gm instruments ltd
Rhinomanometer NR6 User Software Guide and Installation Notes V9
User Software Guide and Installation
30 Pages
Preview
Page 1
Rhinomanometer NR6
Rhinomanometer NR6 User Software Guide And Installation Notes V9
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 1 of 30
Rhinomanometer NR6
Contents Page Software Installation... 3 Hardware Installation ... 6 How To Configure the Software ... 8 Starting A Test ... 20 To prepare for measurements on a new subject... 20 To prepare for measurements on a subject who has already been tested 20 To Make A New Record... 22 To Compare Any Two Test Results ... 23 To Delete A Test Result... 24 Batch Test Facility (only on Executive versions) ... 25 Data Export Facility (only on Executive versions) ... 27
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 2 of 30
Rhinomanometer NR6
Software Installation COMPUTERS SUPPORTED NR6 will work on any PC which can run Windows 2000, XP or Vista and which has a free USB socket. NB units, which comply with EN60950, should be used. PRINTERS SUPPORTED As these instruments operate in a Windows environment, print capability depends on you having installed a printer under Windows. Virtually any printer, which works under the version of Windows you have, will be suitable. INSTALLATION SEQUENCE The software comes on 2 CD’s and should be installed before connecting the NR6 Rhinometer to your PC. It is also advisable to close any other programs you may be running while performing the installation as you may need to close down the PC during this operation. GMI SOFTWARE Put the CD in your drive, navigate to it using My Computer/Windows Explorer, and double click on the file SETUP.EXE. Accept the defaults offered at each stage. INTERFACE BOARD SOFTWARE (Measurement Computing) Ensure the USB cable is not connected to the NR6 unit. Put the CD in your drive. If your PC is set to autorun CD’s then allow this to happen while accepting the defaults offered. The program Instacal and other driver software will be loaded to a folder called Measurement Computing. If your PC does not automatically run the CD’s installation program, click on My Computer, select and open your CD drive, and then double click on mccsetup.exe to install the programs and files. The installation program will offer defaults all of which should be accepted. INTERFACE BOARD HARDWARE The last installation operation is to have your PC identify the USB A/D convertor, which is built into the NR6 unit.. Click on START/ALL PROGRAMS/MEASUREMENT COMPUTING/INSCAL32.exe and Instacal will show no boards installed as shown overleaf.
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 3 of 30
Rhinomanometer NR6
Connect the USB cable between the NR6 and the PC. Wait for a few seconds as the USB interface is recognised and the drivers loaded. Once completed the Instacal screen will offer the opportunity to install your USB board.
Click on OK.
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 4 of 30
Rhinomanometer NR6
When the Instacal program shows the screen above, installation is complete and Instacal can be closed down. The NR6 Rhinomanometer program can now be started by clicking on PROGRAMS/GMI NASAL MEASUREMENTS/NR6 Rhinomanometer
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 5 of 30
Rhinomanometer NR6
Hardware Installation The NR6 hardware installation is concerned with the cable between the NR6 and the PC and the "plumbing" section which is concerned with the pressure and flow tubes and mask assembly. There is a single cable which links the NR6 box to the PC through the interface card. It can only be connected one way, and should already be connected. NR6 “plumbing” 1)
The NR6 box has three nozzles on it, which are colour coded Black is the pressure input. Red is the positive flow input. Green is the negative flow input.
2)
Connect the tube marked black onto the nozzle marked black. The tube marked red onto the nozzle marked red and the tube marked green onto the nozzle marked green. At the free end connect the tube marked red onto the flowhead port marked red. The tube marked green onto the flowhead port marked green and the tube marked black onto the mask port marked black, for posterior tests or on to the anterior tube connector for anterior tests
NR6 front view flowhead and pressure connections
NR6 back panel view showing USB to PC socket and blanked link socket.
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 6 of 30
Rhinomanometer NR6
Opening the Software Start the program by clicking on its desktop icon or by clicking on START/PROGRAMS/GMI NASAL MEASUREMENTS/RHINOMANOMETER NR6 The opening window is as shown below.
A number of elements within the program can be configured to suit your application, and although you may wish to change them again later, it should be noted that fields within the Patient Information screen should not be changed after storing results for a particular subject, as the stored information may become inaccessible after that time. Full information on the alterations possible, and how to make them, are shown on the following pages.
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 7 of 30
Rhinomanometer NR6
How To Configure the Software All changes are made by clicking on TOOLS and OPTIONS. A new window named SETUP is then displayed which shows 4 tabs, each of which contain user configurable items. The four tabs are labelled:Formatting, in which you can change the colour of almost any element of the program on screen or printout. Patient Record, in which you can alter the number of fields available for information input, their size, their names and the type of characters which are acceptable. You can also tell the system to use a particular field entry as the patient file name. Rhinomanometry / Devices lets you tell the program what settings have been used for the interface card and what connections have been made to it. We do not recommend that any of the default settings are altered unless there is a good reason for doing so. Rhinomanometry / Acquisition, is the tab screen to use when you want to preset the type of measurement you want to be offered by default when you start testing. You can alter a number of parameters which are discussed in detail below. Rhinomanometry / Calculation, lets you add the display of Broms angles or Rohrer values when the boxes are ticked. Rhinomanometry / Printout, allows you to select other parameters for printing. Printout, The header text and margin details can be modified from here, along with additional outputs selected A detailed description of each of these Tab screens along with a screen image follows below:
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 8 of 30
Rhinomanometer NR6
Formatting Tab
The formatting tab screen is made up of the following elements:Device “radio” button...from which you can select Screen or Printout for alteration. Item scrolling window which contains a list of the screen or printout elements whose colour can be changed. Pen colour palate, from which you choose the colour you want applied to any element selected in the item window. Pen Style window which can be applied to any line drawing element selected from the item window. Pen Thickness window which allows you to apply different line thickness’ to any line drawing element selected from the item window. Preview window which lets you see the effect of a change. To make a change to the default settings select Screen or Printer, select an Item from the list, select your chosen Pen colour and if appropriate Pen Style or Thickness. You will immediately see the effect of that change in the Preview window.
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 9 of 30
Rhinomanometer NR6
If you click on Apply the change will be recorded and you can move to another element. If you click on OK instead of Apply, the change will be made, and recorded, but you will close the Setup window and return to the main screen.
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 10 of 30
Rhinomanometer NR6
Patient Record Tab
The Patient Record information window can contain many fields. You can however include as many or as few as you want, put them in any order, define what kind of data is acceptable for entry in any field and specify the size of field. The Patient Record tab contains the following elements:List of Fields is a window which shows the current “labels” already selected for inclusion. If you click on one of these it can be edited or deleted or its position moved within the group. When highlighted the information relating to its structure is shown on the right hand side of the window under the headings of:Caption the current label selected for editing. Maximum Length the number of character spaces allocated in the Patient Information window against that name. Input Mask (format) the conditions applied to character entry in the associated Caption field. An example is shown below:-
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 11 of 30
Rhinomanometer NR6
In the above example ID is highlighted in the Lists of fields window. The label ID No is shown in the Caption field and it can be seen that currently space is allowed for 14 characters. The mask applied was one, which only allowed numbers or letters for entry in that field but no spaces or other characters. Masks can be selected or changed by clicking on the button on the extreme right of the Mask field.
The Input Mask window allows selection of the conditions you want to apply and has space below for you to try out or Test the restrictions. If you wish to keep the currently selected field, click on OK, and then on Apply. If you wish to discard any field click on Delete. If you want to add a new one click on New. NR6 Rhinomanometer Software Guide V9 01/10/09 Page 12 of 30
Rhinomanometer NR6
If you want to change the order in which they appear, highlight the one you want to move and then use the UP or Down arrow’d buttons to the right of the Delete button. You can place more than one item on a line by placing it below the item you want it lined up with and clicking on Position on same line as previous field. You can also “hide” a field from printout by highlighting it and clicking on the Don’t Show in Printouts. A field can be used for the test file name by highlighting it and then clicking on Use for suggestion of filename.
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 13 of 30
Rhinomanometer NR6
Rhinomanometry / Devices Tab
The Rhinomanometry / Devices screen shown above contains settings for Board Number, which depends on the number assigned by Instacal during the installation process, but this will normally be 0. The other parameters on this tab should not be altered.
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 14 of 30
Rhinomanometer NR6
Rhinomanometry / Acquisition Tab
The Rhinomanometry / Acquisition tab contains the settings for: Type defines whether you will be offered a Broms or Standard fixed point resistance test. You can always change at the time of testing but this facility allows a standard protocol to be offered. Side gives the option of Posterior or Anterior Right or Anterior Left testing as the default test starting point. Number of Respirations lets you set the number of breathing cycles, or Respiration’s required to complete a test. The default number is 4 cycles. Show Signal Marker, lets you switch on or off the circle which surrounds the Acquisition Spot during an Acquisition. Sampling Interval is a facility to slow down the rate at which the digital meter used during calibration changes is updated. The slide switch can be moved using the mouse, by placing the mouse cursor on the slider control and holding it down while moving the slider to the left or right. Calibrate, this button is a link to the Calibration screen, which is described separately. Show Advanced Settings allows some fundamental parameters to be altered. We strongly advise that these are not altered unless you have clearly NR6 Rhinomanometer Software Guide V9 01/10/09 Page 15 of 30
Rhinomanometer NR6
thought through and understand the consequences. Clicking on the button marked Default returns the settings to the original values.
Under Advanced Settings the step size of flow or pressure change required to force the plotting of a new point is specified in Sampling Difference. The normal value is 30 units. The change in flow or Pressure Signal required to start a measurement recording is set in Sample Threshold. The Maximum Scale Value can be altered at will as it effectively increases the resolution of the displayed curve, but does not affect calibration. The normal value is 800 units. The Lateralisation Calculation can be switched on from here, for example to display Left greater than Right by 15% when Anterior test is performed.
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 16 of 30
Rhinomanometer NR6
Rhinomanometry / Calculation Tab
The Rhinomanometry / Calculation tab contains options about the Broms Angle and the Rohrer curvre and enables you to switch them on and off according to your requirements.
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 17 of 30
Rhinomanometer NR6
Rhinomanometry / Printout Tab
In the Rhinomanometry / Printout tab you can arrange to Plot Averaged Curve for the Respiritory cycles selected for inclusion, or the Acquired Data as actually recorded, when you click on Print. You can also add the Rohrer or Broms Angles – instead of Resistance Values – provided you acquire data with Broms switched on & for Anterior measurements. It is also possible to select printing out of the Diagnostic Values Table at this stage.
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 18 of 30
Rhinomanometer NR6
Printout Tab
The Printout Tab allows the setting of a title which will appear on printouts and the alteration of margins. More detailed control of printer options is available in the normal Windows print driver.
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 19 of 30
Rhinomanometer NR6
Starting A Test To prepare for measurements on a new subject 1. All current patient information and test results are cleared when the program is started or if having performed other measurements, you click on the NEW button. 2. In either event the cursor is set to the first field in the patient information window. 3. You can then enter all patient details. Pressing TAB after entering information in each field will take the cursor to the next one. 4. To start the Acquire procedure you click on the NR6 icon. The NR6 icon which is used to call the acquire screen
New Patient Button
Patient details entry fields
To prepare for measurements on a subject who has already been tested 1) Click on the button marked OPEN and then using the dialog box offered go to the folder you store your results in and click on the subject’s file you want to open. 2) You will then be able to check that you have the correct subject by examining the patient details by clicking on the PATIENT INFORMATION tab. Clicking on the RESULTS tab will show you all previous test dates and results for that patient.
NR6 Rhinomanometer Software Guide V9
01/10/09
Page 20 of 30