Wiring Diagram
3 Pages
Preview
Page 1
4
3
J21.1 J21.2 J21.3 J21.4 J21.5 J21.6 J21.7 J21.8 J21.9 J21.10
Configuration Plug Group 1
PHASE
6600-0800-700
NEUTRAL AC GND
Elect. Enclosure
Elect. Enclosure
PHASE 6600-0517- 602
red white blue green black brown orange yellow
+5 V +5V +5V GND GND GND GND +12V +12 V
6600-1027-600
F1
J41.1 J41.2 J41.3 J41.4 J41.5 J41.6 J41.7 J41.8
6600-0708-700 red
J48.1 . J48.2 J48.3 .
black
NEUTRAL 6600-0517- 602
GND J42.1 +5V J42.2 +12V J42.3 +5VSTBY J42.4
6600-0718-700
Rail Components
6600-0883-700 (X2)
HOOD MIDDLE SW
shield
HOOD DOWN SW
6600-1238-600
brown black
brown black
shield
West Rail
Card Cage
HOOD MOTOR
Card Cage
East Rail
B
East Rail
M
East Rail
East Rail
black
.J49.1 .J49.2 J38.1 J38.2 J38.3 J39.1 J39.2
Base Pedestal
DOWN
Ebase East Foot
1 2 3 4 5
UP
6600-0877-700
ESD shield
red brown black shield
1 2 3 4 5
white orange blue green red black
11 22 33
6600-0750-700
DOWN
Ebase West Foot
1 2 3 4 5
UP
6600-0877-700
ESD shield
SPEAKER
6600-1168-600
Ebase Pedestal
1 2 3 4
ESD shield
6600-0737-701
1 2 3 4
1 2 3 4
West 6600-0737-702
ESD shield
brown black shield
1 2 3 4
J30.1 J30.2 J30.3 J30.4 J30.5 J30.6 J30.7 J30.8 J30.9
J33.1 J33.2 J33.3 J33.4 J33.5
J45.1 J45.2
Ebase Pedestal
. .
PHASE
F3
SW2/NEUTRAL
F4
Ebase Pedestal
out .
Ground Equalization Post
.
6600-0870-700 Heatsink Bed Chassis Cover 2 NO.
REVISED AND REDRAWN REVISIONS
02-7323 8/12/02
Elect. Enclosure Elect. Enclosure
Elect. Enclosure
J54.1 . J54.2 J54.3 J54.4 .
AC Expansion
PHASE
J53.1 J53.2 J53.3 J32.1 J32.2 J32.3
SW/NEUTRAL PHASE NEUTRAL
3 2 1
5 6 7
Humidity Sensor (bottom view) 2
t
1 2
6600-0784-700 C
Heatsink Thermistor 6600-0819-700
Expansion Slots
2
t
6600-0702-700
1
1
1
1
1
black brown red
1 2 3 4 5 6 7 8 9
orange yellow green blue violet
red black green
1 2 3 4 5 6 7 8 9
yellow brown blue orange grey
red brown black shield
1
Black
blue green/yellow brown
6600-1165-600 + -
AC AC
+ -
6600-1057-600 1 2 3 4 5
6600-0729-700
1 2 3 4 5
red black brown orange yellow
HTR DOOR OPEN
HTR DOOR CLOSED HTR DOOR OPEN
Heater Door Switches 6600-1154-600
. .
J51.1 J51.2 J51.3
Radiant Warmer
(Shown with heater doors in closed position) blue green brown
1 2 3
1 2 3
black green black
1 2 3
1
Heater
2
Heater Housing
6600-0867-700
6600-1535-700
6600-1177-600
6600-0747-700 B
Configuration Plug yellow blue
6600-0748-701 ONLY
6600-1003-600
Incubator Airflow Motor
M
1 2 3
6600-0710-700
Solid State Relay
A B C
orange yellow black
11 22 33 44
6600-0882-700
Isolation Transformer
Solid State Relay AC AC
1 2
2
white white
1 2
6600-1041-700 x 2 orange white
6600-0864-700
1 2
1 2 blue
2
1
1
Tstat
2
black
green black
Incubator
Heater (X2)
white white
black white
1
6600-1019-600
6600-0709-700 Optical Fan Sensor
Humidifier Isolation Transformer 6600-1328-600 brown black
No Connection J56.1 J56.2
red orange
115 VAC Configured 1
T1
7
2 5
8 11
6
12
230 VAC Configured 1
yellow
2 5 6
T1
6600-0820-700
Configuration Plug Group 2
7
8 11 yellow 12
grn/yel black brown brown
Elect. Enclosure
6600-0802-700
Humidifier
Add Water Tstat 44 33 22 11
green red white white
1
2
1
2
Heater 6600-1018-600
6600-0714-700
6600-1526-700
brown
11 22
brown
6600-0803-700 1
1
2
black
brown black shield
Overtemp Tstat 6600-1013-600
1 2 3 4
1 2 3 4
red brown black shield
Rh Bottle Open Sw Rh Bracket
6600-0730-700
6600-1519-700
2
J56.5 J56.6 3
1
4
2
P G N A
Power Switch/Circuit Breaker 6600-1014-601 (9 Amp, 230V) 6600-1014-602 (12 Amp, 115V)
brown black red brown black
1
2
1
2
6600-0562-603 X2
brown blue
6600-0583-600 (X3)
3
Elect. Enclosure
Line Filter Elect. Enclosure
6600-1006-600
6600-0731-700 UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES TOLERANCES ARE:
Circuit Breakers 3.5A.
shield
Outlets
4
yellow green blue black
green red white brown orange blue
J6.1 J6.2 J6.3 J6.4 J6.5 J6.6
red black
BY 5
t
6600-0716-701
6600-1003-600
RYP AHR
ECO DATE
blue
Humidifier Isolation Adapter Board 6600-1525-700
PCB SSR
6600-0863-700 (X2)
RT5
black
F4.1 F4.2
6600-1119-6 00 in
1 2
2
t
2
RT4 1
1 2 3
6600-1226-600
Elect. Enclosure
red
F4
A
1 2 3
White
Configuration Plug Group 1
Brown Orange Yellow Blue
F3.1 F3.2
Main Bracket
2
t
6600-0798-700 (reusable) or 6600-0799-700 (disposable) 1
calibrate mode short 5 to 6 and 7 to 8
brown
RT3
Patient Probe
6600-0701-700
black red
F3
6600-0237-700
6600-0743-700
J50.1 J50.2
t
1 2 3 4
2 1
t
HTR DOOR CLOSED
115V=6600-1324-602 230V=6600-1324-601
PHASE SW/NEUTRAL
11 22
5 6 7
6600-0752-700
Fuse F3 and F4
6600-0753-701
3 2 1
RT2
1
red black
1
red brown black shield
6600-0711-700
white brown
EBASE ASSEMBLY
Hood Foot Activate
J40.1 J40.2 J40.3 J40.4 J40.5 J40.6
6600-0713-700
M
Hood Foot Activate East
shield
yellow orange brown green red black
Ebase Pedestal 6600-0715-700
red brown black shield
1 2 3 4 5
11 22 33 44 55 66
RT6
1 2 3 4
6600-0742-700
2
RT7 1
1 2 3
normal mode short 1 to 2 and 3 to 4
Jumper
6600-0816-700
PHASE SW/NEUTRAL
yellow green blue black
orange blue red green
6600-0815-700
J46.1 J46.2 J46.3
red shield
Elect. Enclosure
Elect. Enclosure
J36.1 J36.2 J36.3 J36.4 J36.5 J36.6 J36.7 J36.8 J36.9
brown black
shield
J31.1 J31.2 J31.3 J31.4
1 2 3
JP1.1 JP1.2 JP1.3 JP1.4 JP1.5 JP1.6 JP1.7 JP1.8
1 2 3 4 5 6
HOOD UP SW
1
black brown red
6600-0704-700 F2
1 2 3 4
1 12 23 34 4
Power Supply
J47.1. J47.2 .
1 2 3 4
1 2 3 4 5 6
red black
D
Humidity Sensor (bottom view)
blue green yellow orange shield
1 2 3 4
1N5817
BATTERY 6600-1024-600
J2.1 J2.2 J2.3
6600-0716-702
black brown red
1 2 3 4 5
1 2 3 4 5 6
4
PT 1
green yellow blue shield
1 2 3 4 5
1 2 3 4
3
2
J43.1 J43.2 J43.3 J43.4
J4.1 J4.2 J4.3 J4.4
shield
black brown
J3.1 J3.2
PT 2
RT1 brown black orange red
1 2 3 4 5 6
1
6600-1216-600
1 2 3 4
2
Standby Switch
1 2 3 4
red black red black
300
brown black orange red green yellow blue
1 2 3 4
1
red black red black
red orange yellow green blue
2
t
brown
6600-1512-700
1 2 3 4 5 6
6600-0707-700
brown
6600-1513-700 J1.1 J1.2 J1.3 J1.4 J1.5 J1.6 J1.7
J5.1 J5.2 J5.3 J5.4
blue orange
1
3 2
4
6600-0617-700 yellow brown
Elect. Enclosure
C
NURSE CALL 6600-0243-700
Control Board 6600-0705-700
1 2 3 4
PHASE SW2/NEUTRAL
6 5 4 3 2 1
J52.1 . J52.2 .
6600-0724-700 RELAY BOARD
black green yellow
1 2 3 4 5 6
J44.1 J44.2 J44.3 J44.4
black white
5
7 6
Configuration Plug Group 1 8
Ebase / Hood Transformer 6600-1023-600
red white violet black
RT2 2 1
t
red black
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 8 9
1 2 3 4
Card Cage
J9.1 J9.2 J9.3 J9.4 J9.5 J9.6 J9.7 J9.8 J9.9 J9.10 J9.11 J9.12 J9.13 J9.14 J9.15 J9.16 J9.17 J9.18 J9.19 J9.20 J9.21 J9.22 J9.23 J9.24 J9.25 J9.26 J9.27 J9.28 J9.29 J9.30 J9.31 J9.32 J9.33 J9.34 J9.35 J9.36 J9.37 J9.38 J9.39 J9.40 J9.41 J9.42 J9.43 J9.44 J9.45 J9.46 J9.47 J9.48 J9.49 J9.50
red brown orange
red yellow orange brown black green
J31.1 J31.2 J31.3 J31.4 J31.5 J31.6
1 2 3 4 5 6
6600-0746-700
1 2 3 4 5 6 7 8
6600-0784-700
J30.1 J30.2 J30.3 J30.4 J30.5 J30.6
1 2 3 4
shield
black brown green orange yellow red
J10.1 J10.2 J10.3 J10.4 J10.5 J10.6
1 2 3 4 5 6
1 2 3 4 5
J35.1 J35.2 J35.3 J35.4 J35.5 J35.6 J35.7 J35.8
red brown black black red brown
RT1 1
black
2 1
DOWN
East Hood Hand Control
1 2 3 4 5
J37.1 J37.2 J37.3 J37.4 J37.5 J37.6 J37.7 J37.8 J37.9 J37.10 J37.11 J37.12 J37.13 J37.14 J37.15 J37.16 J37.17 J37.18 J37.19 J37.20 J37.21 J37.22 J37.23 J37.24 J37.25 J37.26 J37.27 J37.28 J37.29 J37.30 J37.31 J37.32 J37.33 J37.34 J37.35 J37.36 J37.37 J37.38 J37.39 J37.40 J37.41 J37.42 J37.43 J37.44 J37.45 J37.46 J37.47 J37.48 J37.49 J37.50
6600-0744-700
Card Cage
UP
6600-0736-701
1 2 3 4 5
orange yellow blue black
5 lemoenv
6600-0783-700
brown black orange red green yellow blue
1 2 3 4 5 6 7
4
6600-1531-700
2 1
shield
1 2 3 4 5
6600-0706-700
1 2 3 4 5 6
DOWN
green
6
6600-0736-702 West Hood Hand Control
6600-0717-700
Environmental Connector viewed from wire termination side of panel connector
2 3
6
6600-1530-700 J7.1 J7.2 J7.3 J7.4 J7.5 J7.6 J7.7
1
8 7
6600-0703-700 red brown
1 2 3 4 5 6 7 8
EL DISPLAY 6600-1025-600
3
UP
6600-0822-700
P G N
SCALE
Configuration Plug Group 2
6600-0739-700
5
230V
1 2 3 4 5 6 7 8
Scale Connector viewed from wire termination side of panel connector
2 4
115V
1 2 3 4 5 6 7 8
4
P G N
1 2 3 4 5 6
Yellow color wires
1 2 3 4 5 6
Orange color wires
5
Systems with Panel Connector on Environmental Sensor
green red blue orange
J23.4 J23.3 J23.2 J23.1
TP2.1 + 5 V TP2.2 AUDIO_FREQ TP2.3 + 5 V STBY TP2.4 GND (D)
3
6
TP3.3 Diplay Brightness B
GRAPHICS DRIVER 6600-0723-700
1
2
7
6600-0749-702
6600-0749-701
1
J26.1 J26.2 J26.3 J26.4 J26.5 J26.6
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
J27.1 J27.2 J27.3 J27.4 J27.5 J27.6 J27.7 J27.8 J27.9 J27.10 J27.11 J27.12 J27.13 J27.14 J27.15 J27.16 J27.17 J27.18 J27.19 J27.20
+vhall S1 S2 S3 -vhall
8
red yellow orange white brown black
J60.1 J60.2 J60.3 J60.4 J60.5 J60.6
LED DISPLAY TP3.1 Display Brightness A TP3.2 LED Voltage 6600-0735-700
D
J24.1 J24.2 J24.3 J24.4 J24.5 J24.6
P G N
6600-1007-600
1 2 3 4 5 6
J25.1 J25.2 J25.3 J25.4
1 2 3 4 5 6
5 6 7 8 9 10
TP1.1 1.2 V ind TP1.2 1.2 V TP1.3 V thref TP1.4 - 5 VAN TP1.5 + 5 VAN TP1.6 GND (A)
230V
1 2 3
TOUCH PANEL 4
J20.1 J20.2 J20.3 J20.4 J20.5 J20.6 J20.7 J20.8 J20.9 J20.10 J20.11 J20.12 J20.13 J20.14
115V
WIRING DIAGRAM FOR GIRAFFE OMNIBED
1
6600-0725-700
J8.1 J8.2 J8.3 J8.4 J8.5 J8.6 J8.7 J8.8 J8.9 J8.10 J8.11 J8.12 J8.13 J8.14
1 2 3 4 5 6
6600-0748-702 Yellow color wires
1 2 3 4 5 6
6600-0748-701 Orange color wires
2
ROTARY ENCODER 6600-1026-600
J22.1 J22.2
5
6600-2323-100
FRACTIONS
COPYRIGHT 2000 BY OHMEDA MEDICAL THIS DOCUMENT CONTAINS CONFIDENTIA L OR PROPRIETARY INFORMATI ON OF OHMEDA MEDICAL. NEITHER THIS DOCUMENT NOR THE INFOR MATION THERIN IS TO BE REPRODUCED, DISTRIBUTED, USED OR DISCLOSED EITHER IN WHOLE OR IN PART, EXCEPT AS SPECIFICALLY AUTHORIZED BY O HMEDA MEDICAL. 2
+ -
DECIMALS
+ .XX - -+ -.XXX + -.XXXX -
APPROVALS
AHR
OHMEDA MEDICAL
6/6/00
LAUREL, MD 20723
CHECKED
STR
WIRING DIAGRAM,
6/12/00
ENGINEERING
SMF CAD
DO NOT SCALE DRAWING SURFACE ROUGHNESS
DATE
DRAWN
RECORDS
6/11/00 Size
GIRAFFE OMNIBED AND INCUBATOR DRAWING NUMBER
D PROJECT
6600-2323-100 Sheet
SCALE
ICP011 1
NONE
of
1
3
4
Elect. Enclosure
6600-0708-700 red
J48.1. J48.2 J48.3.
black
Card Cage
B
PHASE SW/NEUTRAL
J46.1 J46.2 J46.3
.J49.1 .J49.2 J38.1 J38.2 J38.3
Base Pedestal
Ebase East Foot 6600-0877-700
1 2 3 4 5
UP ESD shield
1 2 3 4 5
red brown black shield
white orange blue green red black
11 22 33
6600-0750-700
Ebase West Foot 6600-0877-700
DOWN
1 2 3 4 5
UP ESD shield
1 2 3 4 5
yellow orange brown green red black
11 22 33 44 55 66
J40.1 J40.2 J40.3 J40.4 J40.5 J40.6
shield
J30.1 J30.2 J30.3 J30.4 J30.5 J30.6 J30.7 J30.8 J30.9
Ebase Pedestal 6600-0715-700
red brown black shield
SPEAKER
6600-1168-600 Ebase Pedestal
white brown
J33.1 J33.2 J33.3 J33.4 J33.5
J45.1 J45.2
PHASE SW/NEUTRAL
.J50.1 .J50.2
PHASE
F3
SW2/NEUTRAL
F4
in
Bed Chassis Cover
orange yellow green blue violet
AC Expansion
t
2
1 2
t
6600-1226-600
RT5
5 6 7
Humidity Sensor (bottom view)
RT3 2
t
1 2
6600-0784-700 C
Heatsink Thermistor 6600-0819-700
2
t
Expansion Slots
2
t
6600-0716-701
1 2 3 4 5 6
RT4 1
1 2 3
3 2 1
1 2 3 4 5 6
1 2 3
yellow green blue black
1 2 3 4 5 6
calibrate mode short 5 to 6 and 7 to 8
green red white brown orange blue
J6.1 J6.2 J6.3 J6.4 J6.5 J6.6
6600-0702-700
1
1
1
1
Black
Isolation Transformer 6600-1165-600
6600-1534-700
B
red black green
1 2 3 4 5 6 7 8 9
A B C
yellow brown blue orange grey
M
+ -
6600-1057-600 1 2 3 4 5
6600-0729-700
Configuration Plug
6600-1003-600
Incubator Airflow Motor
PHASE SW/NEUTRAL PHASE NEUTRAL
.J51.1 J51.2 .J51.3
1 2
2
white white
1 2
6600-1041-700 x 2 orange white
6600-0864-700
1 2
1 2 blue black
1
1
Incubator
Heater (X2) 2
white white
green black
2
1
Tstat 6600-1019-600
6600-0709-700 Optical Fan Sensor
Humidifier Isolation Transformer 6600-1328-600 brown black
No Connection J56.1 J56.2
red orange
115 VAC Configured 1
T1
7
2 5
8 11
6
12
230 VAC Configured 1
yellow
2 5 6
T1
7
6600-0820-700
Configuration Plug Group 2
6600-0802-700
8 11 yellow 12
grn/yel black brown brown
Elect. Enclosure
44 33 22 11
1
2
1
2
brown
11 22
brown
6600-0803-700 1
black
Overtemp Tstat
6600-1018-600
1
2
6600-1013-600 Heater
6600-0714-700
6600-1526-700
Humidifier
Add Water Tstat green red white white
brown black shield
1 2 3 4
1 2 3 4
red brown black shield
6600-0730-700
2
Rh Bottle Open Sw Rh Bracket
6600-1519-700
J56.5 J56.6
Humidifier Isolation Adapter Board 6600-1525-700
J32.1 J32.2 J32.3
yellow blue
6600-0748-701 ONLY
AC AC
black white
1 2 3 4 5
red black brown orange yellow
F4.1 F4.2
.J53.1 J53.2 .J53.3
6600-0870-700
Elect. Enclosure
2
blue
1
3
1
4
2
P G N
Power Switch/Circuit Breaker 6600-1014-601 (9 Amp, 230V) 6600-1014-602 (12 Amp, 115V)
out
Ground Equalization Post
Elect. Enclosure Elect. Enclosure
1 2 3 4 5 6 7 8 9
F4
PCB SSR
Heatsink
6600-0742-700
2
6600-0798-700 (reusable) or 6600-0799-700 (disposable)
Solid State Relay
black brown red
F3.1 F3.2
Main Bracket
J54.1. J54.2 J54.3 J54.4.
RT6
1
red black
F3
6600-1119-6 00
6600-0863-700 (X2)
No Connection
115V=6600-1324-602 230V=6600-1324-601
6600-0237-700
Ebase Pedestal
RT7 1
1 2 3
normal mode short 1 to 2 and 3 to 4
Elect. Enclosure Brown Orange Yellow Blue
Fuse F3 and F4
6600-0753-701
No 1 Connection 2
A
1 2 3 4
Patient Probe
White
Configuration Plug Group 1
black red
6600-0711-700
6600-0713-700
M EBASE ASSEMBLY
1 2 3
6600-0701-700
6600-1536-700
J39.1 J39.2 DOWN
brown black red brown black
1
2
1
2
6600-0583-600 (X3)
Elect. Enclosure
Size
3
LAUREL, MD 2072 3
GIRAFFE OMNIBED AND INCUBATOR
D
DRAWING NUMBER
PROJECT 2
6600-2323-100
1
Rev
2
Sheet
SCALE
ICP011 4
6600-1006-600
WIRING DIAGRAM,
6600-0562-603 X2
brown blue
Elect. Enclosure
OHMEDA MEDICAL
Circuit Breakers 3.5A.
shield
A
Line Filter
6600-0731-700
Outlets 5
5 6 7
orange blue red green
6600-0815-700
Card Cage
East Rail
3 2 1
J31.1 J31.2 J31.3 J31.4
J36.1 J36.2 J36.3 J36.4 J36.5 J36.6 J36.7 J36.8 J36.9
6600-0883-700 (X2)
West Rail
NEUTRAL 6600-0517- 602
GND J42.1 +5V J42.2 +12V J42.3 +5VSTBY J42.4
1
black brown red
6600-0704-700 F2
yellow green blue black
11 22 33 44
AC GND
Elect. Enclosure
1 2 3 4
t
1 2 3 4
NEUTRAL
6600-0716-702
Jumper
1 2 3 4 5 6
6600-0800-700
J41.1 J41.2 J41.3 J41.4 J41.5 J41.6 J41.7 J41.8
1 2 3 4
2 1
t
1 2 3 4
PHASE
PHASE 6600-0517- 602
red white blue green black brown orange yellow
+5 V +5V +5V GND GND GND GND +12V +12 V
6600-1027-600
F1
green yellow blue shield
black brown red
JP1.1 JP1.2 JP1.3 JP1.4 JP1.5 JP1.6 JP1.7 JP1.8
1 2 3 4
RT2
1
red black
brown
1 2 3 4
Power Supply
J47.1 . J47.2.
J2.1 J2.2 J2.3
D
Humidity Sensor (bottom view)
blue green yellow orange shield
shield
1 2 3 4 5 6
red black
PT 1
1 2 3 4 5
1 2 3 4
4
J4.1 J4.2 J4.3 J4.4
1 2 3 4 5
1 2 3 4 5 6
2
1N5817
PT 2
RT1 brown black orange red
black brown
J3.1 J3.2
2
t
brown
6600-1512-700
1 2 3 4 5 6
6600-1216-600
BATTERY 6600-1024-600
J43.1 J43.2 J43.3 J43.4
red orange yellow green blue
6600-1513-700 brown black orange red green yellow blue
RT2 2 1
t
red black
1 2 3 4 5 6 7 8
2 1
3
red black red black
green yellow
2 1
1
300
1 2 3 4
2
Standby Switch
1 2 3 4
black
1 2 3 4 5 6
1
red black red black
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 8 9
1 2 3 4
6600-0707-700
red brown orange
6600-0243-700
J1.1 J1.2 J1.3 J1.4 J1.5 J1.6 J1.7
RT1 1
black
brown
6600-0784-700 red yellow orange brown black green
J31.1 J31.2 J31.3 J31.4 J31.5 J31.6
NURSE CALL
6600-0705-700
J5.1 J5.2 J5.3 J5.4
blue orange
1
3 2
4
6600-0617-700 yellow brown
Elect. Enclosure
C
Rail Components
PHASE SW2/NEUTRAL
6 5 4 3 2 1
J52.1. J52.2.
6600-0724-700 RELAY BOARD
CONTROL BOARD
5 lemoenv
6600-1531-700
J30.1 J30.2 J30.3 J30.4 J30.5 J30.6
P G N
black white
5
7 6
8
Configuration Plug Group 1
J44.1 J44.2 J44.3 J44.4
J9.1 J9.2 J9.3 J9.4 J9.5 J9.6 J9.7 J9.8 J9.9 J9.10 J9.11 J9.12 J9.13 J9.14 J9.15 J9.16 J9.17 J9.18 J9.19 J9.20 J9.21 J9.22 J9.23 J9.24 J9.25 J9.26 J9.27 J9.28 J9.29 J9.30 J9.31 J9.32 J9.33 J9.34 J9.35 J9.36 J9.37 J9.38 J9.39 J9.40 J9.41 J9.42 J9.43 J9.44 J9.45 J9.46 J9.47 J9.48 J9.49 J9.50
black brown green orange yellow red
J10.1 J10.2 J10.3 J10.4 J10.5 J10.6
4
6600-0783-700
brown black orange red green yellow blue
1 2 3 4 5 6 7
1 2 3 4 5 6
J37.1 J37.2 J37.3 J37.4 J37.5 J37.6 J37.7 J37.8 J37.9 J37.10 J37.11 J37.12 J37.13 J37.14 J37.15 J37.16 J37.17 J37.18 J37.19 J37.20 J37.21 J37.22 J37.23 J37.24 J37.25 J37.26 J37.27 J37.28 J37.29 J37.30 J37.31 J37.32 J37.33 J37.34 J37.35 J37.36 J37.37 J37.38 J37.39 J37.40 J37.41 J37.42 J37.43 J37.44 J37.45 J37.46 J37.47 J37.48 J37.49 J37.50
J7.1 J7.2 J7.3 J7.4 J7.5 J7.6 J7.7
6
red white violet black
Ebase / Hood Transformer 6600-1023-600
J25.1 J25.2 J25.3 J25.4
orange yellow blue black
6600-0746-700
6600-0706-700
6600-1530-700
green
J35.1 J35.2 J35.3 J35.4 J35.5 J35.6 J35.7 J35.8
6600-0717-700
Environmental Connector viewed from wire termination side of panel connector
2 3
6
6600-0703-700 red brown
1 2 3 4 5 6 7 8
1
8 7
3
SCALE
Configuration Plug Group 2
1 2 3 4 5 6 7 8
6600-0739-700
6600-0822-700
P G N
1 2 3 4 5 6 7 8
4
2 4
5
TP3.3 Diplay Brightness B
Scale Connector viewed from wire termination side of panel connector
TP2.1 + 5 V TP2.2 AUDIO_FREQ TP2.3 + 5 V STBY TP2.4 GND (D)
230V
3
EL DISPLAY 6600-1025-600
green red blue orange
J23.4 J23.3 J23.2 J23.1
5
115V
2
J26.1 J26.2 J26.3 J26.4 J26.5 J26.6
1
1 2 3 4 5 6
Yellow color wires
1 2 3 4 5 6
Orange color wires
1
7 6
red yellow orange white brown black
J60.1 J60.2 J60.3 J60.4 J60.5 J60.6
6600-0723-700
+vhall S1 S2 S3 -vhall
6600-0749-702
LED DISPLAY_1 TP3.1 Display Brightness A TP3.2 LED Voltage 6600-0735-700
GRAPHICS DRIVER
Systems with Panel Connector on Environmental Sensor
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
J27.1 J27.2 J27.3 J27.4 J27.5 J27.6 J27.7 J27.8 J27.9 J27.10 J27.11 J27.12 J27.13 J27.14 J27.15 J27.16 J27.17 J27.18 J27.19 J27.20
P G N
8
6600-0749-701
J24.1 J24.2 J24.3 J24.4 J24.5 J24.6 J21.1 J21.2 J21.3 J21.4 J21.5 J21.6 J21.7 J21.8 J21.9 J21.10
Configuration Plug Group 1 D
1
1 2 3 4
6600-1007-600
1 2 3 4 5 6
1 2 3 4 5 6
5 6 7 8 9 10
TP1.1 1.2 V ind TP1.2 1.2 V TP1.3 V thref TP1.4 - 5 VAN TP1.5 + 5 VAN TP1.6 GND (A)
230V
1 2 3
TOUCH PANEL 4
J20.1 J20.2 J20.3 J20.4 J20.5 J20.6 J20.7 J20.8 J20.9 J20.10 J20.11 J20.12 J20.13 J20.14
115V
WIRING DIAGRAM FOR GIRAFFE INCUBATOR
2
6600-0725-700
J8.1 J8.2 J8.3 J8.4 J8.5 J8.6 J8.7 J8.8 J8.9 J8.10 J8.11 J8.12 J8.13 J8.14
1 2 3 4 5 6
6600-0748-702 Yellow color wires
1 2 3 4 5 6
6600-0748-701 Orange color wires
3
ROTARY ENCODER 6600-1026-600
J22.1 J22.2
5
6600-2323-100
NONE
of
2
3
5
4
6600-2323-100
3
2
OXYGEN MODULE
WIRING DIAGRAM FOR GIRAFFE MODULES
6600-1502-700 6600-0604-700 Sensor Plug
D
J82.1 J82.2 J82.3 J82.4 J82.5 J82.6
6600-1043-700 Servo Oxygen Module
green orange red yellow black violet brown blue
1 2 3 4 5
1 2
1 2
red 11 black orange 22 brown 33 44
Flow Valve Assembly 2
red brown yellow green
J83.1 J83.2 J83.3 J83.4
brown black
P92.1 P92.2 P92.3
1 2
1 2
6600-0618-700
2
PAD 1 PAD 2 PAD 3
red 1 black 2
1 2
O2
Oxygen Sensor 6600-1278-600
Calibration Fan 6600-1510-700
Calibration Valve 6600-0876-400
6600-1501-700 1 2
6600-1278-600 D
P93.1 P93.2 P93.3 P93.4
red black
6600-0602-700 1
Oxygen Sensor
-
P92.1 CONNECTS TO PAD 3 P92.2 CONNECTS TO PAD 2 P92.3 CONNECTS TO PAD 1
brown J91.1 blue J91.2 violet J91.3
J84.1 J84.2
+
O2
6600-1511-700
6600-0602-700
2
RHRH SIGNAL RH+ THERMISTOR2 THERMISTOR1
Sensor Boards
Flow Valve Assembly 1
1
1 2 3 4 5
+
Expansion Slot Connection See Top Level Wiring Diagram
J85.1 J85.2 J85.3 J85.4 J85.5 J85.6 J85.7 J85.8
-
J81.1 J81.2 J81.3 J81.4
1
1 2
6600-1069-700 J86.1 J86.2
brown black
1 2 3
red 1 black 2
1 2 3
Cooling Fan 6600-1523-700
6600-1524-700
C
C
B
B
A
A
OHMEDA MEDICAL
LAUREL, MD 2072 3
WIRING DIAGRAM, Size
GIRAFFE OMNIBED AND INCUBATOR
D
DRAWING NUMBER
PROJECT 4
3
2
1
Rev
2
Sheet
SCALE
ICP011 5
6600-2323-100 NONE
of
3
3