Giraffe Incubator™, Giraffe Incubator Carestation™, Giraffe OmniBed™ and Giraffe OmniBed Carestation™ Replace SE and SW wall latches Service Instruction
5863480 Revision 1 US English © General Electric Company
Replace SE and SW wall latches
Contents
Contents Chapter 1 Introduction ...8 1.1 Giraffe OmniBed and Incubator southeast (SE) and southwest (SW) wall latches ... 8
Chapter 2 Replace latches and receptacles...9 2.1 Warnings and cautions... 9 2.2 SE and SW wall latches and receptacles ...11 2.3 Replace a southeast (SE) or southwest (SW) wall latch ...12 2.3.1 Required tools and equipment...12 2.3.2 Wall latch parts... 12 2.3.3 Replace ... 12 2.4 Replace a southeast (SE) or southwest (SW) receptacle on the OmniBed ...15 2.4.1 Required tools and equipment...15 2.4.2 Receptacle parts for OmniBed ...15 2.4.3 Replace ... 15 2.5 Replace a southeast (SE) or southwest (SW) receptacle on the Incubator...17 2.5.1 Required tools and equipment...17 2.5.2 Receptacle parts for Incubator ...17 2.5.3 Replace ... 17
Chapter 3 Checkout procedure ... 20 3.1 Bedside panel latch checkout procedure ...20
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
7/21
1 Chapter 1 Introduction 1.1 Giraffe OmniBed and Incubator southeast (SE) and southwest (SW) wall latches This Service Installation Instruction applies to the SE and SW wall latches and receptacles on the Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed, and Giraffe OmniBed Carestation. This instruction applies to authorized service personnel to remove the old latch and receptacle, install the new latch and receptacle, and to do a checkout procedure to make sure the hardware is correctly installed. Add this instruction to your service manual for future reference. • Failure to tighten the screws could result in the bedside panel opening unexpectedly. • Take care to not damage the existing labels on the bedside panel. • Install every part included in the assembly kit. • The latch and receptacle come in the same kit.
Applicable models Giraffe Incubator Giraffe Incubator Carestation Giraffe OmniBed Giraffe OmniBed Carestation
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
8/21
2 Chapter 2 Replace latches and receptacles 2.1 Warnings and cautions WARNING INFECTION HAZARD Always wear gloves when you disassemble the bedside panel. Follow local procedures for cleaning or servicing dirty or contaminated devices and components. WARNING PATIENT FALL HAZARD Make sure the latch and receptacle mounting screws are tightened to the specified level. WARNING PATIENT FALL HAZARD You must do the pre-use checkout procedure before you put the patient in the device. Make sure that the 4 bedside panel latches and all porthole latches operate correctly. Do not use the device if a latch is broken or does not correctly close.
CAUTION PERSONNEL OR PATIENT INJURY HAZARD Do not do the pre-use checkout procedure, service, or calibrate the device with the patient in the device. Make sure all pre-use checkout procedure steps are correctly completed before each patient use. If a step fails, do not use the device.
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
9/21
Replace SE and SW wall latches
2.1 Warnings and cautions
WARNING PERSONNEL OR PATIENT INJURY HAZARD This device must only be used by qualified personnel with the direction of a medical professional.
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
10/21
Replace SE and SW wall latches
2.2 SE and SW wall latches and receptacles
2.2 SE and SW wall latches and receptacles Figure 2-1 Giraffe Incubator (Carestation shown)
West
1
3
East
4
2
6
5
SE
SW
Figure 2-2 Giraffe OmniBed (Carestation shown)
East
West 1
4
2
3
5
SW
SE
Item
Description
Item
Description
1
West bedside panel
4
East bedside panel
2
SW receptacle
5
SE receptacle
3
SW latch
6
SE latch
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
6
11/21
Replace SE and SW wall latches
2.3 Replace a southeast (SE) or southwest (SW) wall latch
2.3 Replace a southeast (SE) or southwest (SW) wall latch 2.3.1 Required tools and equipment • #2 cross head screwdriver • Approved cleaning agent (See the user manual for cleaning guidance.) • Gloves and required PPE
2.3.2 Wall latch parts Table 2-1 Replacement parts Assembly kit number:
Description
5697379 (SW OmniBed)
Latch parts:
5697380 (SE OmniBed)
• Latch (1)
5697381 (SE/SW Incubator)
• M2 cross-recessed screw (2) • Backplate (1)
2.3.3 Replace This procedure applies to the Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed, and Giraffe OmniBed Carestation. 1.
Put on new gloves.
2.
Remove the bedside panel that has the latch you will replace.
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
12/21
Replace SE and SW wall latches
3.
2.3 Replace a southeast (SE) or southwest (SW) wall latch
Remove the screws, and then remove the latch. Figure 2-3 Location of screws
4.
Spray a soft dry cloth with cleaning solution (or use pre-saturated wipes) to clean any visible dirt from the latch contact surface. Do not spray directly on the device.
5.
Install the latch as shown below. Tighten the screws manually, plus ¼ turn. Do not tighten the screws too much. Make sure the parts are fully attached.
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
13/21
Replace SE and SW wall latches
2.3 Replace a southeast (SE) or southwest (SW) wall latch
Figure 2-4 Latch installation (SW shown)
1
3 4
2
6.
Item
Description
Item
Description
1
Backplate
3
SW latch
2
West bedside panel (outer wall)
4
M2 cross-recessed screws
Connect the bedside panel to the device. Figure 2-5 Attach the bedside panel
7.
Close the bedside panel.
8.
Do the 3.1 Bedside panel latch checkout procedure on page 20.
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
14/21
Replace SE and SW wall latches
2.4 Replace a southeast (SE) or southwest (SW) receptacle on the OmniBed
2.4 Replace a southeast (SE) or southwest (SW) receptacle on the OmniBed 2.4.1 Required tools and equipment • 2.5 mm hex key • Approved cleaning agent (See the user manual for cleaning guidance.) • Gloves and required PPE
2.4.2 Receptacle parts for OmniBed Table 2-2 Replacement parts Assembly kit number
Description
5697379
OmniBed SW Receptacle parts: • SW Receptacle (1) • M5 hex screws (2) • M5 washer (2) • M5 lock washer (2)
5697380
OmniBed SE Receptacle parts: • SE Receptacle (1) • M5 hex screws (2) • M5 washer (2) • M5 lock washer (2)
2.4.3 Replace You can use this procedure to replace both SE and SW receptacles. The SE and SW receptacles are flipped on opposite sides of the device. They are not interchangeable. You must raise the canopy to get access to the receptacles. 1.
Put on new gloves.
2.
Raise the canopy.
3.
Open the applicable bedside panel to get access to the receptacle.
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
15/21
Replace SE and SW wall latches
4.
2.4 Replace a southeast (SE) or southwest (SW) receptacle on the OmniBed
Remove the screws, and then remove the receptacle. Figure 2-6 Remove the receptacle screws
5.
Spray a soft dry cloth with cleaning solution (or use pre-saturated wipes) to clean any visible dirt from the receptacle contact surface. Do not spray directly on the device.
6.
Install the receptacle as shown below. Tighten the screws manually, plus ¼ turn. Do not tighten the screws too much. Make sure the parts are fully attached. Figure 2-7 Install the receptacle (SW shown) 1
2 3 4
Item Description
7.
1
SW receptacle
2
Flat washer
3
Split lock washer
4
Hex screws
Do the 3.1 Bedside panel latch checkout procedure on page 20.
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
16/21
Replace SE and SW wall latches
2.5 Replace a southeast (SE) or southwest (SW) receptacle on the Incubator
2.5 Replace a southeast (SE) or southwest (SW) receptacle on the Incubator 2.5.1 Required tools and equipment • 2.5 mm hex key • Flat head screwdriver • Approved cleaning agent (See the user manual for cleaning guidance.) • Gloves and required PPE
2.5.2 Receptacle parts for Incubator Table 2-3 Replacement parts Assembly kit number
Description
5697381
SE/SW Receptacle parts: • Receptacle (1) • Corner bracket (1) • M4 hex screw (2) • Screw cover (2) • Bracket nut (1) • Bracket nut cover (1)
2.5.3 Replace You can use this procedure to replace both SE and SW receptacles. The SE and SW receptacles are fully interchangeable. 1.
Put on new gloves.
2.
Open the applicable bedside panel to get access to the receptacle.
3.
Open the screw covers, and remove the receptacle hardware.
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
17/21
Replace SE and SW wall latches
2.5 Replace a southeast (SE) or southwest (SW) receptacle on the Incubator
Figure 2-8 Open the screw covers
4.
Spray a soft dry cloth with cleaning solution (or use pre-saturated wipes) to clean any visible dirt from the receptacle contact surface. Do not spray directly on the device.
5.
Install the receptacle as shown below. Tighten the nut and screws manually, plus ¼ turn. Do not tighten the fasteners too much. Make sure the parts are fully attached.
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
18/21
Replace SE and SW wall latches
2.5 Replace a southeast (SE) or southwest (SW) receptacle on the Incubator
Figure 2-9 Install the receptacle (SW shown)
4
1 2
3
5 6
Item
Description
Item
Description
1
Corner bracket
4
SW receptacle
2
Screw cover (2)
5
Bracket nut cover
3
M4 hex screw (2)
6
Bracket nut
6.
Close the bedside panel.
7.
Do the 3.1 Bedside panel latch checkout procedure on page 20.
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
19/21
3 Chapter 3 Checkout procedure 3.1 Bedside panel latch checkout procedure Perform this procedure on each bedside panel to make sure the latch and receptacle installation and operation is correct. 1.
Open the bedside panel that has the new latch installed, and make sure that it swings all the way down and hangs perpendicular to the bed.
2.
Make sure the latch and receptacle are fully attached to the device.
3.
Close the bedside panel, and make sure the latches operate correctly on both bedside panels: a. Pull on the panel to make sure the latches keep the panel closed (the red tabs will not show). b. Open both wall latches to the point where the red tabs show, but have not cleared the receptacles. See Figure 3-1 Wall latches disengaged. c. Pull on the panel to make sure the latch assembly keeps the panel from fully opening. Figure 3-1 Wall latches disengaged
1
1
2
2
Item
Description
1
SE wall latch shown
2
NE wall latch shown
5863480 Revision 1
Giraffe Incubator, Giraffe Incubator Carestation, Giraffe OmniBed and Giraffe OmniBed Carestation
20/21