User Manual
26 Pages
Preview
Page 1
CardioPerfect PRO ECG Recorder - User Manual
European Regulatory Manager Welch Allyn Ltd. Navan Business Park Dublin Road Navan, County Meath, Republic of Ireland Tel.: +353 46 90 67700 Fax: +353 46 90 67754
0297
DIR 80012763 Ver. D
Welch Allyn 4341 State Street Road Skaneateles Falls, NY 13153-0220 USA www.welchallyn.com
CardioPerfect PRO ECG Recorder - User Manual
Caution
US Federal law restricts this device to sale by or on the order of a physician.
Copyright © Copyright 201010, Welch Allyn. All rights are reserved. No one is permitted to translate, reproduce or duplicate this manual or any part thereof, in any form, without prior written permission from Welch Allyn. Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual. Unauthorized copying of this publication may not only infringe copyright but also reduce the ability of Welch Allyn to provide accurate and up-to-date information to users and operators alike. Welch Allyn®, CardioPerfect® Workstation and SpiroPerfect® are registered trademarks of Welch Allyn. Software in this product is Copyright 2010, Welch Allyn. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software provided on the original distribution medium. The software may not be copied, decompiled, reverse-engineered, disassembled or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title and ownership of the software remains with Welch Allyn. The information contained in this manual is subject to change without notice. All changes will be in compliance with regulations governing manufacture of medical equipment.
User responsibility This product is designed to perform in conformity with the description thereof contained in this manual and accompanying labels and inserts, when assembled, operated, maintained and repaired in accordance with the instructions provided. A defective product should not be used. Parts that are broken, plainly worn, missing or incomplete, distorted or contaminated should be replaced immediately. Should any repair or replacement become necessary, we recommend that service be performed at the nearest approved service center. The user of the product shall have the sole responsibility for any malfunction, which results from improper use, faulty maintenance, improper repair, damage or alteration by anyone other than Welch Allyn or their authorized service personnel.
Accessories The Welch Allyn warranty can only be honored if you use Welch Allyn approved accessories and replacement parts.
Caution
DIR 80012763 Ver. D
Use of accessories other than those recommended by Welch Allyn may compromise product performance.
2 / 26
CardioPerfect PRO ECG Recorder - User Manual Safety and Precautions
Caution
Radio Frequency (RF) interference between the PRO ECG Recorder or cardiograph and any existing RF transmitting or receiving equipment at the installation site, including electrosurgical equipment, in close proximity to the cardiograph should be evaluated before the equipment is operated as they may seriously degrade performance. The Cardio Perfect Cardiograph is susceptible to interference from RF energy sources (lowered RF immunity) which exceed the IEC 60601-1-2 limits, such as power line bursts, other medical devices, cellular products, information technology equipment and radio/television transmission. To reduce EMC interference the cardiograph shall be separated from the emitting source as much as possible. If assistance is needed, call your local Welch Allyn service representative. Artifact on the ECG caused by electromagnetic interference should be evaluated by a physician or physician authorized personnel to determine if it will negatively impact patient diagnosis or treatment. Like all electronic devices, this cardiograph is susceptible to electrostatic discharge (ESD). Electrostatic discharge typically occurs when electrostatic energy is transferred to the patient, the electrodes, or the cardiograph. ESD may result in ECG artifact that may appear as narrow spikes on the cardiograph display or on the printed report. When ESD occurs, the cardiograph’s ECG interpretation may be inconsistent with the physician’s interpretation. Welch Allyn assumes no liability for failures resulting from RF interference between Welch Allyn medical electronics and any radio frequency generating equipment when these levels exceed those established by applicable standards. Patient cables and PC connection cables are intended to be inserted and removed from the Pro-Recorder in a push / pull motion. Do not twist the cables. Damage to the cables and the Pro-Recorder will occur.
Medical Device Directive The CardioPerfect Pro cardiograph complies with the requirements of the Medical Device Directive 93/42/EEC and carries the
DIR 80012763 Ver. D
0297 mark accordingly.
3 / 26
CardioPerfect PRO ECG Recorder - User Manual
Warnings
CardioPerfect devices are an integral part of a personal computer based diagnostic system. The user shall adhere to warnings in order to ensure safe and reliable performance of the system.
The personal computer, computer peripheral and all accessories (nonmedical electrical equipment) connected to the ECG shall be situated outside the patient environment and the system shall meet the requirements of IEC 60601-1-1. The personal computer used should be approved to the appropriate safety standard for non-medical electrical equipment (IEC 60950, or its national variants), and use of an additional protective earth ground or an isolation transformer is required to meet IEC 60601-1-1. Power strips (multiple portable socket outlets) are not allowed for use in connecting the medical electrical equipment or any accessories to ground unless used in concert with a medical approved isolation transformer. Additional equipment that is added to the system may pose a safety hazard. Use of an additional protective earth ground or an isolation transformer is required to meet IEC 60601-1-1. During defibrillation, the ECG signals displayed may show waveform artifacts and cannot be relied on as a true indication of the patient’s physical condition. Accessible metal parts, such as electrode terminations, should not come in contact with other electrically conductive parts, including earth ground. Welch Allyn provides a number of high quality patient leads in varying termination styles. Use of these approved patient leads is required for electrical protection of the patient during cardiac defibrillation. Cardio Perfect devices are not intended for direct cardiac application. An inoperable or damaged electrocardiograph can be identified by abnormal signals on the ECG waveforms. Abnormal signals are characterized by flat lines, excess noise, square waves or other non typical anomalies that appear on the ECG waveform. The electrocardiograph can be periodically tested by connecting the Cardio Perfect to an ECG simulator. Follow manufacturer’s instructions. If there is a requirement for equipment to be connected to a personal computer or other non-medical rated equipment, it is the responsibility of the user to ensure the system provides a level of safety that ensures compliance to IEC 60601-1-1. Other medical equipment-including but not limited to defibrillators, ultrasound machines, pacemakers, and other stimulators-may be used simultaneously with the electrocardiograph. However, such devices may disturb the electrocardiograph signal.
DIR 80012763 Ver. D
4 / 26
CardioPerfect PRO ECG Recorder - User Manual The electrocardiograph has not been designed for use with high frequency (HF) surgical equipment and does not protect against hazards to the patient. Ensure the location of the electrode and associated cables provides maximum separation away from all sources of high-frequency energy. The best way to ensure patient safety is to completely remove all electrodes and cables from the patient when exposed to highfrequency energy.
DIR 80012763 Ver. D
5 / 26
CardioPerfect PRO ECG Recorder - User Manual Patient and Operational Safety The cardiograph isolates all connections to the patient from electrical ground and all other conductive circuits in the cardiograph. This reduces the possibility of hazardous currents passing from the cardiograph through the patient’s heart to ground. To ensure the patient’s safety and your own please observe the following: When integrating CardioPerfect devices with non-medical rated equipment, such as a computer, the addition of an additional protective earth ground or the use of a medically approved isolation transformer is required for compliance to the IEC 60601-1-1 Medical System Safety Standard. Any system components (e.g. treadmill, personal computer, ergo meter) that require to be connected to an outlet socket shall use only grounded power cords (three-wire power cords with grounded plugs). Also make sure the outlet accepts the plug and is grounded. Never adapt a grounded plug to fit an ungrounded outlet by removing the ground prong or ground clip. Do not connect Multiple Portable Socket Outlets (MPSO’s) or extension cords to the system. Do not connect items which are not part of the system. The use of multiple portable socket outlets and other non-medical electrical equipment poses a safety hazard. Refer to the Medical System Safety Standard IEC 60601-1 for the requirements of such attachments. Multiple portable outlet sockets shall not be placed on the floor. Multiple portable outlet sockets or extension cord shall not be connected to the system. Do not connect items which are not part of the system. The use of multiple (non-) medical electrical equipment connected to the same patient may pose a safety hazard due to the summation of leakage currents from each instrument. Any combination of (non-) medical electrical equipment should be evaluated by local safety personnel before put into service. Multiple portable outlet sockets use without an isolation transformer is disapproved unless casual access for additional equipment is impeded or prevented.
Accessories Use of accessories other than those recommended by Welch Allyn may compromise product performance. The Welch Allyn warranty can only be honored if you use Welch Allyn approved accessories and replacement parts.
DIR 80012763 Ver. D
6 / 26
CardioPerfect PRO ECG Recorder - User Manual Warranty, Service, and Spare Parts Warranty All repairs on products under warranty must be performed or approved by Welch Allyn. Unauthorized repairs will void the warranty. In addition, whether or not covered under warranty, any product repair shall exclusively be performed by Welch Allyn certified service personnel. Assistance and Parts If the product fails to function properly or if assistance, service, or spare parts are required, contact the nearest Welch Allyn Technical Support Center. USA Latin America European Call Center United Kingdom France Germany
1-800-535-6663 (+1) 305-669-9003 (+353) 46-90-67790 (+44) 207-365-6780 (+33) 1-55-69-58-49 (+49) 695-098-5132
Canada South Africa Australia Singapore Japan China
1-800-561-8797 (+27) 11-777-7555 (+61) 2-9638-3000 (+65) 6419-8100 (+81) 3-6383-0852 (+86) 21-6327-9631
Before contacting Welch Allyn it is helpful to attempt to duplicate the problem and to check all accessories to ensure that they are not the cause of the problem. When calling, please be prepared to provide: Product name and model number and complete description of the problem The serial number of your product (if applicable) The complete name, address and phone number of your facility For out-of-warranty repairs or spare parts orders, a purchase order (or credit card) number For parts order, the required spare or replacement part number(s)
Repairs If your product requires warranty, extended warranty, or non-warranty repair service, please call first the nearest Welch Allyn Technical Support Center. A representative will assist you troubleshooting the problem and will make every effort to solve it over the phone, avoiding potential unnecessary return. In case the return cannot be avoided, the representative will record all necessary information and will provide a Return Material Authorization (RMA) number, as well as the appropriate return address. A Return Material Authorization (RMA) number must be obtained prior to any return.
Note
Welch Allyn does not accept returned products without an RMA.
Packing Instructions If you have to return goods for service, follow these recommended packing instructions: Remove all hoses, cables, sensors, power cords, and ancillary products (as appropriate) before packing, unless you suspect they are associated with the problem. Wherever possible use the original shipping carton and packing materials. Include a packing list and the Welch Allyn Return Material Authorization (RMA) number. It is recommended that all returned goods be insured. Claims for loss or damage to the product must be initiated by the sender.
DIR 80012763 Ver. D
7 / 26
CardioPerfect PRO ECG Recorder - User Manual Limited Warranty Statement Welch Allyn, Inc. warrants that the Welch Allyn CardioPerfect Workstation computer based product you have purchased meets the labeled specifications of the Product and will be free from defects in materials and workmanship that occur within 1 year after the date of purchase. Accessories used with the Product are warranted for 90 days after the date of purchase. The date of purchase is: 1) the date specified in our records, if you purchased the Product directly from us, 2) the date specified in the warranty registration card that we ask you to send to us, or 3) if you don’t return the warranty registration card, 120 days after the date on which the Product was sold to the dealer from whom you bought the Product, as documented in our records. This warranty does not cover damage caused by: 1) handling during shipping, 2) use or maintenance contrary to labeled instructions, 3) alteration or repair by anyone not authorized by Welch Allyn, and 4) accidents. If a Product or accessory covered by this warranty is determined to be defective because of defective materials, components, or workmanship, and the warranty claim is made within the warranty period described above, Welch Allyn will, at its discretion, repair or replace the defective Product or accessory free of charge. You must obtain a return authorization from Welch Allyn to return your Product before you send it to Welch Allyn’s designated service center for repair. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. WELCH ALLYN'S OBLIGATION UNDER THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF PRODUCTS CONTAINING A DEFECT. WELCH ALLYN IS NOT RESPONSIBLE FOR ANY INDIRECT OR CONSEQUENTIAL DAMAGES RESULTING FROM A PRODUCT DEFECT COVERED BY THE WARRANTY.
DIR 80012763 Ver. D
8 / 26
CardioPerfect PRO ECG Recorder - User Manual Table of contents Safety and Precautions ... 3 Accessories ... 6 Warranty, Service, and Spare Parts... 7 Limited Warranty Statement... 8 1. About the CardioPerfect Pro Cardiograph ... 11 1.1 Accessories ... 12 2. Installation... 14 To connect the cables to the Pro recorder: ... 15 To disconnect the cables from the Pro recorder: ... 15 2.1 Connecting the patient cable to the electrode... 16 3. Maintenance... 18 Cleaning the CardioPerfect Cardiograph and Battery Pack... 18 Cleaning the Patient and PC interface Cables... 18 Battery Led Indicator ... 19 Battery Care ... 19 Battery Charging ... 19 Storing the Battery... 19 4. Technical Specifications ... 21 Discarding the Equipment ... 21 5. Guidance and Manufacturer’s Declarations ... 23
DIR 80012763 Ver. D
9 / 26
CardioPerfect PRO ECG Recorder - User Manual Conventions WARNING - Warning statements describe conditions or actions that can result in personal injury or loss of life. CAUTION - Caution statements describe conditions or actions that can result in damage to the equipment or software. NOTE - Notes contain additional information on cardiograph usage.
Attention (Caution, Warning, Danger, Important, Note, Refer to Accompanying Documentation)
Ingress Protection (ordinary)
Serial Number
Reference Number
Type BF Equipment defibrillation protected
Manufacture Date
CE Mark (according to MDD93/42/EEC)
Temperature Range
Direct Current
Electromagnetic Compatibility When using the Cardio Perfect Pro cardiograph, electromagnetic compatibility with surrounding devices should be considered and evaluated. The Cardio Perfect Pro cardiograph complies with IEC 60601-1-2 limits for EMC.
DIR 80012763 Ver. D
10 / 26
CardioPerfect PRO ECG Recorder - User Manual
1. About the CardioPerfect Pro Cardiograph The Welch Allyn CardioPerfect Pro Cardiograph is specifically intended for the recording of standard 12-lead electrocardiograms of patients. The device is not suitable for direct cardiac application.
A CardioPerfect Pro system typically comes with the following components:
CardioPerfect Pro Cardiograph (SE-PRO) Patient Cable (RE-PC or SE-PC) PC Interface Cable Prolink (PRO-60023, -24, -25) Battery Pack (PRO-60019) CardioPerfect Workstation Software (CPWS-SW)
Caution
Warning
DIR 80012763 Ver. D
The patient cable supplied with the CardioPerfect cardiograph is an integral part of the cardiograph’s safety features. Using any other patient cable may compromise defibrillation protection as well as cardiograph performance. The patient cable should be routed away from power cords and any other electrical equipment. Failure to do so can result in AC power line frequency interference on the ECG trace.
Do not touch the patient, patient cable, PC interface cable (Prolink) or cardiograph during defibrillation. Death or injury may occur from the electrical shock delivered by the defibrillator. It is recommended to check the patient cable and PC interface cable (Prolink) for damage prior to the use of the system. If damage exists do not use the cable, contact your local Welch Allyn Sales Office or your authorized Welch Allyn dealer or distributor to have the cable replaced.
11 / 26
CardioPerfect PRO ECG Recorder - User Manual 1.1
Accessories
Your CardioPerfect Pro cardiograph allows the use of a broad range of accessories; contact your local Welch Allyn Sales Office or your authorized Welch Allyn Dealer or Distributor. NOTE All part numbers may not be available in all countries. Component Part No.
Component Description
Intended Use
CPR-UN-UB-D
PC Based Resting ECG; non interpretive software PC Based Resting ECG; interpretive software PC Based Resting ECG; Excludes patient cables PC Based Resting ECG; Excludes patient cables PC Based Resting ECG; Excludes patient cables; 5 meter USB cable PC Based Resting ECG
The Welch Allyn CardioPerfect Pro Cardiograph is specifically intended for the recording of standard 12-lead electrocardiograms of patients. The device is not suitable for direct cardiac application.
CPR-UI-UB-D CPRL-UN CPR-UN CPR-UN5 CPR-UN-EB CPR-UN-EB-D CPR-UI-EB-D CPR-UN-UB CPR-UI-UB CPR-SN RE-SW-MEANS RE-SW-ECG RE-SW-PEDIA RE-SW-QT RE-SW-VCG RE-SW-RR
PC Based Resting ECG; includes patient cables and electrode set PC Based Resting ECG; Adult interpretive software PC Based Resting ECG PC Based Resting ECG; Adult interpretive software, USB cable PC Based Resting ECG; Adult interpretive software, Serial cable CPWS software disk; with interpretation CPWS software disk CPWS software disk; pediatric version with interpretation CPWS software disk; dispersion option CPWS software disk; vector option
45008-0000
CPWS software disk; interval option CPWS software disk; heart rate variability option CPWS software disk; upgrade from resting to stress CPWS software disk; upgrade from DOS Windows CPWS software disk, update from older CPWS versions CPWS software disk; update from older versions, out of warranty only CPWS software disk; upgrade from resting “lite” resting ECG to normal ECG functionality CPWS software disk; upgrade from POR to normal resting functionality CPWS software disk and cover, upgrade from MDR to normal resting functionality Alligator clips attachment (1k box)
58581-0000
Alligator clips attachment (10 pack)
PRO-60023
PROLINK USB CABLE ASSEMBLY (2M) PROLINK USB CABLE ASSEMBLY (3M) PROLINK USB CABLE ASSEMBLY (5M)
RE-SW-HRV UPG-UN-UC-D-T CCW-UPCPWR UPDT-WACPW-CD UPDT-WACPW-OW CPR-UPG-LTU
CPR-UPG-POR-1 CPR-UPG-MD-1
PRO-60024 PRO-60025
DIR 80012763 Ver. D
Software upgrade from normal resting algorithm to include indicated data analysis options. These options allow data analysis by different methods. Software enables user to view and save electro-cardiograms of patients on a PC.
Software upgrade from originally purchased CPWS. Software enables user to view and save electro-cardiograms of patients on a PC.
Clips allow user to connect banana jack terminated patient leads to electrode tabs needing alligator clip attachment. Cable allows connection of the Pro-Recorder to the PC USB connector.
12 / 26
CardioPerfect PRO ECG Recorder - User Manual RE-PROLINK-SER RE-LINK-SER
SE-PC-IEC-PUSH SE-PC-IEC-PSHL SE-PC-AHA-PSHL SE-PC-IEC-CLIP SE-PC-IEC-CLPL SE-PC-AHA-CLIP SE-PC-AHA-CLPL
RE-PC-IEC-BAN RE-PC-IEC-BANL RE-PC-AHA-BAN RE-PC-AHA-BANL SE-PRO-600-R 08265-0000 RE-SIM
RE-ELEC-SET
RE-GEL RE-GEL-12 RE-ELEC-CUP RE-ELEC-KID
RE-ELEC-CLP RE-ELEC RE-ELEC-20 RE-BAN-ADP PRO-60019
PRO-60039 PRO-60040 PRO-60041 PRO-60042 101920
08282-0000 404008
DIR 80012763 Ver. D
SERIAL PRO RECORDER CABLE W/9P SUB D CONN UNILINK,RS232 FOR RESTING ECG Patient Cables; PUSH style connector, ½ meter PUSH style connector,1.3 meters PUSH style connector, 1.3 meters Patient Cables; CLIP style connector, 1 meter, CLIP style connector, 1.3 meters CLIP style connector, 1 meter CLIP style connector, 1.3 meters Patient Cables; Banana style connector, ½ meter Banana style connector, 2 meters Banana style connector, ½ meter Banana style connector, 2 meters CARDIO PERFECT PRO RECORDER 600HZ, refurbished UTILITY CART ECG SIMULATOR
ELECTRODE SET (6 CUPS,4 CLAMPS,GEL), suction cups used versus the stick on tabs ONE BOTTLE OF ELECTRODES GEL (260G) BOX OF ELECTRODES GEL (12 BOTTLES) WELCH CUP (individual suction cup) PEDIATRIC WELCH CUP ELECTRODS (pediatric version of the cup) LIMB LEAD CLAMPS (IEC) DISP ELECTRODES 50X BANANA CABLE DISPOSABLE ELECTRODES FOR BANANA CABLE ADAPTER SET FOR BANANA TO PUSH STYLE CLIPS BATTERY PACK FOR CP PRO RECORDER; Output 2.85 – 4.5VDC, 700mA BATTERY CHARGER. Friwo P/N FW7400/12; Output 12VDC /1A POWER CORD FOR BATTERY CHARGER,USA POWER CORD FOR BATTERY CHARGER,EUROPE POWER CORD FOR BATTERY CHARGER,UK Additional Protective Earth grounding kit for PC Dell PC, Optiplex GX745 Series CPWS FLEX ARM OPTION
Cable allows connection of the Pro-Recorder to the PC serial connector.
Cable allows connection of the Pro-Recorder to the electrode leads, which adhere to the patient for detection of cardiac signals.
Pro-Recorder as sold with normal resting system, refurbished model. Plastic cart that the PC and associated hardware can sit upon. Simulator of ECG signals that allows the user to confirm operation of the ProRecorder and CPWS system. Electrodes and accessories allow connection of the patient cables to the patient in a variety of attachment methods.
Battery pack and accessories allow the ProRecorder to be run on serial connection and on battery power. Accessories allow battery to be charged upon battery removal from Pro-Recorder.
Kit allows the CPWS system with PC to be complaint with IEC 60601-1-1 from a system perspective. PC that is approved for use with CPWS. Mechanical holder for the Pro-Recorder. Allows better cable management and support of the recorder while it is in use.
13 / 26
CardioPerfect PRO ECG Recorder - User Manual
2. Installation Before using the system, the CardioPerfect Pro cardiograph needs to be connected to the computer and the software must be properly configured.
Patient Cable Connection
Type BF Symbol Led
On/Off Switch
PC Symbol
PC Interface Cable Connection
Led
Figure 1 Front view CardioPerfect Pro Cardiograph Figure 1 is a front view of the CardioPerfect Pro. A description of the items indicated is given in table 1. Table 1 Description of CardioPerfect Pro Cardiograph front panel.
DESCRIPTION Patent Cable Connection Type BF Symbol On/Off Switch Led
PC Symbol PC Interface Cable Connection
EXPLANATION Connection for the patient cable IEC symbol, type BF equipment, defibrillator proof Manual switch to switch on cardiograph Power indicator, red indicates battery is low and needs to be recharged immediately, green indicates that sufficient battery power exists to record an ECG, no led indicates ECG recording is not possible. Input for computer cable Connection for the PC interface cable (Prolink)
Before you can start recording ECG’s with the CardioPerfect Pro cardiograph, you need to: 1. Install the applicable software (refer to the Welch Allyn CardioPerfect Workstation Installation User Manual) 2. Install the hardware, including the required drivers. (Please refer to the Rest manual chapters 11 & 12 or Stress manual chapters 10 & 11) 3. Configure the software (Please refer to the Rest manual chapters 11 &12 or Stress manual chapters 10 & 11)
DIR 80012763 Ver. D
14 / 26
CardioPerfect PRO ECG Recorder - User Manual
Figure 2 PC interface cable (Prolink) Latch release alignment key
Figure 3 Patient cable The Prolink cable has 2 alignment keys and the Patient cable has 1 alignment key. To connect the cables to the Pro recorder: 1. Hold the Pro recorder in one hand; hold the Cable connector with your other hand. 2. Align the alignment keys with the notches in the corresponding Pro recorder receptacle. 3. Push the cable gently into the Pro recorder receptacle until it latches. To disconnect the cables from the Pro recorder: 1. Hold the Pro recorder in one hand; hold the Cable connector with your other hand. 2. Gently pull the latch release away from the Pro recorder.
Warning
The Pro ECG recorder including connectors and cables should be handled with care. Improper use could result in inoperable ECG and or compromising patient safety. Do not drop the Pro ECG recorder; this may result in mechanical failure. Make sure the connectors are properly aligned before mating with the pro recorder. Do not use force to insert the cables into the Pro ECG recorder receptacles. Never pull the cables, only the latch release.
Caution
DIR 80012763 Ver. D
Patient cables and PC connection cables are intended to be inserted and removed from the Pro-Recorder in a push / pull motion. Do not twist the cables. Damage to the cables and the Pro-Recorder will occur.
15 / 26
CardioPerfect PRO ECG Recorder - User Manual
2.1
Connecting the patient cable to the electrode
Please refer to the table below for information on connecting the patient cable to the electrodes. Type of connection Banana
Connection to electrode
Alligator clamp
Resting ECG
Caution: Make sure the metal part of the patient cable connector makes contact with the skin side of the electrode tab.
Exercise ECG or Resting ECG
Push
DIR 80012763 Ver. D
16 / 26
CardioPerfect PRO ECG Recorder - User Manual
Exercise ECG or Resting ECG
Clip
DIR 80012763 Ver. D
17 / 26
CardioPerfect PRO ECG Recorder - User Manual 3. Maintenance NOTE It is recommended to check the performance of the cardiograph regularly. Contact your local Welch Allyn Sales Office or your authorized Welch Allyn dealer or distributor. Only qualified personnel may service the cardiograph.
Cleaning the CardioPerfect Cardiograph and Battery Pack Clean the cardiograph and battery pack with a soft cloth moistened with a recommended disinfectant or cleaning agent. It is advisable to clean the battery pack and ECG recorder using moist cloth. Never clean the recorder excessively with water or other substance so that components inside or connectors on front side of the recorder can be damaged, this may ultimately lead to decreased performance. Cleaning the Patient and PC interface Cables Clean the cables with a soft cloth moistened with a recommended disinfectant or cleaning agent. Cleaning cables: patient cables, plugs and power cords should be kept clean using lukewarm soapy water or a neutral cleaner. Disinfecting the cables: use chemical disinfectants containing ethanol (70%-80%), propanol (70%-80%) or aldehydes (2%-4%).
Caution
Do not clean the patient cable with alcohol. Alcohol can cause the plastic to become brittle and may cause the cable to fail prematurely. Do not autoclave the cable or use ultrasonic cleaners. Do not immerse the patient cable. Do not use abrasive materials to clean metal surfaces-scratches on them can cause artifacts on the ECG. Do not wet the connectors.
DIR 80012763 Ver. D
18 / 26
CardioPerfect PRO ECG Recorder - User Manual NOTE The CardioPerfect Pro cardiograph can be powered by a USB connection or through a battery pack. If you are using a USB connection (dummy battery pack) you do not have to charge the battery pack. If a battery pack is used with a RS232 connection, you have to charge the battery pack with a battery charger supplied by Welch Allyn.
Battery Led Indicator The battery pack is provided with a (green) led indicator that indicates the performance status of the battery pack, the following applies: Led Intermittent – battery pack is charging after complete discharge, it is recommended to not use the battery pack for use with the cardiograph for diagnostic purpose. Led On – battery pack is charging and not completely charged but can be used with the cardiograph for diagnostic purpose. Led Off – battery pack is completely charged and ready to be used with the cardiograph for diagnostic purpose.
Battery Care A fully charged battery will provide a minimum of 8 hours of operation. The CardioPerfect Pro cardiograph can be powered by a battery pack. This battery pack contains rechargeable batteries. Battery performance is maintained when recharging takes place when the battery is almost empty. This can be seen on the recorder when the led on the front lights up in red. This is the best time for recharging the battery pack. After the light has turned red you have still enough power for about ten minutes of recording. Both the internal gauge measurement and the percentage of battery capacity shown in Welch Allyn CardioPerfect Workstation become less accurate.
Battery Charging 1. Pull the battery compartment from the recorder by holding the recorder with one hand and pressing on both sides off the battery compartment with your other hand. 2. Connect the battery to an electrical outlet using the power adapter by inserting the charger cable into the top of the battery compartment and charge your battery. 3. Click the battery compartment back into place after the battery is recharged.
Warning
Charging must be done outside of the patient environment.
Storing the Battery If the cardiograph will not be used for more than three months, remove the battery pack from the unit and place it into storage. While in storage, charge the battery to retain the battery life and capacity. This ensures that the battery does not completely discharge while in storage. DIR 80012763 Ver. D
19 / 26
CardioPerfect PRO ECG Recorder - User Manual
Caution
If the battery pack is discharged to a very low level, the "Battery Time Left" on-screen timer of the CPWS software will not be accurate.
The system must perform a "learning cycle" before accuracy is regained. This "learning cycle" is achieved by fully charging a battery, fully discharging it through use, and recharging before the next use.
Caution
DIR 80012763 Ver. D
Repeated undercharging of the battery will damage the battery and reduce battery life.
20 / 26