Template
4 Pages
Preview
Page 1
CASE 8000
DPMFRM-022 30 Jun 1999
Preventive Maintenance Inspection Report
Page 1 of 4 Customer______________________________ Customer Number______________________ Date___________ FE ___________________________________ FE ID________ Call Number______________________________ Equipment Serial Number ________________ Software Revision _________
Configuration (ä) ___ 12SL ___ Data Storage on File Server
___ Remote Viewing
___ Full Disclosure
Tools 1. Leakage tester 2. Standard hand tools
3. Multimeter 4. Service Manual (p/n 415047-001)
5. ESD Mat and Wrist Strap
Visual Inspection (ä) Inspect the following for excess wear and/or any visual signs of damage. 1. __Coiled patient cable 3.____AC power cord 2. __Patient leadwires 4.____Chassis hardware
Cleaning (ä) 1. ___ Clean all internals of dust
2. ___ Clean external surfaces
Calibration (ä) 1. ___ Writer Tests
2. ___ Key Pad Test
Electrical Safety Checks (ä) Refer to PM Application Form DPMAPP-001-4 class __ for leakage standards for specific voltages. 1. ___ Wall Receptacle Test 2. ___ Ground Continuity (Impedence of protected connection) Test(≤ .1Ω or ≤.2Ω w/power cord) 3. Leakage Ground(Earth) wire leakage to ground test Chassis(Enclosure) leakage to ground test Source Risk(Patient cable leakage to ground) Leakage Sink Risk(Patient cable leakage into patient leads) Leakage
Normal ___ ___ ___ NA
Open Reversed ___ ___ ___ NA
Closed Normal Reversed NA NA ___ ___ ___ ___
___ ___
Checkout Procedure (ä) A.___ Attach ECG simulator, press the RHYTHM key and check the report for proper data, paper tracking and clarity. B.___ Select a protocol and start treadmill(or ergometer) to verify speed and grade( or workload for ergometers) are operating properly. C.___ Depress the Emergency stop button(if installed) while the treadmill(if installed) is running at 3 m.p.h. to verify the treadmill stops correctly. D.___ Depress the Treadmill Stop button on the CASE 8000 stress keyboard while the treadmill(if installed) is running at 3 m.p.h. and verify the treadmill stops correctly. E.___ Perform the Keyboard Test using the Service Screen. F.___ Perform Writer Diagnostics from the Service Screen.
Template Revision D
DPMFRM-022 Page 2 of 4
Comments Briefly describe all repairs/adjustments made and list all parts replaced: ___________________________________________________________________________________________________ ___________________________________________________________________________________________________
Template Revision D
DPMFRM-022 Page 3 of 4
FRU Checkout Procedure FRU Description*
Tools
Visual Inspection
Cleaning
1,3,5
-
-
1,3,5 1,3,5 1,3,5 1,3,5 1,3,5 1,3,5 1,3,5 1,3,5 1,3,5
1,2 -
-
-
-
Calibration
Electrical Safety Tests
Checkout Procedure
1
3
A thru F
1 1 1 1 1 1 1 1 1
3 3 3 3 3 3 3 3 3
A thru F A thru F A thru F A thru F A thru F A thru F A thru F A thru F A thru F
CASE 8000 Stress Test Main CPU Mother Pcb Acquistion I/F Pcb Network Pcb Analog I/O Panel Writer CPU Pcb Writer Power Supply Printhead Hard Drive Floppy Drive Video Pcb
Software Hardware Upgrades (Including Firmware) Supplies Non-Listed FRU’s
1,2 Perform complete PM
Perform applicable Checkout Procedure(s) Perform complete PM
Perform applicable Checkout Procedure(s) No Parts Replaced * The FRU Checkout procedure for any listed FRU also applies to it’s internal PCB’s and parts.
Template Revision D
A thru F
DPMFRM-022 Page 4 of 4
Template Revision D