Hill-Rom Air Shields
Babylog and Neonatal Ventilator Series
Babylog VN800 / VN600 Result Test Sheet Instructions / Service Card IPM-L Rev 4.0
Test Sheet
3 Pages
Preview
Page 1
Result Sheet Test instructions / Service Card IPM-L Babylog VN800 / VN600 Part number: Location:
Serial number:
Department:
Cust. invent. no.:
Maintenance interval:
Other / Delivery date: Key
Applies to Test instructions / Service Card IPM-L Revision 4.0
Test
/ OK
= OK
+
= Spare part used
!
= Error / Report
/
= Accessory not available
-
= Not applicable
Test
Result
Result
3.1.7
1 Device configuration 1.1
Serial numbers
1.1.1
Control unit
txt
1.1.2
Ventilation unit
txt
1.1.3
Inspiratory valve
txt
1.1.4
Neonatal expiratory valve
txt
1.1.5
Second expiratory valve for neonates
txt
1.1.6
(Optional) MCable Mainstream CO2 (CO2 sensor)
txt
Applied part leakage current, mains on applied part, type BF
3.1.7.1
Reference value
μA
3.1.7.2
Recurrent test
μA
3.2
Electrical safety according to IEC 60601-1
3.2.1
3.2.2
Visual inspection
OK
Protective earth resistance
3.2.2.1
Protective earth resistance, ventilation unit
Ω
3.2.2.2
Optional protective earth resistance GS500
Ω
1.1.7
(Optional) GS500
txt
3.2.3
1.1.8
(Optional) PS500
txt
3.2.3.1
Normal condition (N.C.) IEC 60601-1
μA
1.1.9
(Optional) PS500, part number
txt
3.2.3.2
Single fault condition (S.F.C.) IEC 60601-1
μA
1.1.10 (Optional) trolley
txt
3.2.4
1.1.11 (Optional) nebulizer ejector (no serial number) (YES/NO)
txt
3.2.4.1
Normal condition (N.C.) DC IEC 60601-1
μA
3.2.4.2
Normal condition (N.C.) AC IEC 60601-1
μA
Software version
3.2.4.3
Single fault condition (S.F.C.) DC IEC 60601-1
μA
2 Maintenance parts
3.2.4.4
Single fault condition (S.F.C.) AC IEC 60601-1
μA
2.1
4 Function and condition test
1.2
Maintenance intervals, overview
2.1.1
Maintenance intervals and required sets and parts
General condition check
4.1.1
Medical device
OK
dat
4.1.2
Accompanying documents
OK
Ventilation unit, particle filter (O2 filter)
dat
4.1.3
Medical devices book
OK
Ventilation unit, Air filter (medical compressed air filter)
dat
4.1.4
Rating plates, labels and notices on the medical device
OK
(Option) GS500, room air filter (filter mat)
dat
4.1.5
dat
Rating plates, labels and notices on the installed options
OK
(Option) GS500, air filter blower unit (microfilter) (Option) PS500, battery set PS500 (Service) 15Ah
dat
4.1.6
Control unit
OK
4.1.7
Ventilation unit
OK
(Option) PS500, battery set PS500 (Service) 24Ah
dat
(Option) Medication nebulizer, pressure regulator (O2-side)
dat
Maintenance parts by specified interval
2.2.1
Ventilation unit, internal battery M7
2.2.2
2.2.3
2.2.4
2.2.5
2.2.6 2.2.7
2.2.8
2.3
Maintenance parts as required
3 Electrical safety 3.1
Electrical safety according to IEC 62353
3.1.1
3.1.2
Patient leakage current
4.1
2.2
Earth leakage current medical device
Visual inspection
OK
Protective earth resistance
3.1.2.1
Maximum measured value of the ME device
Ω
3.1.2.2
Maximum measured value of ME system.
Ω
3.1.3
Protective earth resistance measuring points
OK
4.2
Condition test, options
4.2.1
(Option) MCable Mainstream CO2 module M11.1 (CO2 sensor)
OK
4.2.2
(Option) Medication nebulizer (black housing)
OK
4.2.3
(Option) Trolley
OK
4.2.4
(Option) Gas cylinder holder
OK
4.2.5
(Option) Hose holder
OK
4.2.6
(Option) GS500
OK
4.2.7
(Option) PS500
OK
4.2.8
(Option) Cable duct
OK
4.3
Fuse rating and readout of software version and operating data
3.1.4
Test adapters for equipment leakage current
4.3.1
Fuse value
OK
3.1.5
Device leakage current
4.3.2
Software version
txt
3.1.5.1
Reference value
μA
4.3.3
3.1.5.2
Recurrent test
μA
4.3.3.1
Battery operation data
txt
4.3.3.2
Battery operating data of the medical device with M7.3 Plus
txt
3.1.6
Applied parts for leakage current from applied part test
������������
Batteries
����������������������
�����
Test
Result
Test
4.4 Preparations for the function tests
4.5
Control unit
4.5.1
Acoustic alarm
OK
4.5.2
Optical alarms
OK
4.16 Nurse call
4.5.3
Rotary encoder lighting
OK
4.6
Ventilation unit module M1.3 (gas delivery module)
OK
Result
4.14.1 Testing
OK
4.15 Ventilation unit, Piezo 4.15.1 Function check 4.16.1 Nurse call with nurse call box (8417370)
OK OK
4.17 Ventilation unit, expiratory valve heating
4.6.1
Self-tests of the ventilation unit
4.6.2
Zeroing all flow sensors
4.6.3
Relative flow difference S2/S3
4.6.4
Relative flow difference S1/S3
4.6.5
Leakage LPSV V3
4.6.6
Valve calibration for workstations with gas supply unit GS500
4.6.7
Comparator shut-off pressure
mbar
4.21 (Option) GS500
4.6.8
Comparator switch-on pressure
mbar
4.6.9
Proportional valve AIR 120 L/min
A
4.22 Medical device, function tests in operation mode
4.6.10 Proportional valve AIR 5 L/min
A
4.22.1 Self-test
4.6.11 Proportional valve O2 120 L/min
A
4.22.2 System test
OK
4.6.12 Proportional valve O2 5 L/min
A
4.22.3 Breathing circuit test
OK
4.6.13 System Leakage Test (M1V3)
OK
4.22.4 Function tests in ventilation mode
OK
4.6.14 Leakage, AIR proportional valve
mbar
4.22.5 Non-return valves in the gas inlet
OK
4.6.15 Leakage, O2 pipeline supply non-return valve
L/min
4.22.6 Display of the ventilation unit
OK
4.6.16 Leakage, O2 proportional valve
mbar
4.22.7 Alarm silence key
OK
4.6.17 Leakage, AIR pipeline supply non-return valve
L/min
4.23 (Option) PS500
4.6.18 Comparative pressure measurement, Ptank and inspiratory pressure sensor
OK
4.23.1 Function check
4.6.19 Comparative flow measurement M1.3
OK
4.7
Ventilation unit module M12.4 (Pato, standalone)
4.7.1
Calibration
4.7.2
Measurement accuracy
4.8
Ventilation unit module M4.1 (pressure-/flow temperature monitoring)
OK L/min L/min OK
4.17.1 Function check
OK
4.18 (Option) Ventilation unit, nebulizer ejector
4.18.1 Function check
OK
4.19(Option) Medication nebulizer (black housing)
OK
4.20 (Option) MCable Mainstream CO2 module M11.1 (CO2 sensor)
4.20.1 Function check 4.21.1 Function check
OK OK OK
4.23.1.1
Mains power supply (LED)
4.23.1.2
Removing fuse F1
OK
4.23.1.3
Fuse rating of fuse F1
OK
OK
4.23.1.4
Inserting fuse F1 and removing fuse F2
OK
OK
4.23.1.5
Fuse rating of fuse F2
OK
4.23.1.6
Power supply
OK
4.8.1
Calibration of pressure sensors
OK
4.8.2
Offset of the inspiratory pressure sensor
mV
4.8.3
Offset of the expiratory pressure sensor
mV
4.8.4
Temperature measurement of module M4.1
OK
4.8.5
Reference voltage of module M4.1
OK
4.9
Safety devices
4.9.1
Emergency expiratory valve SV2
4.9.2
Emergency respiratory valve SV3
4.9.3
Non-return valve in expiratory valve
OK
4.9.4
Expiratory resistance of neonatal expiratory valve
OK
4.9.5
Neonatal flow sensor
OK
OK
4.24 Final procedures
4.24.1 Device status report
OK
4.24.2 Device handover with test label
OK
5 Test equipment 5.1
Test equipment subject to mandatory calibration
5.1.1
Test equipment calibrated in a valid manner used.
OK
OK
5.1.2
Test equipment subject to mandatory calibration
OK
OK
5.2
Test equipment not subject to mandatory calibration
4.10 Ventilation unit power supply module M7.3 Plus
4.10.1 Fan
OK
4.10.2
Supply voltage
OK
4.10.3
Internal temperature
OK
4.11Emergency backup battery of the ventilation unit, part 1
OK
4.12 Ventilation unit PI self-tests 4.12.1 Self-tests of the ventilation unit
4.12.2 Flush flow M12.4
OK
4.12.3 IST neonatal flow sensor
OK
4.12.4 IST Airway Pressure Measurement
OK
4.12.5 Pneumatic Release Logic Test
OK
4.12.6 Pressure sensors S2 and S8
4.12.6.1
S2 - Deviation from the determined ambient pressure
mbar
4.12.6.2
S8 - variation from the measured ambient pressure
mbar
4.13 Emergency backup battery of the ventilation unit, part 2
4.13.1 Battery parameters
OK
4.14 Ventilation unit, interior temperature
������������
����������������������
�����
Report:
����������������������������������������������������������� ����� ����������������������������� ����������������������������������������������������������� ������������������������������ �����������������������������������������������������������
������������
����������������������
�����