrasound AS Strandpromenaden 45 3191 Horten NORWAY Tel: (+47) 3302 1100 Fax: (+47) 3302 1350
9.DEC.2016 GEHC_FRNT_CVR.FM
DIRECTION FR091521, REVISION 3
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Important Precautions TRANSLATION POLICY
-
-i
DIRECTION FR091521, REVISION 3
-ii
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
-
DIRECTION FR091521, REVISION 3
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
-
-iii
DIRECTION FR091521, REVISION 3
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
-
-v
DIRECTION FR091521, REVISION 3
-viii
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
-
DIRECTION FR091521, REVISION 3
-x
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
-
DIRECTION FR091521, REVISION 3
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
DAMAGE IN TRANSPORTATION All packages should be closely examined at time of delivery. If damage is apparent write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested within this 14 day period.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be performed by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall be performed by qualified GE personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes. The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
OMISSIONS & ERRORS If there are any omissions, errors or suggestions for improving this documentation, please contact the GE Global Documentation Group with specific information listing the system type, manual title, part number or direction number, revision number, page number and suggestion details. Mail the information to: Service Documentation, GE Vingmed Ultrasound AS P.O.Box: 141 3191 HORTEN NORWAY GE employees should use Post-Market Quality Management (PQM) to report service documentation issues.
SERVICE SAFETY CONSIDERATIONS DANGER DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING. WARNING WARNING Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety
glasses, and kneeling pad, to reduce the risk of injury. For a complete review of all safety requirements, see the Chapter 1 Safety Considerations section in the Service Manual.
-
-xiii
DIRECTION FR091521, REVISION 3
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
LEGAL NOTES The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE. GE makes no representations or warranties with respect to the information herein. In addition, the information is subject to change without notice. Every precaution has been taken in the preparation of this document. Nevertheless, GE assumes no responsibility for errors, omissions, or any damages, including special or consequential, resulting from the use of this information. GE will issue updates to this information periodically, as needed. If there are any questions regarding the information contained in this manual, please contact your GE Representative.'
TRADEMARKS All products and their name brands are trademarks of their respective holders.
COPYRIGHTS © 2016 by General Electric Company Inc. All Rights Reserved.
-xiv
-
DIRECTION FR091521, REVISION 3
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Revision History Revision
Date
Reason for change
1
2016-DEC-09
Initial Release
List of Effected Pages (LOEP) Pages
Revision
Pages
Revision
Pages
Revision
Title Page
1
4-1 to 4-42
1
10-1 to 10-30
1
Important Precautions -i to -xvi
1
5-1 to 5-30
1
Back Cover
N/A
TOC
1
6-1 to 6-2
1
1-1 to 1-28
1
7-1 to 7-8
1
2-1 to 2-16
1
8-1 to 8-246
1
3-1 to 3-58
1
9-1 to 9-26
1
-
-xv
DIRECTION FR091521, REVISION 3
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
This page was intentionally left blank.
-xvi
-
DIRECTION FR091521, REVISION 1
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Table of Contents CHAPTER 1 Introduction Overview... 1 - 1 Purpose of Chapter 1... 1 - 1 Contents in this Chapter... 1 - 1 Service Manual Overview... Contents in this Service Manual... Typical Users of the Basic Service Manual... Vivid™ S60N/Vivid™ S70N Models Covered in this Manual... Product Description... Overview of the Vivid S60N/Vivid S70N Ultrasound Scanner... Purpose of Operator Manual(s)...
1-2 1-2 1-2 1-3 1-4 1-4 1-4
Important Conventions... Conventions Used in this Manual... Model Designations... Icons... Safety Precaution Messages... Standard Hazard Icons...
1-5 1-5 1-5 1-5 1-5 1-6
Safety Considerations... 1 - 7 Introduction... 1 - 7 Human Safety... 1 - 7 Mechanical Safety... 1 - 9 Electrical Safety... 1 - 12 Probes... 1 - 12 Peripherals... 1 - 13 Safety and Environmental Guidelines... 1 - 13 Vivid S60N/Vivid S70N Battery Safety (part of P/S)... 1 - 15 Patient Data Safety... 1 - 15 Dangerous Procedure Warnings... 1 - 16 Lockout/Tagout (LOTO) Requirements... 1 - 17 Product Labels and Icons... 1 - 18 Universal Product Labels... 1 - 18 System Rating Label... 1 - 18 Table of Contents
-xvii
DIRECTION FR091521, REVISION 1
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
General Label... Label Descriptions... Vivid S60N/Vivid S70N External Labels... GND Label...
1 - 19 1 - 20 1 - 22 1 - 22
Returning/Shipping Probes and Repair Parts... 1 - 23
-xviii
EMC, EMI, and ESD... Electromagnetic Compatibility (EMC)... CE Compliance... Electrostatic Discharge (ESD) Prevention... General Caution...
1 - 24 1 - 24 1 - 24 1 - 25 1 - 25
Customer Assistance... Contact Information... Phone Numbers for Customer Assistance... Phone and Fax Numbers for Manufacturer...
1 - 26 1 - 26 1 - 27 1 - 28
Table of Contents
P R E L I M I N A R Y DIRECTION FR091521, REVISION 1
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
CHAPTER 2 Site Preparations Overview...2 - 1 Purpose of Chapter 2... 2 - 1 Console Requirements...2 - 2 Unit Environmental Requirements... 2 - 2 Cooling Requirements... 2 - 2 Lighting Requirements... 2 - 2 Time and Manpower Requirements... 2 - 2 Electrical Requirements... 2 - 3 Vivid S60N/Vivid S70N Power Requirements... 2 - 3 Inrush Current... 2 - 3 Site Power Outlets... 2 - 3 Mains Power Plug... 2 - 3 Power Stability Requirements... 2 - 4 EMI Limitations... 2 - 4 EMI Prevention/Abatement... 2 - 5 Probe Environmental Requirements... 2 - 5 Facility Needs...2 - 6 Purchaser Responsibilities... 2 - 6 Mandatory Site Requirements... 2 - 7 Site Recommendations... 2 - 7 Recommended Ultrasound Room Layout... 2 - 8 Minimal Floor Plan Suggestion... 2 - 9 Recommended Floor Plan Suggestion... 2 - 10 Suggested Floor Plan with Ultrasound System and EchoPAC in Same Room 2 - 11 Networking Pre-Installation Requirements... 2 - 12 Stand-alone Unit (without Network Connection)... 2 - 12 Unit Connected to Hospital’s Network... 2 - 12 Purpose of the DICOM Network Function... 2 - 12 DICOM Option Pre-Installation Requirements... 2 - 12 Connectivity Installation Worksheet...2 - 13
Table of Contents
-xix
DIRECTION FR091521, REVISION 1
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
CHAPTER 3 System Setup Overview... 3 - 1 Purpose of Chapter 3... 3 - 1 Setup Reminders... 3 - 2 Average Setup Time... 3 - 2 Setup Warnings... 3 - 2 System Acclimation Time... 3 - 3 Safety Reminders... 3 - 4 Receiving and Unpacking the Equipment... 3 - 5 Warnings for Receiving and Unpacking the Equipment... 3 - 5 Overview... 3 - 5 Unpacking the Cardboard Shipping Carton... 3 - 9 Unpacking the Vivid S60N/Vivid S70N Scanner with Peripherals... 3 - 9 Unpacking the Vivid S60N/Vivid S70N Scanner... 3 - 9 Physical Inspection... 3 - 13 System Voltage Settings... 3 - 13 EMI Protection... 3 - 13 Preparing for Setup... Verifying Customer Order... Physical Inspection... Component Inspection... Damage Inspection Checklist... Front View of the Vivid S60N/Vivid S70N Ultrasound Unit... Rear View of the Vivid S60N/Vivid S70N Ultrasound Unit... Peripheral/Accessory Connector Panel... EMI Protection...
3 - 14 3 - 14 3 - 14 3 - 14 3 - 14 3 - 16 3 - 17 3 - 18 3 - 18
Completing the Setup... 3 - 19 Purpose of this Section... 3 - 19 System Specifications... 3 - 19 System Requirements Verification... 3 - 19 Physical Dimensions... 3 - 19 Mass with Monitor and Peripherals... 3 - 19 Acoustic Noise Level... 3 - 19 Electrical Specifications... 3 - 19 Verification of the Vivid™ S60/Vivid™ S70 Voltage Setting... 3 - 19 Electrical Specifications for the Vivid™ S60/Vivid™ S70... 3 - 19 Connections on the I/O Rear Panel... 3 - 20 Connect Ethernet... 3 - 20 -xx
Table of Contents
P R E L I M I N A R Y DIRECTION FR091521, REVISION 1
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Connect USB Flash Card... 3 - 20 Connections on the Patient I/O Panel... 3 - 20 Connect ECG... 3 - 21 Connecting Probes... 3 - 22 Introduction to Connecting Probes... 3 - 22 Connect a Probe... 3 - 23 Disconnect Probes... 3 - 24 Power on/Boot up... 3 - 24 Power Shut Down... 3 - 24 Complete Power Down... 3 - 24 Configuration...3 - 25 Purpose of this Section... 3 - 25 Vivid™ S60/Vivid™ S70 Configuration... 3 - 25 Contents in this Sub-section... 3 - 25 Select System Settings Screen... 3 - 25 Enter Location... 3 - 26 Adjust Date and Time... 3 - 27 Select Language for User Interface and Online Manuals... 3 - 29 Select Units of Measure and Video Format... 3 - 30 Service Screen Setup... 3 - 31 Contents in this Sub-section... 3 - 31 Open Service Screen... 3 - 31 Select Video Format, PAL or NTSC... 3 - 31 Alphanumeric Keyboard Configuration... 3 - 32 Add Printer... 3 - 33 Optional Peripherals/Peripheral Connection... 3 - 34 Contents in this Sub-section... 3 - 34 Approved Internal Peripherals... 3 - 34 Approved External Peripherals... 3 - 34 External Peripherals for Connection to USB... 3 - 34 External Peripherals for Connection to Ethernet (TCP/IP)... 3 - 34 Connecting the HP Laserjet M451 Color Printer... 3 - 34 Software Options Configuration... 3 - 39 Contents in this Sub-section... 3 - 39 Software Option Introduction... 3 - 39 To Install a Software Option... 3 - 39 Remote Check and Configurations... 3 - 39 Connectivity Overview... 3 - 40 Physical Connection... 3 - 40 Stand-alone Vivid™ S60/Vivid™ S70... 3 - 40 “Sneaker Net” Environment... 3 - 40 Wired Ethernet from Vivid™ S60/Vivid™ S70 to a Workstation... 3 - 40 Connection from Vivid™ S60/Vivid™ S70 to a DICOM Server on a Network . 3 - 40 Table of Contents
-xxi
DIRECTION FR091521, REVISION 1
-xxii
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Connectivity Setup... Introduction... Compatibility... Select TCP/IP Screen... Changing the AE Title and/or Port Number (Port No.)... Set the Remote Archive’s Network Information... To Access the Tcpip Screen... To Add a Server in the Server Config List... To Modify the Setup for a Server in the List... To Delete a Server from the List... Save the New Settings... Wireless Network Configuration... Configuring the Wireless Network Adaptor...
3 - 41 3 - 41 3 - 41 3 - 42 3 - 43 3 - 43 3 - 43 3 - 43 3 - 44 3 - 44 3 - 44 3 - 44 3 - 45
Options Setup... Software Options... Color Video Printer Setup... USB Flash Card Setup... Wireless Network Adaptor... Wizard for Wireless Communication Setup... External Monitor Output Resolution Adjustment...
3 - 51 3 - 51 3 - 51 3 - 51 3 - 51 3 - 52 3 - 55
Paperwork After Setup... Installation Acceptance Test Criteria... User’s Manual(s)... Product Locator Installation Card...
3 - 56 3 - 56 3 - 56 3 - 57
Table of Contents
P R E L I M I N A R Y DIRECTION FR091521, REVISION 1
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
CHAPTER 4 General Procedures and Functional Checks Overview...4 - 1 Purpose of Chapter 4... 4 - 1 Special Equipment Required... 4 - 1 General Procedures...4 - 2 Overview... 4 - 2 Power ON/Boot-up... 4 - 3 Warnings... 4 - 3 Connecting AC (Mains) Power to the Ultrasound System... 4 - 4 Switch ON the AC Power to Vivid™ S60/Vivid™ S70... 4 - 5 Power Shut Down... 4 - 7 Logging On to the Vivid™ S60/Vivid™ S70 as “ADM”... 4 - 8 Introduction... 4 - 8 Select Config via the Touch Screen... 4 - 8 Select Config via the Alphanumeric Keyboard... 4 - 8 The Operator Login dialog... 4 - 8 To Log On as ADM... 4 - 9 Data Management... 4 - 9 Deleting Patient Information... 4 - 10 LCD Monitor Position Adjustment... 4 - 11 To Unlock/Lock the LCD Monitor... 4 - 11 To Unlock and Move the LCD Monitor... 4 - 11 To Lock the LCD Monitor... 4 - 11 Moving and Transporting the Vivid™ S60/Vivid™ S70 Ultrasound Scanner . . 4 - 12 The Casters (Wheels) Control... 4 - 12 To Prepare the Vivid™ S60/Vivid™ S70 to be Moved... 4 - 13 To Ensure Safety while Moving the Vivid™ S60/Vivid™ S70... 4 - 13 Transporting the Vivid™ S60/Vivid™ S70 by Vehicle... 4 - 14 At the New Location... 4 - 14 Cleaning the Trackball... 4 - 15 When Cleaning is Needed... 4 - 15 Functional Checks...4 - 16 Overview... 4 - 16 Performance Checks... 4 - 16 Test Phantoms... 4 - 16 2D Mode (B Mode) Checks... 4 - 17 Introduction... 4 - 17 Preparations... 4 - 17 Adjust the 2D Mode Controls... 4 - 19 M Mode Checks... 4 - 20 Table of Contents
-xxiii
DIRECTION FR091521, REVISION 1
VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
M-Mode Overview... Preparations... Using M-Mode... Conventional M-Mode... Anatomical M-Mode... Curved Anatomical M-Mode... Optimizing M-Mode... Color Mode Checks... Introduction... Color 2D Mode Overview... Color M-Mode Overview... Using Color Mode... Color 2D... Color M-Mode... PW/CW Doppler Mode Checks... Introduction... PW and CW Doppler Overview... Using PW/CW Doppler Modes... Alternative 1... Alternative 2... Optimizing PW/CW Doppler Modes... Tissue Velocity Imaging (TVI) Checks... Introduction... TVI Overview... Using TVI... Optimizing TVI... Probe/Connectors Check... ECG Check... Introduction... Parts Needed... Preparations... ECG Check... Cineloop Check... Introduction... Preparation... Using Cineloop... Selection of a Cineloop... Adjustment of Cineloop Playback... Back End Processor Checks... Operator Panel Check... Peripheral Checks... Printer Checks... Windows Print Test Page... Setup and Check a Printer Service... View the Windows Printer Queues... Mechanical Functions Checks... -xxiv
Table of Contents
4 - 20 4 - 22 4 - 22 4 - 22 4 - 22 4 - 22 4 - 23 4 - 23 4 - 23 4 - 23 4 - 25 4 - 27 4 - 27 4 - 27 4 - 28 4 - 28 4 - 28 4 - 30 4 - 30 4 - 30 4 - 30 4 - 31 4 - 31 4 - 31 4 - 33 4 - 33 4 - 34 4 - 35 4 - 35 4 - 35 4 - 35 4 - 35 4 - 36 4 - 36 4 - 36 4 - 36 4 - 36 4 - 37 4 - 37 4 - 37 4 - 38 4 - 38 4 - 38 4 - 38 4 - 39 4 - 40