Software User Manual
24 Pages
Preview
Page 1
Using the Event Viewer Software to download the Event Log from the Sapphire Infusion Pump Table of Contents Overview ...2 System Requirements ...4 Minimum Hardware Requirements ...4 Minimum Software Requirements ...4 Installing the Event Viewer Software on your computer ...5 Uninstalling the Event Viewer Sotware ...6 Working with the Event Viewer Software ...7 Event Log Tables ...12 Power-on Events ...12 Pump Idle Time Events ...13 Pre-infusion and Intra-infusion Events ...13 Operation-related Events ...16 Settings and Options-related Events ...18 General Events ...23
Using the Event Viewer software to download the Event Log from the Sapphire Infusion Pump
1/24
Using the Event Viewer Software to download the Event Log from the Sapphire Infusion Pump The following explains how to download an Event Log from the Sapphire Infusion pump (including but not limited to: Sapphire Multi-Therapy Infusion Pump, Sapphire Epidural Infusion Pump, Sapphire PCA Infusion Pump or the Sapphire TPN Infusion Pump) to a computer (PC). This DFU is applicable to Event Viewer Software Rev13.
Overview The Event Viewer software allows the review of the Sapphire pump operation history (Event Log). The Sapphire pump stores up to 8,000 events in its Event Log memory and the Event Viewer software is used for downloading, sorting and filtering events by description or date.
For a complete list of the events recorded in the Event Log, refer to the Event Log Tables, starting on page 12.
• Make sure to use an Event Viewer software version that is
equivalent to or a higher than the software version of the Sapphire pump. The Event Viewer software version can be viewed by clicking the 'About' option in the 'Help' tab at the top of the Event Viewer main screen.
• The Event Viewer software can only be used with Sapphire pump Rev09 software and above.
Once an Event Log is downloaded, it can be saved in .csv format.
•
I
The .csv file may be imported as a text file by opening it in Excel® spreadsheet softwareI .
Excel is a registered trademark of Microsoft Corporation in the U.S. and/or other countries. Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire 2/24
Event Log files should be sent for evaluation at Q Core Medical Ltd. in the following circumstances:
• • • • •
The pump was involved in the injury or death of a patient. The pump was involved in an event that may have caused injury or death of a patient; or The pump has caused or may have caused injury to the medical staff; or The pump is not performing as expected. Upon Q Core Medical's request.
Using the Event Viewer software to download the Event Log from the Sapphire Infusion Pump
3/24
System Requirements Before installing the Event Viewer software, verify that the following hardware and software requirements are met:
Minimum Hardware Requirements • • •
PC: Intel® Core™2 DuoII - 2GB RAM Screen resolution: 1280 x 1024 pixels Q Core Medical’s Communication Cable for the Sapphire pump.
• •
Distributed by Hospira Worldwide, Inc. (P/N 05020-110-0213).
RS232 connectivity, in either of the following forms:
•
A RS232 port in the PC
Or
•
Q Core Medical’s USB to 4 Port RS232 Adapter
•
Distributed by Hospira Worldwide, Inc. (P/N 15077-000-0001).
Minimum Software Requirements • • •
Operating System: Windows ® XP operating system, Windows ® III Vista operating system, or Windows ® 7 operating system (32 or 64 bit). Excel ® -spreadsheet software (2003 or later). IV
WinRAR ® shareware file archiver and data compression software or other software for handling .zip files.
II
Intel and Intel Core are trademarks of Intel Corporation in the U.S. and/or other countries III Windows is a registered trademark of Microsoft Corporation in the U.S. and/or other countries. IV WinRAR is a registered trademark of win.rar GmbH. Using the Event Viewer software to download the Event Log from the Sapphire Infusion Pump
4/24
Installing the Event Viewer Software on your Computer > To Install the Event Viewer softwa re on your co mput er :
1.
After launching the installer, click Next to continue with the setup wizard.
2.
Read the Q Core Event Viewer End User License Agreement, and select the checkbox if terms are accepted. IMPORTANT NOTE: This product is subject to the End User License Agreement included therein. When installing the product on your hard drive, you will be requested to accept the terms of such License Agreement. If you do not accept the license agreement, you may not install or use the product in any way whatsoever.
3.
If you wish to create a shortcut on your desktop and start menu, make sure that the relevant checkbox is selected.
4.
Click Install to continue with the setup wizard.
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
5/24
6.
Once the installation . is complete click Finish. A folder with the Event Viewer software and the DFU is created.
Uninstalling the Event Viewer Software > To Uninstall the Event Viewer software from your computer:
1.
Open the Windows Control Panel.
2.
In Control Panel, double click Add or Remove Programs / Uninstall a Program.
3.
Select the Event Viewer software, and click Remove/Uninstall.
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
6/24
Working with the Event Viewer Software The following procedure explains how to connect, filter and download events from the Sapphire pump to your PC. > To connect the Sapphire pump to the Event Viewer software
1. 2. 3. 4.
Turn the Sapphire pump on. Then, using the Communication Cable, connect the pump to the computer. Double click the Event Viewer icon, created on your desktop. The "RS232 Communication Ports" window appears with a list of available ports displayed. Select a port from the list, and then click OK. The Event Viewer screen appears, displaying the serial number of the pump on the upper left corner of the window.
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
7/24
If the pump is not prorerly identified properly by the Event Viewer, a "No Device" message will be displayed on the upper left corner of the window.
•
If this occurs: • Verify that the pump is connected to the selected port. • If the pump is connected to the selected port: • Exit the Event Viewer. • Disconnect the pump from the Communication Cable. • Repeat steps 1 through 4 from “To connect the Sapphire pump to the Event Viewer software"in page 7.
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
8/24
> To filter events in the E Event Viewer
Events may be filtered by Event Code, Time, Description or Priority. This action is available before or after the events have been downloaded from the pump. Note that by default, the Event Time filter is activated on the last 7 day If no additional filtering is required, proceed to "To download and export events from the pump" in page 11.
1. In the upper right corner of the Event Viewer screen, click Filters.
2.
To select the desired filters, enter the required condition in the relevant line, and then check the "Is Active" checkbox. In the following example, only the events that occurred between October 5 and October 13, 2015 will be downloaded.
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
9/24
When filtering by Priority, Events can be filtered by the following Priority number:
• 0 (Red) - Fatal error (Events: 1,2,3,4,5,118,119) • 1 (Pink) - Fault (Events: 7,8,9,10,11,12,13,14,17,26,31,129,132,157,158) • 2 (Dark Orange) - Message (Events: 6,15,16,18,19, 22, 24) • 3 (Light Orange) - Service message (Events: 20,21,111) • 4 (Purple) - Notification (Events: 27,28) • 5 (Yellow) - Service notification (Events: 25,30,112) • 6 (Green) - Information (Events: all events which are not included in one of the other categories
3. Click OK. The Filters dialog box closes.
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
10/24
> To download and export events from the pump E:E
1. At the top of the Event Viewer screen, click Load from pump. The caption of the button changes to Stop loading and the download begins. As the download progresses, events are displayed on the Event Viewer sceen. Once the download is completed, the caption on the button changes back to Load from pump. The downloaded events can be sorted by clicking Sort or by clicking on the Time, Code, Description or Source column title. 2. To export the downloaded events to Excel, click Export as Excel and then save the file in a folder of your choice. The file is automatically given the serial number of the pump, and the date and time it was saved. However, it is recommended to give the file a relevant name that includes the institution name, serial number of the pump, date etc.
3.
If the Event Log file should be sent to Q Core Medical for evaluation, as defined in the Overview section of this document, please send the file to Q Core Medical by emailing it to [email protected]
4.
Disconnect the pump from the PC, and then exit the Event Viewer software.
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
11/24
Event Log Tables The following tables list all Sapphire infusion pump events that are recorded to the Event Log: Power-on Events ... 12 Pump Idle Time Events ... 13 Pre-infusion and Intra-infusion Events ... 13 Operation-related Events ... 16 Settings and Options-related Events ... 18 General Events ... 23
Power-on Events The following events occur during the power-on period: Event Code
Event Description
Text on pump Screen
2
End of battery life span
End of battery life
3
Watchdog reset
Built in test failure
111
Battery life expires
Battery life expires in 2 days
112
Battery life expires
Battery life expires in 2 weeks
117
Non authorized battery
Non authorized battery inserted
118
Software failure
Software failure. Code xx (1 – Program memory, 2 – Drug library, 3 – Icons, 4 - Fonts, 5 Language 1, 6- Language 2, 7Language 3)
119
External watchdog failure
External watchdog failure
140
Time changed to
Time changed to: <xx:xx> AM/PM
193
Battery usage
Battery usage error
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
12/24
Event Code
Event Description
Text on pump Screen
246
WiFi malfunction
WiFi module communication malfunction
250
System monitoring used to monitor unrecovered failures in client site
System monitoring
Pump Idle Time Events The following events occur while the pump is idle (not during an infusion): Event Code
Event Description
Text on pump Screen
18
Pump unattended
Pump unattended
248
Time synchronization
Difference of +/- xx days and xx:xx:xx h:min:sec Server time from displayed time
249
Auto program
Auto program was accepted by user
257
Auto program
Auto program was rejected by user
Pre-infusion and Intra-infusion Events The following events occur before and during an infusion:
Event Code
Event Description
Text on pump Screen
5
Inconsistent flow/Pumping mechanism
Faulty pumping mechanism or inconsistent flow <xx>
7
Low input pressure
Upstream occlusion
9
Output Occlusion
Downstream occlusion
10
Administration set misplaced
No administration set loaded
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
13/24
Event Code
Event Description
Text on pump Screen
12
Air in Line
Air in line
13
Faulty air Detector
Air detector failure
14
Battery depleted
Depleted battery
16
Door open
Door of administration cassette open
17
External Communication Error
External communication error
20
Certification due within 2 days
Calibration due in <xx> days.
21
Certification due now Certification date arrived/ passed
Calibration due now
25
Certification due within 2 weeks
Next calibration due in 2 weeks.
26
Low battery voltage for high rates
Low battery for rate over 800 mL/h
28
Treatment near end
Treatment near end
29
Delayed KVO (may appear only when ADP/ EDHR is in control of a pump)
KVO during Delayed Start changed to: <xx.x> mL/h
31
Incorrect pressure value reading
Incorrect pressure value reading
36
Wireless network connection
Pump is connected to wireless network
40
Accumulated Air in Line
Accumulated air in line
42
VI Reset
VI Reset XXXX mL
82
Continuous rate, Soft limit override
Continuous rate, soft limit override: <xxx> mL/h
102
Pump Bolus key pressed
Pump bolus key pressed
103
Bolus handle key pressed
Bolus handle key pressed
127
Wireless network connection
Pump was disconnected from wireless network
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
14/24
Event Code
Event Description
Text on pump Screen
129
Check for Occlusion
Possible occlusion; infusion started with clamps closed
131
Auto-KVO changed
Auto-KVO changed to: <xx.x> mL/h
132
Unknown occlusion
Possible occlusion; infusion started with clamps closed
136
Secondary VI Reset
Secondary VI Reset XXXX mL
139
Infusion near end
Infusion near end alarm set to: <xx> min
169
Power save mode state
Power save mode state: <error code> (0 – Entering stop mode, 1 – Failed to enter stop mode, 2 – Exit stop mode On/Off button pressed, 3 – Exit stop mode Stop button pressed, 5 – Exit stop mode Pump request, 6 – Exit stop mode One Minute Before Step/ State Changed)
184
Automatic restart
Automatic restart after downstream occlusion
234
Catheter bucket changing
Catheter bucket changed from XX to YY
240
Catheter hysteresis
Catheter hysteresis conditions found: XX A/D
244
MedNet connection
Pump is connected to MedNet
245
MedNet connection
Pump was disconnected from MedNet
261
Loading dose complete
Loading Dose complte. Volume infused: XX.X mL
262
Loading Dose quit
Loading Dose quit prematurely. Volume infused: XX.X mL
263
Loading Dose paused
Loading Dose paused
264
Loading Dose started
Loading Dose started
265 270
Accumulated VI Reset
Accumulated VI Reset XXXX mL
Environmental light
Excessive environmental light
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
15/24
Operation-related Events The following events occur during specific pump operations, such as priming, starting an infusion, delivering a bolus, etc.: Event Code
Event Description
Text on pump Screen
24
Treatment completed
<Delivery mode> treatment completed
32
Bolus started Rate: XXX mL/h
Bolus started rate: <xxx> mL/h
33
Bolus stopped Volume infused: XX.X mL
Bolus stopped; volume infused: <xx.x> mL
34
Bolus finished Volume infused: XX.X mL
Bolus finished; volume infused: <xx.x> mL
41
Delayed start
Delayed start set to: <xx:xx:xx>
56
Power off
Power off
57
Power on
Power on
68
Clinician Bolus started
Clinician bolus started
71
Clinician Bolus complete. Volume infused: XX.X mL
Clinician Bolus complete. Volume infused: XX.X mL
73
New Infusion started
New infusion started
104
Bolus started
Bolus started
105
Demand Bolus Complete. Volume infused: XX.X mL
Demand Bolus complete. Volume infused: XX.X mL
108
Prime started
Prime started
109
Prime stopped
Prime stopped
110
Prime finished
Prime finished
122
Started from secondary bag
Started from secondary bag
124
Started from primary bag
Started from primary bag
130
Infusion Info
Infusion ended info: VTBI left = xxxx, VI=xxx, Time=xx:xx:xx, Rate=xxx.x (or last rate).
147
Infusion started with delay
Infusion started with delay (KVO = X mL/h)
159
Continued from primary bag
Continued from primary bag
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
16/24
Event Code
Event Description
Text on pump Screen
160
Continued from secondary bag
Continued from secondary bag
161
Secondary line added
Secondary line added
163
Power on (battery supply)
Power on (battery supply)
164
Demand Bolus quit prematurely. Volume infused: XX.X mL
Demand Bolus quit prematurely. Volume infused: XX.X mL
165
Clinician Bolus quit prematurely. Volume infused: XX.X mL
Clinician Bolus quit prematurely. Volume infused: XX.X mL
166
Demand Bolus paused
Demand Bolus paused
167
Clinician Bolus paused
Clinician Bolus paused
172
Certification succeeded
Pump passed certification
173
Certification failed
Pump failed certification
175
Pausing bolus was not confirmed
Pausing bolus administration was attempted but not confirmed
176
Pausing infusion was not confirmed
Pausing infusion was attempted but not confirmed
177
Turn off while running was not confirmed
Turn off while running was attempted but not confirmed
178
Turn off was not confirmed
Turn off was attempted but not confirmed
179
Unlock screen was not confirmed
Unlock screen was attempted but not confirmed
182
Continue infusion was not confirmed
Continue infusion was attempted but not confirmed
183
Continue bolus was not confirmed
Continue bolus administration was attempted but not confirmed
202
Repeat last infusion
Last treatment was repeated
209
Infusion started with delay
Infusion started with delay (No KVO)
210
KVO of x mL/h was ended
KVO of x mL/h was ended
217
Intermittent dose complete
Intermittent dose complete Volume infused: XX.X <units>
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
17/24
Event Code
Event Description
Text on pump Screen
219
Standby mode started
Standby mode started
220 221 222
Standby mode ended Immediate Taper down New treatment wizard initiated
Standby mode ended Pump stopped with immediate taper down New treatment wizard initiated
223
Bolus wizard initiated
Bolus wizard initiated
224
Prime wizard initiated
Prime wizard initiated
225
Secondary line wizard initiated
Secondary line wizard initiated
226
Programing was cancelled
Programing was cancelled
231
Delayed start finished
Delayed start finished
232
Preset Programs removal
Preset Programs were removed
238 239
External battery External battery
External Btteries Connected External Btteries disconnected or depleted
243
Drug Library removal
Drug Library was removed
Settings and Options-related Events The following events occur when pump configuration options are set: Event Code
Event Description
Text on pump Screen
23
Backlight
Backlight set to: On/Off/Partial
35
Key volume
Keys volume set to: <x>
37
Allow delayed start
Allow Delayed start set to: On/Off
38
Bolus rate
Default bolus rate set to: <xxx> mL/h
39
Accumulated Threshold
Accumulated threshold set to: <xxx>
46
Programmed treatment
Allow load pre-programmed treatments set to: <xx>
48
Pump unattended
Pump unattended set to: <xx>
51
Delivery mode changed to: X
Delivery mode changed to: <x>
52
Occlusion level changed to: XX.X X
Occlusion level changed to XX.XX
53
Occlusion units set to: X
Occlusion units set to: <x>
54
Single Bubble set to: X
Single bubble set to: <x>
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
18/24
Event Code
Event Description
Text on pump Screen
55
Accumulated Bubble set to: X
Accumulated bubble set to: <x>
58
Authorization level set to: X
Authorization level set to: <x>
59
Alarm Volume set to: X
Alarm Volume set to: <x>
60
Speaker interval set to: X
Speaker interval set to: <x>
61
Reset system set to factory defaults
Reset system set to factory defaults
62
Language changed to: X
Language changed to: <x>
63
Time changed to: XX:XX
Time changed to: <xx:xx>
64
Date changed to: XX/XX/XX
Dated changed to: <xx/xx/xx>
65
New drug library
New drug library file uploaded: xxxxxx
67
Dose units changed to: X
Dose unit changed to: <x>
69
Clinician Bolus Amount set
Clinician bolus amount set to: <xx.xxx>
72
Bolus rate set to: XXX.X mL/h
Bolus rate set to: <xxx.x> mL/h
76
VTBI set to: XXXX mL
VTBI set to: <xxxx> mL
77
Rate set to: XXX mL/h
Rate set to: <xxx.x> mL/h
78
Dose set to: XXX
Dose set to: <xxx>
79
Interval set to: XX:XX h:min
Interval set to: <xx:xx> hr:min
81
Infusion Period set to: XX:XX h:min
Infusion period set to: <xx:xx> hr:min
83
KVO set to: XX.X mL/h
<Delivery mode> KVO set to: <xx.x> mL/h
84
KVO options set to XX.X [units] (for intermittent)
KVO options set to XX.X [units]
85
Taper Up set to: XX:XX h:min
Taper up set to: <xx:xx> hr:min
86
Taper Down set to: XX:XX h:min
Taper down set to: <xx:xx> hr:min
87
Plateau Rate calculated to: XXX.X mL/h
Plateau rate calculated to:< xxx.x> mL/h
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire
19/24
Event Code
Event Description
Text on pump Screen
90
Concentration set to: XX.X X /mL
Concentration set to: <xx.xx> /mL
92
Patient Weight set to: XXX kg
Patient weight set to: <xxx> kg
95
Drug concentration selected to: XX.X X
Drug concentration selected to: <xx.xx>
96
Bolus Lockout set to: XX:XX h:min
Bolus lockout set to: <xx:xx> h:min
97
Max Bolus per hour set to: XXX
Number of Boluses per x h(s) set to: xxx (x = 1 or 4)
98
Continuous rate set to: XXX.XX mL /h
Continuous rate set to: <xxx.xxx> /h
99
Demand Bolus set to: XXX.XX X
Demand bolus set to: <xx.xxx>
100
Demand Bolus calculated
Demand Bolus calculated to: <xxx.xx>
101
Rate calculated
Continuous rate calculated to: <xxx.xx>/h
120
Secondary VTBI set to: xxxx mL
Secondary VTBI set to: <xxxx> mL
121
Secondary rate set to: xxx.x mL/h
Secondary rate set to: <xxx.x> mL/h
125
Password changed
Password(s) were changed
134
Infusion type
Infusion type set to <x>
135
Limit Period settings
Limit Period set to: x h(s) (x = 1 or 4)
137
Patient ID
Patient ID set to: xxxxxxxxxx PATIENT for 'null' Patient ID
138
US format
US format set to: On/Off
141
Air detection has been set to Off
Air detection has been set to Off
142
New patient
New patient set to: On/Off
143
Loading dose
Loading dose set to: <xx>
144
Allow loading dose
Allow loading dose set to: <xx>
Using the Event Viewer software to download the Event Log Infusion Pump from the Sapphire