IMH-10 20 IMAGE MANAGEMENT HUB Service Bulletin Sept 2013

Service Bulletin

3 Pages

SERVICE BULLETIN  OEM:  Preventative action to avoid failure of the HDD due to incorrect disconnection from the primary power supply Olympus Medical Systems Corp. TYPE: Medical  MODEL:  IMH-10/20  KM REF:  OKMSV-13-093  STATUS:  Information & Action  DATE:  05-09-2013  SUBJECT:  BACKGROUND There is a potential for the user to cause damage or failure of the internal HDD. This is caused by following the incorrect “Power Off” procedure. To reduce the chance of occurrence a Caution Label is now available and can be applied to the front panel, and an accompanying instruction sheet has been produced to provide guidance of the power off procedure. These are now supplied with all new products.  DETAILS The cause of failure is the IMH-10/20 being powered off using the main power switch of the trolley stack system or directly at the external mains supply. The power switch on the front of the IMH should always be used to power down the product. HDD failure can also occur if the power is removed before the shutdown procedure of the IMH-10/20 has completed.  continued…. Page 1 of 3