AccuScreen Otoacoustic Emissions and Automated ABR Screener Service Manual
Copyright © GN Otometrics A/S, 2 Dybendalsvænget, P.O. Box 119, DK-2630 Taastrup, Denmark. Edition: 2002/09. Part No. 7-26-95200/00 Doc. No. 7-26-9520/00 Although the information in this document has been thoroughly checked, no liability is assumed by GN Otometrics for any discrepancies. Specifications are subject to change without notice.
1
TABLE OF CONTENTS 1. WHEN IS SERVICE NECESSARY ? ...3 2. WHO CAN CARRY OUT WHAT SERVICE ? ...3 3. SHIPPING AND TEST PROCEDURES ...4 4 SOFTWARE TOOLS FOR SERVICE ...5 4.1 PC software ...5 4.2 device software ...5 5 TESTING PROCEDURES...6 5.1 Device check procedure after SW configuration ...6 5.2 Annual test...6 5.3 Calibration and Final test ...6 5.4 Probe calibration (PRE) / Check Probe calibration ...6 5.5 ABR trunk cable test ...6 6. FINAL ASSEMBLY ...7 7. TROUBLESHOOTING ...8 7.1 Error codes ...8 7.2 Systematic troubleshooting...10 7.2.1 Device ...11 7.2.2 Probe...12 7.2.3 Optional configuration ...13 8. SUPPLIES...14 8.1 Accessory ...14 8.2 Spare parts ...15 8.3 Test equipment ...15 9. MAINTENANCE...16 9.1 Probe ...16 9.1.1 Cleaning and maintenance of the probe tip...16 9.1.2 Changing the acoustic filter disc...17 9.1.3 Cleaning the attachable colored ear tips ...17 9.2 Cleaning the Accu Screen ...18 10. LABELS AND OPENING THE DEVICE ...19 11. APPENDIX ...20
2
1. WHEN IS SERVICE NECESSARY ? •
According to the ‘SW/HW Information’ menu on the AccuScreen device the ‘annual test’ service is necessary.
•
after each repair or opening of the device Æ final test
•
if problems appear during proper use, as well as if the measurement procedure or the results are faulty
2. WHO CAN CARRY OUT WHAT SERVICE ? User: - short functional tests (Self tests) - cleaning of device and accessories - troubleshooting according to the „troubleshooting guide“ For additional repair actions, services and maintenance refer to your distributor
Service-centres that have been authorized by the manufacturer: - extensive functional tests - repair of device and accessories - software upgrades - service including calibration and final testing Manufacturer - extensive repair including exchange of essential parts and components - probe calibration - software upgrades - service including calibration and final testing Please notice: Repairs and service including calibration of the AccuScreen may be carried out only by the manufacturer or by services that have been authorised by the manufacturer. In case of repair actions by anyone else, the manufacturer reserves the right to refuse any liability for the operational safety, the reliability or the efficiency of the device
3
3. SHIPPING AND TEST PROCEDURES production AccuScreen device
'ES-Tool' 'IRDA Test'
production Probe
'final test'
Probe calibration
final assembly
S H I P P I N G fz 'up1.exe' 'ES-Tool'
software upload
Internet & 'license1.exe'
licencing
'set_cal.exe'
'device check after SW configuration /firmware update'
final assembly madsen
S H I P P I N G madsen ANNUAL
SERVICE
'annual test' includes firmware upgrade test
'Release Manager'
'set_cal.exe'
'probe test'
REPA IR
open device / change hardware unit
low level repair (e.g. change battery)
probe failure
replace probe
'Release Manager-Tool'
'IRDA Test' 'set_cal.exe'
software upgrade
'Release Manager-Tool'
software upgrade
'final test'
'device check after SW configuration /firmware update'
'probe test'
'probe test'
'Release Manager-Tool'
testing the probe calibration
software upgrade
'device check after SW configuration /firmware update'
.
4
4 SOFTWARE TOOLS FOR SERVICE 4.1 PC software up1.exe
DOS tool for software upload
set_code1.exe
DOS tool to send the license code to the device to unlock the software
set_cal.exe
DOS tool to set a new calibration date on the device
ES-Tool
Software to configure some settings like sound, backlight…
Release Manager
Tool which updates the newest software release automatically
IRDA Test
Tool to test the IRDA link between device and PC
A more detailed description of this configuration tools you find in appendix H.
4.2 device software Service Menu
Main Menu -> Options -> Self test (press and keep down button pressed and press right button) -> Service Menu Mic cal
Microphone calibration resp. test
Spk1 1000Hz
Speaker 1 calibration resp. test
Spk2 1000Hz
Speaker 2 calibration resp. test
Show test counters measurement counter reading
5
5 TESTING PROCEDURES 5.1 Device check procedure after SW configuration The test procedure after software configuration is needed after every software upgrade. You find the whole specification for the software check in appendix A.
5.2 Annual test The annual test is necessary for service issue. The final test includes the device check after software configuration. In appendix B you see the specification for the test.
5.3 Calibration and Final test The final test is necessary after repairing by opening the device. The final test includes the annual test and the check after software upgrade. The final test consists:
final test instruction final test flowchart final test protocol
You find the whole documents in the appendix C.
5.4 Probe calibration (PRE) / Check Probe calibration This procedure is for probe calibration resp. check probe calibration. This procedure can be done with an AccuScreen device or OAE-ABR Simulator You find the whole guideline in appendix D.
5.5 ABR trunk cable test This procedure is for testing the ABR trunk cable. You find the whole guideline in appendix E.
6
NOTE: The clock battery of the device must be replaced every three years to make sure that the date and time functions work properly. In order to change the battery you need a soldering gun and you have to open the device therefore a ‘final test’ is necessary after replacing the battery. The work instruction for changing the battery is in appendix I.
6. FINAL ASSEMBLY Check the manufacturing and serial number sticker. Check the corresponding case equipment. Final visual check of device, probes and accessories.
7
7. TROUBLESHOOTING 7.1 Error codes Error Message Probe Error!
Cause Probe error - probe not connected? - probe occluded?
Stopped Cal FAIL
Solution Remove the probe Check whether the probe is connected. Clean probe if occluded.
Measurement aborted
Repeat measurement on the same ear Stimulus level calibration Assure stable probe placefailed ment; avoid noise
Testing not possible!
Impedance of electrodes or Check electrodes and balance is to high renew them if necessary. EEG values are constantly Repeat measurement if the high baby is quiet.
Testing not possible! Please disconnect the label printer! No printer connected
There is no measuring Disconnect the label printer possible while the printer from the device. is connected to the device. - Printer is not connected - Connect the printer with with the device the device. - Printer is turned off. - Turn printer on.
Printer error
-
Paper jam Out of labels
Labels not adjusted No labels in printer
Change Battery
Battery empty < 5.5 Volt
Check manual of label printer correctly Adjust labels correctly
Battery charge low <5.8 Volt
clock bat low
clock battery discharged
Codec failure Codec is defect Please note that after “Codec failure” was displayed the device will be automatically turned off.
Refill labels Replace the battery before continuing with the measurement Replace the battery before continuing with the measurement Contact your service department (battery lifetime > 2 years Please contact service department.
8
Error Message LCD test failed
Cause
LED blinks, number code is Internal error code showed in upper left corner System Error System malfunction The calibration date has expired. Please have the device calibrated soon.
Solution Please contact service department Please note code and contact service department Contact your service department Please contact service department.
9
7.2 Systematic troubleshooting For troubleshooting use the given flowcharts: On the left you see the steps you should follow. Beside is described how to do the steps. And on the right side you see the procedure to handle the problems.
How to work with the flowcharts: •
Follow the steps if you find no failure go to the next step.
•
If you find a failure try to repair it with the given hints on the right side. o If the trouble shooting hints force you to open the device to replace e. g. the processor board follow first all steps of the flowchart to ensure that you replace the right unit. o If you can’t repair the device with the given hints go to the next step (if possible) maybe the failure can be solved in a following step!
10
7.2.1 Device steps
procedure
external demage check
- sockets - display - housing
probe test
power up device
check ifLED flash short red and green
notice error messages
do the selftests
troubleshooting
- goto probe test flowchart
- insert charged battery - check battery contact - check fuse - replace keyboard - replace processor board
- OK - button
- directly after switch on - change flash-eprom (boot program on flash-eprom is out of order) - observe LCD display e.g. Clock batt low !!!
- Menu (Options -> Selftest)
LCD -test
keyboard - test
- check connection LCD display - check soldering points on LCD - replace LCD panel - replace processor board
- check plug-and-socket connection - replace keyboard - replace processor board
codec - test
- replace processor board
probe - test (witch LDA)
- check plug- and-socket connection (cable, soldering points) - replace analog board (use LDA (Loopback diagnostic adapter))
open device
- remove battery - remove screws (6 pices)
visual check
- check plug- and-socket connection - check soldering points - check for short-circuits and loose connections .
11
7.2.2 Probe steps
optical check
procedure
troubleshooting
- check for loose or damaged cable - check probe housing - check filter disk & probe tip
OK
probe - test
probe - info
probe - test
check probe calibration
Spk1 1000Hz test
- use a proper working Accu-Screen or OAE-ABR Simulator - Menu (Options -> Selftest) - probe in test cavity
- connect LDA to Ear probe connector - does AccuScreen identify LDA? yes: no: - eprom defect - device failure - replace probe -> device flowchart
- display the probe data
probe OK
ok
- The following tests are in the 'Service menu' ('Self test' menu [press down-button keep button pressed and press additional right-button] -> Service menu) - cp. 'Testing the calibration of PRE probes'
79 +-3dB
- speaker defect resp. recalibration - replace probe
fail
ok
Spk2 1000Hz test
79 +-3dB
Spks out of phase test
<70dB
- speaker defect resp. recalibration - replace probe
Spk polarity test (OAE-ABR Simulator)
(Akku Screen)
~79dB
- speakers out of phase (wrong polarity) - replace probe
.
.
Mic cal test
94 +-2dB
- microphone defect resp. recalibration - replace probe
12
7.2.3 Optional configuration steps
procedure
troubleshooting
IRDA
- scratched IR window on top screen?
- replace analog board
- data exchange with working IRDA PC - make sure that PC IRDA adapter is working
- check internal IR wiring - replace analog board
- test printer with PC or working AccuScreen - notice printer manual
- buy a new printer
- e. g. data exchange with EchoLink
- replace processor board
- data exchange printer - AccuScreen with a spare printer cable
- replace printer cable
visual check
test link
printer
test printer
test serial port
test printer cable
.
ABR trunk cable
leakage resistance test
-connect the cable with the 'ABR trunc cable test adapter' -follow the ABR trunk cable test document (appendix)
- replace ABR trunk cable
13
8. SUPPLIES 8.1 Accessory Part No. Units Accessory box 1389 1069 1040 1/5 1339 50 / 200 1340 50 / 200 1341 50 / 200 1342 50 / 200 1343 50 / 200 1344 50 / 200 1345 50 / 200 1346 50 / 200 1347 50 / 200 1368 50 / 200 1369 50 / 200 1414 Case 1415 1/5 1387 1388 1372 1112 1089 1113 1114 1115 1117 1116
Reference Description
blue red brown green iceblue transparent white blue red redbrown green yellow
Accessory box AS complete Accessory box Cleaning wire Ear tip DP 4mm Ear tip DP 4.5mm Ear tip DP 5.0mm Ear tip DP 5.5mm Ear tip DP 6.0mm Ear tip DP 7.0mm Ear tip DP 8.0mm Ear tip DP 9.0mm Ear tip DP 10.mm Ear tip DP mushroom Ear tip DP tree AS Accessory box label Case green "AS" + Inlay T Mad Case green "AS" printed Inlay T Madsen Test cavity DP Charger FW 7206/09 Akku T Main adapter f. char. Euro Main adapter f. char. USA/Japan Main adapter f. char. Australia Main adapter f. char. WW Main adapter f. char. UK
Cables 1307 1367 1370 1371 1326 1/5 1425
Transfer cable PC serial Label printer cable ABR trunk cable Electrode cable set IR - adapter USB/Serial converter
Probe 1402 1401 1361 1362 1353
Probe DP Probe TE Probe tip TEOAE+DPOAE Probe tip TEOAE Acoustic filter DP
1/5 1/5 1/5
blue green yellow
Design
Different accessories 1384 1405 1420 1440 400
Label printer Label for Label printer AccuLink SW Electrodes
Manuals 1421 1422 1423 1424
User Manual Basic User Manual PRO Quick guide Basic Quick guide PRO
Manufacturer
GN part number
1-08-038 8-68-31200 8-68-31201 8-68-31202 8-68-31203 8-68-31204 8-68-31205 8-68-31206 8-68-31207 8-68-31208 8-68-31209 8-68-31210
4.0mm 4,5mm 5,0mm 5,5mm 6,0mm 7,0mm 8,0mm 9,0mm 10,0mm
Ni MH 6V/1500 mAh
on request
1-21-62600 1-21-62700 8-68-31300
Dymo LabelWriter 330 T 54x101mm 220 pc
Dymo Dymo
3/pkg
14
8.2 Spare parts Part No. Qty Reference Description 1392 1393 1184 1272 1437 1310 1296 1297 1410 1273 1274 1394 1413 1276
1 1 1 1 1 6 1 1 1 1 1 1 1 1
Housing upper part AS Display window printed AS LCD - graphic display LCD connection cable LCD display with cable Screws for housing Keyboard 5 keys assembled Keyboard cover 5 keys Front label AS 5 keys Keyboard 17 keys assembled Keyboard cover 17 keys Front label AS 17 keys manufacturing label serial number label
1289 1182 1309 1311
1 1 10 1
shaft cover housing base part rounded head screws for pcb’s foam self-adhesive
1429 1328 1329 1330 1331
1 1 1 1 1
foam ES Analog B. TD ES Analog B. TDA ES Analog B. TD Plus ES Analog B. TDA Plus
1255 1436 1439 1278
1 1 G1 1 1 IC3
ES Proc. B. T past equipped Clock battery piezo buzzer Flash EPROM
Design
Manufacturer
PG-12864 LRU-FTA-H
Rose / Haas Rose / Haas Powertip
Rose Haas Rose Haas
Taguan 230 / RAL 7035
Rose
30 x 9 x 6mm 55 x 30 x 10mm
PLCC 32 S
8.3 Test equipment Part No. Qty Description
Design
Manufacturer
1459 1460
1 OAE-ABR Simulator 1 Battery symbol threshold adapter
fz fz
1461 1462 1463 1464
1 1 1 1
fz fz fz fz
1465 1466 1467
1 coupler IEC711 1 sound level meter 1 sound level calibrator
Adapter for IEC711 coupler (probe PRE) Adapter for sound level meter calibrator (probe PRE) LDA (loopback diagnostic adapter) ABR trunk cable test adapter
model:SIP95S calss:1 model: 513A
ACLAN Cirrus Research
15
9. MAINTENANCE 9.1 Probe 9.1.1 Cleaning and maintenance of the probe tip It is necessary to clean the probe frequently, especially since even small amounts of cerumen can occlude one of the three channels of the probe.
Note B
A
B C A
Remove the probe tip (B) by pressing the button (A) while shoving the probe tip carefully upwards out of the probe case.
D By means of the cleaning wire (C) you remove the acoustic filter disc (D) from the probe tip. C
Use the cleaning wire (C) to clean the sound channels of the probe from the rear.
The probe tip can be cleaned with a disinfectant or alcohol (70%). Ensure the sound channels are completely dry before replacing the tip onto the body of the probe. Even the tiniest amount of moisture may dissolve any residual cerumen and thus contaminate the sensitive micro filter (black) in the body of the probe.
16
NOTE
The probe body contains sensitive components. Never attempt to clean the sound apertures in the probe housing mechanically or with liquids. Doing so may cause damage to the probe. Please contact your authorized service department for a periodical cleaning of the probe body.
9.1.2 Changing the acoustic filter disc
Note
If the calibration procedure fails and the message “Probe Error!” is displayed, the micro filter of the probe may be damaged or obstructed. In such a case, the filter disc needs to be changed. Therefore remove the probe tip (please see chapter 8.1.1) and by means of the cleaning wire (C) you take out the acoustic filter disc from the probe tip and insert a new disc. Test proper probe functions ‘Self test’ -> ‘Probe test’
9.1.3 Cleaning the attachable colored ear tips
Note
An ultrasonic bath is recommended for cleaning the silicone ear tips. Sterilization in alcohol or a disinfectant is also possible, as is the use of steam sterilizers up to a temperature of 150° C. Avoid exposing the ear tips to intense sun light or UV light. Disposable articles (like ear tips or cleaning pins) should be disposed according to local regulations.
The datasheet of the probe is in appendix F.
17
9.2 Cleaning the Accu Screen
•
Before cleaning, the Accu-Screen must be turned off and separated from any current from an external source (e.g. PC). Prior to the actual cleaning, separate the probe and the data transfer plug from the device.
•
Use only warm water and a mild detergent (do not use alcohol or disinfectants.)
•
Clean the surface with a damp cloth.
•
Always ensure that no moisture gets inside the probe or into the device).
•
Never immerse the Accu-Screen into water or other cleaning solution.
Note
18
10. LABELS AND OPENING THE DEVICE
To open the device unscrew the 6 screws (*), after repair screw up the device and put on one screw sealing wax (cp. picture).
19
11. APPENDIX A. Device check procedure after SW configuration / firmware update B. Annual test C. Final test D. Testing the calibration of PRE probes E. ABR trunk cable test F. Probe datasheet (probe handling and cleaning) G. Software upload & configuration (licencing) procedure H. AccuScreen configuration tools I. Change battery (work instruction) J. OAE-ABR Simulator instruction manual K. Schematics